| Session leader(s): | Bruce Lipschultz |
| Physics operator(s): | Jim Irby |
| Engineering operator(s): | Bill Cochran |
| Engineering Operator Run Comment |
| Shot Cycle Gas System Tests |
| Session Leader Plans |
| Entered: Feb 25 2005 08:22:38:847AM |
| Author: Bruce Lipschultz |
| MP364 - sort of.
The idea is to check out the RGA and other data (MKS gauges, IGT, IGP gauges, valves statuses) in terms of their archiving as part of the before/after shot deposition in the tree. We also hope to check whether the trend data, archived each midnight is also being archived properly. The plan is to inject both Ar and D2 into the machine at the time of the shot and keep the gate valves closed for the standard time - 5 minutes. Intitially, we will have the gate valves closed for 6 minutes and see whether that is enough. |
| Physics Operators Plans |
| Entered: Feb 25 2005 07:37:02:830AM |
| Author: Jim Irby |
Overnight ECDC in D2, bake at 60C Engineering setup: Gas setup: ----------- > fill B-side lower with 10 psi Ar Hybrid enabled (PG1) > fill (former) C-side with 30 psi D2 Hybrid Enabled (PG3) Torvac Setup: -------------- GV1 closed, Bypass to RGA open at 180 clicks Torus gate valves closed in Pulse for 5min after shot |
| Session Leader Summaries |
| Entered: Feb 25 2005 03:53:06:230PM |
| Author: Bruce Lipschultz |
| 1050225 - trying to undestand the before/after data
We made progress today. The before and after data appears to be being written. The issues we encccountered are: 1) When the dwell time of the rga/mass is reduced to 15 ms (standard during a run) we continually get 'checksum' errors indicating a communication error from the RGA in the cell to the PC in the control room. This checksum error apparently forces the spectrum to start over again. Thus, if it occurs before the end of a spectrum trace continually one never gets the higher masses (e.g 40 as we were injecting Ar and D2). The solution was to lengthen the dwell to 120 ms. This seemed to reduce the error frequency, but not to zero frequency. The spectra took longer, somewhere between 1.5 and 2 minutes. This necessitated moving the triggers to far enough after the shot (5 minutes) and before (at check, but with 4 minutes of gate valve closing before pulse to allow for 2 full scans). We are assuming 2 full scans are needed becasue if the pulse (or gate valve closure before the shot) occurs near the beginning of a scan, we will want to wait for that scan to finish and a full one before acquiring the data. 2) The files are reversed. We do see that the filenames appearing on the PC have names - TRIGG000.DAT and TRIGG001.DAT. Today it looks like the after data was appearing in the 000 file. This correlates with the reversal in the tree. This may be due to the MDS storage occurring before the after data is taken. Thus, the after data sits around till the next shot and is paired with the before daata of that shot. We need to make sure that the MDS archiving is timed properly. In addition it would be good if the MDS archiving checked the times of the files and deleted the files before the shot (at INIT). This also might eliminate something that Jim Irby has noticed - namely that there are 2 mass spectra, instead of 2, for the 1st shot of the day. 3) The pressure data has some minor issues. It turns out that the mantissa of the IGP and IGT data is being stuck into both the mantissa and exponent when stored in the tree. We have comunicated with TWF about this and have suggested a solution. There is also an issue wwith offsets in the MKS data injected by the AFOLs. The amount, approx. .206 V, translates into 2.06 mT for G-side (multiply the voltage by 10 to get mT) and 20.6 mT for B_Bot MKS (x 100 to get mT). We can live with this for now but it would be good to get this fixed. 4) Potential probblem with other processes going on. During runs the standard RGA archiving ('midnight' data) is set to occur every minute. This could be leading to further confusing of the RGA. For now I recommend that this be stopped or kept at 15 minute intervals, during the run. More tests could be run. In any case it would be good to see if the checksum errors could be reduced to 0. That might help considerably. |
| Physics Operator Summaries |
| Entered: Feb 25 2005 03:21:46:310PM |
| Author: Jim Irby |
| Found many problems with RGA triggering and gatevalve operation. Shots 12 and 14 seem to have the correct before/after data but reversed in the expected order: m(0, 0:100) gives the after data m(1, 0:100) gives the before data Looks like we will have to manually close the gatevalves to get the right before data. See session leader summary for more details. |
