Alcator C-Mod Run 1100324 Information

Back to runs

Miniproposals
Miniproposal:602
Date Filed: 2/26/2010
Title:Core Fluctuations and Transport in Two -Frequency ICRF Heated ITB Plasmas with Jogs and Modulated IC
First Author:Darin Ernst
Session Leader:Darin Ernst (shots 1-24)

Operators
Session leader(s):Darin Ernst
Physics operator(s):Steve Wolfe
Engineering operator(s):Bill Parkin,Bill Byford,Ed Fitzgerald

Engineering Operator Run Comment
MP#602 Core fluctuations and Transport in two-frequency ICRF-heated ITB

Session Leader Plans
Entered: Mar 22 2010 11:03:27:690AM
Author: Darin Ernst
coming soon - dernst@psfc.mit.edu
Entered: Mar 23 2010 06:40:28:050PM
Author: Darin Ernst
1. Begin with 1100322007 but with 5.5 T 800 kA (TBD)
target density to nl_04 1x10^20 m^-2
J-port 2.5 MW LFS off-axis
cryopump on, Ar puff
Move RF 0.1 sec earlier as in 1100323005.

2. Try lowering current to 700 kA to get lower ne_ped. If have trouble
getting ITB's, increase Ip to as much as 900 kA.

3. If successful, add D&E at 0.6 MW each with full on-off modulation
(period TBD, i.e. 10 ms on, 10 ms off) starting at 1.1 - 1.5 sec, 50%
duty cycle. If not successful, return to 800 kA.

4. Adjust D&E timing and modulation to optimize.

5. Increase D&E power to 1.6 MW each.

6. Decrease D&E power to 1.0 MW each.

7. Add vertical jogs 0.8 - 1.5 sec. (see below), turn off modulation.

8. Try jogs with modulation if time permits.

9. Thomson scattering calibration shots if time permits. Locked mode
shot for HIREX Sr.

Most desirable timing (alternatively, we can start DPCS jogs lasting 4
ms each at 0.811 s with a 29 ms flat top for asynchronous). The
interval between YAG 1 pulses is 33.333333 ms. There are three YAG 1
pulses in 100 ms.


0.910253 s: YAG 1
0.911 s: begin vertical jog
0.915 s: end vertical jog
YAG 1 +5ms: YAG 2

0.943586 s: YAG 1
0.944 s: begin vertical jog
0.948 s: end vertical jog
YAG 1 +5ms: YAG 2

0.976918 s: YAG 1
0.977 s: begin vertical jog
0.981 s: end vertical jog
YAG 1 +5ms: YAG 2

1.01025 s: YAG 1
1.011 s: begin vertical jog
1.015 s: end vertical jog
YAG 1 +5ms: YAG 2

1.04359 s: YAG 1
1.044 s: begin vertical jog
1.048 s: end vertical jog
YAG 1 +5ms: YAG 2

1.07692 s: YAG 1
1.077 s: begin vertical jog
1.081 s: end vertical jog
YAG 1 +5ms: YAG 2

1.11025 s: YAG 1
1.111 s: begin vertical jog
1.115 s: end vertical jog
YAG 1 +5ms: YAG 2

1.14358 s: YAG 1
1.144 s: begin vertical jog
1.148 s: end vertical jog
YAG 1 +5ms: YAG 2

1.17692 s: YAG 1
1.177 s: begin vertical jog
1.181 s: end vertical jog
YAG 1 +5ms: YAG 2

1.21025 s: YAG 1
1.211 s: begin vertical jog
1.215 s: end vertical jog
YAG 1 +5ms: YAG 2

1.24358 s: YAG 1
1.244 s: begin vertical jog
1.248 s: end vertical jog
YAG 1 +5ms: YAG 2

1.27692 s: YAG 1
1.277 s: begin vertical jog
1.281 s: end vertical jog
YAG 1 +5ms: YAG 2

1.31025 s: YAG 1
1.311 s: begin vertical jog
1.315 s: end vertical jog
YAG 1 +5ms: YAG 2

1.34358 s: YAG 1
1.344 s: begin vertical jog
1.348 s: end vertical jog
YAG 1 +5ms: YAG 2

1.37692 s: YAG 1
1.377 s: begin vertical jog
1.381 s: end vertical jog
YAG 1 +5ms: YAG 2

1.41025 s: YAG 1
1.411 s: begin jog
1.415 s: end jog
YAG 1 +5ms: YAG 2

1.44358 s: YAG 1
1.444 s: begin vertical jog
1.448 s: end vertical jog
YAG 1 +5ms: YAG 2

1.47691 s: YAG 1
1.477 s: begin vertical jog
1.481 s: end vertical jog
YAG 1 +5ms: YAG 2

1.51025 s: YAG 1
1.511 s: begin vertical jog
1.515 s: end vertical jog
YAG 1 +5ms: YAG 2

1.54358 s: YAG 1
1.544 s: begin vertical jog
1.548 s: end vertical jog
YAG 1 +5ms: YAG 2

1.57691 s: YAG 1
1.577 s: begin vertical jog
1.581 s: end vertical jog
YAG 1 +5ms: YAG 2

1.61025 s: YAG 1
1.611 s: begin vertical jog
1.615 s: end vertical jog
YAG 1 +5ms: YAG 2

1.64358 s: YAG 1
1.644 s: begin vertical jog
1.648 s: end vertical jog
YAG 1 +5ms: YAG 2


Physics Operators Plans
Entered: Mar 23 2010 06:40:58:240PM
Author: Steve Wolfe
Physop Plan and Engineering setup for Wed, Mar 24, 2010
MP#602 Core fluctuations and Transport in two-frequency ICRF-heated ITB
plasmas with jogs and modulated ICRH
SL: Ernst
PO: wolfe

Overnight Prep:
----------------
Boronization (20% mix, 60psi deposited)
Standard 20% prescription:
TC6 ~ 50mTorr,
rmin/rdwell/rmax: 44/52/103 cm
tdwell/tsweep: 60/120 s
EF4 current: 0 A
Post boronization ECDC
----------------------------
1st Gas Setup:
Gas: He
sweep: 44/103/45 cm
scan: 20/120 s
Pressure: 5e-5 Torr
Duration: 2hrs

2nd Gas Setup
Gas: D2
sweep: 44/103/45 cm
scan: 20/120 s
Pressure: 2e-4 Torr
Duration:1hr
-----------------
Engineering Setup
-----------------

Run begins at 09:00 and ends at 17:00

Power systems as on: 1100323005

Acoil: +Dtop -Dbot -Jtop +Jbot (standard) Hybrid Enabled

Gas setup:

Fill B-Top with 6 psi D2 Hybrid enabled (PG4)
fill B-side lower with 1 psi Ar Hybrid Enabled (PG1)
leave B-side upper as is Hybrid DISABLED (PG2)
fill B-main (C-side) with 40 psi D2 Hybrid enabled (PG3)
leave H-bottom as is (pumped out) DISABLED (PG5)
fill NINJA with 10 psi D2 DISABLED

