Alcator C-Mod Run 1070628 Information

Back to runs

Miniproposals
Miniproposal:477
Date Filed: 5/29/2007
Title:Formation of Ohmic H-Mode ITB with Improved Diagnostic Set
First Author:Bill Rowan
Session Leader:Catherine Fiore (shots 1-32)
Session Leader:Bill Rowan (shots 2-32)

Operators
Session leader(s):Catherine Fiore,Bill Rowan
Physics operator(s):Steve Wolfe
Engineering operator(s):Bill Cochran,Bill Byford,Ed Fitzgerald

Engineering Operator Run Comment
MP#477 Formation of Ohmic H-mode ITB with Improved

Session Leader Plans
Entered: Jun 27 2007 01:18:54:190PM
Author: Bill Rowan
Session Leader Plan for Run 1070628
SL: Fiore and Rowan

Purpose
MP#477 Formation of Ohmic H-Mode ITB with Improved Diagnostic Set
The purpose of this experiment is to study the conditions which lead to the
spontaneous formation of Ohmic H-mode ITBā??s with an extended diagnostic set.

Shot plan
Set up the Ohmic H-mode target shots with an ITB using a field ramp which
starts at 5.3 T, decreases to 3.2 T, then increases to 4.6 T. Confirm that an ITB
forms. Reference shot is 1020926030

Acquire profile data on reproducible shots.

If time allows, apply on-axis RF (increase field to 5T) in increments of 200 kW up to a probable
maximum of 1 MW. This will allow an opportunity to verify the TEM mode
suppression of the density peaking as determined for the RF generated ITB.

Auxiliary Systems:
DNB at 50 kV and 6 A.
ICRF: 600 kW may be requested after the run begins


Diagnostics:
Hirex for Ti profiles and rotation
Thomson Scattering
CXRS
Visible Bremsstrahlung Array
PCI
Reflectometry
GPC 1 and 2
Neutron diagnostics
MSE
Bolometry
FRCECE

Physics Operators Plans
Entered: Jun 27 2007 02:38:00:750PM
Author: Steve Wolfe
Physop Plan and Engineering Setup for Thursday, June 28

MP#477 Formation of Ohmic H-mode ITB with Improved 
       Diagnostic Set
SL: Fiore, Rowan
PO: wolfe

Overnight boronization (ending at 05:00)
using the "new standard parameters":
  sweep 75/80/85 (min/dwell/max)
  dwell/period 20/120
  Use EF4 vertical field, 200 Amps

Following boronization:
ECDC for two hours
Start 06:00, finish 8:00
He4: 5e-5 Torr
sweep 44/45/90 (min/dwell/max)
dwell/period 20/120

Engineering Setup
------------------
Power Systems as on 1070627001
A-coil configuration standard (+Dtop -Dbot -Jtop +Jbot) enabled
  start time 0.2sec

Gas Setup:
   fill B-Top            with  6 psi D2  Hybrid enabled  (PG4)
   fill B-side lower     with  1 psi Ar  Hybrid Enabled (PG1)
  leave B-side upper     as is           Hybrid DISABLED (PG2)   
   fill B-main (C-side)  with 40 psi D2  Hybrid enabled  (PG3)
   fill NINJA            with 10 psi D2  Enabled 
Enable the following gate valves and shutters, assuming no vacuum problems:
  ECE, VUV, DNB, Z-bolo shutters
Open HIREX shutter/gate valve

Torvac setup: Gate valves open

ICRF: Ready for RUN-ON (will be used later in run)
LH Setup: NO LH for this run
Cryopump Setup: NO CRYOPUMP for this run

Physop Run Plan:
-----------------

The SL plan calls for re-producing the plasma of 1020926030,
which pre-dates DPCS and the installation of the A-coils. The standard
ohmic H-mode/ITB prescription (ramp TF down to 3.2T to get H-mode, then
back up to 4.6T will be used, as programmed on the reference shot. 
According to the logbook entries and summaries for 1020926, it 
is important to get the triangularity match of the reference shot #30
in order to get a good EDA which will develop into ITB. Gaps for the
reference shot were leftgap~.02 and rightgap~.015m, delta_u>0.32,
delta_l~-.54. The high delta_u was only obtained after 0.9sec, and
there is a comment that it might be better to increase this value
(reduce RXU) earlier.  Becaulse in the changes in diagnostics and
offsets since 2002, some tweaking is expected to be required to recover
the target shape. Also, I'll put in the B21_PROJ feedback on wire 9.

Start from a shot from 1070627 and import seg2 from 1020926030. Re-calculate
all the magnetics predictors (including Ip since the rogowski got changed). 
Add B21_PROJ to wire 9. Note that CLEARIN and probably RCUR will almost
certainly need to be tweaked up. The CLEARIN waveform from the original
has a ramp to +.01; I think the right program should be flat at about 
0.01. The RCUR waveform is flat at .659; I think that needs to be 
changed to .662. I'll leave the xpoint programming as is and adjust as 
we go to try to match the triangularity. Will also need to put in a
100msec argon puff for HIREX at 0.3 sec (PG1). 

Once the target shot is established and hopefully the H-mode and ITB's
are obtained, consider moving the RXU sweep earlier to get an earlier 
H-mode/EDA transition. In this case, we may want to move the TF ramp-up
phase earlier as well. We may be able to speed up the TF ramp-down as well,
since in 2002 we were restricting the ramp rates.

According to the SL_PLAN, the only other PCS change would be to increase
the final TF target field to 5T so that the 80MHz ICRF would be near-axis
for tests of on-axis ICRF to arrest the density rise and test TEM model. 











Session Leader Summaries
Entered: Jun 28 2007 05:36:23:293PM
Author: Catherine Fiore
Session leader summary

We had numerous short lived ITBs today but were frustrated in attempts to
lengthen them or make them consistently from shot to shot. The best shots were 17, 19, and 27. Fluctuations were rampant on many shots, showing up on magnetics data as well as reflectometry and PCI. Unfortunately, the DNB was unavailable, foiling our plans to collect TI and rotation data from CXRS. Also, we were concerned about the effects of argon on being able to achieve EDA H-modes, so we had to cut the argon source as low as we could stand. This may limit the core data from HIREX and its clones.

Session Leader Plan for Run 1070628
SL: Fiore and Rowan

Purpose
MP#477 Formation of Ohmic H-Mode ITB with Improved Diagnostic Set
The purpose of this experiment is to study the conditions which lead to the
spontaneous formation of Ohmic H-mode ITB\ufffd??s with an extended diagnostic set.

Shot plan
Set up the Ohmic H-mode target shots with an ITB using a field ramp which
starts at 5.3 T, decreases to 3.2 T, then increases to 4.6 T. Confirm that an ITB
forms. Reference shot is 1020926030

Acquire profile data on reproducible shots.

If time allows, apply on-axis RF (increase field to 5T) in increments of 200 kW up to a probable
maximum of 1 MW. This will allow an opportunity to verify the TEM mode
suppression of the density peaking as determined for the RF generated ITB.

