Alcator C-Mod Run 950526 Information

Back to runs

Miniproposals
Miniproposal:104
Date Filed: 3/13/1995
Title:Investigation of plasma flows using gas injection plumes
First Author:Bruce Lipschultz
Session Leader:Gary McCracken (shots 1-38)

Operators
Session leader(s):Gary McCracken
Physics operator(s):Robert Granetz
Engineering operator(s):Joe Daigle,Frank Silva

Engineering Operator Run Comment
Granetz

Session Leader Plans

Physics Operators Plans

Session Leader Summaries
Entered: Jul 7 2004 03:49:13:803PM
Author: To Be Determined
Session Leader Run Summary 9505026

Engineering operator: F Silva/J Daigle
Physics operator: R Granetz
Session Leader: G M McCracken

Run for mini-proposals #104
Objectives: Inject impurities at the inner wall and at the divertor with the
NINJA system and follow the spatial spreading with the CCD cameras and filters
from B-side and B-top. Compare with limiter plasma. Measure impurity screening
in parallel

A very successful day. We started off with a lower x-point discharge
and puffed CH4 from the inner wall capillary (B16-001) over a range of
densities from nl04= 0.9 to 1.8e20 m^-2. CII and CIII were observed with the
B-side CCD camera in alternate discharges: in both cases the flow was clearly
seen going towards the divertor. This is in contrast to the results obtained
on 950414 when the flow reversed direction at a density of 1.5e20. The flow
direction generally reverses at the start of current rampdown as the plasma
moves back to being limited rather than diverted. Detachment was observed on
shots with densities nl04=1.3e20 or higher. The detachment occurred before
the CH4 was injected.
An upper X-point discharge was then set up with some difficulty, there
being a series of disruptions. The reasons for this were not very clear, but
it may have been due to the need for conditioning the upper divertor surfaces.
Data was eventually taken for densities from 0.9e20 up to 1.7e20. With the
upper X-point we also injected CH4 from capillaries B26-019 (lower private flux
region) and B22-022 (lower inner cone above the inner nose), shots 025 and
029. In both cases clear evidence of toroidal flow was observed in positions
which, with the upper X-point , should be close to the stagnation point. Good
FSP data were obtained for these discharges, which should allow us to compare
the flow estimated from the plumes with that estimated from the ratio of the
parallel and antiparallel Isat on the FSP
Next we studied the injection of CH4 from B16-001 into inner-wall,
limited discharges. As was to be expected the impurities were ionised rapidly
and the CII distribution was a small (10-20 mm) diameter symmetric blob. The
CIII appeared to spread all round the inner column but was also roughly
symmetric. CH4 was injected from the the B-side lower valve for the limiter
and lower x-point discharges to compare with the NINJA injections.
Many of the shots will be useful for screening estimates since in
general only one valve was used for each condition. Comparisons will be
possible for different injection conditions, different densities and different
plasma configurations, limiter, upper x-point and lower x-point. In general
good data were obtained on the midplane CII monitor at C-port and a consistent
set of data were measured with the McPherson (CVI) and Chromex spectrometers
(CII and CIII). The signals on the spectrometerrs increased very little when
the CH4 was injected, but good background shots were taken for most cases so
that at least upper limits to the screening efficiency should be obtained.
Good FSP data obtained for most conditions, including Limiter
discharges.

Shot summary:
CH4 plenum pressure 6.2 psi (330 torr) throughout. NINJA valves triggered at
0.45 s for shots #02 to 07 and at 0.55 s therafter. Pulse length 0.3 s. When
the B-side lower valve was used it was triggered from 0.6 to 0.95 s. the plenum
pressure in this case was 3.4psi.
The filter stated below refers only to the B-side camera. The B-top camera was
only recorded for a limited number of shots, eg 025,029,036, when puffing at the
bottom. The CII filter was used.

Lower X-point
01. Good background shot, No CH4, nl04=0.8e20.
02. Good shot, CH4 injected at B16-001 and B22-022, CII filter, nl04=1e20.
03. Good shot, CH4 injected at B16-001 only, CIII filter, nl04=1e20.
04. Disruption at 0.6 s.
05. Good background shot, No CH4, nl04=01.3e20.
06. Good shot, CH4 injected at B16-001, CII filter, nl04=1.3e20.
07. Good shot, CH4 injected at B16-001, CIII filter, nl04=1.3e20.
08. Good background shot, No CH4, nl04=1.64e20.
09. Good shot, CH4 injected at B16-001, CIII filter, nl04=1.64e20.
10. Good shot, CH4 injected at B16-001, CII filter, nl04=1.64e20
11. Good shot, CH4 injected at B16-001, CII filter, nl04=1.76e20: disrupted at
0.85 s.
35. Good shot, CH4 injected at B-side lower, CIII filter, nl04=1.10e20.
36. Good shot, CH4 injected at B26-019, CII filter, nl04=1.24e20. Toroidal
flow just barely detectable. Compare with shot 025 with upper X-point.
37. Good background shot, No CH4, nl04=1.15e20.

