Engineering Operator Run Comment
Gas System Testing - B & H Hor Div

Session Leader Plans

Physics Operators Plans

Session Leader Summaries

Physics Operator Summaries

Session Leader Comments

Physics Operator Comments
Jan 13 2012 10:27:12:607AM1120113001Steve Wolfe
Setting up for gas tests using the new Piezo Valves
on B-DIV and H-div. 

Bring up PCS (live) and turn seg2 off,
rename SEG01 to "Gas tests 2012/01/13"

On ALL_Vs screen:
   Draw PG6 (B-DIVERTOR) waveform as 50V trapezoid followed by step to 100V,
     and ramp down to 0 at -2. 
   For PG7 (H-DIVERTOR) drag same waveform to -1.5sec
These will test the waveform generation, with no gas. 
For gas tests, I will want to look at turn on delays, so I'll change the
onsets to vertical. 

Load at 13-Jan-2012 10:26:58.00
Open tree /home/wolfe/pcs_scratch      -1
Open tree done



Jan 13 2012 10:51:41:993AM1120113001Steve Wolfe
Shot#1 with no pressure in the H-bot plenum
B-DIV and H-DIV Hybrid enabled at the PLC
Piezo boards set for chopper mode. 
Ratiomatic on

Shouldn't have any gas, just look at waveforms.
Still setting up scopes

Result: Gas fault on TORVAC PLC,
Got waveforms on demand signal outputs 22 and 23 as programmed.
No, i.e. zero, drive monitor or plenum signals. 

Jan 13 2012 11:05:26:280AM1120113002Steve Wolfe
Shot#2: don't understand why there was a gas fault.
Try again with no changes, see if it goes this time.
It may need the H-bottom plenum enabled? 

Result: HV enable light comes on briefly in pulse,
then gets a gas fill fault and turns off

Yuri is going to look at the logic

Jan 13 2012 11:15:13:573AM1120113003Steve Wolfe
Shot#3: Yuri is going to look at the plc logic during the cycle,
no changes yet, so it's still going to fail.

Yuri had  bypassed the fault, so it didn't come up, 
but the HV enable light came on and went off, as before, 
no permissive no HV

Jan 13 2012 11:21:36:050AM1120113004Steve Wolfe
Shot#4: Yuri is debugging, taking out one possible fault rung 
at a time. 

Result: gas fault, enable removed

Jan 13 2012 11:33:44:460AM1120113005Steve Wolfe
Shot#5: Yuri is debugging, taking out one possible fault rung 
at a time. 

Result: It looks like there is logic in the plc associated with
the non-existent main plenums for bdiv and hdiv that has not been 
bypassed. Yuri is going to try to bypass these. 

Jan 13 2012 11:43:19:603AM1120113006Steve Wolfe
Shot#6: Yuri bypassed the rungs with the nonexistent plenums

Result: Hv enable came on and stayed on. 
Signal seen from H-div, digitizer starts after -2sec, so misses the
B-div signal. 

Next: change waveforms a little, move bdiv

Jan 13 2012 11:57:07:160AM1120113007Steve Wolfe
Shot#7: Moved the waveforms, put in a back porch at 10V to 
see if it still chops near the bottom

Result: success, all looks normal
Next: put in gas (4psi neon in H-bottom plenum)
      draw sharp onset 50V .2sec waveform on B-div, starting at -1sec

Jan 13 2012 12:18:21:763PM1120113008Steve Wolfe
Shot#8:  put in gas (4psi neon in H-bottom plenum)
draw sharp onset 50V .5sec waveform on B-div, starting at -1sec
Copy to H-bot and H-div
Enable only B-div 
Note: H-Bot plenum output enabled, not B-div or H-div
I think this refers to the main plenum, not the Goetz plena. 

Load at 13-Jan-2012 12:08:00.00
Open tree /home/wolfe/pcs_scratch      -1
Open tree done

Result: Got gas, on the five scale on IGT
Plenum signal on B-DIV plenum is in the bit noise, and is coming out
negative. -.02V. I guess we should have used the 10Torr gauges? 
The MKS gauges show a peak of 0.84 Torr-liter
Rise in the ratiomatic starts at -.74sec, about 200msec delay
peaks about -.23 sec. Seems long, maybe it's reasonable.