Enable gatevalves and shutters: ECE, VUV, HiREX Sr, Z-bolo, B-spectrometer

Torvac gatevalve toggle (yes/no): no

Boronization(yes/no): Yes
Overnight ECDC (yes/no): Yes
ICRF(yes/no): Yes (D,E, and J)
LH(yes/no): no
Cryopump (yes/no): Yes (ready for first shot)

Vessel temperature: 35/35/35

---------------------------------
Physop Plan: (to be edited)
--------------
1. Start from whatever 1100322005,
Change Ip to 800kA, nl04 to 1e20. RF should move 100msec earlier (start
at 0.7sec). Probably modify startup? Run a few shots for boronization
recovery before proceeding

2. Decrease Ip to 700kA to try to lower ne_ped. If this doesn't
work, try increasing to 900kA.

3. Add D&/or E at total 1.2MW, modulated, starting at 1.1-1.5sec
Possibly go back to 800kA, if we aren't there already

4. Adjust on-axis timing and modulation, hopefully no DPCS changes
needed in this step

5. Increase on-axis power to max available

6. Decrease on-axis power to complete power scan

7. Add vertical jog modulation (with RF modulation) Start with
ModStart = 0.811 (or after start of H-mode), period=.067, duration=.680
RampUp = .004, RampDown=.005
ModMax=-.01, ModMin=0., DutyCycle=.50
Adjust if necessary to optimize relation to
Yag pulses at [.810253,.81523]+N/30.
Adjust Max and Min to avoid going USN,
putting too much power on W tiles, losing ITB.
Keeping this much perturbation from killing the ITB is likely going
to be the most important. Probably don't want to spend too much
time optimizing this.

8. If time and nature permit, put back RF modulation at some
compatible period.

9. Thomson scattering calibration (cut-off shot). Need a sample shot
to base this on.

10. If requested, locked mode calibration shot for HIREX_SR,
e.g. 1100316030 seg 4.





Session Leader Summaries
Entered: Mar 24 2010 09:46:44:120PM
Author: Darin Ernst
Given the limited time, and two afternoons plagued by tungsten injections in
the preceding two days, we decided to proceed agressively. If a set of
parameters didn't work, we agressively changed them to reduce wall loading.
This was hit and miss, but we were able to avoid ELM-free H-modes through small
increases in target density and current.

A computer problem prevented data access until 9:52 AM: LDAP server was
down. First shot 10:33 AM.

Overall, we obtained three strong ITB cases with central densities
reaching 5.0e20 m^-3, and successfully arrested two ITB cases with
modulated on-axis ICRH, maintaining the same central density.
There were a couple of other interesting cases with too much on-axis
heating.

We used a 50 ms on, 50 ms off modulation in the best cases 019, 020,
for which Te at the edge lagged Te on-axis by 5 ms. Total rise time
is about 17 ms for Te0, producing a 0.5 keV excursion in Te0, to reach
1.9 keV. Pedestal densities were in the range of 2.0 - 2.5e20 m^-3.

There was not enough time available to reduce the H-mode pedestal
density significantly for reflectometry access to the ITB. Operation
at lower currents and densities suffered from tungsten injections and
ELM-free onset. The cyropump may have helped with recovery from
overnight boronization, as thought during the last few days.

We found that the vertical jogs could in fact be done synchronously
with the YAG 1 TS laser. Even though DPCS allows only integer numbers
of ms, the period is retained as a floating point number. However,
the 60 Hz jog frequency required could be resonant with EF power
supplies, creating havoc. Jogs overshot and were unsuccessful in the
two attempts we made. Further development is needed.

HIREX Sr., ECE systems, PCI, Thomson scattering, reflectometry, and
cyropump were operational. The DNB was not operational. Radiated
power suffered from pickup from J-port. Given the 100 ms time
constant, radiated power measurements are possible at the end
of the H-Mode phase. Core thomson densities were calibrated
to ECE cutoffs by Yunxing two days ago. The edge thomson will
be calibrated to the core and Rayleigh scattering by Jerry.

In this experiment, J-port was operational up to 3.0 MW at 70 MHz, and
D-port up to 1.6 MW at 80 MHz. E-port was available intermittently,
but not needed.

The next step is to look for signs of increased fluctuations in PCI
during the on-axis heating, as in 1040323033.

I would especially like to thank Catherine Fiore, Arturo Dominquez,
Steve Wolfe, Yunxing Ma, Jerry Hughes, Naoto Tsujii, Matt Reinke,
Amanda Hubbard, Steve Wukitch, Martin Greenwald, and Yuri Podpaly for
their help.


Interesting shots:

shot nel04_tgt Ip ne0 ne_ped TOI (s) Pon-axis Comment
1100324002 0.8 0.8 5.0 2.5 1.213 none Rad. collapse @ 1.22s
1100324003 0.8 0.7 5.0 2.5 1.213 none Rad. collapse @ 1.24s

We tried lowering the the target density to 0.7 to get lower ne_ped
for reflectometry, but got an ELM-free H-mode. So we returned to nl04_tgt=0.8.
We also returned to 0.7 MA, since 0.8 MA did not give as clear an ITB foot.

1100324006 0.8 0.7 4.0 1.166 >1.15 s, 15/15 ms on/off, peak 1.2 MW
Appears to arrest at ne0=4 instead of 5.
Probably terminated by impurity injection.

Move on-axis power earlier and reduce power. We tried D-port 1.0 MW 30 ms
on 30 ms off, starting 1.0 sec. This was still too strong to allow significant
peaking. We reduced the duty cycle to 40 ms off, 20 ms on. Tungsten injections
were becoming more frequent, so we raised the target density to nl04_tgt=0.9 to
try to make more robust H-modes.

1100324010 0.9 0.7 3.4 all >1.0 sec, 20/40 ms on/off, peak 1.0 MW
Nice long duration weak ITB, density
falls steadily during on-axis heating.

This was almost what we wanted, but the ratio of on/off axis heating
powers is too large, and kept the central density from rising
strongly. In fact there was a strong and very interesting overall
density pumpout during on-axis heating. To counter this, we tried
raising the off-axis J-port power to 3.0 MW. Three attempts were all
multiple H-Modes terminated by tungsten injections. We also tried to
no avail jogging the strike point up after the H-mode transition to
get away from the damaged tile, as suggested by Matt Reinke. The three
terminated H-modes seemed to point to the need to reduce the
time-integrated power to the tile. So, we returned the J-port power to
2.5 MW. Given the strong pumpout in shot 10, we delayed the on-axis
D-port power to 1.1 sec. The resulting discharge had two H-modes,
with the first terminated by an impurity injection from the J-port
antenna. The second had a weak ITB. So to reduce loading on the
J-port antenna due to orbit loss, we increased the current to 0.8 MA.