Upper X-point
12. Disrupted at 0.4 s.
13. Disrupted at 0.96 s, No CH4.
14. Disrupted at 0.4 s.
15. Disrupted at 0.4 s.
16. Fizzle.
17. Good background shot, No CH4, nl04=0.83e20.
18. Good shot, CH4 injected at B16-001, CII filter, nl04=0.87e20.
19. Good shot, CH4 injected at B16-001, CIII filter, nl04=0.88e20.
20. Disrupted at 0.4 s.
21. Good shot, CH4 injected at B16-001, CIII filter, nl04=1.28e20.
22. Good shot, CH4 injected at B16-001, CII filter, nl04=1.3e20.
23. Good background shot, No CH4, nl04=1.21e20.
24. Disrupted at 0.4 s.
25. Good shot, CH4 injected at B26-019, CII filter, nl04=1.24e20. Toroidal
flow.
26. Good shot, CH4 injected at B16-001, CII filter, nl04=1.7e20.
27. Good shot, CH4 injected at B16-001, CII filter, nl04=1.66e20.
28. Fizzle.
29. Good shot, CH4 injected at B22-022, CII filter, nl04=1.27e20. Toroidal
flow.

Limiter shots
30. Good shot, CH4 injected at B16-001, CII filter, nl04=1.20e20: disrupted at
0.85 s
31. Good shot, CH4 injected at B16-001, CIII filter, nl04=1.21e20.
32. Disrupted at 0.55 s.
33. Good shot, CH4 injected at B-side lower, CIII filter, nl04=1.17e20.
34. Died early.
38. Good background shot, No CH4, nl04=1.15e20.


Physics Operator Summaries
Entered: Jul 7 2004 04:36:54:537PM
Author: To Be Determined
Physics operator's summary for run 950526.

MP #104, Investigation of Plasma Flows Using Gas Injection Plumes

SL: G. McCracken
PO: R. Granetz

Primary purpose: Inject CH4 and He at the inner wall and divertor with
NINJA system and study plumes in plasmas with B x grad B
drift down. Compare to data from 950414, when the B x grad B
drift was reversed. We will run lower x-pt's, upper x-pt's,
and limited plasmas, at several densities and several puffing
locations.

Secondary purpose: Compare limiter/divertor screening of methane, which is a
non-recycling impurity.

Summary: Today's run plan was very challenging to say the least. Fortunately
things went very well, and we were able to get good methane plume
data in lower x-pt, upper x-pt, and limited plasmas at a range of
densities. In addition, we got good data for comparing the carbon
screening efficiency of limited and lower x-pt plasmas. The session
leader's report presents some of the plume physics results. I have
looked at the carbon screening comparison and find that there is a
factor of 2 difference. (Thanks to J.T. and Gary for showing me the
CVI and CII data.) The shots of relevance for the screening study
are 033 (limited w/methane), 035 (lower x-pt w/methane), 037 (lower
x-pt, no methane), and 038 (limited, no methane). We did not have
time to do any helium studies today. Also see shot 031 for an
example of a limited plasma with methane puffing directly into the
boundary using the NINJA inboard midplane location. It has much
higher carbon signals than the shots that had methane puffed from the
B-side-lower valve.

001-011 Diverted, lower x-pt
012-029 Diverted, upper x-pt
030-033 Limited
034-037 Diverted, lower x-pt
038 Limited


Statistics: 36 Plasmas
2 Fizzles
0 Duds
Power supply problems caused disruptions on 2 shots.
The I^2*t circuit was enabled today.


Setup: Lower x-pt: 950512027, with density changed to nl_04=0.8e20 m-2,
Upper x-pt: 950322029, with density changed to nl_04=0.8e20 m-2,
Limited : 950519021, with density changed to nl_04=1.0e20 m-2.

To speed up PCS operations, I setup a PCS the evening before this
run with segments 2,3, & 4 loaded with the three configurations,
respectively.

Engineering preparations:

Setup up from shot 950512027 (including commutation resistors,
TF start time, TF current, etc).
Normal ECDC in D2.
Set pre-fill at 1.8e-5 torr of D2.
Fill A-side with D2 at 19-20 psi.
Fill B-side-lower with methane at 3 psi, BUT LEAVE DISABLED!
Fill the NINJA plenum with methane at 5 psi.
All other gas valves should be disabled as well.


Shot 01 -- Setup for a lower x-pt baseline shot. No methane puffing.
Plasma, good shot. Density oscillated a bit.

Next shot: repeat with NINJA system puffing methane from the
inboard midplane location and from the inner divertor nose location.


Shot 02 -- Plasma, good shot. Methane plumes were seen (CII). On request from
Brian, we will tweak upper and lower x-pt positions on the next
shot. Fast scanning probe fired on this shot.

Next shot: tweak upper and lower x-pt locations, and puff methane
from midplane only.


Shot 03 -- Plasma, good shot. Good plume (CIII).

Next shot: tweak upper x-pt another 5 mm up. NINJA puff from both
locations enabled.


Shot 04 -- Plasma, disrupted at 0.59 s. VDE. Probably not due to excessive
radiation. Elongation was very high.

Next shot: lower ZCUR by 5 mm, and turn off methane in order to
get a background shot.


Shot 05 -- Plasma, good shot. Density got to the requested value of 1.3e20 m-2.
Zxray did move down.

Next shot: NINJA methane puff from midplane.


Shot 06 -- Plasma, good shot. Good plume data (CII). Jim Terry reports
that carbon background level is gradually rising from shot to shot.

Next shot: NINJA methane puff from midplane, different filter.

Engineering has enabled the I^2*t circuit starting with the next shot.

Shot 07 -- Plasma, good shot. Good plume data (CIII).

Next shot: increase density to nl_04=1.6e20 m-2. Turn off methane.


Shot 08 -- Plasma, good shot. Density got to nl_04=1.6e20 m-2 by 0.75 s.

Next shot: enable NINJA methane puff.


Shot 09 -- Plasma, good shot. Got plume data (CIII).

Next shot: NINJA methane puff with different filter.


Shot 10 -- Plasma, good shot. Good plume data (CII).