Next: repeat with gate valves closed for 25 sec in pulse
to get a good calibration.  

Jan 13 2012 12:27:44:133PM1120113009Steve Wolfe
Shot#9:  repeat with gate valves closed for 25 sec in pulse
to get a good calibration.  

Result: MKS gauges claim 1.05 Torr-liter, which implies 2 T-l/sec for 0.5sec
at 50V demand and 4psi. 
plen_bdiv gives -.024 change, which is in the bit noise.
Well, it's not so bad if you smooth it, modifying the  gas_input scope 
shows about 1Torr-l

Next: Same puff from H-div, others all off.



Jan 13 2012 12:33:33:190PM1120113010Steve Wolfe
Shot#10: Same puff from H-div, others all off.

Result: got gas, 
Seem to have gotten a little less, MKS shows 0.87 Torr-l
plen_hdiv gets 0.84 Torr-liter (with nominal calibration)

Next: try H-bottom, this will be on the software chopper



Jan 13 2012 12:51:27:890PM1120113011Steve Wolfe
Shot#11: try H-bottom, this will be on the software chopper

Result: Hardly any pressure rise. Drive signal looks like it's only 
outputting 60V at the top of the chop. 
Pressure rose to low e-6 scale on IGT, 
torus MKS gauges show ~.02Torr-liter

Looks like there has been a change in the response of the
piezo driver or the AFOL since the last time we used this. 

Willy and Roza will check in the cell. 



Jan 13 2012 01:22:37:837PM1120113013Steve Wolfe
Shot#13: Enable B-top (and its plenum)
to compare outputs of another valve in analog mode 
Should be 16msec at 100V
Result: That gave full amplitde

Next: swap boards downstairs, B-top and H-bot



Jan 13 2012 01:29:22:183PM1120113014Steve Wolfe
Shot#14: Enable B-top and H-bot
 with swapped boards downstairs, B-top and H-bot

Result: same pathology, it didn't swap. Got 5.5V on H-bot and 10V on Btop

Next: Willy and Roza will take the BNC off and check 
       a) the loading from the valve
       b) response to 125kHz at the AFOL input 



Jan 13 2012 01:29:46:490PM1120113014Steve Wolfe
Shot#14: Enable B-top and H-bot
 with swapped boards downstairs, B-top and H-bot

Result: same pathology, it didn't swap. Got 5.5V on H-bot and 10V on Btop

Next: Willy and Roza will take the BNC off and check 
       a) the loading from the valve
       b) response to 125kHz at the AFOL input 



Jan 13 2012 01:35:05:283PM1120113015Steve Wolfe
Shot#15: Enable B-top and H-bot
 with swapped boards downstairs, B-top and H-bot
 but BNC to piezo off of H-bottom to check loading
Repeating same as shot #14

Result: Again H-bot shows 5.55V 
Also the Hybrid enable light on H-div went off after the pulse,
but the B-top enable did not. 



Jan 13 2012 01:38:20:040PM1120113015Steve Wolfe
Shot#15: Enable B-top and H-bot
 with swapped boards downstairs, B-top and H-bot
 but BNC to piezo off of H-bottom to check loading
Repeating same as shot #14

Result: Again H-bot shows 5.55V 
Also the Hybrid enable light on H-div went off after the pulse,
but the B-top enable did not. 

Next: swap the fibers, to test if the H-bot AFOL is not going
full range. That was not verified on earlier test. 



Jan 13 2012 01:54:51:407PM1120113016Steve Wolfe

Shot#16: swap the fibers, to test if the H-bot AFOL is not going
full range. That was not verified on earlier test. 
Disable B-top and Enable H-bot (with the signal from B-top AFOL)

Result: Got 9.4 V on the H-bot monitor. This implies the problem
is the AFOL. 

Repeat with B-top also enabled and the plenum pumped out



Jan 13 2012 02:00:24:807PM1120113017Steve Wolfe

Shot#17: Repeat with B-top also enabled and the plenum pumped out

Result: Get 5.5V out of the B-top signal, which means the 
problem is in the AFOL.