Auxiliary Systems:
DNB at 50 kV and 6 A.
ICRF: 600 kW may be requested after the run begins


Diagnostics:
Hirex for Ti profiles and rotation
Thomson Scattering
CXRS
Visible Bremsstrahlung Array
PCI
Reflectometry
GPC 1 and 2
Neutron diagnostics
MSE

------------------------------------------
Right off the bat there was a problem with an ice dam on the DNB which requires
warming it up. CXRS will not be available until this afternoon at the
earliest.



-------------------------------------------
Shot 1 No power shot--data system problem.

Shot 2 No shot. Init failure.

Shot 3 Good plasma, short H-mode, but no magnetics data, no density. Nl_04
finally got stored. H-mode looks Elm free.

Shot 4 Longer H-mode, but multiple back transitions. Probably no ITB.

Shot 5 Longer H-mode, but multiple back transitions. Last H-mode appears to
hold. No Thomson data, which means the Dalsa data did not invert correctly.
I'm not sure if there was an ITB late--probably not.

Shot 6 No power test.

Shot 7 No power test.

Shot 8 TF only test.

Shot 9 Early disruption.

Shot 10 Short H-mode, multiple back transition- no ITBs except for a couple
of ENs.

Shot 11 Longer H-mode, multiple back transitions- Slight peaking of density
around 1.2 s. We're going to try and start ramping the field up a little
earlier in the hopes of making the discharge more EDA like.


Shot 12 Two short H-modes, no ITB.

Shot 13 No power shot.

Shot 14 Two short H-modes, no ITB.

Shot 15 Finally, an ITB, although it ends at a the first back transition.
What did we do? We turned off the Argon. We'll try one more without argon
then we'll try putting small amounts back in.

Shot 16 Great ITB, unfortunately ending in disruption. No Argon on this shot.

Shot 17 Great ITB, unfortunately ending in disruption. No Argon on this shot.
Very interesting fluctuations on magnetics, PCI, and reflectometer.

We're going to try 15 ms of Argon on the next shot.

Shot 18 Starts a good ITB that back transitions then goes into a regular
H-mode with not ITB. 15 ms of Argon.

Shot 19 Good ITB that lasts until the plasma disrupts at 1.3 s. 15 ms of
Argon again.

Shot 20 Good ITB that lasts until the plasma back transitions at 1 s .
15 ms of Argon again. We are raising the TF to 5 T in prepartion for adding
central rf.

Shot 21 Good ITB that lasts until the plasma back transitions at 1.15 s .
15 ms of Argon again. We will try 0.5 MW ICRF. We are also going to turn off
the argon.

Shot 22 Good ITB that lasts until the plasma back transitions at 1.s . No
Argon again. ICRF did not fire.

Shot 23 No H-mode.

Shot 24 Very similar to shot 22 with a short ITB ending in a back transition.
The ICRF will not be available after all. We are going back to the magnetic
field ramp of shot 20.

Shot 25 Short ITB that ends with a back transition, followed by a 2nd H-mode.
No ITB in the second H-mode.

Shot 26 Short ITB that ends with a back transition, followed by a 2nd H-mode.
No ITB in the second H mode. Higher triangularity (.4-.45) We will go lower
on the next shot. The argon is back on.

Shot 27

Shot 28 Short ITB followed by multiple back transitions, followed by a longer
H-mode. No ITB in the second H mode. The RF didn't fire. The argon is back on.

Shot 29 Long H-mode but no apparent ITB, except a little peaking at 1. s. The
RF net power looks like it's zero, but the plasma acts like there was.
The argon is back on.

Shot 30 H mode with density peaking ends too early, before the rf pulse. RF
power trace is there this time. I suspect the same amount went in on the last
shot. The argon is still on.

Shot 31 ITB survives to 1.06 s. The argon is off. Current is ramped down
after H-mode onset. The ITB didn't make it long enough to catch the RF pulse.

Shot 32 We overfilled the New Trees area on the disk, so we are waiting to
see if the data set can be salvaged.
Unfortunately, we could not salvage the full data set and we missed key density
diagnostics.

Physics Operator Summaries
Entered: Jun 28 2007 06:09:01:473PM
Author: Steve Wolfe
 
Physop summary for Thursday, June 28
MP#377 Ohmic ITB's with improved diagnostics
SL: Fiore, Rowan
PO: wolfe

The run started and ended with data system problems. In between
we had other troubles. The DNB had regenned last night, boiling off
its LHe, and was unavailable today, reducing the number of "improved 
diagnostics" available. 
The first two shots were no power, shot 3
ran but had no magnetics data; I should be able to generate an EFIT
from the DPCS data, though the first attempt to do this failed. 
Shots #3-5 had short, ELM-free H-modes with no ITB. 

We then spent shots #6-8 trying to resolve a high TF joint resistance reading.
The after shot scanner reading reported up to 8uOhm on turn 43 top-middle.
Gary put this joint on the camac and it read under 1.5uOhm on a
150kA TF test (#8), so we decided it was all right. The next plasma 
attempt #9 was a super-fizzle, getting to 150kA at 30msec before snuffing. 
Probably the visiting dignitary effect. 

Shots#10-14 continued the string of non-ITB h-modes, as we played more 
with shapes and Bt waveform without success. For shot #15 we turned off
the argon (potentially removing more of the "improved diagnostics") and were
rewarded with a decent ITB, though the first H-mode didn't last the whole
shot. Continuing with no argon puff, we got what turned out to be the
best ITB's of the day on #16 and 17, longer single H-modes with sustained
ITB's which eventually suffered H-L (radiation collapse I think) and then
disrupted. These had a variety of interesting fluctuations seen on 
multiple diagnostics, and there was still enough argon hanging around
for HIREX_SR to analyze during the H/ITB phase. Putting back a short 15ms 
argon puff on shot #18 coincided with another shorter H-mode/ITB phase, 
although it's not clear the argon was responsible this time. The disruptions
seemed to have taken a toll on the boronization, and moly and iron were
both up. 

We continued trying to extend or improve the ITB's for the rest of the
day, and shot #19, also with a short argon puff, wasn't bad, but it 
was pretty much downhill from there. ICRF was turned on beginning with
shot 28, timing at 1.1sec after the field was back up to 5T in an attempt
to arrest the density peaking, but by that stage the ITB's weren't lasting
long enough so we didn't really get anything but a weak RF H-mode. 

We did do some exploration of triangularity, and it seemed that somewhat
lower triangularity than on 1020926030 actually kept the ITB longer, 
contrary to expectations and previous belief. Also, dropping Ip after
the L-H seemed to be helpful. 