Next shot: raise density to nl_04=1.8e20 m-2. Also puff methane.


Shot 11 -- Plasma, disrupted at 0.86 s. Got good plume data anyway.

Next shot: change to upper x-pt at nl_04=0.8e20 m-2. No methane
puffing.


Switching to upper x-pt configuration starting with the next shot.

Shot 12 -- Plasma, disrupted at 0.41 s. VDE. Upper x-pt.

Next shot: slow down the current rise with a little tweak.


Shot 13 -- Plasma, disrupted at 0.96 s. VDE. Some hard x-rays. Brian wants
the plasma lowered by 1 cm so the scanning probe can reach, but
we're near the vertical stability limit.

Next shot: Turn on EF2_BY_EF4 control, lower ZCUR and ZXU by 1 cm,
increase RCUR by 2 mm, and raise pre-fill from 1.8 to 2.0e-5 torr.
Also turn on methane puff.

Shot 14 -- Plasma, disrupted at 0.42 s. VDE. No hard x-rays.

Next shot: raise ZCUR by 5 mm. Methane puff enabled.


Shot 15 -- Plasma, disrupted at 0.41 s. TF and PF crowbarred. Vinnie thinks
the OH2L had a fault which caused the disruption.

Next shot: repeat with no changes.

There's a break in the run while a fuse is replaced in the TF supply.

Shot 16 -- Fizzle.

Next shot: repeat with no changes.


Shot 17 -- Plasma, good shot. Upper x-pt. No methane. Brian reports that
the fast scanning probe will be able to reach this plasma.

Next shot: enable NINJA methane puff.


Shot 18 -- Plasma, disrupted at 0.99 s, VDE. Good plume data. The plume
reversed direction. Good FSP data.

Next shot: another NINJA methane puff, with different filter.


Shot 19 -- Plasma, good shot. Good plume data. Good FSP data.

Next shot: raise density to nl_04=1.2e20 m-2, with methane puff.


Shot 20 -- Plasma, disrupted at 0.40, which is before I made any change in
the density programming. VDE.

Next shot: repeat with no changes.

Engineering pulled out a slew-rate limiting circuit board from OH1 between
shots.

There was a delay due to a problem with the STATE machine.


Shot 21 -- Plasma, good shot. Good plume data. Density made it to 1.2e20 m-2.

Next shot: repeat with different filter.

Engineering re-inserted the slew-rate limiting circuit board into OH1 between
shots.


Shot 22 -- Plasma, good shot. Good plume data, good probe data.

Next shot: repeat with no methane, to get a background shot.

Change TF invert time from 1.6 to 1.1 s in order to speed up rep rate.


Shot 23 -- Plasma, good shot. Good baseline shot.

Next shot: NINJA methane puff from bottom divertor (private flux
zone).


Shot 24 -- Plasma, runaway screamer, disrupted at 0.39 s.

Next shot: raise pre-fill from 2.0 to 2.2e-5 torr. Puff methane
from private flux zone.


Shot 25 -- Plasma, good shot.

Next shot: raise density to nl_04=1.6e20 m-2. Puff methane from
inboard midplane.


Shot 26 -- Plasma, good shot. Density made it to 1.6e20 m-2. Good plume data,
Good FSP data.

Next shot: repeat with NINJA methane puff and different filter.


Shot 27 -- Plasma, disrupted at 0.83 s. VDE. Probably due to molybdenum
injection. Got plume data.

Next shot: reduce density to nl_04=1.2e20 m-2 and puff from the
bridge of the divertor. Still upper x-pt configuration.


Shot 28 -- Fizzle.

Next shot: repeat with no changes.


Shot 29 -- Plasma, good shot. Good plume data and FSP data. Nl_04=1.2e20 m-2.
This is the last upper x-pt configuration in this series.

Next shot: change to limiter plasma configuration, with
nl_04=1.0e20 m-2, and NINJA methane puff from inboard midplane.


Shot 30 -- Plasma, limited, disrupted at 0.85 s, maybe due to more carbon
getting in. Plume was more circular than plum-ish.

Next shot: repeat with different filter.


Shot 31 -- Plasma, good shot. Good data. Carbon brightness is a factor of
4 to 5 higher than diverted plasmas.

Next shot: raise density to nl_04=1.5e20 m-2. Continue methane
puffing from inboard midplane.


Shot 32 -- Plasma, disrupted at about 0.45 s. Looks like it got very dirty
and resistive.

Next shot: reduce density back to nl_04=1.0e20 m-2, turn off
NINJA, but puff methane through B-side-lower. (B-side-lower plenum
pressure is 3.4 psi)


Shot 33 -- Plasma, good shot. Limited, with methane puff from B-side-lower
valve. This is the shot I need for screening comparison on non-
recycling impurities. nl_04=1.1e20 m-2.

Next shot: switch back to lower x-pt, with methane puff from
B-side-lower valve.


Shot 34 -- Plasma, disrupted at 0.23 s, prior to becoming diverted. Disruption
was caused by an OH2L fault.

Next shot: repeat with no changes.


Shot 35 -- Plasma, good shot. Diverted, lower x-pt, with methane puff from
B-side-lower valve. This is the other shot I need for screening
comparison of non-recycling impurity. nl_04=1.1e20 m-2.

Next shot: repeat with NINJA puffing of methane from the private
flux zone. B-side-valve is disabled.

Comparison of CII signal shows a factor of two difference between the limiter
plasma (033) and the diverted plasma (035). The CII signal is a measure of
carbon flux through the edge. Measurements of CVI brightness taken
by the McPherson show a factor of 2-3 difference.