Next: Try replacing the AFOL transmitter for Ch 17-20, of which 
The H-bot is on channel 19. 



Jan 13 2012 02:47:03:700PM1120113018Steve Wolfe

Shot#18: Willy is going to replace the AFOL board on 17-20
with the one from 29-32 which was unused.

Result: Btop piezo monitor signal reads 8.2V, the
input as digitized on the ACQ196_1 reads 9.85 volts on the H-bot input,
which is feeding the B-top piezo. 
THe H-bottom piezo, which is seeing the pulse for B-top, reads 9.34V

Willy is going to take the original AFOL card to the shop and check
it on the bench at 10V input



Jan 13 2012 04:30:31:727PM1120113018Steve Wolfe

Shot#101: Bench testing failed.
Back in the interface room
I will run a dpcs test shot to leave 10V on 
output 19 and also on output 32. I'll need to turn off the 
software fizzle detector so it doesn't take out the signal

Draw output ending at 100V on output_19 and 10V on output_32
Turn off dpcs_fizzle
Verify I'm only running segment 1
Acknowledge the warning about disabling the fizzle. 

IDL> test_init_dpcs2,1120113101
Shot ran, I think output_19 stayed high,
but nothing from output 32. Maybe the gain is set to zero? 
Yes, the gain is undefined, as are the phys_limits. 

But the result on output_19 is 125kHz right where it should be 
The AFOL (this is the swapped one) is just fine, despitr
the fact that the board downstairs only interprets it as 8.2V (or did)

Next: Set Gains on output_31 and 32 to be the same as on 19,
and copy the waveforms over. 



Jan 13 2012 04:46:28:760PM1120113018Steve Wolfe

Shot#102: Set Gains on output_31 and 32 to be the same as on 19,
and copy the waveforms over. 
Set Gains to 0.1, phys_limits to [0.,100.]
Verify limits =[-10.,10.]
oad at 13-Jan-2012 16:33:14.00
Open tree /home/wolfe/pcs_scratch      -1
Open tree done
Take test shot #102

Worked on 32, stillnot on 31 
And it came out 1.0V on 32, instead of 10V.

I don't see what's wrong and my workstation or kde is dying. 
I can't bring up another scope




Jan 13 2012 04:46:35:233PM1120113018Steve Wolfe

Shot#102: Set Gains on output_31 and 32 to be the same as on 19,
and copy the waveforms over. 
Set Gains to 0.1, phys_limits to [0.,100.]
Verify limits =[-10.,10.]
oad at 13-Jan-2012 16:33:14.00
Open tree /home/wolfe/pcs_scratch      -1
Open tree done
Take test shot #102

Worked on 32, stillnot on 31 
And it came out 1.0V on 32, instead of 10V.

I don't see what's wrong and my workstation or kde is dying. 
I can't bring up another scope




Jan 13 2012 06:46:09:040PM1120113018Steve Wolfe

Shot#102: Tryng to understand what happened to the output_31 and
output_32 waveforms on #102, it looks like the 100V waveform was not
loaded. I had a 10.0 V signal loaded on output_32, with a gain=0.1
and a zero waveform on 31. I think the waveforms are from 101, which seems
to correspond to the timestamp on the node of 13-JAN-2012 16:14:12.32

I guess I didn't load it until #103. Well, I did do a load of something
at 16:33:14, according to the cut and paste in the logbook entry, 
but maybe I didn't hit apply on the ALL_V screen? Or maybe the 
X problem was already setting in. Of course, now I've blown away
the PCS so I can't scroll back to look at the terminal messages. 




Jan 13 2012 05:06:50:670PM1120113019Steve Wolfe

Shot#19: Loaded with a ramping waveform on output_19, 
still have the 31 and 32 as before

Next: Use an analog ramp on 19, i.e. turn off dpcs_piezo_chopper for h-bot



Jan 13 2012 05:40:35:497PM1120113020Steve Wolfe

Shot#20: Loaded with a ramping waveform on output_19, 
but remove H-Bot from dpcs_piezo_chopper list
still have the 31 and 32 as before

Result: Ch19 Ramps up to 9.80V at -.511sec, then FALLS back
abruptly to 8.2V at -.5sec. The corrsponding MOUT voltages are
 9.75 and 9.85V

