Alcator C-Mod Run 1000626 Information

Back to runs

Miniproposals
Miniproposal:273
Date Filed: 6/19/2000
Title:Effects of Glow Discharge Cleaning on Boronization and Wall Conditions
First Author:Jim Irby
Session Leader:Jim Irby (shots 1-17)

Operators
Session leader(s):Jim Irby
Physics operator(s):Jim Irby
Engineering operator(s):Gary Dekow,Bill Cochran,Ed Fitzgerald

Engineering Operator Run Comment
MP 273 Effects of Glow Discharge Cleaning on Boronization and Wall Condi

Session Leader Plans

Physics Operators Plans

Session Leader Summaries
Entered: Jul 7 2004 03:49:24:540PM
Author: To Be Determined
Session Leader Run Summary for 1000626

MP273 Effects of Glow Discharge Cleaning on Boronization and Wall Condition

Particle Control Task Force

Session Leader: Irby
PO: Irby
EO: Cochran, Dekow, Fitzgerald

This experiment investigated another, possibly better, technique for
preparing the walls for plasma discharges following boronization. It
was a first step test of whether or not glow discharge cleaning
in helium is useful for particle control.

Other machines such as Asdex and DIIID use glow discharge cleaning
techniques to both deposit boron and prepare the wall for plasma
discharges. Between shot glow discharges, typically in helium, are
used to clean the walls. The higher energy ions produced during glow are
thought to be required for producing a "good" boron layer. We should
be able to test whether or not such techniques are better than the ECDC
process we have been using.

Though we had several technical glitches during this run, the results
are generally very useful in beginning to understand the effects
of wall conditions on the plasmas we produce. Prior to the run on
Monday, we did the following:

Saturday: boronized the machine (1600 Ang deposition)

ecdc cleaned in helium for a couple of hours

started glow discharge cleaning in helium

Saturday to Monday morning: glow discharge cleaned in
helium at electrode voltages as high as 550 volts,
and total currents up to 3.5 amps.

Monday: stopped the glow at 7:00. No ECDC prior to run.

For the first plasma shot on Monday we reloaded shot 1000616001 (defined
as our fiducial shot), a 1 MA shot with target densities of 1e20 m^2, nl04.
This shot required about 43 T-l of gas to run. The plan was to re-run this
shot and see how much more gas it took after the extensive glow
discharge campaign. Unfortunately, the first shot only lasted 350 ms.
So we loaded the wall somewhat with a shot that could not be easily
compared to the earlier fiducial. The following shot (3) ran full length
and clearly required much more gas than the fiducial (62 T-l). The
next two shots only required 36 and 33 T-l of gas. These values would
probably have been closer to the fiducial shot had the PO asked the
RF to be turned on at 800 rather than 600 ms. The conclusion is that
it takes only 1 or 2 discharges to reload the walls, even after a
very extended period of wall conditioning; probably just 1.

We also did two periods of between shot glow discharge in helium. Both
of these periods had 30 minutes of glow at up to 2.5 amps and 450 volts.
However, we were unable to get to low enough pressures to produce the
uniform, chamber filling, glow we had over the weekend. Following the
glow, we again ran the fiducial discharge, and in both cases 45 T-l
of gas was required. So, 30 minutes of glow is worth about 10 T-l of
gas. The second shot following the glow was right back to the 33 to
35 T-l value; recovery again with 1 shot. Had we more time, we would have
tried another period of glow lasting 60 minutes to see what the wall
inventory depletion rate was.

Finally, it was hoped that glow discharge cleaning in helium over the
weekend would improve the effects of the boronization and possibly
help the recovery process. I do think that the machine recovered
very quickly. The machine was no harder to run on Monday than it had
been the previous week. The carbon levels were very low, and the H/D
was in the 3 to 5% range, even after the last two run of the previous
week which ran very high hydrogen fraction discharges. However, the
radiated power was rather high, even somewhat worse than the fiducial
shot which was run well after the boronization of 5/30. The radiated
power did seem to be coming down during the run, and was much better
the following day after overnight ECDC in D2. The 1000627 run also
produced our lowest ever H/D ratios (1 to 3%). How much the wall
conditioning actually helped, if at all, will have to be determined
over the next few days as we run the machine.

