Alcator C-Mod Run 930623 Information

Back to runs

Miniproposals
Miniproposal:
Date Filed:
Title:
First Author:
Session Leader:Steve Wolfe (shots 1-20)

Operators
Session leader(s):Steve Wolfe
Physics operator(s):John Goetz
Engineering operator(s):Frank Silva,Vinny Bertolino

Engineering Operator Run Comment
Lower diverter strike points

Session Leader Plans

Physics Operators Plans

Session Leader Summaries
Entered: Jul 7 2004 03:49:07:960PM
Author: To Be Determined

Summary of Run 930623

Physics Operator: Goetz
Asst. Operator: Wolfe
Eng. Operator: Silva/Bertolino

The goal of this run was to work on the location of the strike points
during lower divertor operation. A secondary goal was to determine if
plasma breakdown could be initiated without commutation of the OH1 current.
A piggyback run directed by Horne consists of EFC-chopper operation after
the plasma has terminated.

The run showed that control of the outer strike point location during lower
divertor operation is achievable. The strike point can be moved up-down and
in-out using the EF1's, EF2L, and the position controllers presently in use.
The inner strike point location is much more difficult to change and
was not under our control this run. Large oscillations were seen in the
z-position of the discharge and attention must be paid to the PID gains of ZCUR.
Non-commutation of the OH1 current was never attempted. Data were gathered on
the behavior of the chopper-EFC system and will be used for initial EFC
operation during Run 930624.

Problems:
---------
The run began at 10:30, with the delayed start being attributable to three
causes. The TF bus was being repaired because of evidence of melted solder
on the shunt (see run summary 930622). Next, an increase to just below
1.E-9 in the partial pressure of the 28 peak on the RGA occured, coincident
with the cooling of the EFCL magnet. The final delay was due an unexplained
crash and reboot of XHARPO.

Throughout the day there was an unexplained oscillation in the EF3 current.
It was not determined if the hybrid programming was at the root of the problem.
EF3 turn-on timing drifted during the run but it is not clear that the fizzles
were caused by this drift.

Wall Conditioning/Vacuum:
-------------------------
ECDC was run for about an hour prior to the run, ending at 08:30. A small
mass 28 event occured before the run. The base pressure was in the 2.E-6
range with the cryo pump valved shut.

Detailed Summary:
-----------------
The day began with programming similar to shot 930618024. This shot was
a lower null shot with ZCUR programming of -2cm and an FFIT elongation of 1.6.
Changes made to this shot include decreasing the derivative gain on ZCUR and
moving the turn-on time of the IC_EF1L and IC_EF1U proportional gains earlier
to match the EF1 crowbar times.

Shots 1-5 were attempts to move the inner and outer strike points, as
determined by FFIT, by changing the EF1 current programming. The EF1 currents
were made less negative in order to get the strike points to be below the
inner and outer noses. All these shots were diverted (except the fizzle) with
the outer strike point being below the nose. As the "scan" progressed the outer
strike point moved down the outer divertor and on shot 5 the strike point was
actually on the floor tiles. The inner strike point remained above the nose
for all of these shots.

For shots 6-10, the EF1's were removed from the ZCUR controller. The EF1's
were being current controlled by IC_EF1x. These shots showed large excursions
in the z position after the ramp down to -2 cm. Changes were made to the
ZCUR and BR_0 controllers to stop the oscillation. It was difficult to
determine the behavior of the strike points due to these large oscillations.
The large number of tweaks to ZCUR and IC_EF1x led to the fizzles on shots
11-13. During this set of shots the argon puffing was enabled to allow
HIREX to obtain ion temperature measurements.

At this point a retreat was sounded and shot 003 was loaded into the model.
Shots 14-16 were a "scan" of EF2L current to pull the x-point out. The x-point
location as determined by FFIT R_XL showed .54m -> .55m -> .56m through this
series of shots as the EF2L current went 700A -> 900A -> 1100A during the times
of interest. The outer stike point location remained satisfactory during this
scan. The inner strike point have moved slightly lower but never went below
the nose tile.

Shots 17-19 were attempts to move the plasma down with the EF2L current of
shot 16. These shots were programmed with a ZCUR of -3 cm and had large
oscillations in the z positions. The plasma moved down on shot 17 but the
x-point didn't (determined by FFIT Z_XL) Changes in the EF1 currents on shots
18 and 19 were made and led to early disruptions so no analysis of strike point
location could be done.

Shot 20 was a reload of shot 15 with a change in the RCUR controller from
0.638m to 0.65m. The discharge did indeed move out as determined by FFIT_RCUR
( 0.64 -> 0.65m) and the x-point moved from R_XL = 0.55 -> 0.56m. This change
moved the outer strike point location above the nose tile while the inner
strike point remained above the nose as always.


