| Miniproposals | ||||||||||
|
| Operators | |
| Session leader(s): | Gary McCracken |
| Physics operator(s): | Robert Granetz |
| Engineering operator(s): | Joe Daigle,Frank Silva |
| Engineering Operator Run Comment |
| Granetz |
| Session Leader Plans |
| Physics Operators Plans |
| Session Leader Summaries |
| Entered: Jul 7 2004 03:49:13:803PM |
| Author: To Be Determined |
| Session Leader Run Summary 9505026 Engineering operator: F Silva/J Daigle Physics operator: R Granetz Session Leader: G M McCracken Run for mini-proposals #104 Objectives: Inject impurities at the inner wall and at the divertor with the NINJA system and follow the spatial spreading with the CCD cameras and filters from B-side and B-top. Compare with limiter plasma. Measure impurity screening in parallel A very successful day. We started off with a lower x-point discharge and puffed CH4 from the inner wall capillary (B16-001) over a range of densities from nl04= 0.9 to 1.8e20 m^-2. CII and CIII were observed with the B-side CCD camera in alternate discharges: in both cases the flow was clearly seen going towards the divertor. This is in contrast to the results obtained on 950414 when the flow reversed direction at a density of 1.5e20. The flow direction generally reverses at the start of current rampdown as the plasma moves back to being limited rather than diverted. Detachment was observed on shots with densities nl04=1.3e20 or higher. The detachment occurred before the CH4 was injected. An upper X-point discharge was then set up with some difficulty, there being a series of disruptions. The reasons for this were not very clear, but it may have been due to the need for conditioning the upper divertor surfaces. Data was eventually taken for densities from 0.9e20 up to 1.7e20. With the upper X-point we also injected CH4 from capillaries B26-019 (lower private flux region) and B22-022 (lower inner cone above the inner nose), shots 025 and 029. In both cases clear evidence of toroidal flow was observed in positions which, with the upper X-point , should be close to the stagnation point. Good FSP data were obtained for these discharges, which should allow us to compare the flow estimated from the plumes with that estimated from the ratio of the parallel and antiparallel Isat on the FSP Next we studied the injection of CH4 from B16-001 into inner-wall, limited discharges. As was to be expected the impurities were ionised rapidly and the CII distribution was a small (10-20 mm) diameter symmetric blob. The CIII appeared to spread all round the inner column but was also roughly symmetric. CH4 was injected from the the B-side lower valve for the limiter and lower x-point discharges to compare with the NINJA injections. Many of the shots will be useful for screening estimates since in general only one valve was used for each condition. Comparisons will be possible for different injection conditions, different densities and different plasma configurations, limiter, upper x-point and lower x-point. In general good data were obtained on the midplane CII monitor at C-port and a consistent set of data were measured with the McPherson (CVI) and Chromex spectrometers (CII and CIII). The signals on the spectrometerrs increased very little when the CH4 was injected, but good background shots were taken for most cases so that at least upper limits to the screening efficiency should be obtained. Good FSP data obtained for most conditions, including Limiter discharges. Shot summary: CH4 plenum pressure 6.2 psi (330 torr) throughout. NINJA valves triggered at 0.45 s for shots #02 to 07 and at 0.55 s therafter. Pulse length 0.3 s. When the B-side lower valve was used it was triggered from 0.6 to 0.95 s. the plenum pressure in this case was 3.4psi. The filter stated below refers only to the B-side camera. The B-top camera was only recorded for a limited number of shots, eg 025,029,036, when puffing at the bottom. The CII filter was used. Lower X-point 01. Good background shot, No CH4, nl04=0.8e20. 02. Good shot, CH4 injected at B16-001 and B22-022, CII filter, nl04=1e20. 