Another run sometime in the distant future (September) would be
useful in which between shot glow was run with variations in the
length of glow time. We could then get a good measurement of
the depletion rate. It would also be useful to do target density
scans to see if we can reach a condition in which the density during
the h-mode actually decreases.






Entered: Jul 7 2004 03:49:24:553PM
Author: To Be Determined
Session Leader Run Summary for 1000626

MP273 Effects of Glow Discharge Cleaning on Boronization and Wall Condition

Particle Control Task Force

Session Leader: Irby
PO: Irby
EO: Cochran, Dekow, Fitzgerald

This experiment investigated another, possibly better, technique for
preparing the walls for plasma discharges following boronization. It
was a first step test of whether or not glow discharge cleaning
in helium is useful for particle control.

Other machines such as Asdex and DIIID use glow discharge cleaning
techniques to both deposit boron and prepare the wall for plasma
discharges. Between shot glow discharges, typically in helium, are
used to clean the walls. The higher energy ions produced during glow are
thought to be required for producing a "good" boron layer. We should
be able to test whether or not such techniques are better than the ECDC
process we have been using.

Though we had several technical glitches during this run, the results
are generally very useful in beginning to understand the effects
of wall conditions on the plasmas we produce. Prior to the run on
Monday, we did the following:

Saturday: boronized the machine (1600 Ang deposition)

ecdc cleaned in helium for a couple of hours

started glow discharge cleaning in helium

Saturday to Monday morning: glow discharge cleaned in
helium at electrode voltages as high as 550 volts,
and total currents up to 3.5 amps.

Monday: stopped the glow at 7:00. No ECDC prior to run.

For the first plasma shot on Monday we reloaded shot 1000616001 (defined
as our fiducial shot), a 1 MA shot with target densities of 1e20 m^2, nl04.
This shot required about 43 T-l of gas to run. The plan was to re-run this
shot and see how much more gas it took after the extensive glow
discharge campaign. Unfortunately, the first shot only lasted 350 ms.
So we loaded the wall somewhat with a shot that could not be easily
compared to the earlier fiducial. The following shot (3) ran full length
and clearly required much more gas than the fiducial (62 T-l). The
next two shots only required 36 and 33 T-l of gas. These values would
probably have been closer to the fiducial shot had the PO asked the
RF to be turned on at 800 rather than 600 ms. The conclusion is that
it takes only 1 or 2 discharges to reload the walls, even after a
very extended period of wall conditioning; probably just 1.

We also did two periods of between shot glow discharge in helium. Both
of these periods had 30 minutes of glow at up to 2.5 amps and 450 volts.
However, we were unable to get to low enough pressures to produce the
uniform, chamber filling, glow we had over the weekend. Following the
glow, we again ran the fiducial discharge, and in both cases 45 T-l
of gas was required. So, 30 minutes of glow is worth about 10 T-l of
gas. The second shot following the glow was right back to the 33 to
35 T-l value; recovery again with 1 shot. Had we more time, we would have
tried another period of glow lasting 60 minutes to see what the wall
inventory depletion rate was.

Finally, it was hoped that glow discharge cleaning in helium over the
weekend would improve the effects of the boronization and possibly
help the recovery process. I do think that the machine recovered
very quickly. The machine was no harder to run on Monday than it had
been the previous week. The carbon levels were very low, and the H/D
was in the 3 to 5% range, even after the last two run of the previous
week which ran very high hydrogen fraction discharges. However, the
radiated power was rather high, even somewhat worse than the fiducial
shot which was run well after the boronization of 5/30. The radiated
power did seem to be coming down during the run, and was much better
the following day after overnight ECDC in D2. The 1000627 run also
produced our lowest ever H/D ratios (1 to 3%). How much the wall
conditioning actually helped, if at all, will have to be determined
over the next few days as we run the machine.

