Alcator C-Mod Run 1000906 Information

Back to runs

Miniproposals
Miniproposal:268
Date Filed: 4/7/2000
Title:Conditioning of D-, E-, or J-Port ICRF Antenna
First Author:Gerd Schilling
Session Leader:John Goetz (shots 1-23)

Operators
Session leader(s):John Goetz
Physics operator(s):Dmitri Mossessian
Engineering operator(s):Bill Cochran,Bill Byford

Engineering Operator Run Comment
MP 268 ICRF Conditioning. Goetz/Mossessian/Cochran/Byford.

Session Leader Plans

Physics Operators Plans

Session Leader Summaries
Entered: Jul 7 2004 03:49:24:757PM
Author: To Be Determined
Engineering setup for 1000906:

Overnight ECDC in D2, stopping 1-2 hours prior to the run.

Run begins at 9:00 and ends at 17:00.

Power system setup as on 1000901025 (0.8 MA, 5.2 T, inner-wall limited)

Gas setup - fill B-top with 6 psi of D2 Hybrid enabled
fill B-side lower with 1 psi of Ar Hybrid enabled
B-side upper should be set up for between-shot ECDC in D2
leave B-side upper as is Hybrid DISABLED
fill C-side with 30 psi of D2 Hybrid enabled
leave J-bottom as is Hybrid DISABLED

The following gate valves should be enabled, assuming no vacuum problems:
ECE, VUV

DNB gate valve will be DISABLED today, at least for the beginning of the run.

-------------------------------------------------------------------------------

Run plan for 1000906:

MP 268 -- ICRF Conditioning

Go back to the 0.8 MA inner-wall limited discharges from Friday and
continue to condition J-port.

-------------------------------------------------------------------------------

001 OH test

002 dud

003 plasma
D: 0.6 MW
E: 0.6 MW
J: 0.6 MW

phase was off because we changed the reference

next shot: change the phase of J-port by 0.5

004 plasma
D: 0.6 MW
E: 0.6 MW
J: 0.6 MW

phase was off wrapped the other direction
[0,1,3/4,-.25]

next shot: change the phase of J-port by 0.5

005 plasma - early disruption 1.06
D: 0.6 MW
E: 0.6 MW
J: 0.6 MW

phase improved [0,1,0.1,1.1]

next shot: change the phase of J-port by 0.1 pi
decrease density by 10% hopefully to avoid marfe

006 plasma
D: 0.6 MW
E: 0.6 MW
J: 0.6 MW
good comparison shot except
J-port evolves significant gas

next shot: shorten the pulse to 100 msec and increase the power by 200 kW
overlap D and E in the back of the pulse

007 plasma
D: 0.8 MW
E: 0.8 MW
J: 0.8 MW
marfe again

next shot: turn off ninja

008 plasma
D: 0.8 MW
E: 0.8 MW
J: 0.8 MW
good comparison shot
J is significantly less than D and E in stored energy

next shot: 1 MW each

009 dud
010 fizzle
011 fizzle

012 plasma
D: 1 MW
E: 1 MW
J: 1 MW
J3: 1 MW with two faults - glow from corner
J4: 1 MW with two faults - glow from corner

good comparison shot
poor heating effectiveness from J3 and J4 alone

013 plasma
D: 1.2 MW
E: 1.2 MW
J: 1.2 MW - beginning to see glow from corners
J3: 1 MW with multiple faults - glow from corner
J4: 1 MW - glow from corner

good comparison shot
poor heating effectiveness from J3 and J4 alone

next shot: increase to 1.5 MW each
overlap D and E in the latter part of pulse -> 3 MW
014 fizzle

015 plasma
D: 1.5 MW
E: 1.5 MW
J: 1.5 MW
good comparison shot
D and E - 3 MW with some H-modes? limited?

