| Miniproposals | ||||||||||
|
| Operators | |
| Session leader(s): | Matt Reinke |
| Physics operator(s): | Bill Rowan |
| Engineering operator(s): | Bill Parkin,Bill Cochran,Bill Byford |
| Engineering Operator Run Comment |
| MP564 - Power requirements for high confinement H-modes... |
| Session Leader Plans |
| Entered: Oct 26 2009 11:02:57:643AM |
| Author: Matt Reinke |
| Run Plan for MP564 - Power requirements for high confinement H-modes and the role of radiated power spatial distribution
SL: Loarte/Reinke PO: Rowan/Hutch This run plan covers both 1091019 and 1091020 DIAGNOSTICS: ne/Te tools (TCI,TS, GPC1,2 FRCECE). Will be running at 5.4 T for most of both of the days. VUV/SXR spectroscopy. B-SIDE-LOWER will be impurity puffing but if B-SIDE-UPPER can be made to work for Ar puffs a trace can be used. Reflectometer setup for LH trans region (0.7-1.2 in ELMy and 0.5-1.0 in EDA) LBO (Ca/F2) can fire near the end of the H-mode, but I'll have my eye on the perturbation. NINJA can puff for turbulence/CXRS near the end of the H-mode but will likely need to use it for fueling during high density EDAs (1091020) Divertor IR imaging for heat-flux measurements Have N2, Ne and Ar available during the run w/o a cell access. Start with 1.0 psi Ne in B-SIDE-LOWER. The cryopump will be needed for 1091020 (Tuesday) only. The goal of this experiment is to create impurity seeded EDA and ELMy H-modes with net input power (P_NET=P_TOT-P_RAD) matching unseeded H-modes. Target EDA H-modes were made on 1090902 and the ELMy targets were made on 1090902 and 1090922. 1 MW ICRF ramps will be used to vary P_NET to ensure a good match. The 1090922 run focused on Ne and Ar in the EDA and Ar in the ELMy. This run will finish the Ne pressure scan in ELMy, extend the Ne gas pressure scan in EDA and complete an EDA and ELMy scan of N2 puffing. (most if not all of DAY1) Assuming the high confinement w/ seeding is repeatable, we will then attempt to recover high input power operations at the typical C-Mod ODIV strike point location that has been problematic for FY09/10, then proceed to an ITER shape. From there we will increase current and density (possibly Bt) to maintain the EDA H-mode. As a piggyback on this, we will be monitoring the evolution of the molybdenum vs. RF input Joules for MP568 and at the end of 1091020 will take a few shots at conditions repeating 1090827 and/or 1050610. Will do the ELMy with Ne then N2. Will repeat the H-mode threshold scan shot and two higher unseeded target powers then move to seeding. Ninja puffs allowed in late portion of the shot. 1) Load equilibrium from 1090922021 (JFT-2M shape, nl04=0.5, Ip=800 kA). ICRF ramp from 0.5-3 MW from t=0.8-1.5 sec. Verify similar threshold and proper diagnostic/aux system operations. 2) Change RF timing ICRF ramp 2.0-2.9 MW from t=0.8-0.1.0 sec and 2.9 MW from 1.0-1.5. Repeat, replacing peak power of 2.9 MW to 3.4 MW in RF programming. 3) Enable B-SIDE-LOW 1.5 psi Ne puff from 0.6-0.8 sec. ICRF ramp 2.5-4.0 MW from t=0.8-1.0 sec and 4.0-5.0 MW from t=1.0-1.5 sec. 4) Repeat shot, decreasing Ne pressure to 1.0 and then 0.5 5) Fill B-SIDE-LOW with 3.0 psi N2 and repeat target shot from (3) 6) Move N2 pressure up/down, getting points where the ELMy->ELM-Free and then reducing. Move to seeded EDA H-modes. Start with N2 and then Ne 7) Load equilibrium from 1090922007 (high strike point, nl04=0.8, Ip=800 kA). Run threshold scan with RF from 0.5-3.0 MW from 0.6-1.5 sec. 8) Enable N2 puff, choosing pressure based on knowledge from (5) and (6). Puff from 0.4-0.6 sec. Program ICRF to have 2.5-4.0 MW ramp from 0.6-0.8 and a 4.0-5.0 MW ramp from 0.8-1.5 sec. Scan N2 pressure to get a large range of PRAD/PTOT and H89 -> 1.0. 9) Fill B-SIDE-LOW with 1.5 psi Ne and repeat target from (8). Monitor proper change of radiating species. Repeat until clean Ne puff w/o significant N2 present. Increse Ne pressure to absurd pressure (2.5, 3.5 ->etc.) until L-mode confinement (H89~1.0) is demonstrated. 