Another run sometime in the distant future (September) would be
useful in which between shot glow was run with variations in the
length of glow time. We could then get a good measurement of
the depletion rate. It would also be useful to do target density
scans to see if we can reach a condition in which the density during
the h-mode actually decreases.

Physics Operator Summaries
Entered: Jul 7 2004 04:37:11:503PM
Author: To Be Determined
Engineering Setup for 1000626

MP 273: Effects of Glow Discharge Cleaning on Boronization and Wall Conditions

Boronization recovery as required

MP 262: Main Chamber Recycling

Session Leader: Irby/Pitcher
Physics Operator: Irby
Engineering Operators: Bill Cochran, G. Dekow, E. Fitzgerald

---------------------------------------------------------------------

Engineering Setup:

No power tests will be required. However one cycle test with pulsed
gas will be taken before the run.

Glow discharge cleaning should stop at 07:00

Run begins at 9:00 and ends at 17:00.

Power system setup: Same as for 1000616001

Set OH1, OH2U and OH2L fizzle delays at 400ms
Set TF fizzle delay at 500 ms


Gas setup: B-Top 6 psi D2 Hybrid Enabled
B-side lower 1 psi Argon Hybrid DISABLED
C-side 30 psi D2 Hybrid Enabled
J-bottom 30 psi D2 Hybrid Enabled

The following Gate Valves should be ENABLED

ECE, VUV


The first part of the run will involve operating the machine following
boronization and extensive glow discharge cleaning. We will run the
fiducial shot, 1000616001, at 1 MA. RF power at the 2 MW level or better
should be brought up. We will closely monitor the amount of gas required
to fuel the machine as a function of shot number. We hope to see a more
rapid recovery of the machine to good operational parameters. The second
part of the run (MP262) will begin as soon as we get reliable startups,
and we do a 30-45 minute test of between shot glow discharge cleaning.


Prerun conditioning:

On Saturday, 6/24, we deposited about 1600 Ang of diborane
onto the vessel walls using our standard ECDC technique.
ECDC in D2 was run Friday night into Saturday morning to
reduce the H2 levels in the machine following the Thursday
and Friday runs. Starting around 19:00 on Saturday we began
glow discharge cleaning in helium. After an initial breakdown
at 40 mT, we adjusted the feedback to keep us at about 10 mT
from Saturday until about 13:00 on Sunday. This period was used
as an initial glow conditioning time at relative high pressures
and low electrode voltage and current (200-300 volts, 200 mA
per electrode). On Saturday afternoon the current was brought up
to the 1 amp/electrode level and the voltage rose to 350 to
400 volts. The 400 volt level is the frequently stated level at which
the boron hydrides are reduced by the helium ions. Finally at
15:00 on Sunday we stopped the glow, and reduced the bias resistors from
720 ohms to 360 ohms. On restarting and after adjustments to the fill
pressure we were able to bring the total current up to 2.5 to 3.5 amps
at voltages ranging from 450 to 550 volts. This level was reached and
maintained until Monday morning at 07:00 in a virtually arc free
condition.


Results:

From the operational standpoint we had no problems running
a post boronization plasma today. We had only one fizzle. However,
the extensive glow cleaning over the weekend may have changed how
effective the boronization was. From RGA data it's clear we have
plenty of boron in the machine, but the radiated power levels were rather
high today. The RF was working well and delivered 3.1 MW of total power
from 3 antennas.
Because of the long delays caused by CAMAC and power supply
problems, and the time required to do the between shot glow, we did
not get to Spencer's run today. However, all the goals of MP273 were
accomplished (see session leader report).


Problems:

We had CAMAC problems at the beginning of the run. TCI was
not working until Joe put the highway back to last Friday's
configuration. He was trying to get the x-rays working.
We also lost 2PI diode, hybrid, and HiReX data. Rejean found
his Incaa had died and we didn't have a spare. A couple of
hours was spent in the power room testing the EF1U supply and then
doing a power test. No problem was found, but it was okay during
the power test and the rest of the run.

