| Miniproposals | ||||||||||
|
| Operators | |
| Session leader(s): | Alex Ince-Cushman |
| Physics operator(s): | Joseph Snipes |
| Engineering operator(s): | Bill Parkin,Sam Pierson,Bill Parkin,Gary Dekow |
| Engineering Operator Run Comment |
| Session Leader: Alex Ince-Cushman PhysOp: Bill Rowan |
| Session Leader Plans |
| Entered: Mar 12 2008 08:28:04:003AM |
| Author: Alex Ince-Cushman |
| MP 523: Effect of Lower Hybrid Current Drive on Rotation Profile
Session Leader: Alex Ince-Cushman PhysOp: Joe Snipes The goal of this experiment is to add to the LHCD rotation database, particularly at 60 and 120 degree phasing. If time permits we will try for a trophy shot/fiducial pair. Diagnostics: Hirex-sr, Thomson, Edge Thomson, Bolometer arrays, McPherson, PCI, Hard X-ray Camera Step 1: locked mode discharge for Hirex-Sr calibration (2 shots) - Repeat discharge 1080228024 Step 2: Bring LH up to full power (6 shots) - Use shot 1080226006 as a starting point (60 degree phasing) - Ramp up LH power to 1MW Step 3: Density Scan (14 shots) - Run the LH system at the highest power at which it performs without tripping (60 degrees). - Progressively reduce the density to as low a value as possible (4 shots) - Switch to 120 degree phasing and perform density scan from the lowest value obtained on the 60 degree density scan up to 1.0 nl04. (~6 shots) Step 4: Trophy Shot development - Adjust ne, Ip, Bt at maximum LH power to produce the largest possible change in rotation profile. Note we would be going to lower fields, not higher fields. - Once trophy shot is produce, run an identical discharge without any LH for comparison. Other: Depending on the status of the neutral beam we will likely have ~3 �??beam into gas�?� discharges to help calibrate the MSE diagnostic. |
| Physics Operators Plans |
| Entered: Mar 12 2008 07:17:29:440AM |
| Author: Joseph Snipes |
| Engineering setup for Wednesday 12 Mar 2008
MP 523: Effect of Lower Hybrid Current Drive on Rotation Profile Session Leader: Alex Ince-Cushman PhysOp: Joseph Snipes Normal overnight ECDC in D2 and 60 C bake. Run will start at 09:00 and end at 17:00. Power systems as on 1080214002 A-coil configuration: +Dtop -Dbot -Jtop +Jbot; Enabled (standard configuration) Gas Setup: fill B-Top with 6 psi D2 Hybrid enabled (PG4) fill B-side lower with 1 psi Ar Hybrid ENABLED (PG1) leave B-side upper as is Hybrid DISABLED (PG2) fill B-main (C-side) with 40 psi D2 Hybrid enabled (PG3) leave NINJA as is DISABLED Enable the following gate valves and shutters, assuming no vacuum problems: ECE, VUV, Z-bolo Torvac setup: keep gate valves open Diagnostics: Hirex-sr, Thomson, Edge Thomson, Bolometer arrays, McPherson, PCI, Hard X-ray Camera ------------------------------------------------------------------------------- ICRF Setup: not needed LH Setup: yes DNB: yes for calibration Cryopump: not needed ------------------------------------------------------------------------------- The goal of this experiment is to add to the LHCD rotation database, particularly at 60 and 120 degree phasing. If time permits we will try for a trophy shot/fiducial pair. Step 1: locked mode discharge for Hirex-Sr calibration (2 shots) - Repeat discharge 1080214002 