Alcator C-Mod Run 950120 Information

Back to runs

Miniproposals
Miniproposal:
Date Filed:
Title:
First Author:
Session Leader:Yuichi Takase (shots 1-25)

Operators
Session leader(s):Yuichi Takase
Physics operator(s):Steve Horne
Engineering operator(s):Vinny Bertolino,Joe Daigle

Engineering Operator Run Comment
icrf heating

Session Leader Plans

Physics Operators Plans

Session Leader Summaries
Entered: Jul 7 2004 03:49:12:023PM
Author: To Be Determined
Session Leader's Run Summary for 950120 (ICRF Heating)

SL: Takase
PO: Horne
EO: Bertolino/Daigle

The run objective was to explore ICRF heating with 3MW of RF power.
This was the first RF run with two transmitters and two antennas
after the RF conditioning has been completed (Full power injection
of 3.5MW was achieved during the 950118 run).

The run was successful, and several interesting phenomena were observed:
On Tuesday (950118) we piggy-backed on J. Rice's run @ nel_04=0.6e20.
Today, we ran at target densities of nel_04=0.8e20 and 1.0e20.
The outer gap (separatrix - outboard limiter) varied from 0.8 to 1.2cm.
Because of coax reconfiguration, we had to spend some time trying to
figure out the tuning again (for the D-port antenna), resulting in
not-so-perfect RF pulse shapes. However, this actually gave us data
at intermediate power levels, which are useful. We had 23 useful
plasma shots (shot 10 dosrupted early and shot 25 was a power supply dud),
with 15 shots with over 2.5MW for at least a short time. Shots 6 and 14
did not have high power RF, and are good OH comparison shots. Shots 11,
13, 15, 19,21, and 24 had both antennas for 200msec. Virtually all shots
with high RF power (two antennas) transitioned into H-mode. The H-mode
character appears to be different for different densities (nel_04=0.8e20
compared to 1.0e20 or 0.6e20). At both nel_04=0.8e20 and 1.0e20, we were
able to sustain the H-mode (ELM-free then becoming ELMy) for the duration
of high power RF (200msec). Following the density rise in the ELM-free
period, a new steady-state density level is maintained until the RF pulse
is terminated. The H-factor during this quasi-steady-state H-mode was
not very large (only marginally greater than 1). At all three densities,
the stored energy (EFIT) roughly doubles from 50kJ to almost 100kJ with
3MW of RF power (the temperature increases are consequently more spectacular
at lower densities, with central Te and Ti approaching 5keV). Shot 16
had a medium Li pellet 50msec prior to the RF pulse, and a record neutron
rate (almost 1e14/sec) was obtained during the "PEP" phase when the density
profile is highly peaked.

The E-port transmitter was running reliably with 2MW (occationally even
higher) of output power. However, the peak voltage in the 9" coax was
50kV and some evidence of tuner arc was observed. The E-port coax will
be reconfigured over the weekend to lower this voltage. The D-port
transmitter has not been delivering 2MW, and needs to be improved.

P_RF > 2.5MW @ nel=1.0e20 --- Shots 011, 013, 015, 019
P_RF > 2.5MW @ nel=0.8e20 --- Shots 003, 021, 023, 024
Li pellet + ICRF "PEP" --- Shot 016

Physics Operator Summaries
Entered: Jul 7 2004 04:36:52:910PM
Author: To Be Determined
Run 950120 RF group (SL), Horne(PO), Bertolino, Daigle (EO)

Load shot 950118033 (Yuichi specified it), segment 3 from 950119023 (diverted rampdown)
fill at 1.5e--5

"Piggyback" experiments
-- observe hybrid input 81, the RF loading monitor
-- observe new disruption trigger
-- Possibly tweak the rampdown sequence.

Summary -- Very successful run. 24 good shots out of 25; no fizzles.
3-segment operation, with diverted rampdown in segment 3,
starting with shot 1.
shot 25 was a dud to the OH2 switches not closing.

One full-current disruption (shot 10); otherwise worst
disruption was 400 kA; most at less than 200. Granetz became bored.
Shot 16 had RF + lithium pellet; 1.e14 neutron rate.

Shot list

shot

1 plasma. rampdown to 100 ka.
RF 1.5 MW. Feedback signal doesn't work

No changes.