10) Modify equilibrium to see if we can run a lower strike point with seeding. Load 1090811006, lower Ip to 800 kA. Choose gas species and pressure from earlier shots showing peaking WTH and neutron rate. Modify RF programming to allow for maximum input power. Program ICRF to have 2.5-MAX MW ramp from 0.6-0.8 and remain constant until 1.5 seconds. 11) If (10) is run w/o tungsten injections then keep lower strike point otherwise return to high strike point from earlier. Increase Ip/nl04 ~ 900 kA/0.9 and then ~1.0 MA/1.0, keeping the EDA H-mode. Modify impurity gas pressure to adjust for changes in impurity transport. Move to higher current seeded H-modes to 1) demonstrate seeding at ITER target and/or 2) push high current H-mode performance. These will be moving to high densities that usually lead to neutral pressure limits on the ICRF. To start with, the cyropump will be used with LFS fueling to try and maintain a low pressure. If high current EDA's are not obtainable this way the shape will shifted towards DN and HFS fueling via the NINJA will be added. ***TURN ON CRYO*** 12) Move to ITER shape, loading 1080409016 (Ip=1.35, Bt=5.4, nl04=1.2). Tweak the outer strike point location if (10) was still showing tungsten. Do an RF power scan to get the threshold, ramping from 0.5-3.0 MW from t=0.7-1.15. Alternate starting shot is 1091006025 (Ip=1.25, Bt=5.4, nl04=1.25, fast Ip ramp) 13) Develop seeded H-mode using Ne at a reasonable pressure. Set puff timing to be from 0.5-0.7 sec. ICRF ramp from 2.5-4.0 from 0.7-0.85 and 4.0-5.0 from 0.85-1.2. Ramp power back down from 5.0-1.5 MW from 1.2-1.7 to help avoid disruption in the current ramp down. (These times are for the FY08 target. If using FY09 target, shift all start times back to 0.5 sec.) If this recipe is are not giving reliable target EDA H-modes, develop alternate high current/density target. Starting from ITER targets or 1070829029 increase SSEP to make plasma more DN and use NINJA to fuel from the HFS. 14) Vary the Ne gas pressure to scan PRAD/PTOT, showing both a return to L-mode at high pressure and a return to bad RF/PRAD at low pressure. 15) If time switch, N2 and repeat (14) 16) Return to 1 MA target from (11) and run 2-3 shots showing confinement after a substantial RF input energy. |
| Physics Operators Plans |
| Entered: Oct 19 2009 08:17:20:567AM |
| Author: Bill Rowan |
| Engineering setup for 1091019
Run Plan for MP564 - Power requirements for high confinement H-modes and the role of radiated power spatial distribution SL: Loarte/Reinke PO: Rowan ----------------- Engineering Setup ----------------- Run begins at 09:00 and ends at 17:00 Power systems as on: 1090922021 Acoil: +Dtop -Dbot -Jtop +Jbot (standard) Hybrid Enabled Gas setup: Fill B-Top with 6 psi D2 Hybrid enabled (PG4) fill B-side lower with 1 psi Ne Hybrid DISABLED (PG1) leave B-side upper with 1 psi Ar Hybrid DISABLED (PG2) fill B-main (C-side) with 40 psi D2 Hybrid enabled (PG3) fill NINJA with 10 psi D2 DISABLED NOTE: Have N2, Ne and Ar available during the run w/o a cell access. Enable gatevalves and shutters: ECE, VUV, Z-bolo Torvac gatevalve toggle (yes/no): no Boronization(yes/no): yes Overnight ECDC (yes/no): yes ICRF(yes/no): yes LH(yes/no): no DNB(yes/no): no Cryopump (yes/no): no Vessel temperature: 35/35/35 ------------------------------ ECDC Parameters (if requested) ------------------------------ gas and pressure: D2 at 2e-4 Torr sweep: 44/45/103 cm scan: 20/120 s |
| Session Leader Summaries |
| Entered: Oct 19 2009 07:52:05:430PM |
| Author: Matt Reinke |
| A generally successful day. We got a late start (10:30) due to alternator issues and only got 28 shots with 001 being a fizzle. The Ne and N2 EDA H-mode scans were completed (7),(8) and (9).