800kA, LSN, nl04=5e19 dropping negative after 0.7sec, Acoil=4kA, b21_proj off. Step 2: Bring LH up to full power (6 shots) - Use shot 1080215014 as a starting point (60 degree phasing) - Ramp up LH power to 1MW Step 3: Density Scan (14 shots) - Run the LH system at the highest power at which it performs without tripping (60 degrees). - Progressively reduce the density to as low a value as possible (4 shots) - Switch to 120 degree phasing and perform density scan from the lowest value obtained on the 60 degree density scan up to 1.0 nl04. (~6 shots) Step 4: Trophy Shot development - Adjust ne, Ip, Bt at maximum LH power to produce the largest possible change in rotation profile. Note we would be going to lower fields, not higher fields. - Once trophy shot is produce, run an identical discharge without any LH for comparison. Other: Depending on the status of the neutral beam we will likely have ~3 �??beam into gas�?� discharges to help calibrate the MSE diagnostic. |
| Session Leader Summaries |
| Entered: Mar 12 2008 05:48:22:337PM |
| Author: Alex Ince-Cushman |
| MP 523: Effect of Lower Hybrid Current Drive on Rotation Profile
The locked mode portion of the experiment was successful. The third shot of the day was a locked mode discharges with sufficient argon levels to provide a good wavelength calibration for Hirex-Sr. The first half of the day was spent trying to get good LH coupling at 60 degree phasing. We had some trouble getting to the low density required for accessibility at 60 degrees and so after shot 17 we switched to 75 degree phasing. After 4 shots at 75 we changed the phase to 120 and then to 105, but had no success coupling any LH power. We then switched the phasing back to 60 and increased the toroidal field to help accessibility. These conditions provided the best shots of the day (27 & 28 in particular are very nice). The LH effect on central rotation and some key plasma parameters are summarized in the table below. Shot Bt nl04 LH Power LH Phase DeltaV # [T] [1e19m2] [kW] [deg] [-km/s] 2 5.4 0.5 0 60 0 3 locked mode 4 5.4 0.6 900 60 -8 5 5.4 0.6 950 60 -5 6 5.4 0.6 1000 60 -8 7 5.4 0.55 1000 60 -10 8 fizzle 9 fizzle 10 5.4 0.45 900 60 -35 11 5.4 0.4 1000 60 -36 12 5.4 0.4 1000 60 -36 13 5.4 0.35 1000 60 -20 14 5.4 0.42 850 60 -40 15 fizzle 16 5.4 0.44 920 60 -30 17 5.4 0.44 900 60 -30 18 early disruption 19 5.4 0.45 980 75 -20 20 5.4 0.35 1000 75 -30 21 unintentional locked mode 22 5.4 0.40 980 75 -40 23 5.4 0.40 920 75 -40 24 5.4 0.30 0 120 0 25 5.4 0.30 0 105 0 25 5.4 0.30 0 105 0 26 6.2 0.30 0 60 0 27 6.2 0.40 860 60 -40 28 6.2 0.40 940 60 -40 29 6.2 0.40 940 60 -30 |
| Entered: Mar 12 2008 05:48:28:540PM |
| Author: Alex Ince-Cushman |
| MP 523: Effect of Lower Hybrid Current Drive on Rotation Profile