| Session Leader Comments | |||
| Feb 25 2005 08:22:35:127AM | 1050225001 | Bruce Lipschultz | shot 1: D2 and Ar puffed. But bypass valve too far open - 180. Gate valves may have opened too soon as well. Closing the bypass to 180. Setting the gate valve closure to 6 minutes after the shot for the next shot. |
| Feb 25 2005 08:36:16:800AM | 1050225002 | Bruce Lipschultz | shot 2: D2 and Ar puffed. Bypass valve ok but GV2 open. Try again. |
| Feb 25 2005 09:40:13:703AM | 1050225003 | Bruce Lipschultz | shot 3: D2 and Ar puffed. Bypass valve ok GV2 ok
Mass 2: 9.9e-9 Mass 3: 2.71e-9 Mass 3: 1.36e-6 Mass 40: 2.8e-8 TC2 pressure: 26.5 mT B_bot pressure: 32.7 mT G_side pressure: 26.2 Note - B_bot and G_side have offsets of 8.4 and 2.44 mT respectively RGA data appearing in MDS deos not look right:-( Need to check the pressure data. |
| Feb 25 2005 09:39:49:720AM | 1050225004 | Bruce Lipschultz | shot 4: D2 and Ar puffed. Bypass valve ok GV2 ok
Mass 2: 1.4e-8 Mass 3: 3.8e-9 Mass 3: 1.6e-6 Mass 40: 1.97e-7 TC2 pressure: 29.6 mT B_bot pressure: 32.3 mT G_side pressure: 30.8 Note - B_bot and G_side have offsets of 8.4 and 2.44 mT respectively There appears to be a problem with the ionization gauge data the mantissa is being put into both the mantissa and exponent in the tree. Next shot turn off the RGA archiving so we can look directly at the RGA data before it is put in the tree. |
| Feb 25 2005 03:20:44:433PM | 1050225014 | Bruce Lipschultz | shot 14: D2 and Ar puffed. Bypass valve ok GV2 ok
The data seemed to be taken correctly with this setup of: dwell time = 120 ms before trigger - aat check after trigger - 5 minutes after pulse gate valves closed 2 minutes before init gate valves opened 6 minutes after pulse |
| Physics Operator Comments | |||
| Feb 25 2005 08:35:01:830AM | 1050225001 | Jim Irby | puff D2 from 0 to 200 ms at 80% (pg3) puff argon from 800 to 1000 ms at 100% (pg1) got large puff sv4 was too open at 180 clicks: mass 4 to 2e-5 gatevalve toggle was not enabled vessel at 60C RGA dwell at 15 ms Data transfers every minute |
| Feb 25 2005 08:38:56:017AM | 1050225002 | Jim Irby | sv4 to 160 clicks gatevalves enabled for 360 s nice puff --- but still no gatevalve toggle! gv1 was closed but not gv2 |
| Feb 25 2005 08:53:57:610AM | 1050225003 | Jim Irby | try again with gatevalve toggle --- getting up to 20 mTorr in vessel big puff --- about 24 mTorr gatevalve toggle did work --- see both D2 and Ar Bruce wants more argon for next shot |
| Feb 25 2005 09:16:42:500AM | 1050225004 | Jim Irby | Argon pressure to 30 psi Argon puff from 600 to 1000 ms (from 800 to 1000 ms on last shot) 29 mTorr on TC3 Argon increased from 9e-8 to 2e-7 on the RGA mass 40 before/after rga data still not working properly |
| Feb 25 2005 11:52:56:500AM | 1050225005 | Jim Irby | No changes except before/after data store disabled by Steve Wolfe so we can look at the PLC files before they are deleted. mks pressures about 29 mTorr still having problems with before/after |
| Feb 25 2005 12:32:48:377PM | 1050225006 | Jim Irby | gatevalve toggle delay set to 600 s --- give more time for the RGA data to be transfered. Joe Bosco has added extra delay to the transfer time also. before/after may be okay! reading from RGA console: m40 1.7e-7 m04 1.4e-6 m03 3.7e-9 m02 1.1e-8 |
| Feb 25 2005 12:42:20:127PM | 1050225007 | Jim Irby | repeat no rga data? |
| Feb 25 2005 12:57:03:970PM | 1050225008 | Jim Irby | try again got rga before/after: low mass okay, but mass 40 screwed up |
| Feb 25 2005 03:23:32:920PM | 1050225009 | Jim Irby | repeat note that big board shot number is wrong before/after screwed up again --- consultation with Joe |
| Feb 25 2005 03:15:17:390PM | 1050225012 | Jim Irby | before/after data range okay, but before and after arrays reversed: mass(0,*) after mass(1,*) before |
| Feb 25 2005 03:15:27:577PM | 1050225013 | Jim Irby | no before/after data |
| Feb 25 2005 03:17:11:530PM | 1050225014 | Jim Irby | before/after = mass(1,*)/mass(0,*) again |
| Engineering Operator Comments | ||||
| Shot | Time | Type | Status | Comment |
| 1 | 08:08:18:907AM | Test | Ok | |
| 2 | 08:30:17:097AM | Test | Ok | |
| 3 | 08:39:30:907AM | Test | Ok | |
| 4 | 08:57:11:270AM | Test | Ok | |
| 5 | 09:39:55:720AM | Test | Ok | |
| 6 | 12:13:05:657PM | Test | Ok | |
| 7 | 12:28:19:487PM | Test | Ok | |
| 8 | 12:42:47:457PM | Test | Ok | |
| 9 | 12:57:39:830PM | Test | Ok | |
| 10 | 01:13:42:143PM | Test | Ok | |
| 11 | 01:52:28:800PM | Test | Ok | |
| 12 | 02:19:58:720PM | Test | Ok | |
| 13 | 02:39:53:627PM | Test | Ok | |
| 14 | 02:56:28:970PM | Test | Ok | |