Lost trac data on shot 11. No efit

Score:

tests 02
duds 00
fizzles 01
plasmas 14

total 17


Details:

----------------------------------------------------------------------------
1000626xxx irby PHYSICS_OPERATOR Jun 26 2000 8:52:52:420AM
. . . . . . . . . . . . . . . . . . . .
Loaded shot 1000616001

Zeroed Br0 offset consistent with the last shot from Friday
Set EF4 to -1300 amps consistent with last shot from Friday

Did coldstart

rtnode -- data loaded, A matrix responds
rtnode -- data loaded, M matrix responds
rtnode -- data loaded, PID matrix responds

See Discharge Cleaning Entries for details of the helium glow discharge
cleaning done over the weekend. The glow was done following the
deposition of about 1600 Ang of diborane using ECDC on Saturday, 6/24.

The first shot today will be a no power cycle test to see what the
prefill gas level is like with the walls in their current condition.

----------------------------------------------------------------------------
1000626001 irby PHYSICS_OPERATOR Jun 26 2000 8:59:03:746AM
. . . . . . . . . . . . . . . . . . . .
This was a gas test.

Pressure at t=0 low: added 4 ms to prefill.
----------------------------------------------------------------------------
1000626002 irby PHYSICS_OPERATOR Jun 26 2000 9:42:53:780AM
. . . . . . . . . . . . . . . . . . . .
Added 4 ms to prefill

Just noted that TCI did not work on test shot, probably because of changes
to the CAMAC highway made yesterday.

Plasma for about 0.32 s

Late breakdown, fill high:

----------------------------------------------------------------------------
1000626003 irby PHYSICS_OPERATOR Jun 26 2000 5:15:10:153PM
. . . . . . . . . . . . . . . . . . . .
Reduced prefill by 2 ms.
Dropped early density request slightly (before 400 ms)

Joe switched CAMAC highway back to last weeks configuration.

Good full length shot with h-modes

TCI is back!

Looks like we took 20 Torr-l more gas for this shot compared to the
1000616001 fiducial.

Prefill still looks high, but it ran...

Carbon level is low
120 kJ of stored energy
H/D 3% rising to 5% during RF
----------------------------------------------------------------------------
1000626004 irby PHYSICS_OPERATOR Jun 26 2000 5:16:33:280PM
. . . . . . . . . . . . . . . . . . . .
Plasma to rampdown, 137 kJ, H/D 3 to 5%

Only took 36 Torr-l that shot: already back to standard operation

Pressure even higher
----------------------------------------------------------------------------
1000626005 irby PHYSICS_OPERATOR Jun 26 2000 10:46:56:233AM
. . . . . . . . . . . . . . . . . . . .
Note that RGA is showing lots of diborane as we expect, but also large
mass 69 peak.

Reduced prefill again by 2 ms

Plasma to rampdown, better h-modes, 143 kJ, RF cleaner but we did have
serious arc activity on e-port.

Only needed 35 T-l for that shot

Got prefill down a little

H/D is at the 5 or 6% level now

----------------------------------------------------------------------------
1000626006 irby PHYSICS_OPERATOR Jun 26 2000 11:08:59:780AM
. . . . . . . . . . . . . . . . . . . .
Reduced prefill another 2 ms

Fizzle.

shot: 1000626006


may have reduced prefill too much. Br0 getting too positive?

Several people enter cell to debug CAMAC problems
Amanda to ECE lab to adjust grating for current field
----------------------------------------------------------------------------
1000626007 irby PHYSICS_OPERATOR Jun 26 2000 12:42:56:310PM
. . . . . . . . . . . . . . . . . . . .
Added 2 ms back to prefill

Reduced Br0 by 0.2 mT

Took shot after about 55 minutes of delay with CAMAC fixes and
cell entries

Plasma into rampdown, nice h-mode, 150 kJ

2Pi diode is back: Radiation is worse than on 1000616 by 200-300 kW!