The locked mode portion of the experiment was successful. The third shot of the day was a locked mode discharges with sufficient argon levels to provide a good wavelength calibration for Hirex-Sr. The first half of the day was spent trying to get good LH coupling at 60 degree phasing. We had some trouble getting to the low density required for accessibility at 60 degrees and so after shot 17 we switched to 75 degree phasing. After 4 shots at 75 we changed the phase to 120 and then to 105, but had no success coupling any LH power. We then switched the phasing back to 60 and increased the toroidal field to help accessibility. These conditions provided the best shots of the day (27 & 28 in particular are very nice). The LH effect on central rotation and some key plasma parameters are summarized in the table below. Shot Bt nl04 LH Power LH Phase DeltaV # [T] [1e19m2] [kW] [deg] [-km/s] 2 5.4 0.5 0 60 0 3 locked mode 4 5.4 0.6 900 60 -8 5 5.4 0.6 950 60 -5 6 5.4 0.6 1000 60 -8 7 5.4 0.55 1000 60 -10 8 fizzle 9 fizzle 10 5.4 0.45 900 60 -35 11 5.4 0.4 1000 60 -36 12 5.4 0.4 1000 60 -36 13 5.4 0.35 1000 60 -20 14 5.4 0.42 850 60 -40 15 fizzle 16 5.4 0.44 920 60 -30 17 5.4 0.44 900 60 -30 18 early disruption 19 5.4 0.45 980 75 -20 20 5.4 0.35 1000 75 -30 21 unintentional locked mode 22 5.4 0.40 980 75 -40 23 5.4 0.40 920 75 -40 24 5.4 0.30 0 120 0 25 5.4 0.30 0 105 0 25 5.4 0.30 0 105 0 26 6.2 0.30 0 60 0 27 6.2 0.40 860 60 -40 28 6.2 0.40 940 60 -40 29 6.2 0.40 940 60 -30 |
| Physics Operator Summaries |
| Entered: Mar 12 2008 05:06:34:540PM |
| Author: Joseph Snipes |
| We had a slow start to the run due to some problems in the cell that meant we were not ready to close the doors until 9:00 am and then they turned on the cooling and it took almost an hour before the machine was ready. Once the machine was ready, it ran well except for a spurious alternator fault on shot 8 followed by a data system hang up on shot 9. Otherwise, there was only one fizzle and 26 plasmas out of 29 shots. The LH system also ran well and coupled over 1 MW on many shots. See the session leader summary for details of the physics results.
Shot,Duration(s),Ipmax(MA),Result Shot,Duration(s),Ipmax,Result | 01 1.983 0.788 Plasma | 02 1.980 0.782 Plasma | | 03 1.710 0.822 Plasma | 04 1.968 0.778 Plasma | | 05 1.967 0.780 Plasma | 06 1.966 0.777 Plasma | | 07 1.967 0.778 Plasma | 08 0.000 0.002 Dud | | 09 0.000 0.002 Dud | 10 1.975 0.781 Plasma | | 11 1.973 0.784 Plasma | 12 1.975 0.784 Plasma | | 13 1.978 0.782 Plasma | 14 1.888 0.781 Plasma | | 15 0.000 0.043 Fizzle | 16 1.902 0.778 Plasma | | 17 1.897 0.777 Plasma | 18 0.280 0.669 Plasma | | 19 1.898 0.778 Plasma | 20 1.633 0.786 Plasma | | 21 1.610 0.773 Plasma | 22 1.863 0.783 Plasma | | 23 1.955 0.781 Plasma | 24 1.602 0.766 Plasma | | 25 1.690 0.775 Plasma | 26 1.757 0.774 Plasma | | 27 1.559 0.790 Plasma | 28 1.967 0.788 Plasma | | 29 1.948 0.786 Plasma Date 1080312: 29 shots. 26 plasmas. 1 fizzles. 2 duds. 0 System Error. (Plasma = Ip gt 60.0 kA Dud = H_alpha lt 0.40 V) Total plasma duration: 46.89 s (exceeding -78 kA) |
| Session Leader Comments | |||
| Mar 12 2008 10:24:06:410AM | 1080312001 | Alex Ince-Cushman | Successful full length shot. ~705kW of LH power coupled. We will repeat this shot with increased LH power. For the next shot we will also have the DNB firing. |
| Mar 12 2008 11:04:56:317AM | 1080312002 | Alex Ince-Cushman | Another full length shot however there was a problem with the LH system and no power was coupled. The DNB fired successfully and did not seem to cause a change in density. As a result the DNB will be permitted to fire for the rest of the day.