next shot: lengthen J-port to 0.5 msec

016 plasma
D: 1.5 MW
E: 1.5 MW
J: 1.5 MW
good comparison shot
D and J - 3 MW with some H-modes? limited?
250 msec before corners heated up

next shot: D-port to 0.6 sec

017 plasma - early disruption
D: 1.3 MW
E: 1.3 MW
J: 1.5 MW

200 msec before corners overheat

next shot: go diverted

018 plasma
D: 1.25 MW max faults after 1.1 sec
E: 1.25 MW
J: 1.5 MW with H-modes with faults


next shot: raise field to 5.5 T and reduce power to 1.05

019 plasma EDA
D: 1.05 MW single fault
E: 1.25 MW
J: 1.5 MW with H-modes with faults


next shot: raise field to 5.9 T

020 plasma EDA
D: 1.05 MW single fault
E: 1.25 MW single fault
J: 1.5 MW with H-modes with faults
BT~5.8 T

next shot: raise field to 6.0 T

021 abort and test shot for the alternator

022 alternator test shot

023 plasma EDA
D: 1.05 MW single fault
E: 1.25 MW no faults
J: 1.5 MW with no H-modes with faults; lots of injections
BT~6.05

Summary:

We continued to investigate the operational characteristics of J-port
antenna. With short pulses, we had several comparison discharges at
various power levels (up to 1.5 MW). In terms of the impurity
injections, the performance improved compared to last friday. We
obtained ~0.25 sec pulses at 1.5 MW before injections from the corners
occurred. These injections are largely N and Ti (Terry). The
improvement may be a result of conditioning; however, the rate of
improvement is quite slow. More importantly the heating efficiency is
still lower for J than D and E antennas. From 1000906012 and 015,
J-port heating efficiency was ~50% versus ~75% for D and E-port. This
heating efficiency is little better than those measured in June.

We were able to operate D and E-port antennas at fields up to 6.05 T
reliably in diverted, H-mode plasmas. Before the run we had installed
notch filters to remove cross coupled power from the arc detectors.
No density barrier was observed as is the case for 4.5 T operation.
In the case of the 6 T discharge, the H cyclotron resonance is on the
low field side, ~10 cm from the plasma center. For 4.5 T, the H
cyclotron resonance is on the high field side, ~10 cm from the plasma
center. We injected a maximum of 3 MW into inner wall limited plasmas
and obtained short lived H-modes.

To summarize the ICRF operational status, the D and E-port antennas
are operational. These systems still have a few measurement problems
that we will be sorting out in the coming weeks. The primary issue is
the power measurements. The power measurements are not consistent for
D-port when comparing all the directional couplers and E-port may be
systematicly reporting ~20% high at the FPA output (used for feedback
and net power). We are currently installing calibrated and more
reliable power meters in an effort to obtain reliable power
measurements. J-port as a four-strap antenna is not ready for
physics experiments. It has lower heating efficiency and an impurity
generation problem. Switching to a 2-strap antenna will occur over
the next few days.

Physics Operator Summaries
Entered: Jul 7 2004 04:37:11:927PM
Author: To Be Determined
Physop summary for 1000906: MP268 - ICRF conditioning

Plan was: Do 0.8 MA inner wal limited discharges and continue conditioning.

Summary:

Not a bad day from PO point of view. Machine ran reasonably well, we had nice limited plasmas.
It turned out that the density was affected significantly by the He puff (fluctuations imaging thing) and
we couldn't get below 9e19 m^-2 target untill we turned the puff off. After shot 8 alternator was slowed down
which resulted in minor changes in OH voltages. A dud followed, and then alternating plasmas/fizzles, until
the OH1 and OH2 voltages were adjusted to compensate for the change ( after shot 14).
We turned to diverted plasmas after shot 17. Nice EDA H-modes were obtained at Bt =5.2 ,5.45 and 5.8 T.
Alternator lost its excitation voltage on shot 21. After couple of test shots, it was decided that slowing down the alternator
is what caused the problem. The alternator was sped up to 1645 RPM, and the next test was succesful.
The last shot 23 was a good 6.05T, 0.8 MA shot with 2.3 MW RF and two EDAs

23 shots
1 dud
3 fizzles
4 tests
15 plasmas
----------------------------------------------------------------------------
1000906xxx mossessian PHYSICS_OPERATOR Sep 6 2000 9:02:01:590AM
. . . . . . . . . . . . . . . . . . . .
Engineering setup for 1000906:

Overnight ECDC in D2, stopping 1-2 hours prior to the run.