Shot 36 -- Plasma, good shot. Lower x-pt, with methane puff from private
flux zone.

Next shot: repeat, but with no methane puffing, to get a baseline.
Also change TF invert time from 1.1 to 0.9 s so Amanda can get
a calibration.


Shot 37 -- Plasma, good shot. No methane puffing. Good baseline shot for
diverted, lower x-pt configuration. nl_04=1.1e20 m-2.

Next shot: switch to limiter configuration again, but without
any methane puffing, to get a baseline shot at nl_04=1.0e20 m-2.
TF invert time moved back to 1.1 s.


Shot 38 -- Plasma, good shot. No methane puffing. Good baseline shot for
limiter configuration. nl_04=1.1e20, with some wobbles. Brian
got some fast scanning probe data.

Using these two baseline shots (037 and 038) with the two methane shots (033
and 035), Jim Terry reports a factor of two difference in CVI brightness.

End of run.

Session Leader Comments
May 26 1995 08:43:49:090AMGary McCrackenStudy of plasma flows using gas injection plumes (MP 104)

The plan is to generally repeat the program of CH4 injection obtained
on run 950414 but this time with grad B drift down. The CH4 will be
injected principally from the NINJA system at the inner wall midplane
and observed with the CCD cameras using CII CII filters. We will
compare the X-point at the top and at the bottom. We will also look
at the effect of injecting into limiter discharges.
We plan to use these shots to measure carbon screening and to
use B-side lower injection as well as different NINJA positions

May 26 1995 05:33:11:530PMGary McCrackenBrief summary: Good day; We did a density scan with a lower x-point
plasma from nl04=0.8 to 1.8e20 with methane injection through
the inner wall capillary (B16-001)
CII and CI plumes were observed always going towards the divertor
target. This is different to the results from the run950414 with the
grad B drift upwards, where we got the flow reversing at high density
We then did upper X-point discharges.We again got a density scan and
the flow was always towards the upper divertor
We also did some puffing through the lower divertor capillaries
B26-019 and B22-022 with upper X-point. Interestingly we saw clear
toroidal flow viewing from the B-top camera. No apprecaible signal was
observed from the B26-019 cap when the X-point was down

May 26 1995 08:49:04:090AM950526001Gary McCrackenshot 001

Good shot. nlo40.8e20. Inside and outside gaps good.
Plasma position for FSP very good.

Plan to inject CH4 through NINJA valves B16-001 and B22-022 next shot

May 26 1995 09:05:31:040AM950526002Gary McCrackenshot 002
Inject CH4 through NINJA valves B16-001 and B22-022 CII filter
Plenum pressure 6.2 psi. Trigger at 450 ms for 300 ms
Good shot; CH4 injected and observed both on B-side camera and on
A-side (unfiltered) camera. It is clearly seen from both valves and
is quite bright.
Outer strike point at FMP probe 4

May 26 1995 09:18:30:810AM950526003Gary McCrackenshot 003
Inject CH4 through NINJA valves B16-001 only CIII filter
Plenum pressure 6.2 psi. Trigger at 450 ms for 300 ms
Good shot; CH4 injected and observed on B-side camera .
Intensity looks better than previously observed with the
CIII filter. The plasma is clearly being fuelled by the NINJA
system and the density is going up beyond the value demanded
nl04=1e20

May 26 1995 09:32:26:280AM950526004Gary McCrackenshot 004
Inject CH4 through NINJA valves B16-001 and B22-022 CII filter
Plenum pressure 6.2 psi. Trigger at 450 ms for 300 ms
Disruption at 0.6 s CH4 observed on B-side camera for 2 or 3
frames.
Asked for nlo4=1.3e20 reached 1.15e20 before disruption.
We will have a background shot without any CH4 next shot

May 26 1995 09:49:33:830AM950526005Gary McCrackenshot 005
NO CH4 injected. Good shot
Asked for nlo4=1.3e20 and got it.
We will go back to gas puffing next shot

May 26 1995 10:06:37:560AM950526006Gary McCrackenshot 006
CH4 injected.B16-001 only, conditions as before, CIII filter
Good shot. Clear plume. nl04=1.38e20 nice and constant
It is clear in both this shot and 005 that we are getting detachment
at 0.53 s, up to probe 5. shot 005 was without methane.
The strike point is on probe 3 to 4, moving up towards the end of
the shot

May 26 1995 10:18:57:580AM950526006Gary McCrackenshot 006
CORRECTION :USED CII FILTER THIS SHOT NOT CIII AS PREVIOUSLY STATED

May 26 1995 10:17:25:430AM950526007Gary McCrackenshot 007
CH4 injected.B16-001 only, conditions as before, CIII filter
Plume observed but much weaker. nl04=1.4e20

May 26 1995 10:32:17:050AM950526008Gary McCrackenshot 008
Increased demand density to 1.6e20.
No methane injected to obtain background at higher density.
Good shot; nl04 =1.64e20

May 26 1995 10:45:52:000AM950526009Gary McCrackenshot 009
CH4 injected at about 0.7 s (trigger at 0.55s for 0.3 s).
B16-001 valve only ;plenum pressure still 6.15 psi
Good shot; CIII plume clearly towards the divertor still,
though faint. No sign of reversing direction at high density
as seen on 950414 run
nl04 =1.69e20

May 26 1995 10:56:49:130AM950526010Gary McCrackenshot 010
CH4 injected at about 0.7 s (trigger at 0.55s for 0.3 s).
B16-001 valve only ;plenum pressure still 6.15 psi
CII filter (last shot CIII filter)
Good shot; CII plume still downwards ;bright but possibly
decaying faster along the field line. nl04=1.65e20

