Alcator C-Mod Run 1070731 Information

Back to runs

Miniproposals
Miniproposal:491
Date Filed: 5/25/2007
Title:MSE Calibrations
First Author:Steve Scott
Session Leader:Steve Scott (shots 1-1)
Session Leader:Steve Scott (shots 1-20)

Operators
Session leader(s):Steve Scott,Steve Scott
Physics operator(s):
Engineering operator(s):

Engineering Operator Run Comment
MSE beam-into-gas, no fields Steve Scott

Session Leader Plans
Entered: Jul 31 2007 10:09:04:420AM
Author: Steve Scott
MSE beam-into-gas, no fields

Start times: DNB conditioning starts at noon. shots into torus starting 1PM

Purpose: measure the Doppler shift with a spectrometer looking through the MSE optics on a series of
beam-into-gas shots with no toroidal or poloidal fields. Observe whether the Doppler shift
changes shot-to-shot or within a shot. Such changes would imply that the MSE optics change their
viewing angle with respect to the beam and would potentially explain the poor reproducibility
of MSE.

Reference shot: 1070402013. Fill gas = helium (he4). Desired pressure about 3 mTorr as measured on MKS.
My understanding (based on data 5/29/2007) is that the g-side ratiomatic
reads about a factor of 2 less than the MKS. We should check whether the
same ratio holds for today. Assuming that it does, then we would want
a pressure of 1.0 - 1.5 mTorr as indicated on the ratiomatic.

Run plan:

1. If the desired torus pressure is not achieved, we may take a couple of shots to adjust the pressure to
the desired range.

2. 10-20 shots, DNB into torus, no fields. DNB modulated: 80ms on, 20 ms off. start 0.4 sec. total
duration = 1.8 sec. end at about 2.2 sec. total 'on' time = 1.44 sec. Bill Rowan will be operating
the spectrometer remotely. MSE channel 9 (near optical axis) will be used.

3. 10 shots, DNB un-modulated. start 0.4 sec, end 1.8 sec. More spectrum measurements. still MSE ch 9.

4. Swap MSE channel to channel #0 (edge channel). Take about 10 shots, with either a modulated or un-modulated
DNB. Choice will depend on which one provides more photons as measured in steps #2 and #3.

5. (Much lower priority) 5-10 shots where we deliberately reduce te DNB current. purpose is to correlate
the spectral intensity with DNB current.

Physics Operators Plans
Entered: Jul 31 2007 09:29:29:547AM
Author: Steve Wolfe
Physop plan and engineering setup for Tuesday, July 31,2007

Beam into gas (no fields)
MP: MP#460 ? MSE testing
SL: Scott
PO: wolfe
EO: Toland

Gas setup: 
 Use only B-top (PG4) gas feed.
 Load B-top with 6 psi of Helium.
 All other valves disabled. 
TORVAC valves open 

Eng Control panel 
  Load from 1070402013
  Note: All supplies OFF, fizzle detector should be set On=2.0, Off=2.1

Plan: Run repeated beam into gas shots to monitor possible changes in 
 MSE optics. 

Session Leader Summaries
Entered: Jul 31 2007 03:55:47:673PM
Author: Steve Scott
We had a successful day firing the dnb into the torus with no fields.

DNB performance and performance of the gas system were both very reliable.

shots 002 to 012

dnb = modulated (80 ms on, 20 ms off).
fibers from MSE channel 9 into McPherson spectrometer.
gas about 3.0 mTorr (peak), drooping to 1.8 at the end of the dnb,
as measured by the g-side MKS.

only bad shot was shot 10 -- dnb faulted.

DNB current on shots 003 thru 009 very similar. current was a
little lower on shot 002. the current waveform differed on shots 11
and 12 because Dexter tweaked the gas following the faults on shot 10.

this is a good set to look for possible motion of MSE optics

shots 013 thru 020

dnb = unmodulated. current not as high, typically 5. drooping to 4.0
gas still 2.8 mTorr max
this is a good set to look for motion of MSE optics with an
unmodulated beam.

Jinseok also checked the temperature of the DNB drift tube during shots with
a Raytek noncontact thermometer. He saw no changes larger than 1 degree
Farenheit.