This resulted in a nice, long duration H-Mode with steady ITB,
ne0=3.5, ne_ped ~ 2.7. There were two impurity injections during
H-mode.

1100324015 0.9 0.8 3.5 2.7 all >1.1 sec, 30/30 ms on/off, 0.8 MW peak

To observe trapped electron modes, we believe we need stronger density
gradients. So we delayed the on-axis power to 1.2 sec. To help avoid
tungsten injections, we moved the J-port off-axis power 100 ms
earlier, to start at 0.6 sec., just after flat-top begins. My thinking
was to reduce the time-integrated power to the tile by getting the
shot underway as quickly as possible. This was met with a fizzle
and a four H-Mode case, the first 2 or 3 ELM-free, terminated by
radiation collapse. The third was terminated by impurity injection.

So we returned to starting J-port at 0.7 sec instead of 0.6 sec, and
further delayed D-port to 1.3 sec, but increased D-port power to
observe stronger Te0 modulation.

1100324018 0.9 0.8 4.0 <1.54s >1.3 sec, 30/30 ms on/off, 1.5 MW peak

This was the best so far with on-axis heating, but still had too
much on-axis power. The ITB Extends to 1.54 sec, but then begins to
fall on the first on-axis pulse. To allow more time for the slow
density peaking, we delayed the rampdown by 100ms to 1.6 sec., and
started on-axis heating from D-port later at 1.45 sec. We extended
the RF heating phase to 1.6 sec. as well. This produced our best case,
a nice, long duration ITB, strongly peaked, and successfully arrested
with on-axis heating to maintain a steady state ITB. There are
two 50 ms on-axis heating pulses that clearly modulate Te0.

*1100324019 0.9 0.8 5.0 <1.64s >1.45s 50/50 ms on/off, 1.0 MW peak.
On-axis pulses maintain steady ITB.

Next I decided to try 700 kA for stronger peaking. This produced
another great shot, but interestingly, Te0 cools starting at 1.30 sec,
at the same time the density rise ceases, and is then reheated by the
RF pulse at 1.45 sec.

*1100324020 0.9 0.7 5.0 <1.70s >1.45s 50/50 ms on/off, 1.0 MW peak.
Density constant 1.30-1.45 sec, then
falls during on-axis heating. Prad
is flat.

We then changed the on-axis RF modulation to 20 ms on, 20 ms off, starting
1.35 s, with peak power 0.8 MW. This produced a nice, but short-lived
ITB, terminating at 1.30 sec, likely due to radiative collapse.

1100324021 0.9 0.7 5.0 <1.30s Terminated by radiative collapse.

Moving the on-axis heating earlier to 1.2 sec, with 0.8 MW peak prevented
the density from rising significantly, resulting in a weak ITB.

1100324022 0.9 0.7 >1.20s, 20/20 ms on/off, 0.8 MW peak.
Weak ITB. Too much on-axis power too
early.

We then tried to repeat shot 019 with vertical jogs, starting at 0.877s, with
4ms rise, staying in phase with TS, which has 5 ms delay from YAG 1 to YAG 2.
This resulted in two shots with early disruptions.

Physics Operator Summaries
Entered: Mar 25 2010 08:39:32:617AM
Author: Steve Wolfe
Physics Operator Summary for Wednesday, Mar 24, 2010
MP#602
SL: Ernst
PO: wolfe
EO: Parkin, Byford, Fitzgerald

The run started late due to a computer meltdown involving a hang on
Alcdaq2 and failure of the LDAP server (?) on SV4. This was corrected
by Josh, assisted by twf, mrl, and fjk, by around 10:15. There were
some additional computer related glitches early in the day, and problems
with FMIT#2 continued. However, the tokamak ran quite well, considering
we were putting power on the tungsten row, and J and D were all that
were needed from the RF and they also ran reliably. We got a surprising
amount of the run plan done. 

Boronization recovery was again very quick only a couple of shots to
get the RF going full out. The combination of the 20% bottle and 
cryopump seems to be favorable, no idea why, or whether the cryopump
is really a necessary component; the experience from Catherine's first
run is the only evidence that it might be. 

We started from a variant of 1100322007 (higher current), then dropped
back to 700kA which would make it a clone. We were getting ITB's as early
as shot#3. Some tweaking around with density (down to .7, up to .9), and
current (back up to .8MA) produced several very nice target plasmas
with long-lived ITB/EDA with and without modulated central RF, which
did seem to arrest the peaking. Raising nl04 seemed to be good, with 9e19
a suitable target at both 700 and 800kA. Amazingly, shots later in the
day seemed to be better in terms of impurity/tungsten problems than early. 
Good shots included #10, 15, 18, 19, and 20. At shot#19 we extended the
flattop to 1.6sec to let the ITB develop more before turning on the
central modulated heating. #18,19 were at 800kA, #20-22 at 700kA. Either #19
or 20 could qualify as trophy shots. #21 and 22 were trying to chase the
ITB with earlier central heating and different mod frequency, but the plasma didn't cooperate, giving inconsistent peaking rates that frustrated
the intended interventions, though some good data are probably present. 
See SL summary for more details.

On shot#23 I tried setting up the vertical jogs for scanning the plasma
past the TS channels, with the YAG1 and YAG2 staggered to fire 5msec
apart. I had previously run simulations that seemed to indicate that 
a -1cm mod amplitude would be stable, but the simulations used a period
of .060sec, rather than the correct value of .0667 (15 Hz) to match
the TS lasers. The shot disrupted vertically on the second cycle,
having railed the chopper current and the EF2L voltage on the preceding
half-cycle, with an initial excursion of ~2cm. I think probably the
15Hz modulation was coupling to the 60Hz EF2's, which always have a lag,
and driving a vicious instability. I tried cutting back to -.004 amplitude
for shot#24, but it gave a post-disruption 5 o'clock fizzle. 

Runaways weren't too bad but the startups never looked good, except for
a couple of shots (19,20) after I had raised the pg4, increased EF4
and set BR_0 from 0 to .0005. Bounces came back on shot#21 though. 

Starting around shot#16 or so, the NINJA was allowed to fire late in the
shot to take advantage of some natural R modulation with the central heating
A little later (#17) we extended the RF to 1.6 and started ramping the
TF down at 1.5 to allow Ted Golfinopolous to look at variation in 
some alfvenic modes that were showing up during the EDA/ITB. We put the
TF rampdown back to 1.6 for shot#19, but the RF extended out to 1.7sec
so he may have gotten something useful. 

We had a couple of early disruption/"fizzles" on #9 and 16, both apparently
flakes falling in. Neither of these looked like the runaway disruptions
we've been having lately. Shot#24 was a classic post-disruption fizzle,
which might have been avoided if I had dropped the pg4 by 10msec. Aside
from these, and a no-power shot#4, all plasmas ran full length until the
disruption on #23 due to the vertical kicks. 