May 26 1995 11:15:39:060AM950526011Gary McCrackenshot 011
Density demand raised to 1.8e20
CH4 injected at about 0.7 s (trigger at 0.55s for 0.3 s).
B16-001 valve only ;plenum pressure 6.15 psi CII filter
Good shot until 0.85 s then disrupted CII plume still downwards,
looks fine nl04=1.76e20 at disruption, still rising

May 26 1995 11:53:53:090AM950526013Gary McCrackenshot 013
No methane injection; nl04 0.8e20
Disrupted 0.95 s

May 26 1995 11:49:35:760AM950526014Gary McCrackenshot 012
Change to upper X-point
Reduce density to 0.8e20, no methane puff.
Disrupted at 0.4 s; not clear why.

May 26 1995 11:59:20:620AM950526014Gary McCrackenshot 014
Methane injection B16-001 valve, timing and pressure as above
Tried to improve plasma elongation and position for FSP
Disrupted 0.45 s; nl04 =0.6e20 at disruption. Gas in after disruption
but bright spots at capillaries one frame before disruption. No
obvious explanation.

May 26 1995 12:35:07:880PM950526015Gary McCrackenshot 015
No methane injection
Raised plasma and reduced elongation
Early disruption 0.4 s.

May 26 1995 12:35:35:170PM950526015Gary McCrackenshot 016
No methane injection
Fizzle

May 26 1995 12:52:30:260PM950526016Gary McCrackenshot 016
No methane injection
Fizzle

May 26 1995 12:54:41:140PM950526017Gary McCrackenshot 017
No methane injection ;Background shot
Good plasma with upper X-point nl04 =0.84e20

May 26 1995 01:02:15:460PM950526018Gary McCrackenshot 018
Methane injection , CII filter; B16-001 only;
Trigger at 0.55 s for 0.3 s; Plenum pressure still 6.15 psi
Good plasma with upper X-point nl04 =0.84e20.
CII plume clearly visible with reversed direction, as expected.

May 26 1995 01:17:46:910PM950526019Gary McCrackenshot 019
Methane injection , CIII filter; B16-001 only;
Trigger at 0.55 s for 0.3 s; Plenum pressure still 6.15 psi
Good shot; Clear CIII plume towards upper divertor, although faint.
It is however more symmetric than the CII plume
Clear step up on CII signal on C-port at 0.68 s, on this and previous
pulse, compared with shot 17 which had no methane injected

May 26 1995 01:24:21:800PM950526020Gary McCrackenshot 020
Methane injection , CIII filter; B16-001 only;
Trigger at 0.55 s for 0.3 s; Plenum pressure 6.2 psi
Disrupted at 0.4 s

May 26 1995 01:36:38:060PM950526021Gary McCrackenshot 021
Try repeat of shot 20
Methane injection , CIII filter; B16-001 only;
Trigger at 0.55 s for 0.3 s; Plenum pressure 6.2 psi
Did not fire

May 26 1995 01:45:58:110PM950526021Gary McCrackenshot 021
Try again repeat of shot 20
Methane injection , CIII filter; B16-001 only;
Trigger at 0.55 s for 0.3 s; Plenum pressure 6.2 psi
Good shot; clear but faint CIII plume towards upper divertor
Note plume reverses around the time of current ramp down.

May 26 1995 01:54:14:800PM950526022Gary McCrackenshot 022
Repeat shot
Methane injection , CII filter; B16-001 only;
Trigger at 0.55 s for 0.3 s; Plenum pressure 6.2 psi
Good shot; clear CII plume towards upper divertor
Plume again reverses around the time of current ramp down.

May 26 1995 02:10:08:740PM950526023Gary McCrackenshot 023
Repeat shot
No methane injection, background shot
Good shot nl04= 1.22e20. Very nice flat CII signal from C-port

May 26 1995 02:22:29:490PM950526024Gary McCrackenshot 024
Methane injection, B26-019 at lower divertor;
Trigger at 0.55s for 0.3 s ; 6.1 psi
Disruption at 0.4 s Not clear why



t

May 26 1995 02:39:30:680PM950526025Gary McCrackenshot 025
Methane injection, B26-019 at lower divertor; CII filter
Trigger at 0.55s for 0.3 s ; 6.1 psi
Good shot nl04 1.29 ; CII plume; Faint, but toroidal flow observed on
the B-top camera looking at B26-019

May 26 1995 02:45:13:220PM950526026Gary McCrackenshot 026
Methane injection, B16-001 at inner wall; CII filter, higher density
Trigger at 0.55s for 0.3 s ; 6.1 psi
Good shot ; nl04 1.29 ; CII plume;

May 26 1995 02:53:13:260PM950526027Gary McCrackenshot 027
Methane injection, B16-001 at inner wall; CIII filter
Trigger at 0.55s for 0.3 s ; 6.1 psi

Disruption at 0.82 s,but got the CIII plume, faint but
clearly towards the upper divertor ; nl04 1.62

May 26 1995 02:55:23:880PM950526027Gary McCrackenshot 027
Methane injection, B16-001 at inner wall; CIII filter
Trigger at 0.55s for 0.3 s ; 6.1 psi

Disruption at 0.82 s,but got the CIII plume, faint but
clearly towards the upper divertor ; nl04=1.62

May 26 1995 03:03:33:970PM950526028Gary McCrackenshot 028
Methane injection, B22-022 at inner upper divertor with X-point at top
to compare with shot 25 ,(puff from B26-019); CII filter
Trigger at 0.55s for 0.3 s ; 6.1 psi
Fizzle

May 26 1995 03:20:46:040PM950526029Gary McCrackenshot 029
Methane injection, B22-022 at inner upper divertor with X-point at top
to compare with shot 25 ,(puff from B26-019); CII filter
Trigger at 0.55s for 0.3 s ; 6.1 psi
Very clear flow along the field observed in the B-top camera
Good FSP data obtained. This should allow us to compare the flow
derived from the plume with that derived from the scanning probe.

