| Miniproposals | ||||||||||
| ||||||||||
|
| Operators | |
| Session leader(s): | Matt Reinke,Bruce Lipschultz |
| Physics operator(s): | Syunichi Shiraiwa |
| Engineering operator(s): | Bill Parkin,Andy Pfeiffer,Gary Dekow |
| Engineering Operator Run Comment |
| MP 534: Foil/AXUV Bolometer Calibration in Helium |
| Session Leader Plans |
| Entered: Dec 2 2010 12:06:17:643PM |
| Author: Matt Reinke |
| Run Plan for MP 534: Foil/AXUV Bolometer Calibration in Helium
The primary goal of the run will make highly radiative discharges in He main-ion plasmas for calibration of the AXUV and foil bolometry. Because this will likely not take the entire 1/2 day a number of piggyback experiments that require He main ion will be run. ECDC in He overnight B-SIDE-LOWER with 1.5 psi Ar a) LHRF into He plasmas b) scopping for MP 622 - growing W fuzz c) I-Modes in He d) HIREXSR calibration 1) Main-ion change-over (5-6 shots) Load 1090821016 (LSN, 5.4 T, 1 MA, nl04=1.0), no cryopump, no Ar LHRF from 0.7-1.0 (P, n|| as desired) ICRF from 1.05-1.45 (D+E) P=2-3 MW adjust ICRF level to find H-mode threshold. Monitor IR for temperature change, adjust nl04, Bt to change L/H threshold to increase PSOL to divertor. if (b) gets what they need before main-ion is changed over, then adjust nl04 as desired by (a) 2) Bolometry calibration (3-4 shots) Load 1100720015 (USN, 0.8 MA, nl04=0.8). Change Bt=5.4 T enable cryopump ICRF (D+E) at 3 MW from 0.5-1.5 B-SIDE-LOWER steady puff from 0.4-1.3 Adjust ICRF power/Ar puff to get steady highly radiative discharge. 3) Large outer gap scan (2 shots) Load 1100720026 which is a 10 cm outer gap scan at fixed q, cryopump off 4) I-modes in Helium (remaining) Return to equilibrium from (2) and increase ICRF power (add J-ANT) to see if there's any indication of I-modes. Cryopump off, diagnostic Ar puff |
| Entered: Dec 3 2010 09:02:33:653AM |
| Author: Jim Terry |
| For the 2nd half of the run (starting ~1:30 pm)
This is an exploratory experiment - there is no MP Session leader: B. Lipschultz/J. Terry Physics operator: S. Shiraiwa fill NINJA with 8 PSI He4 ENABLE NINJA Enable the following gate valves and shutters, assuming no vacuum problems: ECE, VUV, Z-bolo, HIREX-Sr, B-spectrometer GV no LH needed ICRF: required Cryopump: not needed DNB: not needed Necessary diagnostics: Divertor IR camera Divertor TCs LH hard x-ray array WIDE2, and LH video cameras; GHLIM camera is desirable Thermocouples on limiters ------------------------------------------------------------------------------- Run plan: We will trigger downward-going Vertical Disruption Events (VDEs) in a series of plasmas varying amounts of stored energy. The goal is to determine our capability for measuring the disruption heat-flux onto the outer divertor targets with high (i.e. 1 ms) time resolution. This may allow resolution of the heat flux onto the targets during the disruption's thermal quench from the heat-flux during the current quench. If this is not possible, then at least we will see what time resolution is possible for the heat-flux diagnostics across a disruption. Previous study has shown that the disruption heat-flux does depend on plasma current, presumably through the loss of magnetic energy (the current quench). The diagnostic "exploration" that is required: How long is the disruption radiation "flash" that obscures the target thermal emission? What integration times are needed for the IR camera in order to avoid saturation during the "flash"? Can the image stabilization be accomplished on the sub-full-frame image from the IR camera? (Sub-full-frame imaging must be used in order to get the 1 ms framing period.) Run Plan: For all of the following shots allow an unmitigated, downward-going VDE to occur at ~1.2 sec) for the following discharge conditions: 1) Ip=1.1 MA, Bt=5.4 or if need go to 6.2 T to avoid