The last shot #32 overfilled the data disk, resulting in lots of 
datasystem errors. JOsh managed to recover some but not all the data,
and it may be possible to re-run some analysis routines. It looked like
the ITB lasted a little longer on 32, at 600kA, than it  had on 31
at 700kA. However, thomson raw data apparently couldn't be recovered. 
Tom and Josh should probably look into putting the cmod/trees/new/
directories on a larger disk volume. It seems to be on a 65Gb disk,and
we were taking about 2Gb/shot, so over 30 shots is asking for trouble,
and we got it!

Scorecard
----------
     Tests/Nopower      6
     Duds               0
     Fizzle             1 (actually 150kA,35msec)
     Plasmas           25
     ===========      ====
     Shots             32

Session Leader Comments
Jun 28 2007 08:22:49:420AMBill RowanBeam availability (from Dexter)

The beam cryopumps regenerated last night, despite the fill that Edgar did
after the run. Apparently the humid air has caused ice to form a thermal
short.

Regeneration (defrosting) of the DNB cryo pumps is still in-progress. The last of the liquid helium evaporated around 23:00 last night, and although the fill is off, there is very likely still liquid nitrogen in the outer jackets of the pumps. Once the structure all warms up, we would be able to recool with LN2, wait for the metal structure to cool, and fill with liquid helium. This entire process will take at least four six more hours so operation today is very unlikely.

The reason for the warmup was a very large thermal leak into the helium. We normally can last a week or more on a 100% fill of LHe; we filled last night to 40% and it was evaporated in less than one hour. The thermal leaks are typically caused by H20 ice so a warmup should cure everything.
Jun 28 2007 09:11:47:603AM1070628001Catherine FioreNo power shot.
Jun 28 2007 09:29:27:930AM1070628001Catherine FioreNo shot. Init failure.
Jun 28 2007 09:49:31:273AM1070628003Catherine FioreGood plasma, short H-mode, but no magnetics data, no density.
Jun 28 2007 10:23:40:627AM1070628003Catherine FioreTCI finally stored. Looks like ELM free H-mode.
Jun 28 2007 10:18:10:420AM1070628004Catherine FioreLonger H-mode, but multiple back transitions. Probably no ITB.
Jun 28 2007 10:34:50:183AM1070628005Catherine FioreLonger H-mode, but multiple back transitions. Last H-mode appears to hold. No Thomson data, which means the Dalsa data did not invert correctly. I'm not sure if there was an ITB late--probably not.
Jun 28 2007 10:53:51:103AM1070628006Catherine FioreNo power test shot.
Jun 28 2007 10:59:59:670AM1070628007Catherine FioreNo power test shot.
Jun 28 2007 11:19:10:077AM1070628008Catherine FioreTF only test shot.
Jun 28 2007 11:33:43:540AM1070628009Catherine FioreEarly disruption.
Jun 28 2007 12:00:56:610PM1070628010Catherine FioreShort H-mode, multiple back transition- no ITBs except for a couple of ENs.
Jun 28 2007 12:09:11:927PM1070628011Catherine FioreLonger H-mode, multiple back transitions- Slight peaking of density around 1.2 s. We're going to try and start ramping the field up a little earlier in the hopes of making the discharge more EDA like.
Jun 28 2007 12:13:55:353PM1070628012Catherine FioreTwo short H-modes, no ITB.
Jun 28 2007 12:25:38:940PM1070628013Catherine FioreNo power shot.
Jun 28 2007 12:40:23:733PM1070628013Catherine FioreTwo short H-modes, no ITB.
Jun 28 2007 12:56:03:573PM1070628014Catherine FioreTwo short H-modes, no ITB.
Jun 28 2007 12:58:00:810PM1070628015Catherine FioreFinally, an ITB, although it ends at a the first back transition. What did we do?
We turned off the Argon. We'll try one more without argon then we'll try putting small amounts back in.
Jun 28 2007 01:16:36:307PM1070628016Catherine FioreGreat ITB, unfortunately ending in disruption. No Argon on this shot.
Jun 28 2007 01:30:09:507PM1070628017Catherine FioreGreat ITB, unfortunately ending in disruption. No Argon on this shot. Very interesting fluctuations on magnetics, PCI, and reflectometer.

We're going to try 15 ms of Argon on the next shot.
Jun 28 2007 01:39:04:950PM1070628018Catherine FioreStarts a good ITB that back transitions then goes into a regular H-mode with not ITB. 15 ms of Argon.
Jun 28 2007 01:54:43:163PM1070628019Catherine FioreGood ITB that lasts until the plasma disrupts at 1.3 s. 15 ms of Argon again.
Jun 28 2007 02:07:48:160PM1070628020Catherine FioreGood ITB that lasts until the plasma back transitions at 1 s . 15 ms of Argon again. We are raising the TF to 5 T in prepartion for adding central rf.


Jun 28 2007 02:23:06:170PM1070628021Catherine FioreGood ITB that lasts until the plasma back transitions at 1.15 s . 15 ms of Argon again. We will try 0.5 MW ICRF. We are also going to turn off the Argon.


Jun 28 2007 02:33:35:617PM1070628022Catherine FioreGood ITB that lasts until the plasma back transitions at 1.s . No Argon again. ICRF did not fire.


Jun 28 2007 02:45:03:063PM1070628023Catherine FioreNo H-mode.

Jun 28 2007 03:31:21:240PM1070628024Catherine FioreVery similar to shot 22 with a short ITB ending in a back transition. The ICRF will not be available after all. We are going back to the magnetic field ramp of shot 20.

Jun 28 2007 03:27:06:720PM1070628025Catherine FioreShort ITB that ends with a back transition, followed by a 2nd H-mode. No ITB in the second H-mode.
Jun 28 2007 03:44:53:343PM1070628026Catherine FioreShort ITB that ends with a back transition, followed by a 2nd H-mode. No ITB in the second H mode. Higher triangularity (.4-.45) We will go lower on the next shot. The argon is back on. There may be a slight bit of peaking around 1.5 s.
Jun 28 2007 04:42:32:933PM1070628027Catherine FioreITB survives through the back transition. Weird fluctuation on the density and soft x-rays on second H-mode.
Jun 28 2007 04:11:52:503PM1070628028Catherine FioreShort ITB followed by multiple back transitions, followed by a longer H-mode. No ITB in the second H mode. The RF didn't fire. The argon is back on.
Jun 28 2007 04:21:38:183PM1070628029Catherine FioreLong H-mode but no apparent ITB, except a little peaking at 1. s. The RF net power looks like it's zero, but the plasma acts like there was. The argon is back on.
Jun 28 2007 04:36:11:367PM1070628030Catherine FioreH mode with density peaking ends too early, before the rf pulse. RF power trace is there this time. I suspect the same amount went in on the last shot. The argon is still on.
Jun 28 2007 04:47:43:220PM1070628031Catherine FioreITB survives to 1.06 s. The argon is off. Current is ramped down after H-mode onset. The ITB didn't make it long enough to catch the RF pulse.
Jun 28 2007 05:23:07:977PM1070628032Catherine FioreWe overfilled the New Trees area on the disk, so we are waiting to see if the data set can be salvaged.