Score-card:
-----------
15 "good" shots
5 fizzles
0 no-breakdowns
--
20 total shots

Physics Operator Summaries

Session Leader Comments
Jun 23 1993 05:28:37:910PMSteve WolfeRun 930623

Physics Op: Goetz
Assist. Op: Wolfe
Eng. Op: Silva/Bertolino

Goal: Improve divertor strike point location.
Background task: Try EFC with chopper hooked up, firing
after the plasma pulse; this is being done by Horne.

Jun 23 1993 05:25:29:400PM930623001Steve WolfeShot 930623001

Started with shot 930618024
Lower derivative gain from 4. to 2. on ZCUR
IC_EF1U&L Proportional gain at 2. moved to 30msec from 60msec
Copied S-waveform from IC_EF1L to IC_EF1U even though we aren't controlling on IC_EF1U now.

Set up for EFC tests: EFC is disconnected for this shot because of worries
about its strngth while warm -- it isn't cooling well. EFC control is on Hybrid
on Wire 11.

Fill pressure 4.0e-5

Diverted at 160msec, looks a lot like the original shot.

Jun 23 1993 11:41:50:700AM930623002Steve WolfeRamp ef1 cuurent from 3000A @ 160msec to 2000 Amps at 200msec
Looks good. Outer strike point is right on. Inner is still on upper
side of baffle. Stops being diverted at 212.

EF3 current trace shows some funny steps.

Jun 23 1993 12:04:33:680PM930623003Steve WolfeShot 930623003

Change RCUR programming to 0.638 to get rid of out of range.
No other changes. Should be almost a repeat.

It looks ok. Still got the EF3 glitches. It seems to be coming
from one of the two halves of the EF3 supply



Jun 23 1993 12:16:52:800PM930623004Steve WolfeShot 930623004

Raise P gain on IC_EF1U&L from 2. to 3. from 30 msec on
Try to get closer to programmed currents (we only were
ramping to 2500 instead of 2000 on previous shots)

Fizzle. EF3 shows an early blip of current and may come on early.

Jun 23 1993 12:30:30:690PM930623005Steve WolfeShot 930623005

Raise hold-off voltage on EF3 to +150V
Ok - EF3 gets to 300A around 4.5 msec as opposed to 3.3 last shot
Shot went - hard xrays went up some.
Lasted to 325. Not much different from 3.

Jun 23 1993 12:48:48:230PM930623006Steve WolfeShot 930623006

Turn off Zcur controller on EF1's and turn on IC_EF1U control M=.111
and take off EF1L controller to V_EF1U.

EFC is now connected. Supply on from .45 to .6
Programming from .47 on IC_EFCU on wire 11.
Fizzle. EF3 reaches -300a at 2.9msec

EFC shows up ok.

Jun 23 1993 01:14:43:280PM930623007Steve WolfeShot 930623007

No change. Got it back
EF3 sawteeth are back and it turns on later again.
EF1's now are balanced again. Plasma lasts till 345msec
Hard xrays are up.

Still in a similar position on strike points but lose the divertor
earlier.

Jun 23 1993 01:38:26:070PM930623008Steve WolfeShot 930623008

Ramp on EF1 is now 300A at 160 to 1750 at 200msec (demand)

EF3 looks clean, 300 amp at 4.9 msec.

Disruption at 190. EF1 current traces look the same.
Big vertical oscillation starts at 130 msec.


Jun 23 1993 02:04:17:690PM930623009Steve WolfeShot 930623009

Raise Deriv gain on ZCUR from 2 to 4.
Lots of oscillations on Z position
small disruption at 270 and disrupts at 305.


Jun 23 1993 02:29:27:020PM930623010Steve WolfeShot 930623010

Lowered the V_ef1u&l from 0 to -200V until 200msec
to try to help the current feedback.

Change ZCUR gains: P from 0.5 to 1.0,D from 4. to 3.
EF1 currents peakat 2500 and drop to 1500
Mini disruption at 260-plasma moves in; then goes vertically unstable.
Outer separatrix goes down the slot, but it doesn't stay diverted. Went too
far down on the EF1s.

Jun 23 1993 02:50:39:000PM930623011Steve WolfeShot 930623011

Raised V_ef1u&l back to -100V at 110msec
Reduced BR_0 PID gain from .5 to 0.1 during 50:400msec

Fizzle. EF3 timing looks ok this time:300A at 4.68msec
Bz is too positive and Bz' is negative until 8msec.

Jun 23 1993 03:06:20:860PM930623012Steve WolfeShot 930623012

Raised V_ef1u&l back to 0V at 110msec
Change EF3 pre-pulse voltage back to +100 from +150.