03. Good shot, CH4 injected at B16-001 only, CIII filter, nl04=1e20. 04. Disruption at 0.6 s. 05. Good background shot, No CH4, nl04=01.3e20. 06. Good shot, CH4 injected at B16-001, CII filter, nl04=1.3e20. 07. Good shot, CH4 injected at B16-001, CIII filter, nl04=1.3e20. 08. Good background shot, No CH4, nl04=1.64e20. 09. Good shot, CH4 injected at B16-001, CIII filter, nl04=1.64e20. 10. Good shot, CH4 injected at B16-001, CII filter, nl04=1.64e20 11. Good shot, CH4 injected at B16-001, CII filter, nl04=1.76e20: disrupted at 0.85 s. 35. Good shot, CH4 injected at B-side lower, CIII filter, nl04=1.10e20. 36. Good shot, CH4 injected at B26-019, CII filter, nl04=1.24e20. Toroidal flow just barely detectable. Compare with shot 025 with upper X-point. 37. Good background shot, No CH4, nl04=1.15e20. Upper X-point 12. Disrupted at 0.4 s. 13. Disrupted at 0.96 s, No CH4. 14. Disrupted at 0.4 s. 15. Disrupted at 0.4 s. 16. Fizzle. 17. Good background shot, No CH4, nl04=0.83e20. 18. Good shot, CH4 injected at B16-001, CII filter, nl04=0.87e20. 19. Good shot, CH4 injected at B16-001, CIII filter, nl04=0.88e20. 20. Disrupted at 0.4 s. 21. Good shot, CH4 injected at B16-001, CIII filter, nl04=1.28e20. 22. Good shot, CH4 injected at B16-001, CII filter, nl04=1.3e20. 23. Good background shot, No CH4, nl04=1.21e20. 24. Disrupted at 0.4 s. 25. Good shot, CH4 injected at B26-019, CII filter, nl04=1.24e20. Toroidal flow. 26. Good shot, CH4 injected at B16-001, CII filter, nl04=1.7e20. 27. Good shot, CH4 injected at B16-001, CII filter, nl04=1.66e20. 28. Fizzle. 29. Good shot, CH4 injected at B22-022, CII filter, nl04=1.27e20. Toroidal flow. Limiter shots 30. Good shot, CH4 injected at B16-001, CII filter, nl04=1.20e20: disrupted at 0.85 s 31. Good shot, CH4 injected at B16-001, CIII filter, nl04=1.21e20. 32. Disrupted at 0.55 s. 33. Good shot, CH4 injected at B-side lower, CIII filter, nl04=1.17e20. 34. Died early. 38. Good background shot, No CH4, nl04=1.15e20. |
| Physics Operator Summaries |
| Entered: Jul 7 2004 04:36:54:537PM |
| Author: To Be Determined |
| Physics operator's summary for run 950526. MP #104, Investigation of Plasma Flows Using Gas Injection Plumes SL: G. McCracken PO: R. Granetz Primary purpose: Inject CH4 and He at the inner wall and divertor with NINJA system and study plumes in plasmas with B x grad B drift down. Compare to data from 950414, when the B x grad B drift was reversed. We will run lower x-pt's, upper x-pt's, and limited plasmas, at several densities and several puffing locations. Secondary purpose: Compare limiter/divertor screening of methane, which is a non-recycling impurity. Summary: Today's run plan was very challenging to say the least. Fortunately things went very well, and we were able to get good methane plume data in lower x-pt, upper x-pt, and limited plasmas at a range of densities. In addition, we got good data for comparing the carbon screening efficiency of limited and lower x-pt plasmas. The session leader's report presents some of the plume physics results. I have looked at the carbon screening comparison and find that there is a factor of 2 difference. (Thanks to J.T. and Gary for showing me the CVI and CII data.) The shots of relevance for the screening study are 033 (limited w/methane), 035 (lower x-pt w/methane), 037 (lower x-pt, no methane), and 038 (limited, no methane). We did not have time to do any helium studies today. Also see shot 031 for an example of a limited plasma with methane puffing directly into the boundary using the NINJA inboard midplane location. It has much higher carbon signals than the shots that had methane puffed from the B-side-lower valve. 