May 26 1995 03:29:55:710PM950526030Gary McCrackenshot 030
Changed to limited plasma
Methane injection, B16-001 at inner wall ; CII filter
Trigger at 0.55s for 0.3 s ; 6.1 psi
Disrupted at 0.85 s, but can see the CII at the midplane.
Very localized symmetric spot, presumably because the
ionization is so fast when it enters in the core plasma

May 26 1995 03:43:34:840PM950526031Gary McCrackenshot 031
Limited plasma
Methane injection, B16-001 at inner wall ; CIII filter
Trigger at 0.55s for 0.3 s ; 6.1 psi
Good shot ; CIII bright light spreads uniformly in both directions

May 26 1995 03:52:32:370PM950526032Gary McCrackenshot 032
Limited plasma
Methane injection, B16-001 at inner wall ; CIII filter
Trigger at 0.55s for 0.3 s ; 6.1 psi
Repeat shot except higher density
Disrupted at 0.45 s due to plasma being too resistive. Why?

May 26 1995 04:09:15:280PM950526033Gary McCrackenshot 033
Limited plasma
Methane injection, through B-side lower; 0.6 to 0.95 s;
Plenum pressure 3.4 psi. CIII filter
Good shot . Clear signal on CII at C-port

May 26 1995 04:10:13:950PM950526033Gary McCrackenshot 033
Limited plasma
Methane injection, through B-side lower; 0.6 to 0.95 s;
Plenum pressure 3.4 psi. CIII filter
Good shot . Clear signal on CII at C-port

May 26 1995 04:14:46:450PM950526034Gary McCrackenshot 034
Lower X-point to compare with previous limited plasma
Methane injection, through B-side lower; 0.6 to 0.95 s;
Plenum pressure 3.4 psi. CIII filter
Died early; Power supply problem on lower OH2

May 26 1995 04:27:22:960PM950526035Gary McCrackenshot 035
Lower X-point to compare with previous limited plasma
Methane injection, through B-side lower; 0.6 to 0.95 s;
Plenum pressure 3.4 psi. CIII filter
Good shot clear signal on Cii at C-port. the delta CIi is about
half of the value in the limiter shot

May 26 1995 04:45:04:930PM950526036Gary McCrackenshot 036
Lower X-point
Methane injection, through B26-019; Trigger at 0.55 s for 0.3 s;
Plenum pressure 6.0 psi. CII filters
Good shot ; nl04 =1.1e20
Very faint signal on the B-top camera but not certain it comes
from the gas puffing at the PFZ

May 26 1995 04:51:21:970PM950526037Gary McCrackenshot 037
Lower X-point
No Methane injection, Backgroundshot for spectroscopy
Good shot; nl04=1.08e20. Nice low flat CII background signal

May 26 1995 05:12:23:790PM950526038Gary McCrackenshot 038
Limiter shot; Another background shot for spectroscopy
No Methane injection,
Good shot; nl04=1.15e20.

Physics Operator Comments
May 25 1995 06:27:03:800PMRobert GranetzMP #104, Investigation of Plasma Flows Using Gas Injection Plumes

Primary purpose: Inject CH4 and He at the inner wall and divertor with
NINJA system and study plumes in plasmas with B x grad B
drift down. Compare to data from 950414, when the B x grad B
drift was reversed. We will run lower x-pt's, upper x-pt's,
and limited plasmas, at several densities and several puffing
locations.

Secondary purpose: Compare limiter/divertor screening of methane, which is a
non-recycling impurity

May 25 1995 06:27:33:940PMRobert GranetzSetup: Lower x-pt: 950512027, with density changed to nl_04=0.8e20 m-2,
Upper x-pt: 950322029, with density changed to nl_04=0.8e20 m-2,
Limited : 950519021, with density changed to nl_04=1.0e20 m-2.

To speed up PCS operations, I setup a PCS the evening before this
run with segments 2,3, & 4 loaded with the three configurations,
respectively.

Engineering preparations:

Setup up from shot 950512027 (including commutation resistors,
TF start time, TF current, etc).
Normal ECDC in D2.
Set pre-fill at 1.8e-5 torr of D2.
Fill A-side with D2 at 19-20 psi.
Fill B-side-lower with methane at 3 psi, BUT LEAVE DISABLED!
Fill the NINJA plenum with methane at 5 psi.
All other gas valves should be disabled as well.

May 26 1995 08:56:56:830AM950526001Robert GranetzShot 01 -- Setup for a lower x-pt baseline shot. No methane puffing.
Plasma, good shot. Density oscillated a bit.

Next shot: repeat with NINJA system puffing methane from the
inboard midplane location and from the inner divertor nose location.

May 26 1995 09:05:09:530AM950526002Robert GranetzShot 02 -- Plasma, good shot. Methane plumes were seen. On request from
Brian, we will tweak upper and lower x-pt positions on the next
shot.

Next shot: tweak upper and lower x-pt locations, and puff methane
from midplane only.

May 26 1995 09:20:49:140AM950526003Robert GranetzShot 03 -- Plasma, good shot. Good plume (CIII).

