| Physics Operator Comments |
| Mar 3 2006 07:51:05:003AM | 1060303001 | Steve Wolfe |
Gary is setting up to run some cycle tests.
He has connected the diborane piezo acquisition signal to
an INCAA someplace, so we will want to run some pulses
Using /usr/local/cmod/codes/pcs3-Beta/pcs
Load_from 1060301002 (the pulse that worked on Wed.)
Load/noclean
Filling in configuration / 1-NOV-2005 12:26:36.00/
Filling in DPCS configuration /22-DEC-2004 09:58:59.00/
about to ask alcdata to copy /home/wolfe/pcs_scratch/pcs_model to /cmod/trees/models/pcs/pcs_model
about to ask alcdata to copy /home/wolfe/pcs_scratch/dpcs_model to /cmod/trees/models/dpcs/dpcs_model
Load at 3-Mar-2006 07:50:27.00
Open tree /home/wolfe/pcs_scratch -1
Open tree done
Ready for Shot#1
|
| Mar 3 2006 08:06:34:140AM | 1060303001 | Steve Wolfe |
Shot#1: No puff. Got output from DPCS, no signal
on the INCAA (Input_11) and no puff seen on the
ratiomatic. Maybe PLC not setup correctly or an interlock
issue? Maybe hanging the digitizer off the line is loading?
|
| Mar 3 2006 08:23:29:650AM | 1060303002 | Steve Wolfe |
Shot#2: Hybrid Enable to diborane puff was not on for shot#1
Once more, with feeling!
Got gas. Got signal on input_11. Looks like there is about a
4V offset on the piezo valve (0.4V on the acquisition signal).
According to Gary this is way below the level that admits any
gas at all on this valve.
Save the scope with the new trace in
/home/wolfe/scopes/gas_scope_dpcs.dat
|