Ohmic H-modes), NL04=0.8e20 (to stay sheath-limited at target), no icrf power, similar to shot 1100824029, but strike point held constant (instead of SSEP sweep) on the 3rd tile-row from the bottom (i.e. ~-50.5 cm). I expect this discharge to have Wp~40 kJ. IR camera setup for full-frame images (frame rate =380 Hz - 2.6 ms time resolution) repeat as necessary, modifying integration times to avoid saturation in camera during the disruption "flash" (3 shots anticipated) 2) repeat 1 with sub-windowing for IR camera with 1 kHz frame rate (1 ms time resolution) (3 shots anticipated) 3) Ip=1.0 MA, Bt=5.4 T, NL04 (target)=0.9e20 (to stay sheath-limited at target), 2.5 MW of icrf power, similar to 1100819020, but with RF held constant and the outer strike point held constant at -50.5 cm. I expect this discharge to have Wp~150 kJ. IR camera setup for full-frame images (frame rate =380 Hz - 2.6 ms time resolution) repeat as necessary, modifying integration times to avoid saturation in camera during the disruption "flash" (3 shots anticipated) 4) repeat 3 with sub-windowing for IR camera with 1 kHz frame rate (1 ms time resolution) (5 shots anticipated) 5) if time permits, drop the current to Ip=0.8 MA and NL04 (target)=0.5e20 (to stay sheath-limited at target),2.5 MW of icrf power. |
| Physics Operators Plans |
| Entered: Dec 4 2010 03:00:00:060PM |
| Author: Syunichi Shiraiwa |
| Engineering setup for Friday 2010/12/03
MP534 (1/2 day) - Foil/AXUV Bolometer Calibration in Helium MP??? - Disruption energy deposition on divertor Session leaders: M. Reinke and J. Terry Physics operator: S. Shiraiwa Overnight: ECDC in He overnight Run will start at 09:00 and end at 17:00. Power systems as on: 1090821016 (LSN, 5.4 T, 1 MA, nl04=1.0, no Cryopump) A-coil configuration: +Dtop -Dbot -Jtop +Jbot; Hybrid enabled (standard configuration) Gas Setup: fill B-Top with 6 psi D2 Hybrid enabled (PG4) fill B-side lower with 1.5 psi Ar Hybrid DISABLED(PG1) leave B-side upper as is Hybrid DISABLED (PG2) fill B-main (C-side) with 40 psi He Hybrid enabled (PG3) fill NINJA with 8 psi He4 DISABLED (will be used in the 2nd session) Enable the following gate valves and shutters, assuming no vacuum problems: ECE, VUV, Z-bolo, HIREX-Sr, B-spectrometer GV Torvac setup: keep gate valves open during pulse Specific diagnostics: DNB, HXR array, LH probes, Divertor IR camera, Divertor TCs, WIDE2 and LH video cameras, Thermocouples on limiters, GHLIM camera(Desired) ------------------------------------------------------------------------------ Boronization: no Impurity injector: no ICRF: yes LH: yes DNB: yes Cryopump: yes ------------------------------------------------------------------------------ |
| Session Leader Summaries |
| Entered: Dec 3 2010 04:24:29:513PM |
| Author: Matt Reinke |
| Summary for MP534: Foil/AXUV Bolometer Calibration in Helium (shots 001-018)
We actually got a little of everything done on the MP. Overall the "light bulb" plasmas that were easily made back in FY08 were difficult to make so we didn't get as good of results as 1080411, but we got the job done. Shots: 001-006 were LSN shots to change over the main-ion and tune up the ICRF as well as test LHRF into He plasmas and do some scoping studies for MP 622. LHRF was able to get a steady 700 kW coupled easily at nl04 ~ 1.0. There was no change in Vloop or nonthermals but the H-PORT He II and H I monitor had a strong ~2 kHz signal on it during the pulse. A good start according the Greg. The 1 MA, nl04~1.2 L/H threshold was found to be around 3.5 MW (ICRF+Ohmic) in 004. A 3 point density scan with ICRF should allow MP 662 to develop an acceptable target discharge. Shots: 007-010 were used to make highly radiative discharges for bolometry cross-calibration. We seeded with Ar, but the high RF power was still making the KLIM very hot so we didn't push it for