Scorecard:
--------- 

          No Power                      1 (crit init on FFT DIO2 init)
          Duds                          0
          Fizzles/current rise disrupt  3
          Plasmas                      20
          ---------------              --
          Shots                        24

Session Leader Comments
Mar 24 2010 10:33:59:570AM1100324001Darin ErnstRF badly glitching, got a few H-Modes, about six.

Mar 24 2010 11:07:48:893AM1100324002Darin ErnstDecent ITB, ne0=5, neped=2.5 at 1.213 sec. Drop current to 700 kA to reduce
ne_ped. E-port working (not used). 2.5 MW from J-Port 70 MHz.

Mar 24 2010 02:53:38:327PM1100324003Darin ErnstVery nice ITB, ne0=5, ne_ped ~ 2.5 (bounces around), well defined ITB foot
at ~77 cm. The lower current seemed to narrow the ITB. Try to lower
target density to 0.7 to get lower ne_ped for reflectometry. Clear
impurity accumulation seems to have terminated this ITB, similar to
last shot, at around 1.236 sec. However, Prad_AXUV did not increase,
but apparently distribution of Prad changes. Sawteeth going away
toward end of ITB. TS was set with 3 ms YAG2 - YAG1 interval for
shots 2,3. Sawtooth period during ITB is 6 ms, amplitude 0.16 keV
Mar 24 2010 12:09:51:183PM1100324005Darin ErnstELM-free H-Mode. This target density is too low. NEXT: 0.6 MW on-axis heating.
E not avail.; D-port 80 MHz at 1.15 s, 15ms on, 15ms off, peak power 1.2 MW.
Lowering nel04 to 0.7 gave ELM-free H-mode, so return to nel04 of 0.8. Raising
Ip to 0.8 MA did not give a clear ITB foot, so keep 0.7 MA. We might consider
0.75 MA later on to fatten the ITB.
Mar 24 2010 12:27:29:533PM1100324006Darin ErnstGood ITB, ne0=4. Terminates about same time, but on-axis power seemed to
arrest density at 4 instead of 5 on-axis. Termination does not appear to
be due to radiative collapse. No sign of on-axis modulation in edge Te,
but core Te appears to follow (along with sawteeth).
NEXT: Move on-axis power earlier and reduce power.
Start D-port 1.0 MW 30 ms on 30 ms off, starting 1.0 sec.
Mar 24 2010 12:54:14:763PM1100324007Darin ErnstTwo H-Modes, first one terminated by a W impurity injection.
First one had no ITB before it got zapped. Second one did
not peak either. Modulation shows clearly on Te0, producing
swings from 1.62-2.24 keV, on top of sawtooth swings 1.7-2.1 keV.
Modulation flat top is about three sawtooth periods.

NEXT: repeat with modulation 40 ms off, 20 ms on.

Mar 24 2010 01:02:51:890PM1100324009Darin ErnstFizzle. Steve suggests raising target density from 0.8 to 0.9 next shot
to try to avoid the tungsten.

Mar 24 2010 01:29:17:853PM1100324010Darin ErnstExtended H-Mode with no strong impurity influx. Weak ITB reaching
neo=3.4, but both core and pedestal density decrease with time
starting at 1.0 sec., coinciding with on-axis power. Had
2.5 MW from J-port. Almost what we want, but the ratio of
on-axis to off-axis power is too high.

HIREX SR. using 10ms interval, starting -0.0969000 sec.
Profiles should be useable.

Steve says modulation producing 2.5mm p-p edge position
changes, RE: E. T. Putterich desiring puffs.

NEXT: Increase J-port to 3.0 MW.


Mar 24 2010 01:26:26:743PM1100324011Darin ErnstLate H-mode. RF power increase wasn't loaded, still 2.5 MW from J-port.
No TS this shot. No sign of density rise.
Mar 24 2010 01:38:29:827PM1100324011Darin ErnstLate H-Mode. Had early impurity injections.
No TS this shot.

Mar 24 2010 01:46:46:440PM1100324012Darin ErnstTwo H-Modes. Weak ITB during first with ne0=2.8. Terminated
by W injection? Try jogging strike point up after H-mode forms
to avoid W, as suggested by Matt.
Mar 24 2010 01:54:09:520PM1100324013Darin ErnstFour H-modes. Jogging strike point didn't work. We will extend RF 100ms
into rampdown to observe Alfven modes while the magnetic field changes.

Mar 24 2010 02:02:39:913PM1100324013Darin ErnstFour H-modes. Terminated by W injections. Jogging strike point didn't work.
We will extend RF 100ms into rampdown to observe Alfven modes while the
magnetic field changes.

NEXT: Reduce J-port to 2.5 MW, reduce D-port to 0.8 peak, 30ms on 30ms off, starting 1.1 sec.
Trying to reduce the W influx problem.

GOAL: Strongly peaked ITB with steady ne0 and strong, modulated on-axis RF.


Mar 24 2010 02:28:15:987PM1100324014Darin ErnstTwo H-modes, first terminated by impurity injection from J.
Second shows an ITB. ne0=3.3, ne_ped~ 2.0.

NEXT: Increase Ip to 800 kA to try to avoid particle losses to J.


Mar 24 2010 02:55:14:640PM1100324015Darin ErnstNice, long duration H-Mode with ITB, ne0=3.5, ne_ped ~ 2.7, steady.
Want a stronger ITB. Two impurity injections during H-mode. On-axis
power coming in at 1.1 instead of 1.2 sec. Maybe the higher current
helped. Want higher ne0.

NEXT: Delay on-axis power to 1.2 sec, increase peak power to 1.5 MW from
1.0 MW, 30 ms on, 30 ms off. Move J-port off-axis power to start 100 ms
earlier, at 0.6 sec. Want to get the shot underway as quickly as possible
to reduce time-integral of power to the bad tile.


Mar 24 2010 02:47:22:760PM1100324016Darin ErnstFizzle.

Mar 24 2010 03:12:36:393PM1100324017Darin ErnstFour H-modes, first 2 or 3 ELM-free, terminated by radiation collapse.
Third terminated by impurity injection.

NEXT: Go back to J starting at 0.7 instead of 0.6 sec., start D at 1.3 sec.,
30 ms on, 30 ms off, 1.5 MW peak.


Mar 24 2010 03:31:17:980PM1100324018Darin ErnstBest case so far with on-axis heating. Too much on-axis power.
ITB Extends to 1.54 sec., on-axis density reaches 4 and begins
to fall on first on-axis pulse. Want more peaking.

NEXT: Start rampdown at 1.6 sec, Start D later at 1.45 sec.
Make D 50ms on, 50 ms off, 1 MW peak. Extend RF to 1.6 sec.




Mar 24 2010 03:43:59:703PM1100324019Darin ErnstGood long duration ITB, ne0=5, on-axis power successfully arrests with
two pulses starting at 1.45 sec, maintains steady ITB. Best case so far.

NEXT: Try 700 kA for better peaking.