For the next shot we will run a locked mode discharge to provide a spectral calibration for Hirex-Sr and to give the LH team time to trouble shoot. |
| Mar 12 2008 11:05:39:160AM | 1080312003 | Alex Ince-Cushman | Full length shot. Locked mode came in at ~0.7s and persisted until ~1.5s. The argon signal level is good and so we will go back to LH shots. |
| Mar 12 2008 11:04:20:363AM | 1080312004 | Alex Ince-Cushman | Full length shot. ~880kW of LH coupled at 60 degree phasing. There was an unidentified density increase at 0.87s which seems to reduce the effectiveness of the LHCD. There was a modest change in central toroidal rotation, ~ -10km/s. |
| Mar 12 2008 11:12:23:207AM | 1080312004 | Alex Ince-Cushman | Full length shot. ~880kW of LH coupled at 60 degree phasing. There was an unidentified density increase at 0.87s which seems to reduce the effectiveness of the LHCD. There was a modest change in central toroidal rotation of order -5km/s. |
| Mar 12 2008 11:15:49:643AM | 1080312005 | Alex Ince-Cushman | Full length shot. ~1000kW of LH coupled at 60 degree phasing with one trip. The unidentified density increase seen on the previous shot did not re-appear. The change in central toroidal rotation was still relatively small ~ -7km/s. For the next shot we will keep the LH power the same and attempt to lower the density. |
| Mar 12 2008 11:30:39:190AM | 1080312006 | Alex Ince-Cushman | Full length shot. ~1000kW of LH coupled at 60 degree phasing with one trip. The density is still too high ( nl04 ~0.6) and so for the next shot we will reduce the target density. The outer gap will be increased to try an prevent the LH from tripping. |
| Mar 12 2008 11:51:56:807AM | 1080312007 | Alex Ince-Cushman | Full length shot with LH up to 1.1MW with no trips. The density remains too high, which is reducing LHCD efficiency and accessibility. The DNB may be increasing the density and so for the next shot the DNB start time will be delayed to 0.86s. The LH induced a Delta V of ~ -12k/ms. |
| Mar 12 2008 12:43:06:990PM | 1080312010 | Alex Ince-Cushman | Full length discharge with nl04 ~ 0.4. Some LH trips, but 1MW was achieved. There was an associated change in toroidal rotation of -40km/s. The toroidal rotation did not reach steady state and deltaV would have been larger if the LH pulse had persisted. |
| Mar 12 2008 12:45:25:413PM | 1080312011 | Alex Ince-Cushman | Full length discharge with nl04 ~ 0.45. Multiple LH trips, but 1MW was achieved. There was an associated change in toroidal rotation of -35km/s. |
| Mar 12 2008 12:57:48:960PM | 1080312012 | Alex Ince-Cushman | The LH had multiple trips but still managed to drive a change in rotation of -30km/s. For the next shot we will try to adjust the outer gap to prevent the LH trips. |
| Mar 12 2008 01:11:46:117PM | 1080312013 | Alex Ince-Cushman | The LH is still having severe tripping problems. For the next shot we will try loading 1080215014 which had excellent LH coupling. |
| Mar 12 2008 01:25:21:070PM | 1080312014 | Alex Ince-Cushman | Only one trip on the LH with ~900kW of power. The toroidal rotation achieved steady state with a deltaV of -40km/s. For the next shot we will increase the LH power. |
| Mar 12 2008 01:58:57:477PM | 1080312016 | Alex Ince-Cushman | Clean LH @ ~900kW. The change in toroidal rotation was somewhat smaller (~-35km/s) than shot 14. |