Run begins at 9:00 and ends at 17:00.

Power system setup as on 1000901025 (0.8 MA, 5.2 T, inner-wall limited)

Gas setup - fill B-top with 6 psi of D2 Hybrid enabled
fill B-side lower with 1 psi of Ar Hybrid enabled
leave B-side upper as is Hybrid DISABLED
fill C-side with 30 psi of D2 Hybrid enabled
leave J-bottom as is Hybrid DISABLED

The following gate valves should be enabled, assuming no vacuum problems:
ECE, VUV

DNB gate valve will be DISABLED today, at least for the beginning of the run.
____________________________________________________

8:50 Loaded 1000901025. Coldstart OK

First shot will be OH test

----------------------------------------------------------------------------
1000906001 OH test OK

Next shot: Plasma
-----------------------------------------------------------------------------
1000906002 EF1_upper power supply problem.
-----------------------------------------------------------------------------
1000906003 Plasma. H/D ratio is about 3%. We are not going to do ECDC between shots
unless it gets to above 5%. Plasma is not limited (according to EFIT) between
.56 and 1 sec.

For next shot: drop CLEARIN more to make it limited.
-----------------------------------------------------------------------------
1000906004 Nice limited plasma up to 1.5 sec, with right gap around 1 cm throughout.

Next shot: no PCS changes
-----------------------------------------------------------------------------
1000906005 Plasma. Disrupted at 1.06 sec because of an injection.

Next shot: Drop target to 0.9e20.
-----------------------------------------------------------------------------
1000906006 Plasma.

For next shot: no changes
-----------------------------------------------------------------------------
1000906007 Plasma.

For next shot: lower density to 0.8e20
-----------------------------------------------------------------------------
1000906008 Plasma. Density is around 0.7e20. It looks like the He puff was increasing the
density a lot and did not allow it to go below 0.9e20. It was turned off for
this shot and the situation improved. We'll keep it off for the rest of the day.

For next shot: slow down density ramp-up somewhat

Alternator has some kind of vibration problem - it was slowed down to 1632 rpm
after this shot
-----------------------------------------------------------------------------
1000906009 Dud. Low prepuff ?

For next shot: increase prepuff to 24 msec (from 20)

Start ECDC between shots
----------------------------------------------------------------------------
1000906010 Fizzle. The gas went up a lot 0.062 -> 0.086. Late breakdown. Bz looks too high

For next shot: Drop prepuff to 22 msec, lower Bz -1292 -> -1305 of EF4U
----------------------------------------------------------------------------
1000906011 Fizzle.

For next shot: prepuff 22 msec -> 20 msec, Bz -1305 -> -1320 of EF4U
----------------------------------------------------------------------------
1000906012 Plasma

For next shot: no changes
----------------------------------------------------------------------------
1000906013 Plasma

For next shot: no changes
----------------------------------------------------------------------------
1000906014 FIzzle. No apparent reason.

Next shot: Twicked OH1 and OH2 trying to compensate for changes of
voltage due to alternator slowing down.
----------------------------------------------------------------------------
1000906015 Plasma. Late breakdown (0.007)

For next shot: lower Bz -1320 -> -1335
----------------------------------------------------------------------------
1000906016 Plasma. Breakdown at 0.056.

Next shot: no changes
___________________________________________________________________________________
1000906017 Plasma. Disrupted at 1.25 sec, injection.

Next shot: changed CLEARIN and RCUR to get diverted plasma
----------------------------------------------------------------------------
1000906018 Plasma. Nice diverted shot with 2.6 MW out of D and E (D trips a lot)
Long steady-state EDA.