2 See 1. The right gap is growing slightly starting at about 600 msec.
This is affecting RF loading. There's very little or no change
in aout, but predictor shows the change. Sure enough, shot is programmed
to move rcur in slightly. Fixed for next shot.
Bounce scope (early ECE) shows strange and complicated breakdown behavior.

3 see 1.
Argon enabled.
outer gap is responding to rf beta increase, possibly too much.

4 No change.

5 no change.
There is a z jog at rampdown, due to programming mismatch. Fix it.
Raise line density to 1.0e15.

6 400 KA disruption on rampdown;

7 Change density programming in segment 3;
Adjust (make longer via stretch on gains) IP rampdown in segment 3
Disrupt on rampdown at 300 kA.
Try moving time of negative NL programming earlier. (to .9 sec)
Slow ramp rate some more., seg 3.

8 disruption at 200 ka.
Sharkfin at 1.15, possibly due to becoming limited.

9 lithium pellet at 1.0 sec.
Same sharkfin.
Try limiting earlier.

10 disrupted when ef2 hit rail at .35 sec or so.
Try ramping clearin more slowly

11 That worked; we're off the ef2 rail.
Disruption at 200 ka on rampdown. Bolo shows spikes,
etc. Could the edge be too hot?
Try increasin rampdown rate.

12 ece signals show high frequency (mhd?) wiggles
at disruption time.
Try increasing rate some more. It doesn't seem to be responding to these
gain changes (IP P in segment 3.)

13 Nice RF h-mode. plasm current increases during RF by 20 KA or so.

14 rf faults of some sort.
The density during rampdown is simply not dropping.
Try moving density rampdown to .85.

15 Another RF h-mode. Disrupts below 200 kA.

Try holding clearin diverted longer for next shot.
Medium lithium pellet at .65. PEP mode,...

16 Nice shot, pellet, H-mode, RF, etc.
Record neutron rate?
Disruption earlier on rampdown -- Undo clearin change -- reload 15.

17 disruption on rampdown at 200 kA.
No change for next shot.

18 See 17.
Adjust IP seg 3 gains to stretch and flatten rampdown.

19 No problems.
NL programmed to 8.e19

20 Nice shot.

21 no changes.
For next shot, turn nl feedback off, and program
gas valve, to try to keep source constant during h-mode.
One theory is that density will continue to rise, since the source
won't drop to zero. However, since nl doesn't drop for 150 msec when we turn
off the gas, I doubt this theory.
The shape of the nl curve suggests that instead, a new loss mechanism
turns on when the density rises.


22 not enough RF to test the theory, though the density didn't rise much.
back to density feedback.

23 Nominal shot. For next, change rcur by 2 mm.

24 Nominal shot.

Apparently the RF transmitters have no water flow. Long expedition
into power room to investigate.

25 After expedition returns, shot is a dud. oh2 Switches don't close.





























Session Leader Comments
Jan 20 1995 09:06:38:190AM950120001Yuichi TakaseShot 001:
RF Heating run with 2 transmitters.
Try higher density than Wednesday.

Transmitter #2 (E port) tripped.
About 1.5MW from D port.

Jan 20 1995 09:07:34:490AM950120002Yuichi TakaseShot 002:

Transmitter #2 (E port) tripped again.
About 1.5MW from D port.

nel_04=0.8

Jan 20 1995 09:28:36:430AM950120003Yuichi TakaseShot 003:

Transmitter #2 (E port) worked halfway.
Hitting 50kV.

target nel_04=0.8.
Trying to get into H-mode?

Jan 20 1995 09:44:03:790AM950120004Yuichi TakaseShot 004:

Transmitter #2 (E port) worked (lowered power).
D port mistuned.

target nel_04=0.8.

Jan 20 1995 10:14:56:700AM950120005Yuichi TakaseShot 005:

Transmitter #2 (E port) worked (0.9MW).
D port mistuned.

target nel_04=0.8.

Jan 20 1995 10:37:47:420AM950120006Yuichi TakaseShot 006:

Raised target density to nel_04=1.0
Both antennas mistuned.

Jan 20 1995 10:39:42:880AM950120007Yuichi TakaseShot 007:

target nel_04=1.0
D-port antenna still mistuned.