| Mar 12 2008 03:14:12:023PM | 1080312017 | Alex Ince-Cushman | LH up to 1MW with two trips. There was a -30km/s change in toroidal rotation, but LH waves were on the edge of accessibility. For the next shot we will change the phasing to 75 degrees and try to further reduce the density. |
| Mar 12 2008 02:16:19:570PM | 1080312018 | Alex Ince-Cushman | Early Disruption. |
| Mar 12 2008 03:14:46:727PM | 1080312019 | Alex Ince-Cushman | LH at 75 degree phasing at ~ 900kW. Very little power coupled after t ~0.95s. For the next shot we will bring the density back up a touch. |
| Mar 12 2008 03:15:07:163PM | 1080312020 | Alex Ince-Cushman | LH at 75 degree phasing, ~ 1MW coupled with one extended trip at t = 0.8s. The LH appears to have suppressed the sawtooth oscillations after the trip. The toroidal rotation has a somewhat complicated time history but the deltaV was of order 25km/s. |
| Mar 12 2008 03:13:35:850PM | 1080312021 | Alex Ince-Cushman | LH at 75 degree phasing (1MW) with no trips. It appears that there was an early locked mode which prevented the LH system from driving substantial amounts of current. We will rerun the shot, trying to prevent the locked mode by early puffing. |
| Mar 12 2008 03:26:50:007PM | 1080312022 | Alex Ince-Cushman | LH at 75 degree phasing (1MW) with one large set of trips. The change in toroidal rotation was substantial (~-40km/s). |
| Mar 12 2008 03:39:07:147PM | 1080312023 | Alex Ince-Cushman | LH at 75 degree phasing (1MW) with one early trip. The change in toroidal rotation was substantial (~-35km/s). |
| Mar 12 2008 03:52:35:900PM | 1080312024 | Alex Ince-Cushman | LH at 120 degree phasing (1MW) but effectively not LH coupling. For the next shot reduce the phasing to 105. |
| Mar 12 2008 04:09:00:477PM | 1080312025 | Alex Ince-Cushman | LH at 105 degree phasing with very poor LH coupling. For the next shot we will return to 60 degree phasing and increase the toroidal field to 6.2 T. |
| Mar 12 2008 04:47:18:163PM | 1080312027 | Alex Ince-Cushman | Nice clean LH pulse at 60 degrees (830kW). Bt @ 6.2T. ~-40km/s change in toroidal rotation and substantial density peaking. |
| Mar 12 2008 04:57:36:320PM | 1080312028 | Alex Ince-Cushman | Another beautiful LH pulse at 60 degrees (940kW). Bt @ 6.2T. ~-45km/s change in toroidal rotation and substantial density peaking. |
| Physics Operator Comments | |||
| Mar 12 2008 09:04:52:317AM | 1080312001 | Joseph Snipes | Shot 1: Due to data system problems, the requested shot for starting with a locked mode is not available. So, we are starting with step 2 of the run plan. Loaded shot 1080226006 with no changes. It is a 0.75 MA, 5.4 T SNL plasma with nl04 set to 4.5 x 10^19 m^-2 with LHCD at 60 degree phasing. Brief delay for magnet cooling and a delayed search due to Granetz' DNB needs. |
| Mar 12 2008 10:02:46:207AM | 1080312001 | Joseph Snipes | Shot 1: Good shot after almost a one hour delay for magnet cooling. |
| Mar 12 2008 10:14:22:503AM | 1080312002 | Joseph Snipes | Shot 2: Enabled argon gas and DNB gate valve. Note that the DNB cryopumps are not working because of a lack of liquid helium, so there may be some gas injected into the plasma after the beam fires. The base pressure of the DNB is at 8 x 10^-6 Torr. Also enabled the NESOX diagnostic. Otherwise, it is a repeat. Good shot. DNB injected at 1.3 s. There was no increase in density from the DNB. |
| Mar 12 2008 10:31:35:207AM | 1080312003 | Joseph Snipes | Shot 3: Loaded shot 1080215001 to get a locked mode with the A-coil on at 4 kA and B21_PROJ gain set to zero in segment 3. Set prepuff to 23 ms duration. LH turned off. Disrupted in rampdown at 1.7 s. Got a locked mode starting at 0.65 s. |