For next shot: Bt to 5.5 T (TF -144700 -> -153000)
----------------------------------------------------------------------------
1000906019 Plasma. Long steady-state EDA, RF 2.5 MW, Bt 5.44 T

For next shot: Bt to 5.9 T (TF -153000 -> -165600)
----------------------------------------------------------------------------
1000906020 Plasma. Long steady-state EDA, RF 2.3 MW, Bt 5.8 T

For next shot: Bt to 6 T (TF -165600 -> -171000)
----------------------------------------------------------------------------
1000906021 Shot abborted because of alternator problems - no excitation voltage.
Next shot: no power CAMAC cycle.
----------------------------------------------------------------------------
Same alternator problem

Next shot: No power alternator test
----------------------------------------------------------------------------
1000906022 Test OK

Next shot: Try to get some plasma.
________________________________________________________________________
1000906022 PLasma, 2.3 MW RF, 6.05 T, EDA



Session Leader Comments
Sep 6 2000 09:41:20:153AMJohn GoetzEngineering setup for 1000906:

Overnight ECDC in D2, stopping 1-2 hours prior to the run.

Run begins at 9:00 and ends at 17:00.

Power system setup as on 1000901025 (0.8 MA, 5.2 T, inner-wall limited)

Gas setup - fill B-top with 6 psi of D2 Hybrid enabled
fill B-side lower with 1 psi of Ar Hybrid enabled
leave B-side upper as is Hybrid DISABLED
fill C-side with 30 psi of D2 Hybrid enabled
leave J-bottom as is Hybrid DISABLED

The following gate valves should be enabled, assuming no vacuum problems:
ECE, VUV

DNB gate valve will be DISABLED today, at least for the beginning of the run.

-------------------------------------------------------------------------------

Run plan for 1000906:

MP 268 -- ICRF Conditioning

Go back to the 0.8 MA inner-wall limited discharges from this past
Wednesday and continue to condition J-port.
Sep 6 2000 09:05:19:030AM1000906001Steve Wukitch001 OH test
Sep 6 2000 09:29:01:653AM1000906002Steve Wukitch002 dud
Sep 6 2000 10:07:31:653AM1000906003John GoetzShot 003:
0.6MW from D-port at 1.2 to 1.45s
0.6MW from E-port at 0.9 to 1.15s
0.3MW from both J-ports at 0.6 to 0.85s

Results:
0.62MW from D-port; no faults
0.66MW from E-port; no faults
0.29MW from J3; no faults
0.31MW from J4; no faults
phase was [0,1,-0.6,0.4]*!pi
core Te and neutron rate much lower with this phase
Sep 6 2000 10:17:30:590AM1000906004John GoetzShot 004:
0.6MW from D-port at 1.2 to 1.45s
0.6MW from E-port at 0.9 to 1.15s
0.3MW from both J-ports at 0.65 to 0.85s
change phase from 195 to 310

Results:
0.61MW from D-port; no faults
0.64MW from E-port; no faults
0.29MW from J3; no faults
0.32MW from J4; no faults
phase was [0,1,0.75,-0.25]*!pi; went the wrong way with the phase change
core Te and neutron rate much lower with this phase; lots of injections
Sep 6 2000 11:09:33:950AM1000906005John GoetzShot 005:
0.6MW from D-port at 1.2 to 1.45s
0.6MW from E-port at 0.9 to 1.15s
0.3MW from both J-ports at 0.65 to 0.85s
change phase from 310 to 85

Results:
early disruption from a moly injection
0.64MW from E-port; one fault at disruption
0.29MW from J3; no faults
0.32MW from J4; no faults
phase was [0,1,-0.1,0.9]*!pi
Te, neutron rate, Wmhd are comparable between E and J
Sep 6 2000 11:09:49:437AM1000906006John GoetzShot 006:
lower density to 0.9(nl04) to get rid of MARFE
0.6MW from D-port at 1.2 to 1.45s
0.6MW from E-port at 0.9 to 1.15s
0.3MW from both J-ports at 0.65 to 0.85s
change phase from 85 to 72

