Alcator C-Mod Run 1070702 Information

Back to runs

Miniproposals
Miniproposal:430
Date Filed: 7/26/2005
Title:Localization of the effect of boronization
First Author:Bruce Lipschultz
Session Leader:Bruce Lipschultz (shots 1-1)
Session Leader:Dennis Whyte (shots 2-2)

Operators
Session leader(s):Bruce Lipschultz,Dennis Whyte
Physics operator(s):Jim Irby
Engineering operator(s):

Engineering Operator Run Comment
automatically entered by signon - please replace with real comment

Session Leader Plans
Entered: Jul 2 2007 09:18:52:910AM
Author: Bruce Lipschultz
The goals of today's study are several -

1) repeat deposition profile measurements with resonance at 60 cm.
2) vary pressure to see if that optimizes the deposition rate
3) vary throughput (one vs 2 turbos)
4) Effect of 20% diborane.


We start this day by using He from the regular bottle through the diborane piezo. We can then tell if there are issues with the control of pressure. If we need to go to fixed gas rate (% piezo opening) we will figure it out then.
After that I would propose we open the diborane bottle and proceed with our experiments.


Physics Operators Plans
Entered: Jul 2 2007 07:57:16:363AM
Author: Jim Irby
Cell will be closed at 8:30 today in preparation for the boronization tests

. cool down EF4s
. prepare S^3 for operation
. close cell and power room by 8:30
. secure power room and cell by 9:00 AM
. transport 20% diborone bottle to cell

. test diborane gas feedback system with helium purge gas
. measure profile with resonance at 60 cm.
. pressure scan
. vary throughput (one vs 2 turbos, run RGA-1)
. change to 20% diborane bottle and note effects

Session Leader Summaries

Physics Operator Summaries

Session Leader Comments
Jul 2 2007 10:12:36:980AM1070702605Dennis WhyteFiducial shot with no gas puffing, no ECDC
in order to establish zeros on the gauges, etc.

Jul 2 2007 10:20:00:393AM1070702606Dennis WhyteFiducial shot with no gas puffing, no ECDC

Blip is adjusting zeroes on gauges
Jul 2 2007 11:18:21:173AM1070702607Dennis WhyteThe manometer will not work properly with ECDC on,
so we'll have to rely on the convectron gauge back in S3
and the foreline pressure on TC6