Mar 24 2010 04:06:00:570PM1100324020Darin ErnstAnother nice long duration ITB. Good peaking up to 1.3 sec, reaching
ne0=5, then Te begins to drop, and density stops rising. Flat out
to 1.45 sec, then core density drops with on-axis heating. Prad is
flat. Great shot. Te at the edge responds 5 ms after Te at the core.
Rise time is about 17 ms for Te0. Te0 begins to fall at 1.30 sec.

NEXT: Change on-axis RF modulation to 20 ms on, 20 ms off, starting 1.35 s.
Peak power 0.8 MW.



Mar 24 2010 04:17:51:403PM1100324021Darin ErnstGreat ITB with neo=5, but H-mode terminated at 1.30, likely due to radiative
collapse. Z-meter shows it. No obvious impurity injections. Neutron rate
started to fall at 1.17 sec.

NEXT: Try on-axis heating at 1.2 s, 20 ms on, 20 ms off. 0.8 MW peak.

Mar 24 2010 04:37:15:740PM1100324022Darin ErnstVery weak ITB, just too much on-axis power too early.

NEXT: Repeat shot 019 with vertical jogs, starting at 0.877s, with 4ms rise,
staying in phase with TS, which has 5 ms delay from YAG 1 to YAG 2.



Mar 24 2010 04:58:35:717PM1100324023Darin ErnstSteve says the supply railed on the jog. Try it again, but with less feeling.

NEXT: Repeat shot 019 with vertical jogs, starting at 0.877s, with 4ms rise,
staying in phase with TS, which has 5 ms delay from YAG 1 to YAG 2.



Mar 24 2010 05:03:50:000PM1100324024Darin ErnstDisruption early. Steve thinks the jogs, being locked to thomson
60 Hz, are resonant with the power supplies -> unstable. Rethink
the jogs.

NEXT: Very modest bragging. Thanks to all for your fantastic help.
Looking forward to extensive post-mortem analysis.




Physics Operator Comments
Mar 24 2010 07:13:05:953AMSteve Wolfe
Trying to get setup for Wednesday's run.
The workstations at the front desk seem to be hosed; 
possible network problem. Emailed Josh and Brandon.
Working temporarily from cmodws57. 

Boronization ran late and He ECDC is still going. Instructed
Dave to proceed with D2 ECDC, but cryo cannot start until ECDC is
done, so we will probably be late getting started. I don't want to
shorten D2 ECDC beyond one hour. 

Mar 24 2010 07:30:11:413AMSteve Wolfe
Trying to load 1100322005 into PCS and it hangs. 
Alcdaq2 seems to be hung. I can't ssh to it. 

Josh and Tom notified by email, will wait to phone for a while.
Logbook entries very slow but seem to go through. 

Mar 24 2010 09:00:06:930AMSteve Wolfe
We are having a delay for a computer problem.
No estimated uptime yet.

Mar 24 2010 09:51:05:613AMSteve Wolfe
Things are freeing up, at least temporarily.
I have now got the 1100322005 loaded into pcs.
Start making changes:
 Ip from 700kA to 800kA
 Darin no longer wants to raise the density, so leave it at 8e19
 Confirm that TF_CURRENT is -150700 which should give 5.5T
 PG1 is at 70msec. Turn off dpcs_piezo_chopper
 PG4 is at 52msec; leave for first shot, change after
Try loading it
Load at 24-Mar-2010 09:50:54.00


Mar 24 2010 09:57:22:850AMSteve Wolfe
Moving back to cmodws50 after reboot


Mar 24 2010 10:23:17:893AM1100324001Steve Wolfe
I may have loaded from the wrong shot. 
Try again load from 1100322007 (may have used 5?)
Change current from 700 to 800kA, turn off dpcs_piezo_chopper
Load at 24-Mar-2010 10:09:58.00

Shot#1: plasma full length multiple H-modes.
Amazing it actually ran after all that!
May be some delays in updates and/or analysis jobs?
Coming through now. 
J looks scruffy, D on for a short clean pulse, no E.
Bouncy looking startup, but no runaways, and pressure rises
after breakdown.

Next: repeat, let boronization recovery progress.

Mar 24 2010 10:40:53:993AM1100324002Steve Wolfe
Shot#2: No DPCS changes, rf working on E (still no gate)
plasma full length, two h-modes, first one looks promising.
Not a whole lot of peaking, probably some. L-H at .755sec, H-L at 1.21,
near RF fault. Not sure which comes first
E ran that time, FFT may still not be triggering?

Next: Drop Ip back to 700kA

Mar 24 2010 11:19:55:087AM1100324003Steve Wolfe
Shot#3: Drop Ip back to 700kA
Load at 24-Mar-2010 10:41:33.00
Delay for pcdaqeng1 problem, followed by rf monitor problem
 plasma full length, h-modes, still a fault around 1.23sec
Sawteeth go away, and a little bit of cut-off following the H-L
Soft xrays are climbing but prad_axa is not. Zave is. 
Decent ITB-like profile with well-defined foot. 
SL wants to get lower ped density, but may settle. 

Next: ?


Mar 24 2010 11:27:59:730AM1100324004Steve Wolfe
Shot#4: Decrease nl04 from .8 to .7e20
Load at 24-Mar-2010 11:22:03.00
SHould have increased the pg4 puff, but didn't.
Delay in init, finally freed up after 3minutes
Critical failure in init. Take the no-power shot,
get Josh back. 



Mar 24 2010 12:04:39:917PM1100324005Steve Wolfe
Shot#5: Increase pg4 from 52 to 55msec
relative to shot#3: Decrease nl04 from .8 to .7e20
Load at 24-Mar-2010 11:35:37.00
Delay while RF guys poke at the FFT cpci
E and FFT will be off this shot. Rebooted pcdaqdpcs2 (twice)
It complained about dio2 configuration, seemed to find acq32
on second reboot. 

plasma full length, data came in faster. One short H-mode early,
Looks like elm-free?

Next: Back to 0.8e19, turn on central heating

Mar 24 2010 12:23:18:970PM1100324006Steve Wolfe
Shot#6:  Back to 0.8e19, as on #3
turn on central heating (D a 1.2MW, modulated)
Load at 24-Mar-2010 12:05:56.00
plasma full length
2 h-modes similar to #3, ITB
D comes on too late, hardly makes it before the H-L.
ITB not as good? 

Next: Change modulation timing


Mar 24 2010 12:32:46:647PM1100324007Steve Wolfe
Shot#7: Change modulation timing (earlier, slower, less power)
No DPCS changes. 
plasma full length, 2 h-modes, first one shot, second through EOF
First H-L before modulation, looks like an injection; Tungsten?

Next: repeat and hope injection goes away (or at least later)

Mar 24 2010 12:42:59:320PM1100324008Steve Wolfe
Shot#8: repeat and hope injection goes away (or at least later)
E will try during cycle again, after two successful test shots.
No DPCS changes

plasma full length, some small early hards (again), short early H-mode
doesn't recover. 