002-005 where in the "grazing" JFT-2M shape (1)-(6). The post-boronization and now He ECDC led to significant wall fueling. By shot 005 the fueling gas puff was shut off by 0.15 seconds and the target density got nl04~0.7 when we needed it at 0.5 to get good ELMs. We punted to the EDA portion with hopes to return to the ELMy late in the day but we ran out of time and will do it first thing on 1091020. 006 was a threshold scan showing P_LH ~ 0.7 MW from the ICRF (7) and then we started the seeding. This was an EDA H-mode, repeat of 1090922007, Ip=800 kA, nl04=0.8 with a standard shape although the high strike point. 007-018 were Ne seeded 019-028 were N2 seeded We had trouble repeating the high performance seeded Ne shots for awhile and this may have been due to the evolving wall conditions. Something funky was happening on the H-alpha and Lyman-alpha that needs a closer look. By 012 things had cleaned up and we could scan the gas pressure to vary PRAD. We had to add extra puffs after the LH transition because a 4.0 psi puff disrupted the plasma. Shots 012,013, 015, 017 and 018 can be used to as scan of confinement with 018 having a nice time evolving H89 during the shot. The N2 puffing needed to be much more aggressive than the Ne. After messing around with discrete puffs from 019-021 which amounted to little PRAD increase and a hot divertor and choppy RF we switch to steady-state puffs. 022 had 9 psi puff from 0.4-1.0 and showed a fast N2 pumpout after 1.0 seconds. Also the RF power started to become choppy and the confinement degrade (what is cause and effect I don't know) at this time. 023 moved to a 9 psi 0.4-1.5 puff which resulted in a very steady plasma with high confinement (H89~1.8). From there we modulated the pressure with 20 psi being too much. Also, shots 023 and 024 were repeated showing little to no N2 buildup. Shots 023,024,026 and 028 can be used as a scan of confinement. |
| Physics Operator Summaries |
| Entered: Oct 20 2009 07:19:05:170AM |
| Author: Bill Rowan |
| Physop Summary for 1091019
The experiment was based on MP564 "Power requirements for high confinement H-modes and the role of radiated power spatial distribution." C-Mod ran well today. Two configurations were used, and both were developed during previous runs. The first was the JFT-2M shape for which the model was 1090922021. The other was a more conventional C-Mod shape for which the model was 1090922007. Nitrogen and neon were puffed into the discharges. With the JFT-2M shape, the density was to have been nl_04=0.5e+20. That density was not reached due to wall fueling. The first shot of the day was a fizzle because the prefill was set too high. These can be traced to boronization on Friday (1091016). The only other wall conditioning prior to this run consisted of the usual short post-boronization sessions of He and D2 ECDC. Helium ECDC for the night before the run might have provided better control of density as it did for 1090923. There were spurious injections on several discharges. They may have been responsibile or contributors to early disruptions. Because of the impurity gas puffing, that would be difficulat to sort out. Early disruptions were 014, 025, and 027). There was a delay in starting the run. Quoting Mike Rowell "This morning the alternator drive system failed to start correctly due to a problem with the "soft start " circuit. The problem has been identified and the alternator was successfully restarted @7:20AM. As a result there will be a delay in the start of CMOD operations. We expect the alternator to be ready for CMOD operations run by 10:30 AM." The predicted start time in this message was accurate. The session leader kept excellent notes as the impurity gases were changed and the plenum pressures varied. Look for detail there. An interesting result was the reduction of ICRF faulting on shots with high levels of impurity puffing. May have been coincidence but is probably worth pursuing. Score card 27 plasmas (3 early disruptions due to impurities) 01 fizzle 28 Total |
| Session Leader Comments | |||
| Oct 19 2009 10:36:22:823AM | 1091019001 | Matt Reinke | Started late (10:30) but first shot was a fizzle :( |
| Oct 19 2009 10:58:18:783AM | 1091019002 | Matt Reinke | Lot of outgassing from this shot. The density wasn't under control and only got down to about nl04~0.8. If it doesn't look to be coming down after one or two more shots we'll switch to doing the EDA stuff.