too many shots. It looks like the cross-calibration is still the same as in previous years even with the new foils (see 008). The core emissivity matches and the off-axis is way different presumed to be from the spectral sensitivity of the AXUV diodes. Shots: 011-012 were outer gap scans for HIREXSR calibration. We actually got what we needed on 011, but a physop mistake had us putting ICRF into the 011 plasma which had a RGAP of 2.1->10.0 cm from t 0.63-1.3. Amazingly J+D stay coupled (see Te increase) until RGAP ~ 4.1 and E stays until it was programmed to turn off at 1.1 seconds where RGAP ~ 8.0 cm!!! A pretty good demonstration of the FFT by accident. Shots: 013-018 were attempts to make I-mode plasmas in He. I had intended on starting from an equilibrium from the light bulb plasmas, but since they didn't work out, I loaded an old D2 I-mode shot: 1100910022 which had H89~1.5 w/ 4 MW of RF and PRAD ~ 1.5 MW at nl04=0.8. Despite similar higher PNET ~ 3 MW at a similar density, we were unable to see any kind of WCM-like activity or H89 > 1.0. We started getting too much heat on the limiter so I moved up to higher density but that resulted in MARFEs that troubled the last few shots. |
| Entered: Dec 6 2010 01:12:28:270PM |
| Author: Jim Terry |
| Summary for the 2nd-half of 1101203:
This half-run was aimed at development of the capability to measure heat loads during disruptions. Development was needed for the IR camera as well as being able to explore whether variations in thermal and magnetic stored energy lead to significant changes in heat loads. In pursuit of the IR camera development data was acquired with both the standard image size, leading to 2.6 ms resolution, and smaller image size allowing of order 1 ms time resolution. The achievement of the faster time resolution concurrent with the disruptions was successful for a range of disruptions. This will allow the next step - determining if the images can be stabilized for smaller image size through a disruption. With regard to the other goal - developing a range of thermal and stored energies we were also successful - thermal energies from 50-close to 150 kJ were achieved. The plasma current was also varied from 0.8 to 1.1 MA - corresponding to a roughly (1.1/0.8)2, or 1.9x, change in magnetic energy. summary of disruption shots The disruption time for the following shots (which went) was typically 1.12 s. shot Wp(kJ) Ip IR image IR cam int time(us) "flash" frame # 019 47 1.1 full 100 427 no saturation 021 fizzle 022 41 1.1 camera time base corrupted 024 49 1.1 sub 100 1204 no sat. 025 46 1.1 sub 100 1202 no sat. 026 91 (RF) 1.1 sub 100 1199 slight sat. 027 109 (RF) 1.1 sub 100 1203 slight sat. 028 130 (RF) 1.1 no camera trigger 029 150 (RF) 1.1 no camera trigger 030 157 (RF) 1.1 sub 100 1040 flash frame is sat. 031 fizzle 032 116 (RF) 1.1 sub 100 1167 no sat. 033 121 (RF) 0.8 sub 100 1134 no sat. |
| Physics Operator Summaries |
| Entered: Dec 4 2010 07:43:12:640PM |
| Author: Syunichi Shiraiwa |
| PO Summary for 1101203 (Fri.)
MP 534: Foil/AXUV Bolometer Calibration in Helium MP ???: Determination of disruption heat-flux onto the outer divertor targets with high (i.e. 1 ms) time resolution. Session leader: M. Reinke (1st) B. Lipschultz/J. Terry (2nd) Physics operator: S. Shiraiwa Today, the machine ran very well, though there were a couple of mistake made by PO (Shot 1, 2, 12, and 19, see also SL-summary fro example) . First session was for the bolometry cross-calibration but also piggy back tasks which needs He main ion. outer gap scans for HIREXSR calibration, attempt to create He I-mode, and LHCD for He. For the second half, VDE was created by pulling plasmas first by 1cm and then ( 5ms later) turn-off vertical position control (setting three PID gains to be zero). This procedure made VDE events with a reproducible timing.