Next: raise nl04 to 9e19, hoping to reduce W influx. 

Mar 24 2010 12:55:49:983PM1100324009Steve Wolfe
Shot#9: raise nl04 to 9e19, hoping to reduce W influx. 
Load at 24-Mar-2010 12:44:42.00

Early disruption/fizzle, .03sec 100kA. injection/bounce

Next: raise pg4

Mar 24 2010 01:17:54:740PM1100324010Steve Wolfe
Shot#10: PG4 from 55 to 60msec
EF4 from -1330 to -1310; note rcur offset is at 2000
 still with raised nl04 to 9e19, hoping to reduce W influx. 
Load at 24-Mar-2010 12:58:10.00
plasma full length, but not steady. 
Runaways early, 4e13, bad bounce but survives. 
One long H-mode, decreasing nl04 and prad. Good rf, nice clean modulation
Is there an ITB, and if not why not?
SL thinks it is a weak one. 

Next: decrease on-axis power, raise pg4

Mar 24 2010 01:31:11:950PM1100324011Steve Wolfe
Shot#11: PG4 from 60 to 65msec
Load at 24-Mar-2010 13:19:17.00
Raise J-port power
plasma full length, early runaway burst
single H-mode, late around 1.25sec
Higher power on J didn't get loaded, still at 2.5MW,
shot#10 must have been very marginal. 

Next: really raise J power

Mar 24 2010 01:41:58:800PM1100324012Steve Wolfe
Shot#12:  really raise J power
No dpcs changes, although the startup on 11 was still shaky
plasma full length, runaways down a lot 
H-mode dies at 1sec, then two more late. I'm guessing W

Try moving the strike point up around 0.8sec, see if we can 
get it off the W before it kills us.


Mar 24 2010 01:57:03:913PM1100324013Steve Wolfe
Shot#13: Try moving the strike point up around 0.8sec, see if we can 
get it off the W before it kills us. 
ZXL from 0 to .015, RXL form -.0025 to -.007
Load at 24-Mar-2010 13:47:37.00

plasma full length, but lots of short H-modes. 
May have moved rxl in too much, trying to keep the triangularity.

Next: put rxl,zxl back where it was and lower power?

Mar 24 2010 02:16:08:717PM1100324014Steve Wolfe
Shot#14:  put rxl,zxl back where it was (-.0025,0.)
and lower J power, change D-port mod
Load at 24-Mar-2010 13:59:30.00
Two h-modes, early and late. 
Lots of radiated power in the first one
injection during the second doesn't kill it
Some peaking during second H-mode
Still very bouncy startup, but runaways are not bad.

Next: decrease D power, increase pg4 again

Mar 24 2010 02:37:43:083PM1100324015Steve Wolfe
Shot#15: increase pg4 again, from 65 to 70msec
raise Ip from 700 to 800kA
Load at 24-Mar-2010 14:20:08.00
plasma full length
long eda phase, into modulation, ends 1.4 or so.
Modulation seen on neuts. 
Runaways at 1.5e13 level, still bouncing in startup
Peaking arrested? during D mods.

Next: Change central rf timing later and raise power to 1.5MW


Mar 24 2010 02:51:39:247PM1100324016Steve Wolfe
Shot#16: Change central rf timing later and raise power to 1.5MW
start J at .6sec
No DPCS changes
Early disruption. 300ka, something falls in at .05sec. No runaways
just a brick. Might be  moving vertically before it hits or gets hit?

Next: no startup changes. Move TF rampdown from 1.6 to 1.5sec
for Ted G to look at his alfvenic modes during TF ramp (thought it 
already was at 1.5). 


Mar 24 2010 03:08:32:910PM1100324017Steve Wolfe
Shot#17:  no startup changes. Move TF rampdown from 1.6 to 1.5sec
for Ted G to look at his alfvenic modes during TF ramp (thought it 
already was at 1.5). 
Relative to #15:
Change central rf timing later and raise power to 1.5MW
start J at .6sec
Load at 24-Mar-2010 14:53:31.00

plasma full length, multiple H-modes. First H-mode at .65sec looks ELM-free
Runaways ~1e13, bounce in startup.
Heating up the W again (today's code is ORANGE)

Next: Delay RF to 0.7sec, mod to 1.3sec


Mar 24 2010 03:45:04:220PM1100324019Steve Wolfe
Shot#19: Br_0 offset from 0 to .0005 (because I think it's wrong,
not that it will matter to the bounce)
Extend current and TF flattop to 1.6sec
Load at 24-Mar-2010 15:31:05.00
D delayed to 1.45
Long h-mode through EOF, looks promising, Prad pretty constant.
SL says best one so far.
No bounce in startup, maybe it did do something
Next: Drop Ip to 0.7MA


Mar 24 2010 04:01:06:130PM1100324020Steve Wolfe
Shot#20:  Drop Ip to 0.7MA
Load at 24-Mar-2010 15:44:06.00
Long EDA H-mode, ELMy things after 1sec
Peaking up nicely. May be a new best one of the day. 
Loses sawteeth at 1.3sec, well before D comes on
Prad comes down and stays down. 
W row has heated up (brown), but no obvious injections

Next: change modulation on axis

Mar 24 2010 04:18:10:530PM1100324021Steve Wolfe
Shot#21:  change modulation on axis
No DPCS changes
H-L at 1.3sec, density peaking faster this time,
didn't last to the central heating time. 
Modulations are in the second H-mode.
big bounce in startup is back, so much for the Br tweak.

Next: move mod central heating earlier

Mar 24 2010 04:25:37:673PM1100324022Steve Wolfe
Shot#22:   move mod central heating earlier
Long H-mode phase, density drops after 1.2sec, 
Not as peaked as one might hope . 
Runaways 3e13, still bouncing. 

Next: vertical jogs

Mar 24 2010 04:50:10:260PM1100324023Steve Wolfe
Shot#23:  vertical jogs starting at .877sec
Period=.0667, duration .72, 
Rise/fall=.004
ModMax=0.01, ModMin=0., duty=.5
Raise Ip from 700 to 800
Load at 24-Mar-2010 16:37:28.00

Plasma disrupts at .92sec, early in the vertical jogs
Rails EFC on the first reverse jog. Also voltage rails 
EF2L. 

Next: decrease mod amplitude a lot

Mar 25 2010 08:51:56:153AM1100324024Steve Wolfe
Shot#24:  decrease modMax from -.01 to -.004
Load at 24-Mar-2010 16:51:57.00
I didn't simulate the 15.0Hz modulation, I think it may be coupling
to the EF2's, which run off the mains. This is probably a
bad idea, but it's the last shot of the day anyway. 

Fizzle.Needed to drop the fill, looks like a lot.  
END OF RUN. 