Unfortunately, we could not salvage the full data set.

Physics Operator Comments
Jun 28 2007 07:33:45:223AMSteve Wolfe
Bring up PCS and Import Seg 2 from 1020926030
Acknowledge config and no DPCs tree warnings
Seg 2 on, seg 3 off
Call predictors wires 1-8 (eliminate offset on clearin)
Call Ip predictor (wire 16) (eliminate offset)
Call NL04 predictor (wire 14) - Custom
 fill in nl04 gain=1
 Apply/set factor
Put B21_PROJ on wire 9
 Zero Acoil_cur on V01-16 screen (ack name change)
 Int Gate C to 0.2sec
On All-P's
 Clearin waveform to +.01
 RCUR from .659 to .662
 Note EF2byEF4 is at -2500 until 1.4sec; leave it.
V01-16 screen
 Add 100msec puff to pg1@0.3sec
 Change chopper from 100V to 0.
V17-32
 Touch a wire to get the names update, then apply
Load at 28-Jun-2007 07:33:33.00
Open tree /home/wolfe/pcs_scratch      -1
Open tree done


Jun 28 2007 09:11:53:010AM1070628001Steve Wolfe
Shot#1: Trying to run a shot from 2002. What did I forget?
Data system problem in INIT?
No power shot. 

Jun 28 2007 09:30:54:807AM1070628002Steve Wolfe
Shot#2: Try again to run the shot from 2002. 
Still having data system problems.
Trying anyway after six minutes in INIT.
Nothing went, but we left the breakers in anyway.
I think nothing got inited.

Jun 28 2007 09:49:16:070AM1070628003Steve Wolfe
Shot#3: And around we go again.
Plasma, but it didn't look like an ITB on the TV
Short H-mode, starts at 1sec off by 1.1sec
No magnetics data. Not getting TCI.
Also no Traq data according to Tom; he's looking into
the magnetics problem.

Jun 28 2007 10:15:46:620AM1070628004Steve Wolfe
Shot#4: Still no magnetics from #3.
running efit from dpcs data failed, no time to debug now. 
No changes just try again. 
Longer H-mode, still not steady. Got magnetics this time. 
H-mode starts at 0.83sec, on the way down in TF waveform. 
Triangularity looks ok, 0.3 to 0.4. left gap 2cm, right
gap a bit small, 1.3 dropping to 0.9. I can adjust that.
I don't think it will help the H-mode though. Looks ELM-free
and highly radiating. Minimum q95 is 2.7. 

Jun 28 2007 10:44:11:050AM1070628005Steve Wolfe
Shot#5: RCUR at 1.5sec from .662 to .660
RXU from -.02 to -.03 (higher triang.
multiple H-modes, starting at 0.84sec, still looks ELM-free.
bouncy startup, maybe a few runaways. 
right gap could still be increased some. Upper triang now over 0.4
q95 still 2.7 rising to 4.1
Actually, there is a late H-mode that looks sort of EDA'ish
There is a very broad QC mode on the magnetics. 
Doesn't look ITB'ish though.


Jun 28 2007 10:40:06:470AM1070628006Steve Wolfe
Shot#6: increase pg4 from 18 to 20ms
This will be a no-power shot to check on TF resistance value.

Jun 28 2007 10:53:34:197AM1070628007Steve Wolfe
Shot#7: One more test shot, with longer integration time.
TF turn #43 is showing a high resistance reading on the scanner,
starting from shot#3 this morning. It was ok yesterday. 

Jun 28 2007 11:18:50:153AM1070628008Steve Wolfe
Shot#8: TF only test at 150kA, PLC control.


Jun 28 2007 11:24:11:457AM1070628008Steve Wolfe
Shot#8: TF only test at 150kA, PLC control.
Live resistance looks acceptable. 

Jun 28 2007 12:17:08:403PM1070628009Steve Wolfe
Shot#9: Back to plasma shots (or attempts)
No changes in seg2 this time
Fizzle. Got to 150kA and ran back in and possibly up.



Jun 28 2007 12:17:32:810PM1070628010Steve Wolfe
Shot#10: PG4 from 20ms to 18ms
plasma, short H-modes only, no eda phase.





Jun 28 2007 12:17:49:073PM1070628011Steve Wolfe
Shot#11: Increase CLEARIN from .01 to .012
plasma, still multiple H-modes. 
Maybe the third one is a little better. 

Try ramping tf up a little earlier, since we're getting H-mode by .8sec



Jun 28 2007 12:18:15:357PM1070628012Steve Wolfe
Shot#12: Start reverse TF ramp at 0.85 to .95sec
plasma, more short h-modes, no eda phase this time. 




Jun 28 2007 12:25:45:580PM1070628013Steve Wolfe
Shot#13: Decrease RXU (upper triang) back to -.02 from -.03
No Power shot - PF breaker pulled out with a type 1 fault.




Jun 28 2007 12:41:21:170PM1070628014Steve Wolfe
Shot#14: Decrease RXU (upper triang) back to -.02 from -.03
H-modes still too short. 

Next: turn off argon for a shot or two


Jun 28 2007 12:54:31:430PM1070628015Steve Wolfe
 
Shot#15: turn off argon for a shot or two
One longer h-mode, then a second EDA'ish h-mode. 
Actually the first one doesn't look bad as far as peaking
is concerned. 

Next: repeat

Jun 28 2007 01:10:22:617PM1070628016Steve Wolfe
 
Shot#16:repeat
Longer ITB until 1.15, then disrupts at 1.22sec
Stored energy increasing throughout H-mode.
TS ne0 gets to 7e20.



Jun 28 2007 01:38:15:903PM1070628017Steve Wolfe
 
Shot#17: drop pg4 from 18 to 16msec, otherwise repeat 
 (leave Ar off one more shot)
Possible computer problem, lost Pat's voice
Plasma, another ITB, still disrupting around 1.25sec
Very similar to 16
Interesting fluctuations on magnetics on this shot.
Something at ~700kHz early in H-mode, then multiple 
modes 50 to 150kHz, starting around .9sec, plus a 10kHz tearing? mode.
Big burst at the end of Hmode

Jun 28 2007 01:37:45:247PM1070628018Steve Wolfe
 
Shot#18: Put back a short argon puff, 15ms at 0.3sec
Shorter first h-mode, second one not as good,
but plasma full length. 
ITB phase ends at 1.0sec


Jun 28 2007 01:51:14:850PM1070628019Steve Wolfe
 
Shot#19: Repeat, leaving 15msec of Argon on. 
plasma, longer ITB, disruption at 1.33sec

Next: allow earlier NINJA puff, leave argon at 15ms


Jun 28 2007 02:03:41:420PM1070628020Steve Wolfe
 
Shot#20: allow earlier NINJA puff, leave argon at 15ms
No other changes
Shorter ITB phase, full length plasma
Drops out of H-mode about 1.04 sec. No second H-mode.
It was ITB while it lasted. 