Results:
0.61MW from D-port; no faults
0.64MW from E-port; no faults
0.29MW from J3; no faults
0.32MW from J4; no faults
phase was [0,1,-0.1,0.9]*!pi
Te, neutron rate, Wmhd are comparable between J and D,E
Sep 6 2000 11:08:17:560AM1000906007John GoetzShot 007:
0.8MW from D-port at 1.05 to 1.15s and 0.5MW from 1.2 to 1.45s
0.8MW from E-port at 0.75 to .85s and 0.5MW from 1.2 to 1.45s
0.4MW from both J-ports at 0.9 to 1.0s
change phase from 72 to 65

Results:
0.84 and 0.51MW from D-port; no faults
0.84 and 0.55MW from E-port; no faults
0.38MW from J3; no faults
0.43MW from J4; no faults
phase was [0,1,0,1]*!pi
Te, neutron rate, Wmhd are with J than with D and E; injections coming
from the Faraday screen; no hot spots in the corners
Sep 6 2000 11:29:08:263AM1000906008John GoetzShot 008:
0.8MW from D-port at 1.05 to 1.15s and 1.0MW from 1.2 to 1.45s
0.8MW from E-port at 0.75 to .85s and 1.0MW from 1.2 to 1.45s
0.4MW from both J-ports at 0.9 to 1.0s


Results:
0.79 and 0.96MW from D-port; no faults
0.85 and 1.04MW from E-port; no faults
0.37MW from J3; no faults
0.41MW from J4; no faults
phase was [0,1,0,1]*!pi
Te, neutron rate, Wmhd are lower with J than with E (D is at lower density);
no difference in radiation; no hot spots in the corners
Sep 6 2000 11:39:16:280AM1000906009John GoetzShot 009:
1.0MW from D-port at 1.05 to 1.15s and 1.25MW from 1.2 to 1.45s
1.0MW from E-port at 0.75 to .85s and 1.25MW from 1.2 to 1.45s
0.5MW from both J-ports at 0.9 to 1.0s

Results:
Dud
Sep 6 2000 12:02:48:420PM1000906010John GoetzShot 010:
1.0MW from D-port at 1.05 to 1.15s and 1.25MW from 1.2 to 1.45s
1.0MW from E-port at 0.75 to .85s and 1.25MW from 1.2 to 1.45s
0.5MW from both J-ports at 0.9 to 1.0s

Results:
Fizzle
Sep 6 2000 12:56:29:090PM1000906011John GoetzShot 011:
1.0MW from D-port at 1.05 to 1.15s and 1.25MW from 1.2 to 1.45s
1.0MW from E-port at 0.75 to .85s and 1.25MW from 1.2 to 1.45s
0.5MW from both J-ports at 0.9 to 1.0s

Results:
Fizzle
Sep 6 2000 12:59:04:280PM1000906012John GoetzShot 012:
1.0MW from D-port at 1.05 to 1.15s
1.0MW from E-port at 0.75 to .85s
0.5MW from both J-ports at 0.9 to 1.0s; 1.0MW from J3 at 1.2 to 1.3s; 1.0MW from J4 at 1.35 to 1.45s


Results:
1.08MW from D-port; no faults
1.06MW from E-port; no faults
0.48MW from J3; no faults - 0.97MW from J3; two faults
0.54MW from J4; no faults - 1.09MW from J4; one fault
Te, neutron rate, Wmhd are lower with J than with D and E;
one injection but not mudh difference in radiation; no hot spots in the corners
with only 0.5MW per strap pair; the H-port corners hot spots light up with
J3 running at 1.0MW and the K-port corners light up with J4 running
Sep 6 2000 01:23:39:543PM1000906013John GoetzShot 013:
1.2MW from D-port at 1.05 to 1.15s
1.2MW from E-port at 0.75 to .85s
0.6MW from both J-ports at 0.9 to 1.0s; 1.0MW from J3 at 1.2 to 1.3s; 1.0MW from J4 at 1.35 to 1.45s