Physics Operator Summaries

Session Leader Comments
Jul 31 2007 12:58:33:367PM1070731001Steve ScottFirst shot planned for about 1:00 PM. Starting with reference shot 1070402013.
Expected torus pressure = 3 mTorr (g-side mks), = 1.5 mTorr (g-side-ratiomatic)
MSE channel 9 being fed into the McPherson 1-meter monochrometer.
DNB timing: 80 ms on, 20 ms off. Start 0.4 sec, end 2.2 sec.

Result: 1) actual gas delivered looks very small or non-existant.
2) no DNB either.


Jul 31 2007 01:11:14:287PM1070731002Steve ScottRep rate for these shots will be about 6 minutes - limited by data acquisition.

Result: good pressure: peak = 3.0 on MKS, about 1.3 on g-side ratiomatid
good DNB. current = 6.0 amps, quickly dropping to about 5.2 no faults.

Jul 31 2007 01:22:40:817PM1070731003Steve Scottgood dnb. But it appears that we have a data hangup. investigating ...

Two minutes later: data have arrived. peak pressure same as shot 2.
ther is a little more pressure droop as time progresses, but OK overall.
dnb current: peak = 6.0, but a little less droop. about 5.4 amps
at the end.
Jul 31 2007 01:27:59:743PM1070731004Steve ScottDNB: good. 6.2 amps, decreasing to 5.5
gas: good.
Jul 31 2007 01:37:30:457PM1070731005Steve Scottdnb: good. 6.4 amps, drooping to 5.5
gas: good
Jul 31 2007 01:46:33:233PM1070731006Steve ScottDNB: good. 6.4 --> 5.5 amps.
gas: good
Jul 31 2007 01:58:12:280PM1070731007Steve ScottDNB: good. 6.4 --> 5.5 amps
gas: good
Jul 31 2007 02:04:52:210PM1070731008Steve ScottDNB: good. 6.4 --> 5.6 amps
gas: good
Jul 31 2007 02:16:03:080PM1070731009Steve ScottDNB: good. about 5.6 amps
gas: good.
Jul 31 2007 02:23:41:603PM1070731010Steve ScottDNB: multiple faults. only 3 beam blips. this is not a member of the scan.
gas: ok
Jul 31 2007 02:36:46:493PM1070731011Steve ScottDexter tweaked dnb gas in test cell. Hopefully this will eliminate the faults that we experienced on the
previous shot.

dnb current is lower than on previous shots, particularly at the beginning,
where we got only about 4.5 amps.

gas: good
Jul 31 2007 02:45:06:770PM1070731012Steve ScottDNB: about the same as shot 11. 4.5 amps early, rising to 5.5
gas: good
Jul 31 2007 03:01:37:460PM1070731013Steve ScottStarting this shot, we will go to an UNMODULATED dnb. timing = 0.4 --> 1.8 sec.
still using MSE channel 9 into the spectrometer.

result: unmodulated, 5.0 --> 4.0
ghas: ok.
Jul 31 2007 03:04:26:230PM1070731014Steve ScottDNB: unmodulated. current a little lower than on shot 13. generally OK.
gas: good
Jul 31 2007 03:14:22:990PM1070731015Steve ScottDNB: current drooping some more, to about 4.0 amps for more than half the shot. but no faults.
gas: good
Jul 31 2007 03:23:08:877PM1070731016Steve ScottDNB: current history like previous shot. 1 late fault.
gas: good
Jul 31 2007 03:32:25:083PM1070731017Steve ScottDNB: current similar to shots 15, 16. no faults.
gas: good
Jul 31 2007 03:41:20:077PM1070731018Steve ScottDNB: a little more current than previous shots. good.
gas: good
Jul 31 2007 03:52:21:580PM1070731019Steve ScottDNB: good. like previous shot.
gas: good.

next shot will be the last.
Jul 31 2007 04:00:42:100PM1070731020Steve ScottDNB: one fault, otherwise ok. current is higher.
gas: good

------------------- END OF RUN -------------------------

Physics Operator Comments
Jul 31 2007 11:54:55:120AMSteve Wolfe
Load from 1070402013
Change seg1 title to "beam into gas tests NO FIELDS"
Zero wires 4 and 6 (not used); the zxl and bz_0 predictors still had
f10 f11 and f19 weights. 

The pg4 puff is on from -1.59sec to 0. Supposedly, this gave what 
they wanted before. 