| Mar 12 2008 10:51:12:800AM | 1080312004 | Joseph Snipes | Shot 4: Startup hesitates. Reloading shot 2. Try setting the prepuff to 27 ms duration. Good startup and a good shot, although there are some large low frequency MHD modes between 0.06 and 0.14 s. |
| Mar 12 2008 11:03:34:050AM | 1080312005 | Joseph Snipes | Shot 5: Set nl04 to 4 x 10^19 m^-2 to attempt to reduce the density. Good shot but the density did not drop. |
| Mar 12 2008 11:17:21:817AM | 1080312006 | Joseph Snipes | Shot 6: There was some preprogrammed gas in PG3 from 0.1 s to 0.5 s that I have now removed for this shot. Good shot. Density dropped a tiny bit. |
| Mar 12 2008 11:33:07:880AM | 1080312007 | Joseph Snipes | Shot 7: Reduced nl04 to 3 x 10^19 m^-2 until 0.4 s with a ramp up to 4 x 10^19 m^-2 at 0.55 s. Reduced RCUR by 3 mm to 0.659 m throughout the flattop to increase the outer gap. Reduced prepuff to 23 ms duration. Good shot. Density is a bit lower, but I think to get lower density will require reducing the argon puff. |
| Mar 12 2008 11:50:18:367AM | 1080312008 | Joseph Snipes | Shot 8: Reduced nl04 to 3 x 10^19 m^-2 throughout the flattop. Increased prepuff to 27 ms duration. DNB delayed until 0.86 s. Alternator problem. No power shot. |
| Mar 12 2008 12:02:49:070PM | 1080312009 | Joseph Snipes | Shot 9: Try again. No power shot due to data system problem. |
| Mar 12 2008 12:29:50:007PM | 1080312010 | Joseph Snipes | Shot 10: One more time . . . Good shot. Density is definitely lower. |
| Mar 12 2008 12:32:53:040PM | 1080312011 | Joseph Snipes | Shot 11: Increased RCUR before 1 s with a slight ramp to try to flatten out the outer gap at 1 cm during the flattop. Good shot. |
| Mar 12 2008 12:46:56:697PM | 1080312012 | Joseph Snipes | Shot 12: Set the derivative gain on nl04 to 0. Set the derivative gain on RCUR from 6 to 4 in an attempt to reduce the oscillations in the outer gap. Good shot. |
| Mar 12 2008 01:00:09:633PM | 1080312013 | Joseph Snipes | Shot 13: RCUR oscillations were still there. Try increasing the derivative gain on RCUR to 8. Good shot. |
| Mar 12 2008 01:14:02:070PM | 1080312014 | Joseph Snipes | Shot 14: Loaded shot 1080215014. Set nl04 to 3 x 10^19 m^-2. Good shot. |
| Mar 12 2008 01:25:44:367PM | 1080312015 | Joseph Snipes | Shot 15: Increased EF4 current from -1340 to -1355 A. Reduced prepuff by 3 ms to 24 ms duration. Fizzle! |
| Mar 12 2008 01:45:20:087PM | 1080312016 | Joseph Snipes | Shot 16: Returned EF4 current to -1340 A. Reduced prepuff another 2 ms to 22 ms duration. Good shot. |
| Mar 12 2008 01:58:46:633PM | 1080312017 | Joseph Snipes | Shot 17: Increased prepuff to 25 ms duration. Reduced nl04 to 2.5 x 10^19 m^-2. Good shot. |
| Mar 12 2008 02:14:39:523PM | 1080312018 | Joseph Snipes | Shot 18: Reduced nl04 after 0.25 s to 2 x 10^19 m^-2. Disrupted at 0.27 s. |
| Mar 12 2008 02:27:18:447PM | 1080312019 | Joseph Snipes | Shot 19: Increase the prepuff by 2 ms to 27 ms duration to try to reduce the early runaways. Keep nl04 at 2.5 x 10^19 m^-2 until 0.35 s, then ramp quickly to 2 x 10^19 m^-2 by 0.4 s. Good shot. |
| Mar 12 2008 02:44:47:337PM | 1080312020 | Joseph Snipes | Shot 20: Increased nl04 to 2.5 x 10^19 m^-2 throughout the flattop. Disrupted at 1.6 s. Hard x rays go through the roof right after the LH switches off at 1.2 s.