Engineering Operator Comments
ShotTimeTypeStatusComment
110:16:27:983AMPlasmaOk
210:33:02:490AMPlasmaOk
311:07:32:830AMPlasmaOk
411:26:19:307AMPlasmaBadCritical Failure
511:57:25:570AMPlasmaOk
612:10:37:370PMPlasmaOk
712:25:28:360PMPlasmaOk
812:38:32:553PMPlasmaOk
912:52:31:793PMPlasmaOk
1001:05:51:770PMPlasmaOk
1101:21:19:650PMPlasmaOk
1201:35:12:233PMPlasmaOk
1301:49:36:630PMPlasmaOk
1402:05:21:370PMPlasmaOk
1502:24:33:720PMPlasmaOk
1602:42:57:960PMPlasmaOk
1702:55:48:043PMPlasmaOk
1803:12:07:300PMPlasmaOk
1903:33:39:480PMPlasmaOk
2003:47:26:283PMPlasmaOk
2104:06:17:993PMPlasmaOk
2204:19:37:547PMPlasmaOk
2304:38:25:427PMPlasmaOk
2404:58:14:700PMPlasmaOk

System Availability
Mar 24 2010 10:15:58:170AM
SystemAvailableCommentAvailability
Changed
OwnerDescriptionWho Can
Change
Active MHDnosecured - tagged outJun 02 2008 11:03AM (sears)Joe SnipesActive MHD Antennasears,snipes
BESyes7 horizontal views spread out R_maj=[0.72-0.9m], 3 cm over the tokamak midplane (beam) Jul 09 2009 05:12PM (bespam)Bill RowanBeam Emission Spectroscopybespam,liao,rowan
Bolometersyes2pi (diode+foil), AXA, AXJ, WB3AX, WB4AX, LEDGE-AX working fine. MID-FOILS working fine. The LEDGE/DIV foils still need calibrations appliedAug 27 2009 09:09AM (mlreinke)Matt ReinkeBolometershutch,mlreinke
Boronizationyes Nov 13 2009 06:37AM (dekow)Jim IrbyBoronizationdekow,irby
CNPAnoSingle channel CNPA available as of 12/1707Jun 10 2009 10:22AM (twf)Ron ParkerCompact Neutral Analyzerparker
Cryopump - Upper DivertoryesOperationalNov 09 2009 09:12AM (irby)Brian LaBombardUpper Divertor Cryopumpirby,labombard
CXRSyesThe wide-view toroidal array was installed, spatially and absolutely calibrated. There are 7 toroidal chords intersect the beam at the major_R_range=[74.2:83.6]cm. Spatial resolution: 1.2 cm. The F-top wide-view poloidal array was installed and spatially calibrated. There are 19 chords intersect the beam at the major_R_range=[0.65:0.9.]m. Spatial resolution: 1.22 cm.Jul 09 2009 05:16PM (bespam)Bill RowanCharge Exchange Spectroscopybespam,blip,liao,rachmcd,rowan
D Alphayes\ha2_bright is available; \SPECTROSCOPY::TOP.VUV.VIS_SIGNALS:MCP_VIS_SIG1 is not availableJul 01 2009 09:45AM (terry)Jim TerryBalmer Alpha Monitorirby,terry,wolfe
Dalsayessystem is aligned and operating; not yet calibratedJul 14 2009 02:22PM (marmar)Earl MarmarVisible Bremsstrahlungmarmar
Disruption Mitigationyesoperational, but turned offNov 03 2009 04:59PM (granetz)Robert GranetzGas Jetgranetz,whyte
Divertor IRno Mar 06 2007 10:50AM Glen WurdenDivertor IR Digital Videogwurden,terry
Divertor IR Spectroscopyno Mar 06 2007 10:51AM Glen WurdenDivertor IR Spectroscopygwurden,terry
DNBnoin test mode onlyNov 03 2009 05:01PM (granetz)Robert GranetzDiagnostic Neutral Beamdterry,granetz,gung
ECDCyes Nov 09 2009 09:13AM (irby)Jim Irby2.45 GHz, 3 kW, discharge cleaning systemdekow,irby,pfeiffer
FRC-ECEyesAll channels working and calibrated.Feb 02 2010 09:56AM (phillips)Perry PhillipsHeterodyne ECE-32 channelsphillips
G-Side RatiomaticyesGauge installed. Controller repaired.Aug 23 2009 10:36PM (toland)Brian LaBombardG-side ratiomatic ionization gaugelabombard,toland
Glow Dischargeno6/10/09: Paddles meggered at 1000 v to machine ground (>300 Mohm) prior to closing the vacuum chamber. Plugs for Power Supply and Controls Unplugged and Red Tagged, Safe for In-Vessel AccessJun 10 2009 01:47PM (dekow)Jim Irby1 kV, 5 A, glow discharge systemdekow,dterry,irby
GPCyesOperational.Jul 28 2009 04:23PM (hubbard)Amanda HubbardECE Polychromaterhubbard,schmidt
GPC2yesMost channels operational.Jul 28 2009 04:23PM (hubbard)Amanda HubbardECE Polychromaterhubbard,schmidt
GPIyesThe Xpt-view, the midplane-view cameras are operational. The innerwall GPI (coupled to the fast-diodes) is available. Jul 01 2009 09:45AM (terry)Jim TerryGas Puff Imagingterry,zweben
H/Dyesposes no safety hazard for people working around the machineJun 18 2009 01:04PM (marmar)Earl MarmarH/D Ratiomarmar
HEATyes Oct 27 2009 10:39AM (dekow)Gary Dekow - Machine Heaters -burke,dekow
Hirex JryesHirex-Jr has been installed and testedJun 17 2009 11:54AM (rice)John RiceHigh Resolution X-Rayince,rice
Hirex Sryescurrently only one detector is operational, off-axis He-like argon emissionJul 14 2009 10:01AM (rice)John RiceHigh Resolution X-Rayince,rice
HXRnoIn safe mode (high voltage turned off)May 27 2008 09:19PM (schmidt)Ron ParkerHard X-Ray Arrayparker,schmidt
ICRF D-Antennayesoperational at 80 MHzJul 21 2009 04:49PM (wukitch)Steve WukitchD-port antennawukitch,ylin
ICRF E-Antennayesantenna and FFT operationalJul 21 2009 04:49PM (wukitch)Steve WukitchE-port antennawukitch,ylin
ICRF J-Antennayesavailable at 80 MHz and [0,pi,0,pi] phasingSep 04 2009 10:01AM (wukitch)Steve WukitchJ-port antennawukitch,ylin
Laser Blow-OffyesRequires some advance notice for laser start-upOct 16 2009 12:46PM (g)Martin GreenwaldImpurity Injection Systemg,nthoward
Lower HybridnoBreaker locked and tagged out. Waveguide disassembly started. Open Waveguides have been shorted.May 28 2008 09:47AM (dterry)Ron ParkerLower Hybriddterry,parker,wallaceg
LPInoBeam line removedMay 30 2008 02:09PM (bose)Earl MarmarLithium Pellet Injectorbose,marmar