Next shot: tweak upper x-pt another 5 mm up. NINJA puff from both
locations enabled.

May 26 1995 09:34:35:090AM950526004Robert GranetzShot 04 -- Plasma, disrupted at 0.59 s. VDE. Probably not due to excessive
radiation. Elongation was very high.

Next shot: lower ZCUR by 5 mm, and turn off methane in order to
get a background shot.

May 26 1995 09:50:45:090AM950526005Robert GranetzShot 05 -- Plasma, good shot. Density got to the requested value of 1.3e20 m-2.
Zxray did move down.

Next shot: NINJA methane puff from midplane.

May 26 1995 10:03:08:730AM950526006Robert GranetzShot 06 -- Plasma, good shot. Good plume data (CIII). Jim Terry reports
that carbon background level is gradually rising from shot to shot.

Next shot: NINJA methane puff from midplane, different filter.

May 26 1995 10:03:28:250AM950526006Robert GranetzEngineering has enabled the I^2*t circuit starting with the next shot.

May 26 1995 10:14:06:780AM950526007Robert GranetzShot 07 -- Plasma, good shot. Good plume data (CIII).

Next shot: increase density to nl_04=1.6e20 m-2. Turn off methane.

May 26 1995 10:30:10:470AM950526008Robert GranetzShot 08 -- Plasma, good shot. Density got to nl_04=1.6e20 m-2 by 0.75 s.

Next shot: enable NINJA methane puff.

May 26 1995 10:46:05:040AM950526009Robert GranetzShot 09 -- Plasma, good shot. Got plume data (CIII).

Next shot: NINJA methane puff with different filter.

May 26 1995 10:51:51:760AM950526010Robert GranetzShot 10 -- Plasma, good shot. Good plume data (CII).

Next shot: raise density to nl_04=1.8e20 m-2. Also puff methane.

May 26 1995 11:08:27:120AM950526011Robert GranetzShot 11 -- Plasma, disrupted at 0.86 s. Got good plume data anyway.

Next shot: change to upper x-pt at nl_04=0.8e20 m-2. No methane
puffing.

May 26 1995 11:27:19:930AM950526012Robert GranetzShot 12 -- Plasma, disrupted at 0.41 s. VDE.

Next shot: slow down the current rise with a little tweak.

May 26 1995 11:44:41:820AM950526013Robert GranetzShot 13 -- Plasma, disrupted at 0.96 s. VDE. Some hard x-rays. Brian wants
the plasma lowered by 1 cm so the scanning probe can reach, but
we're near the vertical stability limit.

Next shot: Turn on EF2_BY_EF4 control, lower ZCUR and ZXU by 1 cm,
increase RCUR by 2 mm, and raise pre-fill from 1.8 to 2.0e-5 torr.
Also turn on methane puff.

May 26 1995 11:58:42:710AM950526014Robert GranetzShot 14 -- Plasma, disrupted at 0.42 s. VDE. No hard x-rays.

Next shot: raise ZCUR by 5 mm. Methane puff enabled.

May 26 1995 12:10:40:300PM950526015Robert GranetzShot 15 -- Plasma, disrupted at 0.41 s. TF and PF crowbarred. Vinnie thinks
the OH2L had a fault which caused the disruption.

Next shot: repeat with no changes.

May 26 1995 12:47:34:390PM950526016Robert GranetzShot 16 -- Fizzle.

Next shot: repeat with no changes.

May 26 1995 12:48:29:040PM950526017Robert GranetzShot 17 -- Plasma, good shot. Upper x-pt. No methane.

Next shot: enable NINJA methane puff.

May 26 1995 01:01:16:750PM950526018Robert GranetzShot 18 -- Plasma, disrupted at 0.99 s, VDE. Good plume data. The plume
reversed direction.

Next shot: another NINJA methane puff, with different filter.

May 26 1995 01:09:54:210PM950526019Robert GranetzShot 19 -- Plasma, good shot. Good plume data.

Next shot: raise density to nl_04=1.2e20 m-2, with methane puff.

May 26 1995 01:29:59:980PM950526020Robert GranetzShot 20 -- Plasma, disrupted at 0.40, which is before I made any change in
the density programming. VDE.

Next shot: repeat with no changes.

May 26 1995 01:45:27:580PM950526021Robert GranetzShot 21 -- Plasma, good shot. Good plume data. Density made it to 1.2e20 m-2.

Next shot: repeat with different filter.

May 26 1995 01:55:08:560PM950526022Robert GranetzShot 22 -- Plasma, good shot. Good plume data, good probe data.

Next shot: repeat with no methane, to get a background shot.

May 26 1995 02:07:41:240PM950526023Robert GranetzShot 23 -- Plasma, good shot. Good baseline shot.

Next shot: NINJA methane puff from bottom divertor.

May 26 1995 02:17:31:580PM950526024Robert GranetzShot 24 -- Runaway screamer.

Next shot: raise pre-fill from 2.0 to 2.2e-5 torr.

May 26 1995 02:29:00:250PM950526025Robert GranetzShot 25 -- Plasma, good shot.

Next shot: raise density to nl_04=1.6e20 m-2.

May 26 1995 02:42:49:610PM950526026Robert GranetzShot 26 -- Plasma, good shot. Density made it to 1.6e20 m-2. Good plume data,
Good FSP data.

Next shot: repeat with NINJA methane puff and different filter.

May 26 1995 02:56:52:010PM950526027Robert GranetzShot 27 -- Plasma, disrupted at 0.83 s. VDE. Probably due to molybdenum
injection. Got plume data.