The discharge next to the disruption tend to do fizzle. But, a decrease of pre-puff to 20ms (no cryo-pump) softened this issue (at least for today's disruptions) Score : Upto ramp-down 7, Disruption 20, Fizzle 5, |
| Session Leader Comments | |||
| Dec 3 2010 09:13:45:533AM | 1101203001 | Matt Reinke | Repeat of 1090821016 equilibrium but change main-ion to He
LH from 0.7-1.0, ~300 kW ICRF from 1.05-1.45 2-3 MW ramp RESULT: rough startup but good fkattop, early Mo injection. LHRF and ICRF both coupled. No L/H transition. NEXT: repeat trying to get full RF, see L/H threshold. Lower the outer strike point ~ 5 mm |
| Dec 3 2010 09:39:17:053AM | 1101203002 | Matt Reinke | Trying to tweak the discharge for the fuzz target.
LHRF increasing ICRF trying to get steady power RESULT: x-point was mistakenly moved up, got a brief H-mode at 1.17, LH had more power coupled, still a little ratty ICRF. Getting a fast brighntess modulation on the He II monitor during the LH pulse. NEXT: move x-point down 1 cm from 001 (1.5 from 002) |
| Dec 3 2010 09:52:04:910AM | 1101203003 | Matt Reinke | Still trying for correct equilribium for fuzz test, trying to get IR camera working.
LH at 500 kW, ICRF at 2-3 MW. RESULT: RF still having trouble, IR footprint on a tile crack. LHRF seems to love helium. NEXT: Adjust J/D/E mix to get steady 2-3 MW, move strike point down another 15 mm. Increase LHRF power. |
| Dec 3 2010 10:07:33:810AM | 1101203004 | Matt Reinke | Still trying for correct equilribium for fuzz test.
LH at 800 kW, ICRF at 2-3 MW. RESULT: L/H threshold at 2.5 MW at t=1.3 sec. LH still loves it. NEXT: Move x-point inward 1 cm to try and push strike down a bit more. Increase nl04 target to 1.3 to try and push L/H threshold up. |
| Dec 3 2010 10:18:33:723AM | 1101203005 | Matt Reinke | Still trying for correct equilribium for fuzz test. Increase target density.
LH at 800 kW, ICRF at 2-3 MW. RESULT: W injection during LHRF kills the plasma. NEXT: Turn LH off, undo strike point change from 005. |
| Dec 3 2010 10:32:11:433AM | 1101203006 | Matt Reinke | Repeat equilibrium from 004 but keep target nl04 at 1.2
LH off, ICRF at 2-3 MW ramp RESULT: Steady ICRF w/o D (1.5-2.25 MW) NEXT: Move to (2) USN, Ar seeded ICRF heated |
| Dec 3 2010 10:47:33:573AM | 1101203007 | Matt Reinke | Moving on to bolometry calibration shot. 1100720015. Have ICRF steady from 0.5-1.5 at 2.5 MW (D+E). Have B-SIDE-LOWER puff from 0.4-1.3.
RESULT: Full length shot, no D-PORT (Lin says due to field). Bt was still at 6.0 from loaded shot. PRAD_MAIN < 1 MW. NEXT: Bt to 5.4 and repeat. |
| Dec 3 2010 11:02:16:400AM | 1101203008 | Matt Reinke | Moving to higher ICRF power w/ Bt @ 5.4
RESULT: Steady RF at 2.5 MW, PRAD main still too low. Red spot on K-LIM. NEXT: Increase Ar (3.7 psi in plenum, PG1 to 60 V). Increase ICRF power to 3 MW. |
| Dec 3 2010 11:22:18:807AM | 1101203009 | Matt Reinke | Higher ICRF power, more Ar. Put ICRF into ramp down to help with disruption. Not that big a deal at 0.8 MA, but if going up in Ip for I-modes, we'll want a full rampdown.