001-011 Diverted, lower x-pt 012-029 Diverted, upper x-pt 030-033 Limited 034-037 Diverted, lower x-pt 038 Limited Statistics: 36 Plasmas 2 Fizzles 0 Duds Power supply problems caused disruptions on 2 shots. The I^2*t circuit was enabled today. Setup: Lower x-pt: 950512027, with density changed to nl_04=0.8e20 m-2, Upper x-pt: 950322029, with density changed to nl_04=0.8e20 m-2, Limited : 950519021, with density changed to nl_04=1.0e20 m-2. To speed up PCS operations, I setup a PCS the evening before this run with segments 2,3, & 4 loaded with the three configurations, respectively. Engineering preparations: Setup up from shot 950512027 (including commutation resistors, TF start time, TF current, etc). Normal ECDC in D2. Set pre-fill at 1.8e-5 torr of D2. Fill A-side with D2 at 19-20 psi. Fill B-side-lower with methane at 3 psi, BUT LEAVE DISABLED! Fill the NINJA plenum with methane at 5 psi. All other gas valves should be disabled as well. Shot 01 -- Setup for a lower x-pt baseline shot. No methane puffing. Plasma, good shot. Density oscillated a bit. Next shot: repeat with NINJA system puffing methane from the inboard midplane location and from the inner divertor nose location. Shot 02 -- Plasma, good shot. Methane plumes were seen (CII). On request from Brian, we will tweak upper and lower x-pt positions on the next shot. Fast scanning probe fired on this shot. Next shot: tweak upper and lower x-pt locations, and puff methane from midplane only. Shot 03 -- Plasma, good shot. Good plume (CIII). Next shot: tweak upper x-pt another 5 mm up. NINJA puff from both locations enabled. Shot 04 -- Plasma, disrupted at 0.59 s. VDE. Probably not due to excessive radiation. Elongation was very high. Next shot: lower ZCUR by 5 mm, and turn off methane in order to get a background shot. Shot 05 -- Plasma, good shot. Density got to the requested value of 1.3e20 m-2. Zxray did move down. Next shot: NINJA methane puff from midplane. Shot 06 -- Plasma, good shot. Good plume data (CII). Jim Terry reports that carbon background level is gradually rising from shot to shot. Next shot: NINJA methane puff from midplane, different filter. Engineering has enabled the I^2*t circuit starting with the next shot. Shot 07 -- Plasma, good shot. Good plume data (CIII). Next shot: increase density to nl_04=1.6e20 m-2. Turn off methane. Shot 08 -- Plasma, good shot. Density got to nl_04=1.6e20 m-2 by 0.75 s. Next shot: enable NINJA methane puff. Shot 09 -- Plasma, good shot. Got plume data (CIII). Next shot: NINJA methane puff with different filter. Shot 10 -- Plasma, good shot. Good plume data (CII). Next shot: raise density to nl_04=1.8e20 m-2. Also puff methane. Shot 11 -- Plasma, disrupted at 0.86 s. Got good plume data anyway. Next shot: change to upper x-pt at nl_04=0.8e20 m-2. No methane puffing. Switching to upper x-pt configuration starting with the next shot. Shot 12 -- Plasma, disrupted at 0.41 s. VDE. Upper x-pt. Next shot: slow down the current rise with a little tweak. Shot 13 -- Plasma, disrupted at 0.96 s. VDE. Some hard x-rays. Brian wants the plasma lowered by 1 cm so the scanning probe can reach, but we're near the vertical stability limit. Next shot: Turn on EF2_BY_EF4 control, lower ZCUR and ZXU by 1 cm, increase RCUR by 2 mm, and raise pre-fill from 1.8 to 2.0e-5 torr. Also turn on methane puff. Shot 14 -- Plasma, disrupted at 0.42 s. VDE. No hard x-rays. Next shot: raise ZCUR by 5 mm. Methane puff enabled. Shot 15 -- Plasma, disrupted at 0.41 s. TF and PF crowbarred. Vinnie thinks the OH2L had a fault which caused the disruption. Next shot: repeat with no changes. There's a break in the run while a fuse is replaced in the TF supply. Shot 16 -- Fizzle. Next shot: repeat with no changes. Shot 17 -- Plasma, good shot. Upper x-pt. No methane. Brian reports that the fast scanning probe will be able to reach this plasma. Next shot: enable NINJA methane puff. Shot 18 -- Plasma, disrupted at 0.99 s, VDE. Good plume data. The plume reversed direction. Good FSP data. Next shot: another NINJA methane puff, with different filter. Shot 19 -- Plasma, good shot. Good plume data. Good FSP data. Next shot: raise density to nl_04=1.2e20 m-2, with methane puff. Shot 20 -- Plasma, disrupted at 0.40, which is before I made any change in the density programming. VDE. Next shot: repeat with no changes. Engineering pulled out a slew-rate limiting circuit board from OH1 between shots. There was a delay due to a problem with the STATE machine. Shot 21 -- Plasma, good