Took about 35 T-l of gas again

Br0 tweak seems to have worked
Gas okay
What is going on with Bz0?

Late breakdown though

Steve Wolfe noted 300 amp imbalance in the ef1 currents. Seems the ef1u
is not following demand. This problem probably explains toggling of Bz0
we have seen today and in the past.
----------------------------------------------------------------------------
1000626008 irby PHYSICS_OPERATOR Jun 26 2000 1:23:43:390PM
. . . . . . . . . . . . . . . . . . . .
Tried again.

Plasma for 200 ms

Still have problem with EF1s. Bill and Gary go to check.

----------------------------------------------------------------------------
1000626009 irby PHYSICS_OPERATOR Jun 26 2000 1:58:31:106PM
. . . . . . . . . . . . . . . . . . . .
We have done 30 minutes of glow at 500 volts, 2 amps. I was
unable to get the pressure low enough to produce a good glow
throughout the machine. We ran at about 8 mTorr. Any lower and
the electrode voltage would climb too high.

Bill and Gary found nothing wrong with EF1 supply.

This shot will be an EF1 test.

EF1s do not look very different in the test.

Next shot will be a plasma shot
----------------------------------------------------------------------------
1000626010 irby PHYSICS_OPERATOR Jun 26 2000 2:22:30:483PM
. . . . . . . . . . . . . . . . . . . .
Plasma to rampdown, 112 kJ

Took about 45 T-l to make the plasma --- glow may have mattered

Only had e-port RF, but still a good h-mode

EF1 looks fine

Lots of radiation on 2Pi: 1.3 MW

Still no HiReX


----------------------------------------------------------------------------
1000626011 irby PHYSICS_OPERATOR Jun 26 2000 2:51:31:623PM
. . . . . . . . . . . . . . . . . . . .
Plasma to rampdown

Took about 34 T-l


----------------------------------------------------------------------------
1000626012 irby PHYSICS_OPERATOR Jun 26 2000 2:53:08:593PM
. . . . . . . . . . . . . . . . . . . .
PCS Repeat

Plasma to rampdown, 33 t-l

H/D from 4 to 6%

Plasma into rampdown, but lost d-port RF again, 118 kJ

Took 33 T-l; so no change from last shot

----------------------------------------------------------------------------
1000626013 irby PHYSICS_OPERATOR Jun 26 2000 4:10:19:356PM
. . . . . . . . . . . . . . . . . . . .
Did another 30 minutes of glow in helium. Still not able to drop
pressure low enough to get a good uniform glow.

Elapsed time since last shot was about 1:05

Plasma to rampdown, took 45 T-l again, 152 kJ

Bz0 getting a little too negative
----------------------------------------------------------------------------
1000626014 irby PHYSICS_OPERATOR Jun 26 2000 4:30:28:403PM
. . . . . . . . . . . . . . . . . . . .
Dropped EF4 to -1280 from -1300

Plasma into rampdown, 168 kJ, 3 MW of RF, H to D at 5%

Radiation about 1.8 MW on 2Pi diode

Injection at .81 s probably caused d-port trip

----------------------------------------------------------------------------
1000626015 irby PHYSICS_OPERATOR Jun 26 2000 4:54:43:763PM
. . . . . . . . . . . . . . . . . . . .
Repeat

Plasma to rampdown, solid h-mode with very brief d-port RF interrupt

Took 34 T-l of gas

2Pi peaked at 1.89 but was going down during shot
----------------------------------------------------------------------------
1000626016 irby PHYSICS_OPERATOR Jun 26 2000 5:06:19:356PM
. . . . . . . . . . . . . . . . . . . .
Repeat

Plasma to rampdown, 180 kJ, H to D 4 to 6%

2Pi down to 1.7 MW peak with > 3 MW of RF

33 T-l of gas

----------------------------------------------------------------------------
1000626017 irby PHYSICS_OPERATOR Jun 26 2000 5:32:11:216PM
. . . . . . . . . . . . . . . . . . . .
Removed TF waggle for Rejean