Ly-alphayesMidplane, WB4-LY, LEDGE-LY all collecting data. WB1-LY not attached.Aug 27 2009 09:10AM (mlreinke)Matt ReinkeLyman alpha cameraslabombard,mlreinke
McPhersonyesOperating normallyAug 27 2009 09:10AM (mlreinke)Jim TerryVUV Spectrometermlreinke,terry
MichelsonyesOperational. Recalibrated 7/27/09.Jul 28 2009 04:22PM (hubbard)Amanda HubbardECE Te Profileshubbard,schmidt
MKSyesGauges are on. B-bottom is not operating - bad sensor head?Jul 14 2009 09:12PM (labombard)Brian LaBombardBaratron pressure gaugeslabombard
MLPnoSystem is in edge lab. Version 2 drivers ready to test. Needs a student to help work on it.Jul 14 2009 09:20PM (labombard)Brian LaBombardMirror Langmuir Probe Electronics Systemlabombard
MSEnoAs of 6/10/09, MSE is calibrated & ready to go. Awaiting DNB.Jun 10 2009 10:34AM (sscott)Steve ScottMotional Stark Effectjinseok,sscott
NeutronsyesNeutron detectors are calibrated and ready to go. Voltages will be turned on prior to first plasma.Jun 10 2009 10:35AM (fiore)Catherine FioreNeutron Detectorsfiore
Neuts and HardsyesPower back on morning of 11/09 before runNov 09 2009 09:10AM (irby)Jim IrbyX-Ray/Neutron Detector at K-Horirby
NINJAnoCapillaries were tested in-vessel. Plenum has a leak and needs fixing.Jul 14 2009 09:14PM (labombard)Brian LaBombardNINJA gas pufflabombard,terry
PCIyesOperational.Nov 12 2009 02:36PM (ntsujii)Miklos PorkolabPhase Contrast Imagingarai,ennever,ntsujii,porkolab
Penning GaugesyesHigh voltage is enabled (on during INIT, off during RECOOL).Jun 30 2009 09:42PM (labombard)Brian LaBombardPenning Ionization Gaugeslabombard
Plasma TVyesWIDE1, WIDE2 (chan 33), DANT, JANT and DIV are operational.Jul 01 2009 09:46AM (terry)Jim TerryVideo of plasma via several camerasdekow,irby,terry,wolfe
PolarimeternoSafe for Machine EntrySep 14 2007 12:17PM (irby)Jim IrbyC-Mod Polarimeterirby
Power - Alternator Air SwitchyesChecked and Operational as of 06/18/09Jun 19 2009 08:04AM (irby)Gary Dekowfeeds all the other suppliescochran,dekow,dterry,irby
Power - PEIyes Nov 13 2009 06:38AM (dekow)Gary DekowPEI Systemcochran,dekow,dterry,irby
Power - TMXyesChecked out and ready for Magnet Tests as of 06/18/09Jun 19 2009 08:07AM (irby)Gary DekowEf2s,. EFCs and A-Coilsbyford,cochran,dekow,dterry,irby
Power SystemsyesChecked out and ready for Magnet Tests as of 06/18/09Jun 19 2009 08:07AM (irby)Jim IrbyC-Mod Magnet Suppliesbyford,cochran,dekow,dterry,irby
Probe Array - ISDyesArray is operational with new cPCI data system. Probe 14 has open circuit.Jul 14 2009 09:17PM (labombard)Brian LabombardInner Lower Divertor Probe Arraylabombard
Probe Array - LHnoLH launcher not installed. Cables are unplugged at the A-port rack.Jul 14 2009 09:16PM (labombard)Brian LaBombardLower Hybrid Probe Arraylabombard
Probe Array - OSDyesArray is operational with new cPCI data system.Jul 14 2009 09:16PM (labombard)Brian LaBombardOuter Lower Divertor Probe Arraylabombard
Probe Array - UDIVyesArray is operational with new cPCI data system. Probe 13 has a bad cable (open).Jul 14 2009 09:17PM (labombard)Brian LaBombardUpper Divertor Probe Arraylabombard
ReflectometeryesSince 1100120000 the system is operational except for channel 110 which is not installedJan 20 2010 09:17AM (dominguez)Earl MarmarReflectometerdominguez,marmar
Scanning Probe - ASPyesProbe is operational. Needs plasma conditioning.Jul 14 2009 09:18PM (labombard)Brian LaBombardA-Port Scanning Probelabombard
Scanning Probe - FSPyesProbe is operational. Needs plasma conditioning.Jul 14 2009 09:18PM (labombard)Brian LaBombardF-Port Scanning Probelabombard
Scanning Probe - WASPyesSystem connected to A-port WASP. Needs plasma testing.Jul 14 2009 09:19PM (labombard)Brian LaBombardInner Wall Scanning Probelabombard,nsmick
SparkeryesSparker back in operation before 11/09 runNov 09 2009 09:11AM (irby)Jim IrbyField Emission Preionization for C-Moddekow,irby,parkin
TCIyesSystem restarted and aligned. Several chords available including NL04. Density Feedback available. Ready for plasma.Jun 18 2009 09:15AM (murray)Jim IrbyTwo-Color Interferometerirby,murray
TS - Coreyes2 lasers working. 15 spatial channels operational. Calibrated Oct 13 2009 07:49PM (yma)Jerry HughesCore Thomson scatteringjwhughes,yma,zhang
TS - EdgeyesOperational, and calibrated.Oct 11 2009 03:01PM (jwhughes)Jerry HughesMillimeter resolution edge Thomson scatteringjwhughes,yma,zhang
Vispecno350 and 670 nm, 4 spatial channelsNov 28 2007 09:58AM (agraf)Alex GrafVisible Spectrometeragraf
XTOMOyesOperating normallyAug 27 2009 09:10AM (mlreinke)Bob GranetzX-Ray Tomographygranetz,mlreinke

Apr 26 2010 08:46:43:870PM
Re-fit runs 1100316~26

IDL> restore,'abs_100323.dat'
IDL> restore,'wp100423.dat'
IDL> mdsopen,'electrons',1100322001
% Compiled module: MDSOPEN.
% Compiled module: MDSVALUE.
% Compiled module: MDSCHECKARG.
% Compiled module: MDSISCLIENT.
% Compiled module: MDSIDLIMAGE.
IDL> mdsput,'calib:abs_calib','$',ab
% Compiled module: MDSPUT.
IDL> mdsput,'calib:calib_coef','$',gn_rt
IDL> mdsclose
% Compiled module: MDSCLOSE.


IDL>put_wp, 1100322001, 1100316001, 1100326040
IDL> put_abs_calib, 1100322001, 1100316001, 1100326040
% Compiled module: PUT_ABS_CALIB.
IDL> put_errors, 1100316001, 1100326040
% Compiled module: PUT_ERRORS.
IDL> re_fit_tab, 1100316001, 1100326040