Results:
1.29MW from D-port; no faults
1.21MW from E-port; no faults
0.57MW from J3; no faults - 0.95MW from J3; three faults
0.63MW from J4; no faults - 1.04MW from J4; no faults
Te, neutron rate, Wmhd are slightly lower with J than with E; not much difference in
radiation for the three pulses; D is the best antenna in this bunch
At the high power with single strap pairs, there are lots of injections and not much heating
Sep 6 2000 02:18:36:420PM1000906015John GoetzShot 015:
1.5MW from D-port at 1.05 to 1.15s; 1.5MW from D-port at 1.2 to 1.45s
1.5MW from E-port at 0.75 to .85s; 1.5MW from E-port at 1.2 to 1.45s
0.75MW from both J-ports at 0.9 to 1.0s

Results:
1.6MW from D-port; no faults - 1.6MW; four faults
1.57MW from E-port; no faults - 1.55MW; no faults
0.73MW from J3; no faults
0.8MW from J4; no faults
central Te (.85 vs 1 delta), neutron rate (1.1 vs 1.3 delta), Wmhd (14 vs 18 kJ delta) are slightly lower
with J than with E; not much difference in radiation for the three pulses
no glow in the corners on this one; a small injection from the screen @ 0.98s

Sep 6 2000 02:37:49:560PM1000906016John GoetzShot 016:
1.5MW from D-port at 1.05 to 1.15s
1.5MW from E-port at 0.75 to .85s
0.75MW from both J-ports at 0.9 to 1.4s

Results:
1.6MW from D-port; two faults (due to the limiter H-mode)
1.55MW from E-port; no faults
0.73MW from J3; seven faults
0.8MW from J4; three faults
J has slightly less performance than E (before D turns on)
glow in the corners on this one, starting after about 1.15s (Chromex nitrogen signal rises dramatically
at this time; boron signal is up for all of the pulse)
many injections from the Faraday screen
Sep 6 2000 03:00:53:920PM1000906017John GoetzShot 017:
1.3MW from D-port at 0.6 to 0.7s
1.3MW from E-port at 0.75 to .85s
0.75MW from both J-ports at 0.9 to 1.4s

Results:
1.4MW from D-port; no faults
1.3MW from E-port; no faults
0.73MW from J3; one fault
0.8MW from J4; no faults
early disruption from J this time
J has slightly less performance than E (and D)
glow in the corners on this one, starting after about 1.15s (Chromex nitrogen signal rises dramatically
at this time as does the boron signal)
many injections from the Faraday screen
looks like an energy problem - run longer pulses and get hot spots after ~0.25s
Sep 6 2000 03:18:06:090PM1000906018John GoetzShot 018:
reload shot 1000830025, a diverted 1MA discharge
1.3MW from D-port at 0.6 to 1.25s
1.25MW from E-port at 0.6 to 1.25s
0.75MW from both J-ports at 0.9 to 1.4s

Results:
diverted plasma; 800kA; 5.2T; H-mode
1.3MW from D-port; max faults (all DC0)
1.3MW from E-port; no faults
0.75MW from J3; one fault
0.75MW from J4; one faults
two brief H-modes with J alone; there is a glow in the upper H-port corner
Sep 6 2000 03:36:58:560PM1000906019John GoetzShot 019:
raise toroidal field to 5.5T
tune up #3 and #4
1.05MW from D-port at 0.6 to 1.25s
1.25MW from E-port at 0.6 to 1.25s
0.75MW from both J-ports at 1.3 to 1.45s

Results:
diverted plasma; 800kA; 5.45T; H-mode
1.1MW from D-port; one fault at L->H transition
1.3MW from E-port; no faults
0.7MW from J3; one fault at turn-on
0.7MW from J4; no faults
one H-mode with J alone; there is still a glow in the upper H-port corner
Sep 6 2000 04:00:01:060PM1000906020John GoetzShot 020:
raise toroidal field to 6.0T
tune up all antennas
1.05MW from D-port at 0.6 to 1.25s
1.25MW from E-port at 0.6 to 1.25s
0.75MW from both J-ports at 1.3 to 1.45s