Next: do another test shot to leave the output full on



Jan 13 2012 05:46:47:267PM1120113020Steve Wolfe

Shot#103:do another test shot to leave the output full on on output_19,
so Willy can retune the AFOL 
Copy ch 31 into ch19. 
Load at 13-Jan-2012 17:41:51.00
Open tree /home/wolfe/pcs_scratch      -1
Open tree done

test_init_dpcs2,1120113103

That seems to work, got to 9.85V on 19; should be sitting there.
Also, output_31 and output_32 went to 9.83 and 9.92 Volts this time.
I have no idea what changed from #102. This is worrisome as well. 




Jan 13 2012 06:03:44:527PM1120113020Steve Wolfe

Shot#103:do another test shot to leave the output full on on output_19,
so Willy can retune the AFOL 
Copy ch 31 into ch19. 
Load at 13-Jan-2012 17:41:51.00
Open tree /home/wolfe/pcs_scratch      -1
Open tree done

test_init_dpcs2,1120113103

That seems to work, got to 9.85V on 19; should be sitting there.
Also, output_31 and output_32 went to 9.83 and 9.92 Volts this time.
I have no idea what changed from #102. This is worrisome as well. 


Willy and Roza readjusted the AFOL so the output is 123.9kHz, 
which is not saturating the receiver, according to the led indicator.

Next: Take a final test shot, with the shot from 1120106033 reloaded. 



Jan 13 2012 06:06:50:503PM1120113020Steve Wolfe

Shot#104: Take a final test shot, with the shot from 1120106033 reloaded. 
This should put everything back to zero, restore the dpcs_fizzle and
put H-bot back in the dpcs_piezo_chopper list

Result: I think it worked. Scope MOUT values look normal on all channels
I still have the gain and phys_limits set on output_31 and output_32.  



Engineering Operator Comments
ShotTimeTypeStatusComment
110:46:33:690AMTestBadGas Fault
210:57:18:380AMTestBadGas Fault after the HV came on
311:10:30:410AMTestBadLooking at logic
411:19:27:353AMTestBadLooking at logic
511:27:02:197AMTestBadLogic
611:35:06:197AMTestOk
711:48:36:950AMTestOk
812:08:44:847PMTestOk
912:18:53:600PMTestOk
1012:27:47:930PMTestOk
1112:35:10:283PMTestOk
1212:58:27:363PMTestOk
1301:15:11:563PMTestOk
1401:23:22:213PMTestOk
1501:30:44:800PMTestOk
1601:41:53:373PMTestOk
1701:55:32:680PMTestOk
1802:35:36:980PMTestOk
1905:02:37:353PMTestOk
2005:08:56:090PMTestOk