|
| Mar 12 2008 03:11:45:273PM | 1080312021 | Joseph Snipes | Shot 21: Repeat. Disrupted at 1.6 s. Locked mode from at least 0.26 s. |
| Mar 12 2008 03:13:34:430PM | 1080312022 | Joseph Snipes | Shot 22: Increased nl04 to 3 x 10^19 m^-2 up to 0.3 s then ramp back down to 2.5 x 10^19 m^-2. Good shot. |
| Mar 12 2008 03:30:47:087PM | 1080312023 | Joseph Snipes | Shot 23: Increased prepuff by 2 ms to 29 ms duration to try to eliminate the runaways early in the startup. Good shot. |
| Mar 12 2008 03:50:00:197PM | 1080312024 | Joseph Snipes | Shot 24: Repeat with 120 degree phasing on the LH. Disrupted at 1.59 s. LH tripped for most of the pulse but still generated a huge fast electron tail. |
| Mar 12 2008 04:01:17:210PM | 1080312025 | Joseph Snipes | Shot 25: Repeat. Disrupted at 1.66 s. Hard x rays again go through the roof. LH trips several times. |
| Mar 12 2008 04:26:10:853PM | 1080312026 | Joseph Snipes | Shot 26: Increased TF current to 172 kA to go to 6.2 T. Started TF rampdown at 1.4 s. Li pellet enabled and set to fire at 1.5 s. Good shot. Disrupted in rampdown at 1.71 s. LH tripped. |
| Mar 12 2008 04:31:29:807PM | 1080312027 | Joseph Snipes | Shot 27: Repeat. Disrupted at 1.5 s. Large hard x ray emission throughout and after the LH pulse up to the disruption. |
| Mar 12 2008 04:57:13:320PM | 1080312028 | Joseph Snipes | Shot 28: Repeat. Good shot. |
| Mar 12 2008 05:03:08:180PM | 1080312029 | Joseph Snipes | Shot 29: Repeat. Good shot. |
| Engineering Operator Comments | ||||
| Shot | Time | Type | Status | Comment |
| 1 | 09:50:10:973AM | Plasma | Ok | |
| 2 | 10:10:19:363AM | Plasma | Ok | |
| 3 | 10:26:30:097AM | Plasma | Ok | |
| 4 | 10:44:52:630AM | Plasma | Ok | |
| 5 | 11:00:10:583AM | Plasma | Ok | |
| 6 | 11:14:41:363AM | Plasma | Ok | |
| 7 | 11:29:55:990AM | Plasma | Ok | |
| 8 | 11:47:33:693AM | Plasma | Bad | Alt. Inhibit, fault on bearing #2 |
| 9 | 11:58:49:583AM | Plasma | Bad | looks like one of the camac servers died. |
| 10 | 12:14:34:150PM | Plasma | Ok | |
| 11 | 12:30:29:337PM | Plasma | Ok | |
| 12 | 12:45:01:867PM | Plasma | Ok | |
| 13 | 12:57:37:303PM | Plasma | Ok | |
| 14 | 01:10:52:947PM | Plasma | Ok | |
| 15 | 01:23:56:413PM | Plasma | Ok | |
| 16 | 01:39:58:133PM | Plasma | Ok | |
| 17 | 01:56:53:540PM | Plasma | Ok | |
| 18 | 02:12:21:240PM | Plasma | Ok | |
| 19 | 02:25:29:617PM | Plasma | Ok | |
| 20 | 02:39:08:930PM | Plasma | Ok | |
| 21 | 02:53:16:243PM | Plasma | Ok | |
| 22 | 03:11:48:540PM | Plasma | Ok | |
| 23 | 03:26:56:900PM | Plasma | Ok | |
| 24 | 03:39:31:413PM | Plasma | Ok | |
| 25 | 03:53:08:337PM | Plasma | Ok | |
| 26 | 04:14:58:227PM | Plasma | Ok | |
| 27 | 04:27:28:663PM | Plasma | Ok | |
| 28 | 04:44:46:320PM | Plasma | Ok | |
| 29 | 04:59:57:867PM | Plasma | Ok | |
| System Availability | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Mar 12 2008 09:50:01:503AM | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||