WIDE1 looks defocused or possible boronized over. NEXT: Repeat shot. |
| Oct 19 2009 11:11:49:947AM | 1091019003 | Matt Reinke | Good threshold shot (1), although for a higher density nl04~0.75. Still don't have density control.
NEXT: Move to running the 2.9 MW ICRF pulse from (2). 2.0-2.9 MW from 0.8-1.0 and 2.9 MW from 1.0-1.5 MW. |
| Oct 19 2009 11:22:58:340AM | 1091019004 | Matt Reinke | Bad ICRF, no ELMs, still have nl04~0.75 at the LH trans.
NEXT: Repeat once more then we'll move the EDA target while walls get degassed. |
| Oct 19 2009 11:35:45:203AM | 1091019005 | Matt Reinke | Still no density control, nl04~0.70. This is not giving me what I need so we'll punt to the EDA and come back to this later.
NEXT: Move to (7) on the plan - Ip=800 kA, nl04=0.8, normal shape, high strike point. RF power ramp. |
| Oct 19 2009 11:55:05:697AM | 1091019006 | Matt Reinke | Steady EDA H-mode until a large Cu/Ni injection hits at 1.15 seconds. Don't have good edge TS data as no one's around to fix it.
NEXT: Move to (9) with Ne seeding since it's in the plenum. 0.4-0.6 sec puff at 1.7 psi. RF 2.5-4.0 MW from 0.6-0.8 and 4.0-5.0 from 0.8-1.5. |
| Oct 19 2009 12:07:20:700PM | 1091019007 | Matt Reinke | The Ne made it in just fine. and the tile temperatures a lower than 006 despite more ICRF. ICRF has some trips and there is a large Mo injection.
NEXT: Repeat shot, but IMPINJ at 1.4 and LFS/HFS NINJA puff before LH trans |
| Oct 19 2009 12:20:27:707PM | 1091019008 | Matt Reinke | Not getting similarly high H89's and with these shots compared to 1090922. JANT running into a HV limit, try reducing and increase D/E.
NEXT: Repeat target plasma, increase Ne pressure to 2.5 psi |
| Oct 19 2009 12:36:48:213PM | 1091019009 | Matt Reinke | D and E ran well, J poorly. Confinement isn't as good as previous experiments, but the RF isn't at full power yet. If still not the same turn of AUX systems.
NEXT: lower pressure 2.0 psi and repeat. |
| Oct 19 2009 12:54:16:907PM | 1091019010 | Matt Reinke | Performance was better with a higher WTH and neutron rate, but the J-ANT power information is stuck in limbo. Yijun says it was clean, though. Despite a puff pressure of 1.9 psi we're still getting a large amount of divertor heat flux. PRAD is down compared with similar puffs, so that's consistent.
We're getting ELMs which is weird. NEXT: Repeat shot with puff pressure at 3.0 psi |
| Oct 19 2009 01:06:50:630PM | 1091019011 | Matt Reinke | Ratty J-ANT, but heat flux lower, PRAD higher. No ELMs on this one.
NEXT: Repeat Ne gas pressure down to 2.3 psi. |
| Oct 19 2009 01:24:14:137PM | 1091019012 | Matt Reinke | Good shot, repeated the high WTH, H89 and neutron rate of the 1090922 shots. The IMPINJ perturbs at the end.
NEXT: Turn of IMPINJ, leave off NINJA and repeat with Ne pressure at 3.0 psi |
| Oct 19 2009 01:30:55:077PM | 1091019013 | Matt Reinke | Large injection kills the early H-mode, but it recovers nicely. PRAD not too much higher and H89 drops from 1.8-1.7 from 012.
NEXT: Repeat increase puff pressure to 4.0 |
| Oct 19 2009 01:49:36:770PM | 1091019014 | Matt Reinke | The plasma disrupted before the RF and PRAD/edge Te traces look like the puff was too much, similar to 1090909 exploratory shots. Seems like we can't "pre-load"
the impurity to the right level. NEXT: Lower pressure back to 3.0 psi and puff from 0.4-0.6 and 1.0-1.2 |
| Oct 19 2009 02:03:26:607PM | 1091019015 | Matt Reinke | The 2nd gas puff didn't get loaded so this was a repeat of 013 (programming wise) but a cleaner version.