Jun 28 2007 02:16:33:620PM1070628021Steve Wolfe
 
Shot#21: Bring TF up to 5T (139kA)
Delay NINJA to 1.3sec
ITB until 1.16sec, no second H-mode. 
Plasma lasts full length. 

Next: turn on ICRF at 1.1sec 500kW



Jun 28 2007 02:35:19:350PM1070628022Steve Wolfe
 
Shot#22: Start TF back up at 0.8sec, argon back off for a shot
 turn on ICRF at 1.1sec 500kW
ITB still shorter than earlier, ends at 1sec
Didn't seem to get ICRF. 

Next: raise nl04 from 7 to 8e19

Jun 28 2007 02:46:06:907PM1070628023Steve Wolfe
 
Shot#23:raise nl04 from 7 to 8e19, leave argon off
RF is having a demand problem, off for this shot. 

NO H-mode, too much density I guess. 


Jun 28 2007 03:11:57:570PM1070628024Steve Wolfe
 
Shot#24:Nl04 from 8 to 6e19; this is really finicky. 
Back to a short Hmode, ends at 1.04sec 

Back to TF program as on 20

Jun 28 2007 03:24:22:360PM1070628025Steve Wolfe
 
Shot#25:Back to TF program as on 20, ramp up to 4.6 T starting at 0.85sec
Two H-modes, some ITB character on first. 


Jun 28 2007 03:34:45:057PM1070628026Steve Wolfe
 
Shot#26: Turn argon back on (15msec)
RXU from -.02 to -.03
Maybe a little worse, ITB lives to 1.0sec

Next: try less triangular


Jun 28 2007 04:05:49:123PM1070628028Steve Wolfe
 
Shot#28: TF back up to 5T (139kA), try to turn on RF at 1.1sec
Argon puff from 15 to 25msec
ITB in first H-mode ?, then junk during rf faulting, then a third hmode 
with no ITB. 



Jun 28 2007 04:19:18:183PM1070628029Steve Wolfe
 
Shot#29: Repeat, try to get rf to run
Two hmodes again, something stops the density rise.
No net power trace from D, but it looks like there was some
power there. 

There is a "system fault" message on the rf screen. 



Jun 28 2007 04:32:47:770PM1070628030Steve Wolfe
 
Shot#30: Raise nl04 back from 6 to 7e19
First h-mode only to 0.95sec, has ITB
Got the rf trace this time, looks fine but second h-mode is just 
central rf no ITB. 



Jun 28 2007 05:02:10:590PM1070628032Steve Wolfe
 
Shot#32: Drop Ip to 600kA after 0.8sec
small back transition at 1sec, then keeps going 
looks like a lot of data errors. 
EFIT data didn't show up. Ran it manually and it still didn't
show up, although it ran. 

Jerry is reporting tree failure problem
New tree area has zero free blocks.
Josh is working on it.

Jun 28 2007 05:24:13:093PM1070628032Steve Wolfe
 
Shot#32: Data from the last shot were partially restored by
Josh, after he made room. However, thomson seems to be gone 
forever, also DALSA. 
It's too late to try another shot, and there doesn't seem to 
be enough room anyway. 

END OF CRAWL. 

Engineering Operator Comments
ShotTimeTypeStatusComment
109:02:00:970AMTestBad
309:40:02:063AMPlasmaOk
410:06:08:880AMPlasmaOk
510:23:53:423AMPlasmaOk
610:40:27:173AMTestOkTF Scanner Test
710:55:34:480AMTestOkTF Scanner test
811:16:16:467AMTestOkTF 150kA Scanner Test
911:29:57:757AMPlasmaOk
1011:43:09:173AMPlasmaOk
1111:57:23:653AMPlasmaOk
1212:10:19:180PMPlasmaOk
1312:23:00:203PMPlasmaBadOH2L type 1 fault in check
1412:36:32:230PMPlasmaOk
1512:48:53:600PMPlasmaOk
1601:02:04:733PMPlasmaOk
1901:45:17:080PMPlasmaOk
2001:57:55:730PMPlasmaOk
2102:10:51:770PMPlasmaOk
2202:23:35:797PMPlasmaOk
2302:41:46:450PMPlasmaOk
2403:06:13:203PMPlasmaOk
2503:18:50:917PMPlasmaOk
2603:31:22:660PMPlasmaOk
2703:43:49:250PMPlasmaOk
2803:57:13:667PMPlasmaOk
2904:09:49:677PMPlasmaOk
3004:27:38:330PMPlasmaOk
3104:39:57:900PMPlasmaOk
3204:52:22:567PMPlasmaOk