RESULT: PRAD has a break in slope after 1.0 sec (Mo) rises cools the edge. Density is way above target. Limiters even hotter on this one. NEXT: Increase right gap to help move power from limiters (maybe?). Increase Ar and ICRF. |
| Dec 3 2010 11:39:11:537AM | 1101203010 | Matt Reinke | Increasing Ar, ICRF and outer gap to try to find the light bulb.
RESULT: Get a MARFE at 0.74 which raises the radiation and kills the shot. Still getting high heating on K-LIM despite larger gap. Not working as planned... NEXT: Move to outergap scan (3). B-SIDE LOWER to 50 V from 0.4-0.5. No ICRF. |
| Dec 3 2010 12:03:09:460PM | 1101203011 | Matt Reinke | Large outergap scan for HIREXSR calibration.
RESULT: Great got 15 cm of gap scan. NEXT: Move to an I-mode target (4) w/ seeding (1100910022 w/ no LH). Ar puff at 0.3-1.0 @ 30 V |
| Dec 3 2010 12:01:09:620PM | 1101203012 | Matt Reinke | The target for the light bulbs wasn't working at all, so move to an I-mode target and try to optimize seeding there.
4 MW ICRF - 0.5 sec pulse but large Ar seeding. RESULT: Physop error kept the Ip from 011 (gap scan) somehow. Probably set a record for outergap distance for ICRF operation NEXT: Do what we intended for 12..for reals this time. |
| Dec 3 2010 12:19:30:043PM | 1101203013 | Matt Reinke | I-mode target and try to optimize seeding there.
4 MW ICRF - 0.5 sec pulse but large Ar seeding. RESULTS: Late disruption due to too much Ar. ICRF works fine, nl04 over target, getting some density from ICRF. P_NET is about the same as 1100910022 but solid H89=1.0. NEXT: Repeat with lower Ar. 1.5 psi in B-SIDE lower. |
| Dec 3 2010 12:29:48:430PM | 1101203014 | Matt Reinke | Repeat of 013 but w/ less Ar.
RESULT: Density still overshooting I think the feedback control can be tweaked to help stop this. PRAD a little higher, but Mo x2 higher, Ar x10 lower. KLIM deltaT increases from 150 to 250 degC. Still solid L-mode. NEXT: Adjust He gas feedback, shorten RF length (0.6-1.0), increase RF power to 5 MW. Increase Ar (PG1 to 70 V) |
| Dec 3 2010 12:47:17:270PM | 1101203015 | Matt Reinke | More RF, lower density, more Ar
RESULT: Density undershoots, is 0.6 at ICRF turn on. Mo way up, KLIM energy flux WAY up. This is not good. NEXT: Increase nl04 target to 1.0. |
| Dec 3 2010 01:01:27:033PM | 1101203016 | Matt Reinke | Higher target density
RESULT: Density overshoots, strong MARFE after 0.5 when Ar increases quickly leading to disruption. nl04 gets to 1.5x10^20. At least no heat on the limiters... NEXT: Move Ar puff to 0.4-0.5 |
| Dec 3 2010 01:13:13:570PM | 1101203017 | Matt Reinke | Reduce Ar, hopefully give more time for ICRF to heat and prevent MARFE.