NEXT: load the puff timing for the 1.0-1.2 puff |
| Oct 19 2009 02:19:13:580PM | 1091019016 | Matt Reinke | Early part of the plasma was very repeatable but the 2nd puff from 1.0-1.2 was too much for it.
NEXT: Try x2 smaller puffs, 1.0-1.05 and 1.3-1.3.05. Also NINJA back on looking at LH transition. |
| Oct 19 2009 02:30:58:240PM | 1091019017 | Matt Reinke | REALLY nice ICRF power trace but not enough radiation.
NEXT: puffs at 0.95-1.05, 1.25-1.35 in the H-mode |
| Oct 19 2009 02:43:53:603PM | 1091019018 | Matt Reinke | Excellent shot. With the 2nd puff see a trajectory of decreasing confinement and then a HL transition.
NEXT: repeat but use 3.0 psi N2 instead of Ne. Keep same puff timing and work from there. |
| Oct 19 2009 02:55:58:593PM | 1091019019 | Matt Reinke | Clearly not enough PRAD on this shot, possibly causing the bad RF?
NEXT: Increase N2 pressure to 5.0 psi |
| Oct 19 2009 03:12:42:823PM | 1091019020 | Matt Reinke | Again, not enough prad..needs more N2 puff
NEXT: 9 psi pressure on B-SIDE-LOW |
| Oct 19 2009 03:29:11:190PM | 1091019021 | Matt Reinke | See the early radiation rise this time at 9 psi but no large rise during the shot. I thought that this would disrupt it but it still seems stable.
NEXT: Make a big step. Steady puff at 9 psi from 0.4-1.0. |
| Oct 19 2009 03:39:33:347PM | 1091019022 | Matt Reinke | A steady state N2 puff works well and it pumps out quickly after the gas valve shuts off. This is good news that we've got a knob to turn on the radiation but bad news because we're probably making the machine quiet dirty.
NEXT: Increase the length of the N2 puff from 0.9-1.5 |
| Oct 19 2009 04:00:02:230PM | 1091019023 | Matt Reinke | Great shot. Steady RF now that seeding is during the whole shot. Confinement is good, H89 < 1.80. Still have lowish PRAD and high heat flux to the divertor.
NEXT: Repeat to check for N2 buildup then raise the pressure. |
| Oct 19 2009 04:06:56:970PM | 1091019024 | Matt Reinke | Very repeatable
NEXT: 14 psi puff, start puff at 0.5 seconds to prevent early disruption |
| Feb 9 2010 08:54:48:533AM | 1091019024 | Matt Reinke | Looking back at this shot there's appears to a be a change in transport that occurs ~1.20-1.25. Despite the ramped ICRF there is no evolution of Te, ne neutron rate or PRAD. By the H98 vs. PNET arguement confinement should be improving
Looking at the fluctuations there is a mode at around 55 kHz seen on magnetics and the AXUV diodes from 0.97 < t < 1.23. Whatever this is might be responsible for enhanced energy transport. Once it dies down confinement continues to improve with RF input power. |
| Oct 19 2009 04:25:19:303PM | 1091019025 | Matt Reinke | Shot had problems before the N2 puff started, MARFE and injections.
NEXT: Repeat |
| Oct 19 2009 04:41:07:200PM | 1091019026 | Matt Reinke | Good shot with higher PRAD. Confinement down slightly but not much.
NEXT: Increase to 20 psi N2 puff. Let RF ramp from 4.0-6.0 MW from 0.8-1.5. |
| Oct 19 2009 05:01:04:240PM | 1091019027 | Matt Reinke | Early disruption which seems to be from the N2 puff
NEXT: lower the puff pressure to 17 psi and repeat. |
| Oct 19 2009 07:24:31:497PM | 1091019028 | Matt Reinke | This showed a time evolving H89 making for a great last shot on the N2 scan in EDA.
Also, the ICRF was programmed for 6 MW and did OK but probably because the confinement started to degrade resulting in transient H-modes. This makes me excited for tomorrow. |
| Physics Operator Comments | |||
| Oct 19 2009 08:13:48:393AM | Bill Rowan | There will be a delay of approximately 1 hour in the start of the run. Alternator got a late start. Lock up delayed until 9:30 | |
| Oct 19 2009 08:22:37:130AM | Bill Rowan | Boronization completed on Friday night (1091016)
| |
| Oct 19 2009 08:53:50:923AM | Bill Rowan | Loaded shot 1090922021.