Plasma to rampdown, but had many injections from e-port antenna

Radiated power down until the antenna started to act up

183 kJ, H to D 4 or 5%

Session Leader Comments
Jun 27 2000 08:59:08:780AM1000626003Spencer Pitcherright gap too large >>>>>> reduce next shot
no H-mode

Physics Operator Comments
Jun 26 2000 08:52:52:420AMJim IrbyLoaded shot 1000616001

Zeroed Br0 offset consistent with the last shot from Friday
Set EF4 to -1300 amps consistent with last shot from Friday

Did coldstart

rtnode -- data loaded, A matrix responds
rtnode -- data loaded, M matrix responds
rtnode -- data loaded, PID matrix responds

See Discharge Cleaning Entries for details of the helium glow discharge
cleaning done over the weekend. The glow was done following the
deposition of about 1600 Ang of diborane using ECDC on Saturday, 6/24.

The first shot today will be a no power cycle test to see what the
prefill gas level is like with the walls in their current condition.

Jun 26 2000 08:59:03:747AM1000626001Jim IrbyThis was a gas test.

Pressure at t=0 low: added 4 ms to prefill.
Jun 26 2000 09:42:53:780AM1000626002Jim IrbyAdded 4 ms to prefill

Just noted that TCI did not work on test shot, probably because of changes
to the CAMAC highway made yesterday.

Plasma for about 0.32 s

Late breakdown, fill high:
Time (ms): 0 5 10 25
----------------------------------------
Ip (kA): 2.1699 -8.2067 -18.5606 -114.5672
dIp/dt (kA/ms): -1.0184 -2.0335 -3.5606 -7.7570
pressure (mTorr): 0.0668 0.0675 0.0683 0.0564
Bz0 (mT): -3.8917 -1.5407 0.3038 19.2264
Br0 (mT): -0.3081 -0.9577 -0.5916 -1.9719

Jun 26 2000 05:15:10:153PM1000626003Jim IrbyReduced prefill by 2 ms.
Dropped early density request slightly (before 400 ms)

Joe switched CAMAC highway back to last weeks configuration.

Good full length shot with h-modes

TCI is back!

Looks like we took 20 Torr-l more gas for this shot compared to the
1000616001 fiducial.

Time (ms): 0 5 10 25
----------------------------------------
Ip (kA): 2.2154 -10.7782 -26.3658 -106.7847
dIp/dt (kA/ms): -1.2717 -2.8009 -3.4357 -8.2679
pressure (mTorr): 0.0668 0.0653 0.0675 0.0590
Bz0 (mT): -3.3258 -0.3721 1.4085 18.7939
Br0 (mT): -0.3330 -0.6437 -0.7864 -2.2774

Prefill still looks high, but it ran...

Carbon level is low
120 kJ of stored energy
H/D 3% rising to 5% during RF
Jun 26 2000 05:16:33:280PM1000626004Jim IrbyPlasma to rampdown, 137 kJ, H/D 3 to 5%

Only took 36 Torr-l that shot: already back to standard operation

Time (ms): 0 5 10 25
----------------------------------------
Ip (kA): -0.4015 -10.7554 -23.7262 -91.1971
dIp/dt (kA/ms): -1.2727 -2.5433 -3.3063 -6.3582
pressure (mTorr): 0.0722 0.0706 0.0722 0.0698
Bz0 (mT): -2.6649 -0.6366 0.7886 18.5044
Br0 (mT): -0.8198 -0.7528 -0.7980 -1.5098

Pressure even higher
Jun 26 2000 10:46:56:233AM1000626005Jim IrbyNote that RGA is showing lots of diborane as we expect, but also large
mass 69 peak.

Reduced prefill again by 2 ms

Plasma to rampdown, better h-modes, 143 kJ, RF cleaner but we did have
serious arc activity on e-port.