shot. Good plume data. Density made it to 1.2e20 m-2. Next shot: repeat with different filter. Engineering re-inserted the slew-rate limiting circuit board into OH1 between shots. Shot 22 -- Plasma, good shot. Good plume data, good probe data. Next shot: repeat with no methane, to get a background shot. Change TF invert time from 1.6 to 1.1 s in order to speed up rep rate. Shot 23 -- Plasma, good shot. Good baseline shot. Next shot: NINJA methane puff from bottom divertor (private flux zone). Shot 24 -- Plasma, runaway screamer, disrupted at 0.39 s. Next shot: raise pre-fill from 2.0 to 2.2e-5 torr. Puff methane from private flux zone. Shot 25 -- Plasma, good shot. Next shot: raise density to nl_04=1.6e20 m-2. Puff methane from inboard midplane. Shot 26 -- Plasma, good shot. Density made it to 1.6e20 m-2. Good plume data, Good FSP data. Next shot: repeat with NINJA methane puff and different filter. Shot 27 -- Plasma, disrupted at 0.83 s. VDE. Probably due to molybdenum injection. Got plume data. Next shot: reduce density to nl_04=1.2e20 m-2 and puff from the bridge of the divertor. Still upper x-pt configuration. Shot 28 -- Fizzle. Next shot: repeat with no changes. Shot 29 -- Plasma, good shot. Good plume data and FSP data. Nl_04=1.2e20 m-2. This is the last upper x-pt configuration in this series. Next shot: change to limiter plasma configuration, with nl_04=1.0e20 m-2, and NINJA methane puff from inboard midplane. Shot 30 -- Plasma, limited, disrupted at 0.85 s, maybe due to more carbon getting in. Plume was more circular than plum-ish. Next shot: repeat with different filter. Shot 31 -- Plasma, good shot. Good data. Carbon brightness is a factor of 4 to 5 higher than diverted plasmas. Next shot: raise density to nl_04=1.5e20 m-2. Continue methane puffing from inboard midplane. Shot 32 -- Plasma, disrupted at about 0.45 s. Looks like it got very dirty and resistive. Next shot: reduce density back to nl_04=1.0e20 m-2, turn off NINJA, but puff methane through B-side-lower. (B-side-lower plenum pressure is 3.4 psi) Shot 33 -- Plasma, good shot. Limited, with methane puff from B-side-lower valve. This is the shot I need for screening comparison on non- recycling impurities. nl_04=1.1e20 m-2. Next shot: switch back to lower x-pt, with methane puff from B-side-lower valve. Shot 34 -- Plasma, disrupted at 0.23 s, prior to becoming diverted. Disruption was caused by an OH2L fault. Next shot: repeat with no changes. Shot 35 -- Plasma, good shot. Diverted, lower x-pt, with methane puff from B-side-lower valve. This is the other shot I need for screening comparison of non-recycling impurity. nl_04=1.1e20 m-2. Next shot: repeat with NINJA puffing of methane from the private flux zone. B-side-valve is disabled. Comparison of CII signal shows a factor of two difference between the limiter plasma (033) and the diverted plasma (035). The CII signal is a measure of carbon flux through the edge. Measurements of CVI brightness taken by the McPherson show a factor of 2-3 difference. Shot 36 -- Plasma, good shot. Lower x-pt, with methane puff from private flux zone. Next shot: repeat, but with no methane puffing, to get a baseline. Also change TF invert time from 1.1 to 0.9 s so Amanda can get a calibration. Shot 37 -- Plasma, good shot. No methane puffing. Good baseline shot for diverted, lower x-pt configuration. nl_04=1.1e20 m-2. Next shot: switch to limiter configuration again, but without any methane puffing, to get a baseline shot at nl_04=1.0e20 m-2. TF invert time moved back to 1.1 s. Shot 38 -- Plasma, good shot. No methane puffing. Good baseline shot for limiter configuration. nl_04=1.1e20, with some wobbles. Brian got some fast scanning probe data. Using these two baseline shots (037 and 038) with the two methane shots (033 and 035), Jim Terry reports a factor of two difference in CVI brightness. End of run. |
| Session Leader Comments | |||
| May 26 1995 08:43:49:090AM | Gary McCracken | Study of plasma flows using gas injection plumes (MP 104)