Results:
diverted plasma; 800kA; 5.8T; H-mode
1.1MW from D-port; one fault late
1.2MW from E-port; one fault late
0.7MW from J3; two faults
0.7MW from J4; two faults
no H-mode with J alone; no glow in the upper H-port corner but plenty of injections from the screen
Sep 6 2000 05:16:08:997PM1000906023John GoetzShot 023:
raise toroidal field to 6.0T
tune up all antennas
1.05MW from D-port at 0.6 to 1.25s
1.25MW from E-port at 0.6 to 1.25s
0.75MW from both J-ports at 1.3 to 1.45s

Results:
diverted plasma; 800kA; 6.05T; 2 H-modes
1.1MW from D-port; one fault in the middle; tune off a bit
1.3MW from E-port; no faults; tune off as the shot progresses
big injection @ 0.88s that eventually causes the H->L
0.7MW from J3; three faults
0.7MW from J4; one fault
no H-mode with J alone; but plenty of injections from the screen (Ti observed with the McPherson)

Physics Operator Comments
Sep 6 2000 09:02:01:590AMDmitri MossessianEngineering setup for 1000906:

Overnight ECDC in D2, stopping 1-2 hours prior to the run.

Run begins at 9:00 and ends at 17:00.

Power system setup as on 1000901025 (0.8 MA, 5.2 T, inner-wall limited)

Gas setup - fill B-top with 6 psi of D2 Hybrid enabled
fill B-side lower with 1 psi of Ar Hybrid enabled
leave B-side upper as is Hybrid DISABLED
fill C-side with 30 psi of D2 Hybrid enabled
leave J-bottom as is Hybrid DISABLED

The following gate valves should be enabled, assuming no vacuum problems:
ECE, VUV

DNB gate valve will be DISABLED today, at least for the beginning of the run.
____________________________________________________

8:50 Loaded 1000901025. Coldstart OK

First shot will be OH test

Sep 6 2000 09:07:07:513AM1000906001Dmitri MossessianEngineering setup for 1000906:

Overnight ECDC in D2, stopping 1-2 hours prior to the run.

Run begins at 9:00 and ends at 17:00.

OH test OK

Next shot: Plasma
Sep 6 2000 09:32:25:903AM1000906002Dmitri MossessianEF1_upper power supply problem.
Sep 6 2000 10:09:58:250AM1000906003Dmitri MossessianPlasma. H/D ratio is about 3%. We are not going to do ECDC between shots
unless it gets to above 5%. Plasma is not limited (according to EFIT) between
.56 and 1 sec.

For next shot: drop CLEARIN more to make it limited.
Sep 6 2000 10:16:38:200AM1000906004Dmitri MossessianNice limited plasma up to 1.5 sec, with right gap around 1 cm throughout.

Next shot: no PCS changes
Sep 6 2000 10:35:42:153AM1000906005Dmitri MossessianPlasma. Disrupted at 1.06 sec because of an injection.

Next shot: Drop target to 0.9e20.
Sep 6 2000 10:46:35:997AM1000906006Dmitri MossessianPlasma.

For next shot: no changes
Sep 6 2000 11:03:23:357AM1000906007Dmitri MossessianPlasma.

For next shot: lower density to 0.8e20
Sep 6 2000 04:28:02:733PM1000906008Dmitri MossessianPlasma. Density is around 0.7e20. It looks like the He puff was increasing the
density a lot and did not allow it to go below 0.9e20. It was turned off for
this shot and the situation improved. We'll kep it off for the rest of the day.

For next shot: slow down density ramp-up somewhat

Alternator has some kind of vibration problem - it was slowed down to 1632 rpm
after this shot
Sep 6 2000 11:47:16:437AM1000906009Dmitri MossessianDud. Low prepuff ?