Warning appeared: need to recalculate the magnetics predictors for shots after 9/24/09 Recalculated magnetics predictors | |
| Oct 19 2009 09:13:08:603AM | Bill Rowan | Load segment 1 from 1091014033
reduced prefill by 2 ms from 25 to 23 | |
| Oct 19 2009 09:22:11:997AM | Bill Rowan | ratiomatic is expected to be working today.(Willie and Tommy) | |
| Oct 19 2009 10:40:24:667AM | 1091019001 | Bill Rowan | 001. Fizzle.
Delay before this shot due to late startup of alternator. Next shot. Reduce the prefill from 23 ms to 19 ms Ratiomatic is not working well. May need to fiddle to find the breakdown. |
| Oct 19 2009 10:54:33:127AM | 1091019002 | Bill Rowan | 002. Plasma.
Injections at 1.0 and 1.3 s Impurity injection not in operation. SL not concerned nl_04=0.76e+20 (high) and valve is off. Next shot No PCS changes |
| Oct 19 2009 11:07:45:040AM | 1091019003 | Bill Rowan | 003. Plasma
No injections H-mode at 0.85 s nl_04 = 0.75e+20 |
| Oct 19 2009 11:26:09:793AM | 1091019004 | Bill Rowan | 004. Plasma
|
| Oct 19 2009 11:26:19:607AM | 1091019004 | Bill Rowan | Next shot. No PCS changes
|
| Oct 19 2009 11:31:54:280AM | 1091019005 | Bill Rowan | 005. Plasma
|
| Oct 19 2009 11:40:44:440AM | 1091019005 | Bill Rowan | Next shot.
Loaded 1090922007. Corrected predictors Loaded 1091019005 segment 1 into segment 1 |
| Oct 19 2009 11:47:56:037AM | 1091019006 | Bill Rowan | 006. Plasma
Next shot. Enabled b-side lower (neon) |
| Oct 19 2009 12:00:08:090PM | 1091019007 | Bill Rowan | 007. Plasma. Full length
Disruption during ramp down 1.7 s Next shot. No PCS changes |
| Oct 19 2009 12:16:13:440PM | 1091019008 | Bill Rowan | 008. Plasma
|
| Oct 19 2009 12:19:43:160PM | 1091019008 | Bill Rowan | 008. Plasma
Next shot No PCS changes |
| Oct 19 2009 12:34:52:447PM | 1091019009 | Bill Rowan | 009. Plasma
Next shot. No PCS changes |
| Oct 19 2009 12:51:44:810PM | 1091019010 | Bill Rowan | 010. Plasma
Next shot No PCS changes |
| Oct 19 2009 12:58:52:457PM | 1091019011 | Bill Rowan | 011. Plasma
disruption at 1.6 s |
| Oct 19 2009 01:08:38:913PM | 1091019011 | Bill Rowan | Next shot
No PCS changes |
| Oct 19 2009 01:24:26:793PM | 1091019012 | Bill Rowan | 012. Plasma
Next shot. no pcs changes |
| Oct 19 2009 01:36:41:470PM | 1091019013 | Bill Rowan | 013. Plasma
Few injections between 0.7 and 1 s next shot. No PCS changes |
| Oct 19 2009 01:51:58:850PM | 1091019014 | Bill Rowan | 014. Plasma
Disruption at 0.57 s Likely due to impurity injection. bp_04bc magnetics alarm Bob is watching it Next shot. Added a pulse to pg1 (b-side lower) from 1.0 to 1.2 s |
| Oct 19 2009 01:56:36:430PM | 1091019015 | Bill Rowan | 015. Plasma
Neglected to load the pcs changes Next shot. Added a pulse to pg1 (b-side lower) from 1.0 to 1.2 s |
| Oct 19 2009 02:13:07:873PM | 1091019016 | Bill Rowan | 016. Plasma
|
| Oct 19 2009 02:18:36:813PM | 1091019016 | Bill Rowan | 016. Plasma
Next shot Reduced the second pulse on pg1. Now from 1.0 to 1.05s Added a pulse to pg1 (b-side lower) from 1.3 to 1.35 s |