RESULT: MARFE still affecting density control. KLIM energy flux back under control. NEXT: Adjust feedback of He fueling to try and get better density control. |
| Dec 3 2010 01:27:07:047PM | 1101203018 | Matt Reinke | Steve suggesting many density feedback changes to help w/ control
RESULT: Still crazy MARFE NEXT: Switch to PM miniproposal |
| Dec 3 2010 02:25:53:707PM | 1101203020 | Bruce Lipschultz | Disruption as programmed at 1.1 MA and nl04=0.8. Strike point on tile 4 beforehand.IR camera in full frame view. Next shot switch to narrower view and longer integration time (vut faster framing time). |
| Dec 3 2010 02:46:14:040PM | 1101203022 | Bruce Lipschultz | good disruption but some issues with camera timing. Repeat |
| Dec 3 2010 03:15:36:820PM | 1101203023 | Bruce Lipschultz | fizzle |
| Dec 3 2010 03:17:07:663PM | 1101203024 | Bruce Lipschultz | good disruption, good data. toroidal stripes on the IR view possibly indicating 'homoclinic tangles' according to B. LaBombard. |
| Dec 3 2010 03:25:47:987PM | 1101203025 | Bruce Lipschultz | good disruption, good data. toroidal stripes on the IR again. |
| Dec 3 2010 03:37:58:803PM | 1101203026 | Bruce Lipschultz | good disruption, good data. ELM free H-modes, 100 kJ (50 kJ during previous Ohmic shots). |
| Dec 3 2010 03:50:08:900PM | 1101203027 | Bruce Lipschultz | good disruption, good data. ELM free H-modes, 100 kJ 3.5MW for ELM-free H-mode at disruption. Next shot raise the density. |
| Dec 3 2010 04:45:18:960PM | 1101203028 | Bruce Lipschultz | good disruption, problems with IR camera. ELM free H-modes, 130 kJ 3.0MW for ELM-free H-mode at disruption. |
| Dec 3 2010 04:46:19:020PM | 1101203029 | Bruce Lipschultz | good disruption, no camera data. ELM free H-modes, 150 kJ peak, 130 at disruption using 4.0MW for ELM-free H-mode. |
| Dec 3 2010 04:40:07:050PM | 1101203030 | Bruce Lipschultz | good disruption, good data. ELM free H-modes, 130kJ at disruption using 4.0MW for ELM-free H-mode. |
| Dec 3 2010 04:44:20:820PM | 1101203031 | Bruce Lipschultz | fizzle |
| Dec 3 2010 04:58:21:297PM | 1101203032 | Bruce Lipschultz | H-mode but RF faulting - 130 kJ, IR data ok. Next shot try for lower current and density. |
| Dec 3 2010 05:23:54:103PM | 1101203033 | Bruce Lipschultz | H-mode 120 kJ, IR data ok. 800 kA, nl04=0.8. |
| Physics Operator Comments | |||
| Dec 3 2010 08:42:43:790AM | 1101203001 | Syunichi Shiraiwa | Load 1090821016 |
| Dec 3 2010 09:40:50:047AM | 1101203002 | Syunichi Shiraiwa | Reload 1101202030 for all segment
reload 1090821016 into the segment 2 update magnetic signal coeff. move lower X point up by 5 mm ( <- PO's mistake) Next: move lower X point down by 15 mm |
| Dec 3 2010 09:53:21:150AM | 1101203003 | Syunichi Shiraiwa | X point moves by 10mm from shot 2.
Next: moves X point lower by another 15 mm |
| Dec 3 2010 10:07:15:223AM | 1101203004 | Syunichi Shiraiwa | X point did not move much
Next moves lower X inward by 10 mm, raise target density to 1.3e19 |
| Dec 3 2010 10:17:15:533AM | 1101203005 | Syunichi Shiraiwa | Disrupted (Injection at 0.9s)
X point moved 1cm inward. Next : |
| Dec 3 2010 10:20:33:747AM | 1101203005 | Syunichi Shiraiwa | Disrupted (Injection at 0.9s)
X point moved 1cm inward. Next : LH off. Same density as shot 5 and same X point as shot 4 |
| Dec 3 2010 10:29:49:770AM | 1101203006 | Syunichi Shiraiwa | Full length
Next : Load 1100720015 |
| Dec 3 2010 10:36:01:677AM | 1101203006 | Syunichi Shiraiwa | Full length
Next : Load 1100720015, Increase pre-puff 50ms for USN. Cryoon, PG1 enabled. B-SIDE-LOWER steady puff from 0.4-1.3 |
| Dec 3 2010 10:51:36:763AM | 1101203007 | Syunichi Shiraiwa | TF field was wrong. Early X-ray
Next : set to 5.4 T and increase pre-puff from 50 ms to 55ms |
| Dec 3 2010 11:02:16:553AM | 1101203008 | Syunichi Shiraiwa | Full length. 5.4 T this time.