System Availability
Jan 13 2012 10:46:08:290AM
SystemAvailableCommentAvailability
Changed
OwnerDescriptionWho Can
Change
Active MHDnosecured - tagged outJun 02 2008 11:03AM (sears)Joe SnipesActive MHD Antennagolfit
BESyesBES system is onlineNov 22 2011 02:21PM (bespam)Bill RowanBeam Emission Spectroscopybespam,liao,rowan
Bolometersyes2pi and midplane AXUV diodes awaiting first plasma. Divertor diodes installed, but amplifiers need to be installed. Core and divertor bolometry not yet hooked upNov 17 2011 02:45PM (mlreinke)Matt ReinkeBolometershutch,mlreinke
BoronizationnoBottle Removed. System has been tested and will be ready once bottle is installedJan 09 2012 06:32AM (dekow)Jim IrbyBoronizationdekow,irby
CECEno Jan 06 2012 03:48PM Anne WhiteCorrelation ECEwhitea
CNPAyesMultiple channel CNPA available, no safety risk for in-vessel accessApr 14 2011 12:01PM (parker)Ron ParkerCompact Neutral Analyzerparker
Cryopump - Upper DivertoryesOperational. He level meter not working on 500L dewar. Frank will take manual readings.Jan 11 2012 11:46AM (labombard)Brian LaBombardUpper Divertor Cryopumpdekow,irby,labombard
CXRSyescore-CXRS system is onlineNov 22 2011 02:21PM (bespam)Bill RowanCharge Exchange Spectroscopybespam,blip,liao,rachmcd,rowan
CXRS edgeyesMichael CHurchill has primary responsibility at the momentApr 19 2011 11:44AM (blip)Bruce LipshultzMeasurement of CXR light in the pedestal region at both the high- and low-field sidesblip,rmchurch
D Alphayes\ha2_bright is available; \SPECTROSCOPY::TOP.VUV.VIS_SIGNALS:MCP_VIS_SIG1 is not availableJul 01 2009 09:45AM (terry)Jim TerryBalmer Alpha Monitorirby,terry,wolfe
Dalsanosystem is is installed and aligned, but not functional (appears to be a broken cable)Dec 05 2011 02:41PM (marmar)Earl MarmarVisible Bremsstrahlung Imagingmarmar
Disruption Mitigationnogas jet valve disabled on 2011/04/13Apr 13 2011 02:04PM (granetz)Robert GranetzGas Jetgranetz,whyte
Divertor IRno Mar 06 2007 10:50AM Glen WurdenDivertor IR Digital Videogwurden,terry
Divertor IR Spectroscopyno Mar 06 2007 10:51AM Glen WurdenDivertor IR Spectroscopygwurden,terry
DMBolo AXUV diodesnoDiodes are taking data, but system is not fully operational yet. Post-acquisition data processing code is not yet functional.Dec 02 2011 03:27PM (golynyk)Geoff OlynykSet of six AXUV diodes in a toroidal array of midplane wall boxes.golynyk,granetz,mlreinke,whyte
DNByes Dec 01 2011 11:46AM (granetz)Robert GranetzDiagnostic Neutral Beamdterry,granetz,mochung
ECDCyesOperational as of 11/19/11Nov 22 2011 03:01PM (irby)Jim Irby2.45 GHz, 3 kW, discharge cleaning systemdekow,irby,pfeiffer
FRC-ECEnoPower off to RF and IF sections- Power on to DAApr 11 2011 02:54PM (phillips)Perry PhillipsHeterodyne ECE-32 channelsbespam,liao,phillips,rowan,whitea
FTCIno Jan 06 2012 04:36PM Jim IrbyFast two-color interferometerirby
G-Side Ratiomaticyes Dec 02 2011 12:24PM (toland)Brian LaBombardG-side ratiomatic ionization gaugelabombard,toland
Glow DischargenoPower supply unplugged and redtagged, HV cables removedApr 14 2011 08:03AM (dekow)Jim Irby1 kV, 5 A, glow discharge systemdekow,dterry,irby
GPCyesUsing new preamps.Nov 30 2011 07:51AM (hubbard)Amanda HubbardECE Polychromaterhubbard
GPC2yesDewar cold. Nov 21 2011 09:55AM (hubbard)Amanda HubbardECE Polychromaterhubbard
GPIyesThe Xpt-view, the midplane-view cameras are operational. The innerwall GPI (coupled to the fast-diodes) is available. Jul 01 2009 09:45AM (terry)Jim TerryGas Puff Imagingterry,zweben
H/Dyessystem is functional and taking dataDec 05 2011 02:42PM (marmar)Earl MarmarH/D Ratiomarmar