Jan 20 1995 11:06:55:040AM950120008Yuichi TakaseShot 008:

target nel_04=1.0
D-port antenna still mistuned.
E-port 1.1MW

Jan 20 1995 12:02:14:430PM950120009Yuichi TakaseShot 009:

target nel_04=1.0
D-port antenna better, but trips after transition into H-mode.
E-port OK.

Jan 20 1995 12:05:03:840PM950120010Yuichi TakaseShot 010:

Bad shot, disrupted @ 0.4sec.

Jan 20 1995 12:28:16:490PM950120011Yuichi TakaseShot 011:

Good shot. Both antennas.
E-port still needs tuning.

Jan 20 1995 12:40:56:320PM950120012Yuichi TakaseShot 012:

Good shot. Both antennas.
E-port tripping @ H-mode transition.

Jan 20 1995 12:58:50:400PM950120013Yuichi TakaseShot 013:

Good shot. Long (200msec) H-mode.
Both antennas stayed on (2.8MW).
Density saturates after ELM starts.

Jan 20 1995 01:14:57:480PM950120014Yuichi TakaseShot 014:

No RF (comparison shot).

Jan 20 1995 01:41:46:170PM950120015Yuichi TakaseShot 015:

H-mode (ELMy).

Jan 20 1995 01:54:29:440PM950120016Yuichi TakaseShot 016:

Li pellet @ .65sec.
H-mode @ .73sec (nel_04=1.65).

Jan 20 1995 02:13:44:590PM950120017Yuichi TakaseShot 017:

Short H-mode.
D-port tripping.

Jan 20 1995 02:14:14:740PM950120017Yuichi TakaseShot 017:

Short H-mode.
E-port (not D-port) tripping.

Jan 20 1995 02:27:41:440PM950120018Yuichi TakaseShot 018:

Short H-mode.
D-port tripped.

Jan 20 1995 02:50:33:570PM950120019Yuichi TakaseShot 019:

H-mode, both antennas on.
target nel_04=1.0, increases to 1.4 during H-mode.

Jan 20 1995 03:14:08:640PM950120020Yuichi TakaseShot 020:

no H-mode, D-port tripped.
target nel_04=0.8

Jan 20 1995 03:34:35:440PM950120021Yuichi TakaseShot 021:

H-mode, both antennas.
target nel_04=0.8

Jan 20 1995 03:42:07:210PM950120022Yuichi TakaseShot 022:

Turned off ne feedback.
no H-mode, D-port tripped.
target nel_04=0.8, increases to 0.9 during RF.

Jan 20 1995 04:09:02:150PM950120023Yuichi TakaseShot 023:

Turned ne feedback on.
H-mode, ELM-free for 40msec.
target nel_04=0.8, increases to 1.1 during H-mode.

Jan 20 1995 04:24:34:480PM950120024Yuichi TakaseShot 024:

outer gap increased to 1.2cm (was 0.8cm).
H-mode, shorter ELM-free period.
target nel_04=0.8, increases to 1.0 during H-mode.

Physics Operator Comments

Engineering Operator Comments
ShotTimeTypeStatusComment
108:34:10:440AMPlasmaOkno faults
208:51:49:660AMPlasmaOkno faults
309:07:47:650AMPlasmaOkno faults
409:31:17:060AMPlasmaOkno faults
509:54:54:070AMPlasmaOkno faults
610:16:27:600AMPlasmaOkno faults
710:34:38:580AMPlasmaOkno faults
810:54:22:230AMPlasmaOkno faults
911:11:44:470AMPlasmaOkno faults
1011:39:00:690AMPlasmaOkno faults
1111:58:46:580AMPlasmaOkno faults
1212:21:24:420PMPlasmaOkno faults
1312:42:14:430PMPlasmaOkno faults
1412:57:33:230PMPlasmaOkno faults
1501:16:08:330PMPlasmaOkno faults
1601:36:07:790PMPlasmaOkno faults
1701:55:39:810PMPlasmaOkNo Faults.
1802:14:14:490PMPlasmaOkNo faults
1902:33:28:500PMPlasmaOkNo Faults.
2002:53:35:240PMPlasmaOkNo Faults
2103:11:03:350PMPlasmaOkNo Faults.
2203:29:34:120PMPlasmaOkno faults
2303:51:15:500PMPlasmaOkNo Faults.
2404:10:22:680PMPlasmaOkNo Faults.
2504:49:55:000PMPlasmaOkNo Faults