Only needed 35 T-l for that shot

Time (ms): 0 5 10 25
----------------------------------------
Ip (kA): 2.2154 -13.3496 -26.3204 -96.3627
dIp/dt (kA/ms): -1.5271 -2.7965 -3.3063 -6.1039
pressure (mTorr): 0.0668 0.0660 0.0675 0.0660
Bz0 (mT): -3.1602 -0.5830 1.1563 18.0803
Br0 (mT): -0.3708 -0.6407 -0.2059 -1.5199

Got prefill down a little

H/D is at the 5 or 6% level now

Jun 26 2000 11:08:59:780AM1000626006Jim IrbyReduced prefill another 2 ms

Fizzle.

shot: 1000626006

Time (ms): 0 5 10 25
----------------------------------------
Ip (kA): -0.4015 -13.3723 -28.9373 -18.5606
dIp/dt (kA/ms): -1.5282 -2.9226 -2.9259 0.2532
pressure (mTorr): 0.0603 0.0610 0.0603 0.0660
Bz0 (mT): -3.4843 -1.3232 0.4096 16.9044
Br0 (mT): 0.1166 -0.2905 -0.4154 -1.4200

may have reduced prefill too much. Br0 getting too positive?

Several people enter cell to debug CAMAC problems
Amanda to ECE lab to adjust grating for current field
Jun 26 2000 12:42:56:310PM1000626007Jim IrbyAdded 2 ms back to prefill

Reduced Br0 by 0.2 mT

Took shot after about 55 minutes of delay with CAMAC fixes and
cell entries

Plasma into rampdown, nice h-mode, 150 kJ

2Pi diode is back: Radiation is worse than on 1000616 by 200-300 kW!

Took about 35 T-l of gas again

Time (ms): 0 5 10 25
----------------------------------------
Ip (kA): 2.1926 -10.7554 -21.1548 -60.0899
dIp/dt (kA/ms): -1.2705 -2.2890 -2.1641 -5.0888
pressure (mTorr): 0.0631 0.0617 0.0624 0.0617
Bz0 (mT): -0.1096 1.9161 4.0471 19.7439
Br0 (mT): -0.3971 -0.7952 -1.3480 -1.5249

Br0 tweak seems to have worked
Gas okay
What is going on with Bz0?

Late breakdown though

Steve Wolfe noted 300 amp imbalance in the ef1 currents. Seems the ef1u
is not following demand. This problem probably explains toggling of Bz0
we have seen today and in the past.
Jun 26 2000 01:23:43:390PM1000626008Jim IrbyTried again.

Plasma for 200 ms

Still have problem with EF1s. Bill and Gary go to check.

Jun 26 2000 06:09:35:357PM1000626009Jim IrbyWe have done 30 minutes of glow at 500 volts, 2 amps. I was
unable to get the pressure low enough to produce a good glow
throughout the machine. We ran at about 8 mTorr. Any lower and
the eletrode voltage would climb too high.

Bill and Gary found nothing wrong with EF1 supply.

This shot will be an EF1 test.

EF1s do not look very different in the test.

Next shot will be a plasma shot
Jun 26 2000 02:22:30:483PM1000626010Jim IrbyPlasma to rampdown, 112 kJ

Took about 45 T-l to make the plasma --- glow may have mattered

Only had e-port RF, but still a good h-mode

EF1 looks fine

Lots of radiation on 2Pi: 1.3 MW

Still no HiReX


Time (ms): 0 5 10 25
----------------------------------------
Ip (kA): 2.2154 -8.1840 -23.7262 -114.5445
dIp/dt (kA/ms): -0.8924 -3.0531 -4.3225 -8.5200
pressure (mTorr): 0.0539 0.0552 0.0558 0.0498
Bz0 (mT): -2.1706 0.6757 2.0490 20.7540
Br0 (mT): -0.2923 -0.6124 -1.2448 -1.9476
Jun 26 2000 02:51:31:623PM1000626011Jim IrbyPlasma to rampdown

Took about 34 T-l


Jun 26 2000 02:53:08:593PM1000626012Jim IrbyPCS Repeat

Plasma to rampdown, 33 t-l

H/D from 4 to 6%

Plasma into rampdown, but lost d-port RF again, 118 kJ

Took 33 T-l; so no change from last shot

Jun 26 2000 04:10:19:357PM1000626013Jim IrbyDid another 30 minutes of glow in helium. Still not able to drop
pressure low enough to get a good uniform glow.