HEATyes Jan 09 2012 06:32AM (dekow)Gary Dekow - Machine Heaters -burke,dekow
Hirex JryesHirex-Jr has been installedJun 28 2010 03:36PM (rice)John RiceHigh Resolution X-Rayince,rice
Hirex SryesInstalled and aligned. Will be available for Ti/Vr profiles after focusing tests and locked modes are taken.Dec 06 2011 08:21AM (ypodpaly)John RiceHigh Resolution X-Raycgao,ldelgado,rice,ypodpaly
HXRyes32 Channels operational,, no safety risk for in-vessel accessApr 14 2011 12:05PM (parker)Ron ParkerHard X-Ray Arrayfaust,parker,schmidt
ICRF D-AntennanosecuredApr 14 2011 08:55AM (wukitch)Steve WukitchD-port antennawukitch,ylin
ICRF E-AntennanosecuredApr 14 2011 08:55AM (wukitch)Steve WukitchE-port antennawukitch,ylin
ICRF J-AntennanosecuredApr 14 2011 08:56AM (wukitch)Steve WukitchJ-port antennawukitch,ylin
Impurity SpectroscopyyesBeing upgraded with better optics and new spectrographs. Dan Miller has primary responsibility.Apr 19 2011 11:46AM (blip)Bruce LipshultzMeasurement of recycling light and impurity source rates at multiple surfaces around the tokamakblip,miller
Laser Blow-OffyesOperational and testedJan 06 2012 03:50PM (g)Martin GreenwaldImpurity Injection Systemg,nthoward
Lower HybridnoLH HV Breaker Opened and Tagged, TPS Upgrade UnderwayNov 10 2011 03:51PM (dterry)Ron ParkerLower Hybriddterry,parker,wallaceg
Lower Hybrid ReflectometernooffApr 15 2011 08:24AM (dominguez)Arturo DominguezLower Hybrid Reflectometerdominguez,marmar,parker,sgbaek
LoWEUSyesLoWEUS working with the shot cycle. Viewing Mo line-emission normally covered by McPherson. Raw data only in tree.Dec 01 2011 02:54PM (mlreinke)Matthew L. Reinke Long Wavelength Extreme Ultraviolet Spectrometermarkchil,mlreinke
LPInoBeam line removedMay 30 2008 02:09PM (bose)Earl MarmarLithium Pellet Injectorbose,marmar
Ly-alphanoMidplane Lyman awaiting first plasma. B-PORT and LEDGE arrays installed but do not yet have amplifiers hooked up.Nov 17 2011 02:42PM (mlreinke)Matt ReinkeLyman alpha camerasfaustian,labombard,mlreinke
Magneticsyes Dec 01 2011 11:51AM (g)Robert GranetzBp coils and flux loops for equilibrium controlgranetz,wolfe
McPhersonnonew turbo pump, installed, McP pumping down. Almost ready to open to machine.Dec 06 2011 12:16PM (mlreinke)Jim TerryVUV Spectrometermarkchil,mlreinke,terry
MichelsonyesFront end at 4.5e-8 T. Reduced speed for initial operation.Nov 30 2011 07:51AM (hubbard)Amanda HubbardECE Te Profileshubbard,schmidt
MKSyes Dec 02 2011 12:27PM (toland)Brian LaBombardBaratron pressure gaugeslabombard,toland
MLPnoSystem is in edge lab. Version 2 drivers ready to test. Needs a student to help work on it.Jul 14 2009 09:20PM (labombard)Brian LaBombardMirror Langmuir Probe Electronics Systemlabombard
MSEno1/3/2011: push-pull mechanism broken. MSE inoperativeJan 03 2012 11:47AM (sscott)Steve ScottMotional Stark Effectjinseok,sscott
NeutronsyesHigh voltage is on. We're ready for plasma.Nov 21 2011 12:58PM (fiore)Catherine FioreNeutron Detectorsfiore
Neuts and HardsyesAll detectors Powered On --- Waiting for neutrons for calibrationNov 22 2011 01:05PM (irby)Jim IrbyX-Ray/Neutron Detectors at K-Horirby
NINJAyesC-bottom and new F-bottom system ready.Dec 02 2011 12:28PM (toland)Brian LaBombardNINJA gas pufflabombard,terry,toland
PCIyesOperational.Nov 30 2011 12:45PM (ntsujii)Miklos PorkolabPhase Contrast Imagingemd,ennever,ntsujii,porkolab
Penning GaugesyesOperational. High voltage comes on during INIT and off at RECOOL.Dec 09 2011 09:48AM (labombard)Brian LaBombardPenning Ionization Gaugeslabombard