Next : Increase Ar. (Increase PG1 pressure from 1.5 to 4 psi.) Change PG1 voltage from 100V to 60V Put ICRF power near the end of pulse to help avoiding disruption |
| Dec 3 2010 11:21:23:530AM | 1101203009 | Syunichi Shiraiwa | Disrupted at 1.3 s
Next : Rcur reduce by 2mm to increase the outer gap to 2cm from 1.6cm Set PG1 to 100V |
| Dec 3 2010 11:48:00:863AM | 1101203010 | Syunichi Shiraiwa | Disrupted at 1.0 s. Right gap moved by roughly 4mm.
Moving to part 3. Load 1100720026 Seg 3 to Seg 4. Decrease pre-puff since the next is Cryo-off Change PG1 50V from 0.4 to 0.5 s |
| Dec 3 2010 11:54:48:853AM | 1101203011 | Syunichi Shiraiwa | Full length. Good outer gap scan.
Nest : Cryo-pump on (pre-puff increased) Load 1100910022 seg 2 to seg. 2 PG1 30V from 0.3 to 1.0s |
| Dec 3 2010 12:03:31:527PM | 1101203012 | Syunichi Shiraiwa | PO mistake. Shot was not loaded correctly....
Next : What is planned for shot 11. |
| Dec 3 2010 12:20:54:100PM | 1101203013 | Syunichi Shiraiwa | Disrupted at 1.3 s
Next : Change Bside Lower Pressure. No change in PCS |
| Dec 3 2010 12:30:49:390PM | 1101203014 | Syunichi Shiraiwa | Full length
Next: Reduce the density demands around 0.3 s Increase PG1 to 70V |
| Dec 3 2010 12:47:49:777PM | 1101203015 | Syunichi Shiraiwa | Disurpted at 1.25 s
Next : Set target density at 1.0e20m^-3 |
| Dec 3 2010 01:02:56:147PM | 1101203016 | Syunichi Shiraiwa | Disrupted at 0.63 s
Next : Change B-side lower. 70V from 0.4s to 0.5 s Increase density feedback gain |
| Dec 4 2010 03:16:43:443PM | 1101203017 | Syunichi Shiraiwa | Disrupted to 1.05 s
Next : move to the second session. Load 1100824029 and change to a constant RXL = -0.005 Pre-puff 25ms, Cryo-off fill B-main (C-side) with 40 psi D2 Hybrid enabled (PG3) fill B-side lower with 1.5 psi Ar Hybrid DISABLED(PG1) |
| Dec 3 2010 02:10:12:117PM | 1101203019 | Syunichi Shiraiwa | PO mistake. Bt request was 6T instead of 5.4T |
| Dec 3 2010 02:23:22:540PM | 1101203020 | Syunichi Shiraiwa | Disrupted at 1.12 s
Early run-away. Increase pre-puff by 3ms (25-> 27ms) |
| Dec 3 2010 02:28:54:600PM | 1101203021 | Syunichi Shiraiwa | Fizzle |
| Dec 3 2010 02:46:17:050PM | 1101203022 | Syunichi Shiraiwa | Disruption at 1.115s
Next : pre-puff 24ms (-1ms from this shot) |
| Dec 3 2010 02:54:11:640PM | 1101203023 | Syunichi Shiraiwa | Fizzle
Next : pre-puff 25ms |
| Dec 3 2010 03:13:24:650PM | 1101203024 | Syunichi Shiraiwa | Disrupted at 1.115s
Next: Pre-puff 20ms |
| Dec 3 2010 03:18:03:523PM | 1101203025 | Syunichi Shiraiwa | Disrupted at 1.115s
Next: Pre-puff 20ms |
| Dec 3 2010 03:31:11:283PM | 1101203026 | Syunichi Shiraiwa | Disrupted at 1.115s
Next: Pre-puff 20ms |
| Dec 3 2010 03:48:35:127PM | 1101203027 | Syunichi Shiraiwa | Disrupted at 1.12 s
Next: Pre-puff 20ms. Increase ne0 to 1.1 from 0.8e20 |
| Dec 3 2010 04:03:13:120PM | 1101203028 | Syunichi Shiraiwa | Disrupted at 1.12 s
Density 1.0-1.1e20 Next: Pre-puff 20ms. |
| Dec 3 2010 04:16:54:083PM | 1101203029 | Syunichi Shiraiwa | Same as before.