Load into model
Load at 31-Jul-2007 11:54:05.00
Open tree /home/wolfe/pcs_scratch      -1
Open tree done

Should be ready for first shot, whenever the beam guys are ready.

Jul 31 2007 01:00:27:477PMSteve Wolfe
SHot#1: Gas wasn't enabled correctly. Was on fill select
instead of hybrid enable.

Try again. 
Jul 31 2007 01:07:51:440PMSteve Wolfe
SHot#2: Try again. 
Got gas this time. Just over 1mTorr on ratiomatic
Got beam. 

Steve is satisfied with puff. 

Engineering Operator Comments
ShotTimeTypeStatusComment
112:55:43:550PMTestBadHybrid Select fof Gas Missed
201:04:37:437PMTestOk
301:13:48:447PMTestOk
401:25:15:193PMTestOk
501:34:15:563PMTestOk
601:43:58:607PMTestOk
701:52:45:710PMTestOk
802:02:32:457PMTestOk
902:11:39:920PMTestOk
1002:21:10:600PMTestOk
1102:33:50:570PMTestOk
1202:42:38:643PMTestOk
1302:52:55:920PMTestOk
1403:01:59:553PMTestOk
1503:11:54:597PMTestOk
1603:20:36:187PMTestOk
1703:29:53:120PMTestOk
1803:38:59:657PMTestOk
1903:48:55:843PMTestOk
2003:57:55:067PMTestOk