The plan is to generally repeat the program of CH4 injection obtained on run 950414 but this time with grad B drift down. The CH4 will be injected principally from the NINJA system at the inner wall midplane and observed with the CCD cameras using CII CII filters. We will compare the X-point at the top and at the bottom. We will also look at the effect of injecting into limiter discharges. We plan to use these shots to measure carbon screening and to use B-side lower injection as well as different NINJA positions | |
| May 26 1995 05:33:11:530PM | Gary McCracken | Brief summary: Good day; We did a density scan with a lower x-point
plasma from nl04=0.8 to 1.8e20 with methane injection through the inner wall capillary (B16-001) CII and CI plumes were observed always going towards the divertor target. This is different to the results from the run950414 with the grad B drift upwards, where we got the flow reversing at high density We then did upper X-point discharges.We again got a density scan and the flow was always towards the upper divertor We also did some puffing through the lower divertor capillaries B26-019 and B22-022 with upper X-point. Interestingly we saw clear toroidal flow viewing from the B-top camera. No apprecaible signal was observed from the B26-019 cap when the X-point was down | |
| May 26 1995 08:49:04:090AM | 950526001 | Gary McCracken | shot 001
Good shot. nlo40.8e20. Inside and outside gaps good. Plasma position for FSP very good. Plan to inject CH4 through NINJA valves B16-001 and B22-022 next shot |
| May 26 1995 09:05:31:040AM | 950526002 | Gary McCracken | shot 002
Inject CH4 through NINJA valves B16-001 and B22-022 CII filter Plenum pressure 6.2 psi. Trigger at 450 ms for 300 ms Good shot; CH4 injected and observed both on B-side camera and on A-side (unfiltered) camera. It is clearly seen from both valves and is quite bright. Outer strike point at FMP probe 4 |
| May 26 1995 09:18:30:810AM | 950526003 | Gary McCracken | shot 003
Inject CH4 through NINJA valves B16-001 only CIII filter Plenum pressure 6.2 psi. Trigger at 450 ms for 300 ms Good shot; CH4 injected and observed on B-side camera . Intensity looks better than previously observed with the CIII filter. The plasma is clearly being fuelled by the NINJA system and the density is going up beyond the value demanded nl04=1e20 |
| May 26 1995 09:32:26:280AM | 950526004 | Gary McCracken | shot 004
Inject CH4 through NINJA valves B16-001 and B22-022 CII filter Plenum pressure 6.2 psi. Trigger at 450 ms for 300 ms Disruption at 0.6 s CH4 observed on B-side camera for 2 or 3 frames. Asked for nlo4=1.3e20 reached 1.15e20 before disruption. We will have a background shot without any CH4 next shot |
| May 26 1995 09:49:33:830AM | 950526005 | Gary McCracken | shot 005
NO CH4 injected. Good shot Asked for nlo4=1.3e20 and got it. We will go back to gas puffing next shot |
| May 26 1995 10:06:37:560AM | 950526006 | Gary McCracken | shot 006
CH4 injected.B16-001 only, conditions as before, CIII filter Good shot. Clear plume. nl04=1.38e20 nice and constant It is clear in both this shot and 005 that we are getting detachment at 0.53 s, up to probe 5. shot 005 was without methane. The strike point is on probe 3 to 4, moving up towards the end of the shot |
| May 26 1995 10:18:57:580AM | 950526006 | Gary McCracken | shot 006
CORRECTION :USED CII FILTER THIS SHOT NOT CIII AS PREVIOUSLY STATED |
| May 26 1995 10:17:25:430AM | 950526007 | Gary McCracken | shot 007
CH4 injected.B16-001 only, conditions as before, CIII filter Plume observed but much weaker. nl04=1.4e20 |
| May 26 1995 10:32:17:050AM | 950526008 | Gary McCracken | shot 008
Increased demand density to 1.6e20. No methane injected to obtain background at higher density. Good shot; nl04 =1.64e20 |
| May 26 1995 10:45:52:000AM | 950526009 | Gary McCracken | shot 009
CH4 injected at about 0.7 s (trigger at 0.55s for 0.3 s). B16-001 valve only ;plenum pressure still 6.15 psi Good shot; CIII plume clearly towards the divertor still, though faint. No sign of reversing direction at high density as seen on 950414 run nl04 =1.69e20 |
| May 26 1995 10:56:49:130AM | 950526010 | Gary McCracken | shot 010