System Availability
Jun 28 2007 09:01:29:783AM
SystemAvailableCommentAvailability
Changed
OwnerDescriptionWho Can
Change
Active MHDyesFully operationalMar 28 2007 09:19PM (sears)Joe SnipesActive MHD Antennasears,snipes
BESyesvertical array of six views at R ~ 87.3 cm, radial array of three views out to R ~ 89 cm, both at outboard midplane. Three discrete fiber views of beam at R ~ 70.0, 76.5 and 82.5 cm.Feb 06 2007 10:00AM (bravenec)Ron BravenecBeam Emission Spectroscopybravenec
Bolometersyes2Pi diode operational, 2Pi foil being upgrade for noise resistance. Midplane foils calibrated for senstivity, still need better time constants. Midplane and divertor AXUV being digitized, processing calibration data.May 29 2007 11:22AM (mlreinke)Matt ReinkeBolometershutch,mlreinke
BoronizationyesFirst Boronization 5/9/07May 10 2007 04:55PM (irby)Jim IrbyBoronizationdekow,irby
CNPAno Jan 14 2007 02:06PM Ron ParkerCompact Neutral Analyzerparker
Cryopump - Upper DivertoryesThe cryopump is operational in a manual mode. We will be working on interfacing its control system with the C-Mod state machine for semi-automatic operation.May 30 2007 07:58PM (labombard)Brian LaBombardUpper Divertor Cryopumpirby,labombard
CXRSyesThe pedestal toroidal array was installed, spatially and absolutely calibrated. There are 10 chords intersect the beam at the major_R_range=[75,4:87.4]cm. Spatial resolution: 1.2 cm. The F-top poloidal array was installed and spatially calibrated. There are 19 chords intersect the beam at the major_R_range=[67.8:91.0]cm. Spatial resolution: 1.22 cm.Jun 20 2007 05:35PM (bespam)Bill RowanCharge Exchange Spectroscopybespam,blip,rachmcd,rowan
D Alphayes\ha2_bright and \SPECTROSCOPY::TOP.VUV.VIS_SIGNALS:MCP_VIS_SIG1 are availableMar 11 2007 10:28PM (terry)Jim TerryBalmer Alpha Monitorirby,terry,wolfe
Dalsayescamera is re-installed, and view has been calibratedFeb 26 2007 03:28PM (marmar)Earl MarmarVisible Bremsstrahlungmarmar
Disruption Mitigationno Jan 14 2007 02:02PM Bob GranetzGas Jetgranetz,whyte
Divertor IRno Mar 06 2007 10:50AM Glen WurdenDivertor IR Digital Videogwurden,terry
Divertor IR Spectroscopyno Mar 06 2007 10:51AM Glen WurdenDivertor IR Spectroscopygwurden,ldorf,terry
DNByesAll systems are operational, but as of 28 March 2007, the beam needs additional conditioning for full performance shotsMar 28 2007 09:36AM (beals)Bob GranetzDiagnostic Beambeals,granetz
FRC-ECEnoOnly high frequency channels now working (17 to 32). LO looks good. Most likely it is the 5V supply for mixer bais and amp that is the problem.Jun 20 2007 09:23AM (phillips)Perry PhillipsHeterodyne ECE-32 channelsphillips
G-Side RatiomaticyesThe gauge is operational.Mar 30 2007 10:54AM (labombard)Brian LaBombardG-side ratiomatic ionization gaugelabombard,toland
GPCyesChs 1-8 operational. May 24 2007 08:40AM (hubbard)Amanda HubbardECE Polychromaterhubbard,schmidt
GPC2yesMost channels (except 1,5,8,10,17) operational. Jun 08 2007 08:29AM (hubbard)Amanda HubbardECE Polychromaterhubbard,schmidt
GPIyesNew view just outside of the lower X-point is operational, but data is being archived locally , not to the tree. Midplane GPI is not operational.Mar 27 2007 01:59PM (terry)Jim TerryGas Puff Imagingterry,zweben
H/Dyesspectral calibration needs to be checked, but system should be ready for the runFeb 26 2007 03:28PM (marmar)Earl MarmarH/D Ratiomarmar
Hirex Jryes Mar 06 2007 08:59AM (rice)John RiceHigh Resolution X-Rayince,rice
Hirex SryesStoring Data, but profiles not yet availableApr 27 2007 01:20PM (ince)John RiceHigh Resolution X-Rayince,rice
HXRyesSomething might be wrong with channel 26; all others working fine.Mar 23 2007 10:43AM (schmidt)Ron ParkerHard X-Ray Arrayparker,schmidt
ICRF D-Antennayes80.5 MHz. Max power is about 1.5 MWJun 20 2007 09:23AM (ylin)Steve WukitchD-port antennawukitch,ylin
ICRF E-Antennayes80 MHz. Maxc power is about 1.5 MWJun 20 2007 09:23AM (ylin)Steve WukitchE-port antennawukitch,ylin
ICRF J-Antennayes78 MHz. dipole configuration. Using #4 transmitter only. Max power about 1.5 MW instead of 3 MW.Jun 20 2007 09:22AM (ylin)Steve WukitchJ-port antennawukitch,ylin
Lower Hybridno6B Pressure window failed. Arcing in 6B, 8B,9B & 23B. May 04 2007 03:44PM (dterry)Ron ParkerLower Hybriddterry,parker
LPIyes Apr 26 2007 08:35PM (bose)Earl MarmarLithium Pellet Injectorbose,marmar
Ly-alphayesMidplane data being digitized, processing calibration data. Divertor views showing significant noise, and not available. Upper view being debugged.May 29 2007 11:23AM (mlreinke)Matt ReinkeLyman alpha cameraslabombard,mlreinke
McPhersonyesWorking normally.Apr 19 2007 09:52AM (mlreinke)Jim TerryVUV Spectrometermlreinke,terry
MichelsonyesOperational. Mar 28 2007 08:33AM (hubbard)Amanda HubbardECE Te Profileshubbard,schmidt
MKSyesE-Top, E-Bottom, F-top (cryo), G-side, and B-bottom are operational. F-top gauge has a 'sniffer tube' to measure pressures in cryopump gas box.Mar 20 2007 06:42PM (g)Brian LaBombardBaratron pressure gaugeslabombard
MLPnoInitial tests on the A-port scanning probe went well. We need to take care of some control glitches that caused some component failures. May 30 2007 08:01PM (labombard)Brian LaBombardMirror Langmuir Probe Electronics Systemlabombard
MSEno(1) Bandpass filters replacement underway (2) Beam-into-gas spectrum measurement underway (3) Shutter mechanism currently unstable; remains open. (4) Expect beam-into-gas calibrations complete about 4/24. Apr 03 2007 12:05PM (sscott)Steve ScottMotional Stark Effectjinseok,sscott
NeutronsyesBoth neutron systems are fully operationalApr 05 2007 11:19AM (fiore)Catherine FioreNeutron Detectorsfiore
Neuts and HardsyesoperationalMar 21 2007 04:29PM (irby)Jim IrbyX-Ray/Neutron Detector at K-Horirby
NINJAyesAll capillaries displayed by the NINJA_CONTROL GUI have been tested. NINJA puffed on 1070329009 (A-B limiter).Mar 29 2007 11:47AM (terry)Brian LaBombardNINJA gas pufflabombard,terry
PCIyesThe system is to be optimized.May 30 2007 01:12PM (lianglin)Miklos PorkolabPhase Contrast Imagingedlund,lianglin,ntsujii,porkolab
Penning Gaugesyes8 in-situ Penning ionization gauges appear to be operational: CB_DivU, CB_OutL, CB_DivL, KA_OutU, KA_OutL, B_DivL, Cryo_1, Cryo_2. Cryo_1 and Cryo_2 are new and installed in the upper cryopump 'gas-box'.Mar 13 2007 10:01PM (labombard)Brian LaBombardPenning Ionization Gaugeslabombard
Plasma TVyesWIDE1, WIDE2 (chan 33), DANT, JANT and DIV (without an ND filter) are operational.Mar 27 2007 01:53PM (terry)Jim TerryVideo of plasma via several camerasdekow,irby,terry,wolfe
PolarimeternoRetro shutter closedMar 15 2007 11:00AM (irby)Jim IrbyC-Mod Polarimeterirby
Probe Array - ISDyesProbes at 15 of the 16 poloidal locations ('ramped' geometry) are operational. n,Te, particle flux reported every 10 ms.Mar 13 2007 09:47PM (labombard)Brian LabombardInner Lower Divertor Probe Arraylabombard
Probe Array - LHyesAll 6 probes on the lower hybrid launcher (tungsten, 1.5 mm projection) are operational. Mar 13 2007 09:46PM (labombard)Brian LaBombardLower Hybrid Probe Arraylabombard
Probe Array - OSDyesProbes at 10 poloidal locations (one 'ramped' and one 'flush') are operational. n,Te, particle flux reported every 10 ms. Mar 13 2007 09:45PM (labombard)Brian LaBombardOuter Lower Divertor Probe Arraylabombard
Probe Array - UDIVyes14 probes are operational ('ramped' geometry). n,Te, particle flux reported every 10 ms.Two high-gain channels are installed and operational for ECDC plasma diagnosis.Mar 13 2007 12:46PM (labombard)Brian LaBombardUpper Divertor Probe Arraylabombard
ReflectometeryesAll low frequency channels (50, 60, 75, 88, 88u, 88l and 110) have been installed (for the first time!) on 1070626. Thery are being tested. The PPPL variable frequency channel is installed and being run (for now) at fixed frequency (Voltage set at 6V). 140GHz channel still down. Jun 28 2007 08:32AM (dominguez)Earl MarmarReflectometerdominguez,marmar
Scanning Probe - ASPyesPerpendicular Mach probe head is installed. The drive is operational.Mar 15 2007 09:09PM (labombard)Brian LaBombardA-Port Scanning Probelabombard
Scanning Probe - FSPyesPerpendicular Mach probe head is installed. The probe drive is now operational.Mar 13 2007 09:51PM (labombard)Brian LaBombardF-Port Scanning Probelabombard
Scanning Probe - WASPyesA-port WASP successfully collected data on 04/04. Some intermittent open circuit on SE Electrode. K-WASP took data on 04/11. Both ready to scan as needed.Apr 12 2007 04:18PM (nsmick)Brian LaBombardInner Wall Scanning Probelabombard,nsmick
TCIyesNL_04 at 71.04 cm from 3/9/07 calibrationMay 23 2007 07:38AM (irby)Jim IrbyTwo-Color Interferometerirby,murray
TS - Coreyes2 lasers, 11 spatial channels. Density is calibrated based on ECE cutoffs obtained on 4 May 2007. Additional spatial channels will be added over the next several weeks.May 10 2007 10:37AM (jwhughes)Jerry HughesCore Thomson scatteringjwhughes
TS - Edgeyes2 lasers, 20 spatial channels. Fiber holder position is -124.5mm (standard). Density is calibrated according to Raman scattering results from 26 May, supplemented with ECE cutoffs obtained on shot 1070531015.Jun 14 2007 06:21PM (jwhughes)Jerry HughesMillimeter resolution edge Thomson scatteringjwhughes
Vispecyes350 and 670 nm, 4 spatial channelsMay 04 2007 01:02PM (agraf)Alex GrafVisible Spectrometeragraf