Elapsed time since last shot was about 1:05

Plasma to rampdown, took 45 T-l again, 152 kJ

Time (ms): 0 5 10 25
----------------------------------------
Ip (kA): -0.4015 -10.7782 -31.5314 -80.8204
dIp/dt (kA/ms): -1.0162 -3.4335 -2.5455 -6.9952
pressure (mTorr): 0.0583 0.0583 0.0583 0.0539
Bz0 (mT): -2.7483 -1.5168 0.0541 17.4267
Br0 (mT): -0.3291 -0.4670 -0.5012 -1.7993

Bz0 getting a little too negative
Jun 26 2000 04:30:28:403PM1000626014Jim IrbyDropped EF4 to -1280 from -1300

Plasma into rampdown, 168 kJ, 3 MW of RF, H to D at 5%

Radiation about 1.8 MW on 2Pi diode

Injection at .81 s probably caused d-port trip

Time (ms): 0 5 10 25
----------------------------------------
Ip (kA): 2.2381 -10.7327 -23.7489 -106.7620
dIp/dt (kA/ms): -1.1434 -2.7998 -3.9421 -6.9963
pressure (mTorr): 0.0583 0.0583 0.0597 0.0590
Bz0 (mT): -1.6447 0.1488 2.4201 19.0838
Br0 (mT): -0.5586 -0.5878 -1.2669 -2.1381
Jun 26 2000 06:08:36:043PM1000626015Jim IrbyRepeat

Plasma to rampdown, solid h-mode with very brief d-port RF interrupt

Took 34 T-l of gas

2Pi peaked at 1.89 but was going down during shot
Jun 26 2000 05:06:19:357PM1000626016Jim IrbyRepeat

Plasma to rampdown, 180 kJ, H to D 4 to 6%

2Pi down to 1.7 MW peak with > 3 MW of RF

33 T-l of gas


Time (ms): 0 5 10 25
----------------------------------------
Ip (kA): -0.3788 -13.3496 -28.9146 -101.5737
dIp/dt (kA/ms): -1.5271 -2.7976 -3.8150 -5.9756
pressure (mTorr): 0.0610 0.0610 0.0597 0.0603
Bz0 (mT): -2.1960 1.1165 1.4291 17.6579
Br0 (mT): -0.5737 -0.4770 -0.8426 -1.6799
Jun 26 2000 05:32:11:217PM1000626017Jim IrbyRemoved TF waggle for Rejean

Plasma to rampdown, but had many injections from e-port antenna

Radiated power down until the antenna started to act up

183 kJ, H to D 4 or 5%

Engineering Operator Comments
ShotTimeTypeStatusComment
108:49:52:200AMTestOkGas Test
209:33:47:263AMPlasmaOk
309:50:30:013AMPlasmaOk
410:10:26:843AMPlasmaOk
510:30:25:090AMPlasmaOk
610:51:20:623AMPlasmaOk
711:47:42:873AMPlasmaOk
812:12:23:090PMPlasmaBadEF1U Voltage Response Problem
901:42:41:340PMTestOkEF1U-L PLC, 5Kamp ,100V Test
1002:01:00:780PMPlasmaOk
1102:22:04:030PMPlasmaOk
1202:41:50:200PMPlasmaOk
1303:48:51:657PMPlasmaOk
1404:16:47:420PMPlasmaOk
1504:35:43:733PMPlasmaOk
1604:56:14:607PMPlasmaOk
1705:17:45:437PMPlasmaOk