Another fizzle. This time it's earlier: 300A at 4.01 msec

Jun 23 1993 03:15:59:970PM930623013Steve WolfeShot 930623013

Lower gas to 3.7e-5 (was 4.0)
Three fizzles in a row. Bz' is negative and Bz positive relative
to shot #3.

Jun 23 1993 03:29:24:390PM930623014Steve WolfeShot 930623014

Reloaded Shot#3. Still at 3.7e-5
Got it back. More hard xrays.
Strike points look good. EF3 junk is back.

Jun 23 1993 03:52:26:230PM930623015Steve WolfeShot 930623015

Start a "scan" of EF2L current, trying to move the x-point out.
Raised IC_EF2L to 900 A from 50msec to 400msec (used to have a
dip and then 700A, then ramp down to 0.)
Also gave them back the Argon pulse.

X-point out about a cm (.54 to .55) and outer strike still ok.

Jun 23 1993 04:13:18:640PM930623016Steve WolfeShot 930623016

Continue a "scan" of EF2L current, trying to move the x-point out.
Raised IC_EF2L to 1100 A from 50msec to 250msec
Also pulse gas increase 38 to 42V from 80msec to 180msec
Raise fill pressure to 4e-5

Disrupts at about 250msec. Moved x-point out to .56 but then swings back
Strike point now on the nose-piece.

Jun 23 1993 04:28:55:100PM930623017Steve WolfeShot 930623017

Change ZCUR demand to -3cm and moved Ar puff 50msec earlier.

Oscillations seem worse. Disrupt at 265ms, first mini at 250.
4cm p-p vertical oscillations, centered at -3cm.
Z of x-point oscillates about -.35 (previous was steady at -.35)
Outer strike was on nose piece.

Jun 23 1993 04:44:22:120PM930623018Steve WolfeShot 930623018

Load IC_EF1 demands from shot #8
(lower EF1 currents should lower elongation and bring
x-point down).

Moved Ar puff 5msec earlier and 5msec wider.
Disrupted around 170. Higher P_rad.
Vertical instability right away on the step down.

Jun 23 1993 04:52:50:140PM930623019Steve WolfeShot 930623019

Changed EF1 demand to 2000amps from 1750.
Disruption at 175ms

Jun 23 1993 05:13:11:080PM930623020Steve WolfeShot 930623020

Loaded shot 15 and moved RCUR from 0.638 to 0.65
Raise pressure to 4.4e-5

Disrupts at 230msec
Hard xrays saturated.
It did seem to move out by 1cm (based on FFIT)
Also moves the x-point out by 1 cm.

Jun 24 1993 09:36:33:670AM930623020Gerry TiniosStart with shot 930623017 from yesterday.
EFC-rogowski AFOL transmitter was replaced, so the EFC current offset
from yesterday shoud be fixed.
Limits on EFC-chopper voltage trace set to [-1000.,1000.]
Start with negative EFC-chopper demand voltage so as to smooth out
the kicking in of the EFC without having to program the EFC-TMX.

Log of partial pressure of mass-28: -8.58 (normal cool leak?)
Fill pressure: 4.e-5 like yesterday.

Dispupted at 184 msec. Not many hard X-rays.
EFC-current offset is zero.

Physics Operator Comments

Engineering Operator Comments
ShotTimeTypeStatusComment
110:44:31:600AMPlasmaOkat last
211:08:40:170AMPlasmaOk
311:44:14:680AMPlasmaOk330ka plasma -- no faults
412:06:19:790PMPlasmaOkfizzle
512:19:35:050PMPlasmaOk330ka plasma -- efc+chopper will be connected for next shot
612:38:13:540PMPlasmaOkfizzle
712:50:24:430PMPlasmaOk330ka plasma -- no faults
801:22:04:500PMPlasmaOk310ka plasma -- no faults
901:47:00:700PMPlasmaOk330ka plasma
1002:09:59:670PMPlasmaOk320ka plasma -- no faults
1102:33:11:270PMPlasmaOkfizzle
1202:55:26:960PMPlasmaOkfizzle
1303:07:26:070PMPlasmaOkfizzle
1403:18:28:830PMPlasmaOk340ka plasma
1503:38:45:180PMPlasmaOk340ka plasma -- no faults
1603:56:14:980PMPlasmaOk320ka plasma
1704:14:48:920PMPlasmaOk320ka plasma -- no faults
1804:33:11:620PMPlasmaOk280ka plasma -- no faults
1904:46:22:650PMPlasmaOk300ka plasma
2004:59:26:390PMPlasmaOk300ka plasma