Jun 28 2007 09:21:45:533AM
SystemAvailableCommentAvailability
Changed
OwnerDescriptionWho Can
Change
Active MHDyesFully operationalMar 28 2007 09:19PM (sears)Joe SnipesActive MHD Antennasears,snipes
BESyesvertical array of six views at R ~ 87.3 cm, radial array of three views out to R ~ 89 cm, both at outboard midplane. Three discrete fiber views of beam at R ~ 70.0, 76.5 and 82.5 cm.Feb 06 2007 10:00AM (bravenec)Ron BravenecBeam Emission Spectroscopybravenec
Bolometersyes2Pi diode operational, 2Pi foil being upgrade for noise resistance. Midplane foils calibrated for senstivity, still need better time constants. Midplane and divertor AXUV being digitized, processing calibration data.May 29 2007 11:22AM (mlreinke)Matt ReinkeBolometershutch,mlreinke
BoronizationyesFirst Boronization 5/9/07May 10 2007 04:55PM (irby)Jim IrbyBoronizationdekow,irby
CNPAno Jan 14 2007 02:06PM Ron ParkerCompact Neutral Analyzerparker
Cryopump - Upper DivertoryesThe cryopump is operational in a manual mode. We will be working on interfacing its control system with the C-Mod state machine for semi-automatic operation.May 30 2007 07:58PM (labombard)Brian LaBombardUpper Divertor Cryopumpirby,labombard
CXRSyesThe pedestal toroidal array was installed, spatially and absolutely calibrated. There are 10 chords intersect the beam at the major_R_range=[75,4:87.4]cm. Spatial resolution: 1.2 cm. The F-top poloidal array was installed and spatially calibrated. There are 19 chords intersect the beam at the major_R_range=[67.8:91.0]cm. Spatial resolution: 1.22 cm.Jun 20 2007 05:35PM (bespam)Bill RowanCharge Exchange Spectroscopybespam,blip,rachmcd,rowan
D Alphayes\ha2_bright and \SPECTROSCOPY::TOP.VUV.VIS_SIGNALS:MCP_VIS_SIG1 are availableMar 11 2007 10:28PM (terry)Jim TerryBalmer Alpha Monitorirby,terry,wolfe
Dalsayescamera is re-installed, and view has been calibratedFeb 26 2007 03:28PM (marmar)Earl MarmarVisible Bremsstrahlungmarmar
Disruption Mitigationno Jan 14 2007 02:02PM Bob GranetzGas Jetgranetz,whyte
Divertor IRno Mar 06 2007 10:50AM Glen WurdenDivertor IR Digital Videogwurden,terry
Divertor IR Spectroscopyno Mar 06 2007 10:51AM Glen WurdenDivertor IR Spectroscopygwurden,ldorf,terry
DNByesAll systems are operational, but as of 28 March 2007, the beam needs additional conditioning for full performance shotsMar 28 2007 09:36AM (beals)Bob GranetzDiagnostic Beambeals,granetz
FRC-ECEnoOnly high frequency channels now working (17 to 32). LO looks good. Most likely it is the 5V supply for mixer bais and amp that is the problem.Jun 20 2007 09:23AM (phillips)Perry PhillipsHeterodyne ECE-32 channelsphillips
G-Side RatiomaticyesThe gauge is operational.Mar 30 2007 10:54AM (labombard)Brian LaBombardG-side ratiomatic ionization gaugelabombard,toland
GPCyesChs 1-8 operational. May 24 2007 08:40AM (hubbard)Amanda HubbardECE Polychromaterhubbard,schmidt
GPC2yesMost channels (except 1,5,8,10,17) operational. Jun 08 2007 08:29AM (hubbard)Amanda HubbardECE Polychromaterhubbard,schmidt
GPIyesNew view just outside of the lower X-point is operational, but data is being archived locally , not to the tree. Midplane GPI is not operational.Mar 27 2007 01:59PM (terry)Jim TerryGas Puff Imagingterry,zweben
H/Dyesspectral calibration needs to be checked, but system should be ready for the runFeb 26 2007 03:28PM (marmar)Earl MarmarH/D Ratiomarmar
Hirex Jryes Mar 06 2007 08:59AM (rice)John RiceHigh Resolution X-Rayince,rice
Hirex SryesStoring Data, but profiles not yet availableApr 27 2007 01:20PM (ince)John RiceHigh Resolution X-Rayince,rice
HXRyesSomething might be wrong with channel 26; all others working fine.Mar 23 2007 10:43AM (schmidt)Ron ParkerHard X-Ray Arrayparker,schmidt
ICRF D-Antennayes80.5 MHz. Max power is about 1.5 MWJun 20 2007 09:23AM (ylin)Steve WukitchD-port antennawukitch,ylin
ICRF E-Antennayes80 MHz. Maxc power is about 1.5 MWJun 20 2007 09:23AM (ylin)Steve WukitchE-port antennawukitch,ylin
ICRF J-Antennayes78 MHz. dipole configuration. Using #4 transmitter only. Max power about 1.5 MW instead of 3 MW.Jun 20 2007 09:22AM (ylin)Steve WukitchJ-port antennawukitch,ylin
Lower Hybridno6B Pressure window failed. Arcing in 6B, 8B,9B & 23B. May 04 2007 03:44PM (dterry)Ron ParkerLower Hybriddterry,parker
LPIyes Apr 26 2007 08:35PM (bose)Earl MarmarLithium Pellet Injectorbose,marmar
Ly-alphayesMidplane data being digitized, processing calibration data. Divertor views showing significant noise, and not available. Upper view being debugged.May 29 2007 11:23AM (mlreinke)Matt ReinkeLyman alpha cameraslabombard,mlreinke
McPhersonyesWorking normally.Apr 19 2007 09:52AM (mlreinke)Jim TerryVUV Spectrometermlreinke,terry
MichelsonyesOperational. Mar 28 2007 08:33AM (hubbard)Amanda HubbardECE Te Profileshubbard,schmidt