| Oct 19 2009 02:22:31:283PM | 1091019017 | Bill Rowan | 017. Plasma
|
| Oct 19 2009 02:32:32:833PM | 1091019017 | Bill Rowan | 017. Plasma
Next shot Changed the last two pg1 pulses Pulse 1: 0.4 to 0.6 Pulse 2: 0.95 to 1.05 Pulse 3: 1.25 to 1.35 |
| Oct 19 2009 02:39:00:603PM | 1091019018 | Bill Rowan | 018. Plasma
Next shot. B-side lower: nitrogen |
| Oct 19 2009 03:07:07:043PM | 1091019019 | Bill Rowan | 019. Plasma
Next shot. No PCS changes |
| Oct 19 2009 03:15:58:040PM | 1091019020 | Bill Rowan | 020. Plasma
Next shot. No PCS changes |
| Oct 19 2009 03:28:21:063PM | 1091019021 | Bill Rowan | 021. Plasma
Next shot. Changed pg1. Now just one puff from 0.4 to 1.0 |
| Oct 19 2009 03:41:43:210PM | 1091019022 | Bill Rowan | 022. Plasma
next shot. PG1 puff. From 0.4 s to 1.5 s |
| Oct 19 2009 03:53:59:697PM | 1091019023 | Bill Rowan | 023. Plasma
injection at 1.25 s Next shot. No pcs changes |
| Oct 19 2009 04:04:24:733PM | 1091019024 | Bill Rowan | 024. Plasma
|
| Oct 19 2009 04:08:39:030PM | 1091019024 | Bill Rowan | Next shot.
PG1 puff. From 0.5 s to 1.5 s |
| Oct 19 2009 04:23:41:647PM | 1091019025 | Bill Rowan | 025.Plasma
Disrupted at 1.25 s Next shot. No PCS changes |
| Oct 19 2009 04:35:50:057PM | 1091019026 | Bill Rowan | 026. Plasma
|
| Oct 19 2009 04:41:52:733PM | 1091019026 | Bill Rowan | Next shot.
No PCS changes |
| Oct 19 2009 04:57:52:190PM | 1091019027 | Bill Rowan | 027. Plasma. Disruption at 0.6 s
Next shot. No PCS changes |
| Oct 19 2009 05:07:04:287PM | 1091019028 | Bill Rowan | 028. Plasma
Last shot. |
| Engineering Operator Comments | ||||
| Shot | Time | Type | Status | Comment |
| 1 | 10:31:00:007AM | Plasma | Ok | |
| 2 | 10:46:36:967AM | Plasma | Ok | |
| 3 | 11:02:38:770AM | Plasma | Ok | |
| 4 | 11:15:53:510AM | Plasma | Ok | |
| 5 | 11:28:57:890AM | Plasma | Ok | |
| 6 | 11:41:32:720AM | Plasma | Ok | |
| 7 | 11:56:39:883AM | Plasma | Ok | |
| 8 | 12:09:35:750PM | Plasma | Ok | |
| 9 | 12:22:13:033PM | Plasma | Ok | |
| 10 | 12:40:51:200PM | Plasma | Ok | |
| 11 | 12:55:50:440PM | Plasma | Ok | |
| 12 | 01:11:57:350PM | Plasma | Ok | |
| 13 | 01:25:17:357PM | Plasma | Ok | |
| 14 | 01:40:35:563PM | Plasma | Ok | |
| 15 | 01:53:40:897PM | Plasma | Ok | |
| 16 | 02:06:29:247PM | Plasma | Ok | |
| 17 | 02:19:23:893PM | Plasma | Ok | |
| 18 | 02:33:19:680PM | Plasma | Ok | |
| 19 | 02:48:09:810PM | Plasma | Ok | |
| 20 | 03:02:40:947PM | Plasma | Ok | |
| 21 | 03:17:05:310PM | Plasma | Ok | |
| 22 | 03:32:44:500PM | Plasma | Ok | |
| 23 | 03:48:16:837PM | Plasma | Ok | |
| 24 | 04:00:50:810PM | Plasma | Ok | |
| 25 | 04:15:24:987PM | Plasma | Ok | |
| 26 | 04:31:50:930PM | Plasma | Ok | |
| 27 | 04:47:23:000PM | Plasma | Ok | |
| 28 | 05:04:27:570PM | Plasma | Ok | |
| System Availability | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Oct 19 2009 10:30:39:100AM | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||