System Availability
Jul 31 2007 12:55:16:940PM
SystemAvailableCommentAvailability
Changed
OwnerDescriptionWho Can
Change
Active MHDyesFully operationalMar 28 2007 09:19PM (sears)Joe SnipesActive MHD Antennasears,snipes
BESyesvertical array of six views at R ~ 87.3 cm, radial array of three views out to R ~ 89 cm, both at outboard midplane. Three discrete fiber views of beam at R ~ 70.0, 76.5 and 82.5 cm.Feb 06 2007 10:00AM (bravenec)Ron BravenecBeam Emission Spectroscopybravenec
Bolometersyes2Pi diode operational, 2Pi foil being upgrade for noise resistance. Midplane foils calibrated and operational, data in the tree. Outboard midplane AXUV calibrated and operational, data in the tree. Full midplane and divertor views being digitized.Jul 19 2007 04:48PM (mlreinke)Matt ReinkeBolometershutch,mlreinke
BoronizationyesFirst Boronization 5/9/07May 10 2007 04:55PM (irby)Jim IrbyBoronizationdekow,irby
CNPAno Jan 14 2007 02:06PM Ron ParkerCompact Neutral Analyzerparker
Cryopump - Upper DivertoryesThe cryopump is operational in a manual mode. We will be working on interfacing its control system with the C-Mod state machine for semi-automatic operation.May 30 2007 07:58PM (labombard)Brian LaBombardUpper Divertor Cryopumpirby,labombard
CXRSyesThe pedestal toroidal array was installed, spatially and absolutely calibrated. There are 10 chords intersect the beam at the major_R_range=[75,4:87.4]cm. Spatial resolution: 1.2 cm. The F-top poloidal array was installed and spatially calibrated. There are 19 chords intersect the beam at the major_R_range=[67.8:91.0]cm. Spatial resolution: 1.22 cm.Jun 20 2007 05:35PM (bespam)Bill RowanCharge Exchange Spectroscopybespam,blip,rachmcd,rowan
D Alphayes\ha2_bright and \SPECTROSCOPY::TOP.VUV.VIS_SIGNALS:MCP_VIS_SIG1 are availableMar 11 2007 10:28PM (terry)Jim TerryBalmer Alpha Monitorirby,terry,wolfe
Dalsayescamera is re-installed, and view has been calibratedFeb 26 2007 03:28PM (marmar)Earl MarmarVisible Bremsstrahlungmarmar
Disruption Mitigationno Jan 14 2007 02:02PM Bob GranetzGas Jetgranetz,whyte
Divertor IRno Mar 06 2007 10:50AM Glen WurdenDivertor IR Digital Videogwurden,terry
Divertor IR Spectroscopyno Mar 06 2007 10:51AM Glen WurdenDivertor IR Spectroscopygwurden,ldorf,terry
DNByesAll systems are operational, but as of 28 March 2007, the beam needs additional conditioning for full performance shotsMar 28 2007 09:36AM (beals)Bob GranetzDiagnostic Beambeals,granetz
FRC-ECEnoOnly high frequency channels now working (17 to 32). Ch 22 has bad video amp. LO is OK. Problem looks to be in mixer amp unit.Jul 11 2007 10:52AM (phillips)Perry PhillipsHeterodyne ECE-32 channelsphillips
G-Side RatiomaticyesThe gauge is operational.Mar 30 2007 10:54AM (labombard)Brian LaBombardG-side ratiomatic ionization gaugelabombard,toland
GPCyesChs 1-8 operational. May 24 2007 08:40AM (hubbard)Amanda HubbardECE Polychromaterhubbard,schmidt
GPC2yesMost channels (except 1,5,8,10,17) operational. Jun 08 2007 08:29AM (hubbard)Amanda HubbardECE Polychromaterhubbard,schmidt
GPIyesNew view just outside of the lower X-point is operational, but data is being archived locally , not to the tree. Midplane GPI is not operational.Mar 27 2007 01:59PM (terry)Jim TerryGas Puff Imagingterry,zweben
H/Dyesspectral calibration needs to be checked, but system should be ready for the runFeb 26 2007 03:28PM (marmar)Earl MarmarH/D Ratiomarmar
Hirex Jryes Mar 06 2007 08:59AM (rice)John RiceHigh Resolution X-Rayince,rice
Hirex SryesStoring Data, but profiles not yet availableApr 27 2007 01:20PM (ince)John RiceHigh Resolution X-Rayince,rice
HXRyesSomething might be wrong with channel 26; all others working fine.Mar 23 2007 10:43AM (schmidt)Ron ParkerHard X-Ray Arrayparker,schmidt
ICRF D-Antennayes80.5 MHz. Max power is about 1.5 MWJun 20 2007 09:23AM (ylin)Steve WukitchD-port antennawukitch,ylin
ICRF E-Antennayes80 MHz. Maxc power is about 1.5 MWJun 20 2007 09:23AM (ylin)Steve WukitchE-port antennawukitch,ylin
ICRF J-Antennayes78 MHz. 4-strap configuration. Max ~ 3 MW.Jul 17 2007 03:52PM (ylin)Steve WukitchJ-port antennawukitch,ylin
Lower Hybridno6B Pressure window failed. Arcing in 6B, 8B,9B & 23B. May 04 2007 03:44PM (dterry)Ron ParkerLower Hybriddterry,parker
LPIyes Apr 26 2007 08:35PM (bose)Earl MarmarLithium Pellet Injectorbose,marmar
Ly-alphayesMidplane calibrated and operational, data being stored in tree. Divertor views showing significant noise, and not available. Upper view being digitized and analyzed.Jul 19 2007 04:49PM (mlreinke)Matt ReinkeLyman alpha cameraslabombard,mlreinke