Next: Repeat Pre-puff 20ms. |
| Dec 3 2010 04:29:54:730PM | 1101203030 | Syunichi Shiraiwa | Same as before.
Next: Pre-puff 20ms. No change in PCS. RF power will be increased. |
| Dec 3 2010 04:42:17:750PM | 1101203031 | Syunichi Shiraiwa | Fizzle
Next : Prepuff 25 ms |
| Dec 3 2010 04:59:52:590PM | 1101203032 | Syunichi Shiraiwa | Normal disruption
Next :prepuff 20ms, 800 kA, ne0 demand =0.75e19 |
| Dec 3 2010 05:10:27:280PM | 1101203033 | Syunichi Shiraiwa | Start up OK. Density and Ip as programmed. Disrupted at 1.128 s
This is the last shot. |
| Engineering Operator Comments | ||||
| Shot | Time | Type | Status | Comment |
| 1 | 09:02:49:113AM | Plasma | Ok | |
| 2 | 09:27:32:103AM | Plasma | Ok | |
| 3 | 09:42:50:360AM | Plasma | Ok | |
| 4 | 09:57:47:927AM | Plasma | Ok | |
| 5 | 10:11:19:907AM | Plasma | Ok | |
| 6 | 10:24:42:053AM | Plasma | Ok | |
| 7 | 10:37:35:793AM | Plasma | Ok | |
| 8 | 10:52:49:373AM | Plasma | Ok | |
| 9 | 11:09:09:757AM | Plasma | Ok | |
| 10 | 11:23:15:177AM | Plasma | Ok | |
| 11 | 11:41:36:020AM | Plasma | Ok | |
| 12 | 11:55:24:540AM | Plasma | Ok | |
| 13 | 12:07:54:803PM | Plasma | Ok | |
| 14 | 12:20:47:087PM | Plasma | Ok | |
| 15 | 12:35:57:883PM | Plasma | Ok | |
| 16 | 12:50:08:870PM | Plasma | Ok | |
| 17 | 01:04:03:257PM | Plasma | Ok | |
| 18 | 01:23:45:527PM | Plasma | Ok | |
| 19 | 01:53:02:360PM | Plasma | Ok | |
| 20 | 02:09:26:703PM | Plasma | Ok | |
| 21 | 02:23:43:910PM | Plasma | Ok | |
| 22 | 02:36:59:150PM | Plasma | Ok | |
| 23 | 02:49:31:077PM | Plasma | Ok | |
| 24 | 03:02:05:307PM | Plasma | Ok | |
| 25 | 03:14:50:907PM | Plasma | Ok | |
| 26 | 03:27:25:570PM | Plasma | Ok | |
| 27 | 03:40:32:530PM | Plasma | Ok | |
| 28 | 03:58:32:997PM | Plasma | Ok | |
| 29 | 04:11:27:000PM | Plasma | Ok | |
| 30 | 04:25:00:287PM | Plasma | Ok | |
| 31 | 04:39:22:423PM | Plasma | Ok | |
| 32 | 04:52:26:127PM | Plasma | Ok | |
| 33 | 05:06:01:290PM | Plasma | Ok | |
| System Availability | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Dec 3 2010 09:02:32:247AM | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||