Plasma TVyesWIDE1, WIDE2 (chan 33), DANT, JANT and DIV are operational.Jul 01 2009 09:46AM (terry)Jim TerryVideo of plasma via several camerasdekow,irby,terry,wolfe
Polarimeteryes3 chords operational but with high noise levelJan 06 2012 04:22PM (irby)Jim IrbyC-Mod Polarimeterirby
Power - Alternator Air Switchyes Jan 09 2012 06:33AM (dekow)Gary Dekowfeeds all the other suppliescochran,dekow,dterry,irby
Power - PEIyes Jan 09 2012 06:33AM (dekow)Gary DekowPEI Systemcochran,dekow,dterry,irby
Power - TMXyes Jan 09 2012 06:33AM (dekow)Gary DekowEf2s,. EFCs and A-Coilsbyford,cochran,dekow,dterry,irby
Power Systemsyes Jan 09 2012 06:33AM (dekow)Jim IrbyC-Mod Magnet Suppliesbyford,cochran,dekow,dterry,irby
Probe Array - ISDyesOperational, except for probe #1 (open circuit) and probe #16 (not installed).Dec 09 2011 09:50AM (labombard)Brian LabombardInner Lower Divertor Probe Arraylabombard
Probe Array - LHyesOperational, except for bottom-D probe (open circuit).Dec 09 2011 09:52AM (labombard)Brian LaBombardLower Hybrid Probe Arraylabombard
Probe Array - OSDyesF-divertor ok. J-divertor ok, except for JDIV_06 which is shorted out.Dec 09 2011 09:54AM (labombard)Brian LaBombardOuter Lower Divertor Probe Array (F and J)labombard
Probe Array - UDIVyesOperational except for probes #4, 5 and 13 (open)Dec 09 2011 09:56AM (labombard)Brian LaBombardUpper Divertor Probe Arraylabombard
ReflectometeryesAs of 1120105xxx, all channels are connectedJan 05 2012 09:14AM (dominguez)Earl MarmarReflectometerdominguez,marmar
Scanning Probe - ASPyesIon Sensitive Probe head installed. Testing in piggy-back mode.Dec 20 2011 12:33PM (labombard)Brian LaBombardA-Port Scanning Probelabombard
Scanning Probe - FSPyesOperational.Dec 09 2011 09:57AM (labombard)Brian LaBombardF-Port Scanning Probelabombard
Scanning Probe - WASPyesA-port WASP is operational. K-port WASP is dead (all probes melted together).Dec 09 2011 09:58AM (labombard)Brian LaBombardInner Wall Scanning Probelabombard,nsmick
Shoelace AntennayesFixed frequency operation available (100kHz) at moderate power (< 0.5kW). Otherwise in receiver mode.Jan 11 2012 11:50AM (labombard)Brian LaBombardHigh-k active MHD antenna, for QCM, WCM studiesgolfit,labombard
SOL reflectometeryesc-port waveguides operational, h-port waveguides are not.Nov 29 2011 12:04PM (cornwall)Cornwall LauSOL reflectometercornwall,wukitch,ylin
SparkeryesNew Type II sparker replaced Type I on 11/17/2011Nov 23 2011 08:53AM (irby)Jim IrbyField Emission Preionization for C-Moddekow,irby,parkin
Surface Science StationnoSSS is up to air, no safety issuesApr 14 2011 12:06PM (ochoukov)Bruce LipschultzRadially scanning probe located on K-portblip,ochoukov
TCIyesTCI Operational. Density feedback operational.Nov 10 2011 11:47AM (irby)Jim IrbyTwo-Color Interferometerirby,murray
TS - CorenoAll lasers are shut down and cannot be fired into cell.Apr 11 2011 01:20PM (jwhughes)Jerry HughesCore Thomson scatteringjrwalk,jwhughes,yma
TS - EdgenoAll lasers are shut down and cannot be fired into cell.Apr 11 2011 01:20PM (jwhughes)Jerry HughesMillimeter resolution edge Thomson scatteringjrwalk,jwhughes,yma
Vispecno350 and 670 nm, 4 spatial channelsNov 28 2007 09:58AM (agraf)Alex GrafVisible Spectrometeragraf
XEUSyesXEUS working with the shot cycle. Viewing low-Z emission.Dec 01 2011 02:55PM (mlreinke)Matt ReinkeX-ray Extended Ultraviolet Spectrometermarkchil,mlreinke
XTOMOyesARRAY1->4 waiting first plasma, new core ARRAY5 needs debugging and the calibration data checkedNov 17 2011 02:44PM (mlreinke)Robert GranetzX-Ray Tomographyfaustian,granetz,mlreinke