MKSyesE-Top, E-Bottom, F-top (cryo), G-side, and B-bottom are operational. F-top gauge has a 'sniffer tube' to measure pressures in cryopump gas box.Mar 20 2007 06:42PM (g)Brian LaBombardBaratron pressure gaugeslabombard
MLPnoInitial tests on the A-port scanning probe went well. We need to take care of some control glitches that caused some component failures. May 30 2007 08:01PM (labombard)Brian LaBombardMirror Langmuir Probe Electronics Systemlabombard
MSEno(1) Bandpass filters replacement underway (2) Beam-into-gas spectrum measurement underway (3) Shutter mechanism currently unstable; remains open. (4) Expect beam-into-gas calibrations complete about 4/24. Apr 03 2007 12:05PM (sscott)Steve ScottMotional Stark Effectjinseok,sscott
NeutronsyesBoth neutron systems are fully operationalApr 05 2007 11:19AM (fiore)Catherine FioreNeutron Detectorsfiore
Neuts and HardsyesoperationalMar 21 2007 04:29PM (irby)Jim IrbyX-Ray/Neutron Detector at K-Horirby
NINJAyesAll capillaries displayed by the NINJA_CONTROL GUI have been tested. NINJA puffed on 1070329009 (A-B limiter).Mar 29 2007 11:47AM (terry)Brian LaBombardNINJA gas pufflabombard,terry
PCIyesThe system is to be optimized.May 30 2007 01:12PM (lianglin)Miklos PorkolabPhase Contrast Imagingedlund,lianglin,ntsujii,porkolab
Penning Gaugesyes8 in-situ Penning ionization gauges appear to be operational: CB_DivU, CB_OutL, CB_DivL, KA_OutU, KA_OutL, B_DivL, Cryo_1, Cryo_2. Cryo_1 and Cryo_2 are new and installed in the upper cryopump 'gas-box'.Mar 13 2007 10:01PM (labombard)Brian LaBombardPenning Ionization Gaugeslabombard
Plasma TVyesWIDE1, WIDE2 (chan 33), DANT, JANT and DIV (without an ND filter) are operational.Mar 27 2007 01:53PM (terry)Jim TerryVideo of plasma via several camerasdekow,irby,terry,wolfe
PolarimeternoRetro shutter closedMar 15 2007 11:00AM (irby)Jim IrbyC-Mod Polarimeterirby
Probe Array - ISDyesProbes at 15 of the 16 poloidal locations ('ramped' geometry) are operational. n,Te, particle flux reported every 10 ms.Mar 13 2007 09:47PM (labombard)Brian LabombardInner Lower Divertor Probe Arraylabombard
Probe Array - LHyesAll 6 probes on the lower hybrid launcher (tungsten, 1.5 mm projection) are operational. Mar 13 2007 09:46PM (labombard)Brian LaBombardLower Hybrid Probe Arraylabombard
Probe Array - OSDyesProbes at 10 poloidal locations (one 'ramped' and one 'flush') are operational. n,Te, particle flux reported every 10 ms. Mar 13 2007 09:45PM (labombard)Brian LaBombardOuter Lower Divertor Probe Arraylabombard
Probe Array - UDIVyes14 probes are operational ('ramped' geometry). n,Te, particle flux reported every 10 ms.Two high-gain channels are installed and operational for ECDC plasma diagnosis.Mar 13 2007 12:46PM (labombard)Brian LaBombardUpper Divertor Probe Arraylabombard
ReflectometeryesAll low frequency channels (50, 60, 75, 88, 88u, 88l and 110) have been installed (for the first time!) on 1070626. 88's, 75 and 110 have been working properly, 50 hasn't up to date (but I went in today and it may be fixed) and 60 has been a little noisy (also went in today and took a look)....I'll post follow ups on their status. The PPPL variable frequency channel is installed and being run (for now) at fixed frequency (Voltage set at 6V). 140GHz channel still down. Jun 28 2007 09:08AM (dominguez)Earl MarmarReflectometerdominguez,marmar
Scanning Probe - ASPyesPerpendicular Mach probe head is installed. The drive is operational.Mar 15 2007 09:09PM (labombard)Brian LaBombardA-Port Scanning Probelabombard
Scanning Probe - FSPyesPerpendicular Mach probe head is installed. The probe drive is now operational.Mar 13 2007 09:51PM (labombard)Brian LaBombardF-Port Scanning Probelabombard
Scanning Probe - WASPyesA-port WASP successfully collected data on 04/04. Some intermittent open circuit on SE Electrode. K-WASP took data on 04/11. Both ready to scan as needed.Apr 12 2007 04:18PM (nsmick)Brian LaBombardInner Wall Scanning Probelabombard,nsmick
TCIyesNL_04 at 71.04 cm from 3/9/07 calibrationMay 23 2007 07:38AM (irby)Jim IrbyTwo-Color Interferometerirby,murray
TS - Coreyes2 lasers, 11 spatial channels. Density is calibrated based on ECE cutoffs obtained on 4 May 2007. Additional spatial channels will be added over the next several weeks.May 10 2007 10:37AM (jwhughes)Jerry HughesCore Thomson scatteringjwhughes
TS - Edgeyes2 lasers, 20 spatial channels. Fiber holder position is -124.5mm (standard). Density is calibrated according to Raman scattering results from 26 May, supplemented with ECE cutoffs obtained on shot 1070531015.Jun 14 2007 06:21PM (jwhughes)Jerry HughesMillimeter resolution edge Thomson scatteringjwhughes
Vispecyes350 and 670 nm, 4 spatial channelsMay 04 2007 01:02PM (agraf)Alex GrafVisible Spectrometeragraf