McPhersonyesWorking normally.Apr 19 2007 09:52AM (mlreinke)Jim TerryVUV Spectrometermlreinke,terry
MichelsonyesOperational. Mar 28 2007 08:33AM (hubbard)Amanda HubbardECE Te Profileshubbard,schmidt
MKSyesE-Top, E-Bottom, F-top (cryo), G-side, and B-bottom are operational. F-top gauge has a 'sniffer tube' to measure pressures in cryopump gas box.Mar 20 2007 06:42PM (g)Brian LaBombardBaratron pressure gaugeslabombard
MLPnoInitial tests on the A-port scanning probe went well. We need to take care of some control glitches that caused some component failures. May 30 2007 08:01PM (labombard)Brian LaBombardMirror Langmuir Probe Electronics Systemlabombard
MSEno(1) Bandpass filters replacement underway (2) Beam-into-gas spectrum measurement underway (3) Shutter mechanism currently unstable; remains open. (4) Expect beam-into-gas calibrations complete about 4/24. Apr 03 2007 12:05PM (sscott)Steve ScottMotional Stark Effectjinseok,sscott
NeutronsyesBoth neutron systems are fully operationalApr 05 2007 11:19AM (fiore)Catherine FioreNeutron Detectorsfiore
Neuts and HardsyesoperationalMar 21 2007 04:29PM (irby)Jim IrbyX-Ray/Neutron Detector at K-Horirby
NINJAyesAll capillaries displayed by the NINJA_CONTROL GUI have been tested. NINJA puffed on 1070329009 (A-B limiter).Mar 29 2007 11:47AM (terry)Brian LaBombardNINJA gas pufflabombard,terry
PCIyesThe system is to be optimized.May 30 2007 01:12PM (lianglin)Miklos PorkolabPhase Contrast Imagingedlund,lianglin,ntsujii,porkolab
Penning Gaugesyes8 in-situ Penning ionization gauges appear to be operational: CB_DivU, CB_OutL, CB_DivL, KA_OutU, KA_OutL, B_DivL, Cryo_1, Cryo_2. Cryo_1 and Cryo_2 are new and installed in the upper cryopump 'gas-box'.Mar 13 2007 10:01PM (labombard)Brian LaBombardPenning Ionization Gaugeslabombard
Plasma TVyesWIDE1, WIDE2 (chan 33), DANT, JANT and DIV (without an ND filter) are operational.Mar 27 2007 01:53PM (terry)Jim TerryVideo of plasma via several camerasdekow,irby,terry,wolfe
PolarimeternoRetro shutter closedMar 15 2007 11:00AM (irby)Jim IrbyC-Mod Polarimeterirby
Probe Array - ISDyesProbes at 15 of the 16 poloidal locations ('ramped' geometry) are operational. n,Te, particle flux reported every 10 ms.Mar 13 2007 09:47PM (labombard)Brian LabombardInner Lower Divertor Probe Arraylabombard
Probe Array - LHyesAll 6 probes on the lower hybrid launcher (tungsten, 1.5 mm projection) are operational. Mar 13 2007 09:46PM (labombard)Brian LaBombardLower Hybrid Probe Arraylabombard
Probe Array - OSDyesProbes at 10 poloidal locations (one 'ramped' and one 'flush') are operational. n,Te, particle flux reported every 10 ms. Mar 13 2007 09:45PM (labombard)Brian LaBombardOuter Lower Divertor Probe Arraylabombard
Probe Array - UDIVyes14 probes are operational ('ramped' geometry). n,Te, particle flux reported every 10 ms.Two high-gain channels are installed and operational for ECDC plasma diagnosis.Mar 13 2007 12:46PM (labombard)Brian LaBombardUpper Divertor Probe Arraylabombard
Reflectometeryes50, 60, 75 88, 88l,88u and 110 are operational. 60GHz is a little noisy. Variable frequency channel set at 6.473V (~130GHz), not a clear signal.Jul 26 2007 07:14AM (dominguez)Earl MarmarReflectometerdominguez,marmar
Scanning Probe - ASPyesPerpendicular Mach probe head is installed. The drive is operational.Mar 15 2007 09:09PM (labombard)Brian LaBombardA-Port Scanning Probelabombard
Scanning Probe - FSPyesPerpendicular Mach probe head is installed. The probe drive is now operational.Mar 13 2007 09:51PM (labombard)Brian LaBombardF-Port Scanning Probelabombard
Scanning Probe - WASPyesA-port WASP successfully collected data on 04/04. Some intermittent open circuit on SE Electrode. K-WASP took data on 04/11. Both ready to scan as needed.Apr 12 2007 04:18PM (nsmick)Brian LaBombardInner Wall Scanning Probelabombard,nsmick
TCIyesNL_04 at 71.04 cm from 3/9/07 calibrationMay 23 2007 07:38AM (irby)Jim IrbyTwo-Color Interferometerirby,murray
TS - Coreyes2 lasers, 10 spatial channels. Density is calibrated based on ECE cutoffs obtained on 4 May 2007. Additional spatial channels will be added over the next several weeks.Jul 11 2007 09:18AM (jwhughes)Jerry HughesCore Thomson scatteringjwhughes
TS - Edgeyes2 lasers, 20 spatial channels. Fiber holder position is -120.0mm. Density is calibrated according to Raman scattering results from 26 May, supplemented with ECE cutoffs obtained on shot 1070531015.Jul 24 2007 06:49PM (jwhughes)Jerry HughesMillimeter resolution edge Thomson scatteringjwhughes
Vispecyes350 and 670 nm, 4 spatial channelsMay 04 2007 01:02PM (agraf)Alex GrafVisible Spectrometeragraf