Next shot: reduce density to nl_04=1.2e20 m-2 and puff from the
bridge of the divertor. Still upper x-pt configuration.

May 26 1995 03:04:17:250PM950526028Robert GranetzShot 28 -- Fizzle.

Next shot: repeat with no changes.

May 26 1995 03:16:05:360PM950526029Robert GranetzShot 29 -- Plasma, good shot. Good plume data and FSP data. Nl_04=1.2e20 m-2.
This is the last upper x-pt configuration in this series.

Next shot: change to limiter plasma configuration, with
nl_04=1.0e20 m-2.

May 26 1995 03:33:14:850PM950526030Robert GranetzShot 30 -- Plasma, limited, disrupted at 0.85 s, maybe due to more carbon
getting in.

Next shot: repeat with different filter.

May 26 1995 03:42:17:140PM950526031Robert GranetzShot 31 -- Plasma, good shot. Good data. Carbon brightness is a factor of
4 to 5 higher than diverted plasmas.

Next shot: raise density to nl_04=1.5e20 m-2. Continue methane
puffing from inboard midplane.

May 26 1995 03:54:45:940PM950526032Robert GranetzShot 32 -- Plasma, disrupted at about 0.45 s. Looks like it got very dirty
and resistive.

Next shot: reduce density back to nl_04=1.0e20 m-2, turn off
NINJA, but puff methane through B-side-lower. (B-side-lower plenum
pressure is 3.4 psi)

May 26 1995 04:04:18:490PM950526033Robert GranetzShot 33 -- Plasma, good shot. Limited, with methane puff from B-side-lower
valve. This is the shot I need for screening comparison on non-
recycling impurities. nl_04=1.1e20 m-2.

Next shot: switch back to lower x-pt, with methane puff from
B-side-lower valve.

May 26 1995 04:09:03:050PM950526034Robert GranetzShot 34 -- Plasma, disrupted at 0.23 s, prior to becoming diverted. Disruption
was caused by an OH2L fault.

Next shot: repeat with no changes.

May 26 1995 04:31:05:100PM950526035Robert GranetzShot 35 -- Plasma, good shot. Diverted, lower x-pt, with methane puff from
B-side-lower valve. This is the other shot I need for screening
comparison of non-recycling impurity. nl_04=1.1e20 m-2.

Next shot: repeat with NINJA puffing of methane from the private
flux zone. B-side-valve is disabled.

May 26 1995 05:27:30:320PM950526035Robert GranetzComparison of CII signal shows a factor of two difference between the limiter
plasma (033) and the diverted plasma (035). The CII signal is a measure of
carbon flux through the edge. Measurements of CVI brightness taken
by the McPherson show a factor of 2-3 difference.

May 26 1995 04:35:39:050PM950526036Robert GranetzShot 36 -- Plasma, good shot. Lower x-pt, with methane puff from private
flux zone.

Next shot: repeat, but with no methane puffing, to get a baseline.

May 26 1995 04:45:21:180PM950526037Robert GranetzShot 37 -- Plasma, good shot. No methane puffing. Good baseline shot for
diverted, lower x-pt. nl_04=1.1e20 m-2.

Next shot: switch to limiter configuration again, but without
any methane puffing, to get a baseline shot at nl_04=1.0e20 m-2.

May 26 1995 05:15:22:300PM950526038Robert GranetzShot 38 -- Plasma, good shot. No methane puffing. Good baseline shot for
limiter configuration. nl_04=1.1e20, with some wobbles. Brian
got some fast scanning probe data.

May 26 1995 05:15:38:210PM950526038Robert GranetzUsing these two baseline shots (037 and 038) with the two methane shots (033
and 035), Jim Terry reports a factor of two difference in CVI brightness.

Engineering Operator Comments
ShotTimeTypeStatusComment
108:35:16:400AMPlasmaOk
208:51:58:830AMPlasmaOk
309:07:25:090AMPlasmaOk
409:23:33:930AMPlasmaOk
509:37:10:750AMPlasmaOk
609:51:30:660AMPlasmaOk
710:06:35:220AMPlasmaOk
810:20:33:300AMPlasmaOk
910:34:35:380AMPlasmaOk
1010:48:32:930AMPlasmaOk
1111:01:28:480AMPlasmaOk
1211:13:17:840AMPlasmaOk
1311:29:22:270AMPlasmaOk
1411:45:00:600AMPlasmaOk
1512:00:48:140PMPlasmaBadoh2l comm fault
1612:31:41:170PMPlasmaOk
1712:41:37:860PMPlasmaOk
1812:53:39:100PMPlasmaOkComm fault OH2U OH1
1901:05:59:020PMPlasmaOk
2001:17:58:500PMPlasmaOk
2101:30:35:000PMPlasmaOk
2201:49:07:340PMPlasmaOk
2302:03:18:430PMPlasmaOk
2402:14:40:420PMPlasmaOk
2502:24:27:020PMPlasmaOk
2602:35:21:150PMPlasmaOk
2702:46:53:380PMPlasmaOk
2803:00:11:640PMPlasmaOk
2903:09:47:930PMPlasmaOk
3003:20:38:320PMPlasmaOk
3103:32:18:640PMPlasmaOk
3203:42:31:610PMPlasmaOk
3303:54:52:900PMPlasmaOk
3404:04:34:530PMPlasmaBadComm fault OH2L caused disruption.
3504:16:29:620PMPlasmaOk
3604:27:51:450PMPlasmaOk
3704:40:34:020PMPlasmaOk
3804:51:52:330PMPlasmaOk