For next shot: increase prepuff to 24 msec (from 20)

Start ECDC between shots
Sep 6 2000 12:05:30:953PM1000906010Dmitri MossessianFizzle. The gas went up a lot 0.062 -> 0.086. Late breakdown. Bz looks too high

For next shot: Drop prepuff to 22 msec, lower Bz -1292 -> -1305 of EF4U
Sep 6 2000 12:23:53:123PM1000906011Dmitri MossessianFizzle.

For next shot: prepuff 22 msec -> 20 msec, Bz -1305 -> -1320 of EF4U
Sep 6 2000 01:26:52:577PM1000906012Dmitri MossessianPlasma

For next shot: no changes
Sep 6 2000 01:26:57:937PM1000906013Dmitri MossessianPlasma

For next shot: no changes
Sep 6 2000 01:54:41:530PM1000906014Dmitri MossessianFIzzle. No apparent reason.

Next shot: Twicked OH1 and OH2 trying to compensate for changes of
voltage due to alternator slowing down.
Sep 6 2000 02:06:29:840PM1000906015Dmitri MossessianPlasma. Late breakdown (0.007)

For next shot: lower Bz -1320 -> -1335
Sep 6 2000 02:23:48:140PM1000906016Dmitri MossessianPlasma. Breakdown at 0.056.

Next shot: no changes
Sep 6 2000 02:47:54:840PM1000906017Dmitri MossessianPlasma. Disrupted at 1.25 sec, injection.

Next shot: no changes
Sep 6 2000 02:59:00:607PM1000906017Dmitri MossessianPlasma. Disrupted at 1.25 sec, injection.

Next shot: changed CLEARIN and RCUR to get diverted plasma
Sep 6 2000 03:16:02:170PM1000906018Dmitri MossessianPlasma. Nice diverted shot with 2.6 MW out of D and E (D trips a lot)
Long steady-state EDA.

For next shot: Bt to 5.5 T (TF -144700 -> -153000)
Sep 6 2000 03:31:54:827PM1000906019Dmitri MossessianPlasma. Long steady-state EDA, RF 2.5 MW, Bt 5.44 T

For next shot: Bt to 5.9 T (TF -153000 -> -165600)
Sep 6 2000 03:55:56:090PM1000906020Dmitri MossessianPlasma. Long steady-state EDA, RF 2.3 MW, Bt 5.8 T

For next shot: Bt to 6 T (TF -165600 -> -171000)
Sep 6 2000 04:13:00:247PM1000906021Dmitri MossessianShot abborted because of alternator problems - no excitation voltage.
Next shot: no power CAMAC cycle.
Sep 6 2000 04:42:09:497PM1000906021Dmitri MossessianSame alternator problem

Next shot: No power alternator test
Sep 6 2000 04:46:24:983PM1000906022Dmitri MossessianTest OK

Next shot: Try to get some plasma.

Engineering Operator Comments
ShotTimeTypeStatusComment
109:02:53:373AMTestOkOH Power Supply Test
209:21:13:140AMPlasmaBadEF1U did not fire
309:51:27:090AMPlasmaOk
410:09:59:420AMPlasmaOk
510:22:55:140AMPlasmaOk
610:39:28:153AMPlasmaOk
710:54:27:123AMPlasmaOk
811:11:16:733AMPlasmaOk
911:34:56:140AMPlasmaOk
1011:57:51:840AMPlasmaOk
1112:17:52:327PMPlasmaOk
1212:39:39:670PMPlasmaOk
1301:02:56:483PMTestOk
1401:28:02:780PMTestOk
1501:56:04:373PMTestOk
1602:17:24:107PMTestOk
1702:37:38:747PMTestOk
1803:00:40:390PMTestOk
1903:22:13:467PMPlasmaOk
2003:44:46:937PMPlasmaOk
2104:07:32:170PMTestBadNO EXCITATION VOLTAGE
2204:42:52:247PMTestOkAlternator speed increased and excitation came on.
2304:56:33:733PMPlasmaOk