CH4 injected at about 0.7 s (trigger at 0.55s for 0.3 s). B16-001 valve only ;plenum pressure still 6.15 psi CII filter (last shot CIII filter) Good shot; CII plume still downwards ;bright but possibly decaying faster along the field line. nl04=1.65e20 |
| May 26 1995 11:15:39:060AM | 950526011 | Gary McCracken | shot 011
Density demand raised to 1.8e20 CH4 injected at about 0.7 s (trigger at 0.55s for 0.3 s). B16-001 valve only ;plenum pressure 6.15 psi CII filter Good shot until 0.85 s then disrupted CII plume still downwards, looks fine nl04=1.76e20 at disruption, still rising |
| May 26 1995 11:53:53:090AM | 950526013 | Gary McCracken | shot 013
No methane injection; nl04 0.8e20 Disrupted 0.95 s |
| May 26 1995 11:49:35:760AM | 950526014 | Gary McCracken | shot 012
Change to upper X-point Reduce density to 0.8e20, no methane puff. Disrupted at 0.4 s; not clear why. |
| May 26 1995 11:59:20:620AM | 950526014 | Gary McCracken | shot 014
Methane injection B16-001 valve, timing and pressure as above Tried to improve plasma elongation and position for FSP Disrupted 0.45 s; nl04 =0.6e20 at disruption. Gas in after disruption but bright spots at capillaries one frame before disruption. No obvious explanation. |
| May 26 1995 12:35:07:880PM | 950526015 | Gary McCracken | shot 015
No methane injection Raised plasma and reduced elongation Early disruption 0.4 s. |
| May 26 1995 12:35:35:170PM | 950526015 | Gary McCracken | shot 016
No methane injection Fizzle |
| May 26 1995 12:52:30:260PM | 950526016 | Gary McCracken | shot 016
No methane injection Fizzle |
| May 26 1995 12:54:41:140PM | 950526017 | Gary McCracken | shot 017
No methane injection ;Background shot Good plasma with upper X-point nl04 =0.84e20 |
| May 26 1995 01:02:15:460PM | 950526018 | Gary McCracken | shot 018
Methane injection , CII filter; B16-001 only; Trigger at 0.55 s for 0.3 s; Plenum pressure still 6.15 psi Good plasma with upper X-point nl04 =0.84e20. CII plume clearly visible with reversed direction, as expected. |
| May 26 1995 01:17:46:910PM | 950526019 | Gary McCracken | shot 019
Methane injection , CIII filter; B16-001 only; Trigger at 0.55 s for 0.3 s; Plenum pressure still 6.15 psi Good shot; Clear CIII plume towards upper divertor, although faint. It is however more symmetric than the CII plume Clear step up on CII signal on C-port at 0.68 s, on this and previous pulse, compared with shot 17 which had no methane injected |
| May 26 1995 01:24:21:800PM | 950526020 | Gary McCracken | shot 020
Methane injection , CIII filter; B16-001 only; Trigger at 0.55 s for 0.3 s; Plenum pressure 6.2 psi Disrupted at 0.4 s |
| May 26 1995 01:36:38:060PM | 950526021 | Gary McCracken | shot 021
Try repeat of shot 20 Methane injection , CIII filter; B16-001 only; Trigger at 0.55 s for 0.3 s; Plenum pressure 6.2 psi Did not fire |
| May 26 1995 01:45:58:110PM | 950526021 | Gary McCracken | shot 021
Try again repeat of shot 20 Methane injection , CIII filter; B16-001 only; Trigger at 0.55 s for 0.3 s; Plenum pressure 6.2 psi Good shot; clear but faint CIII plume towards upper divertor Note plume reverses around the time of current ramp down. |
| May 26 1995 01:54:14:800PM | 950526022 | Gary McCracken | shot 022
Repeat shot Methane injection , CII filter; B16-001 only; Trigger at 0.55 s for 0.3 s; Plenum pressure 6.2 psi Good shot; clear CII plume towards upper divertor Plume again reverses around the time of current ramp down. |
| May 26 1995 02:10:08:740PM | 950526023 | Gary McCracken | shot 023
Repeat shot No methane injection, background shot Good shot nl04= 1.22e20. Very nice flat CII signal from C-port |
| May 26 1995 02:22:29:490PM | 950526024 | Gary McCracken | shot 024
Methane injection, B26-019 at lower divertor; Trigger at 0.55s for 0.3 s ; 6.1 psi Disruption at 0.4 s Not clear why t |
| May 26 1995 02:39:30:680PM | 950526025 | Gary McCracken | shot 025
Methane injection, B26-019 at lower divertor; CII filter Trigger at 0.55s for 0.3 s ; 6.1 psi Good shot nl04 1.29 ; CII plume; Faint, but toroidal flow observed on the B-top camera looking at B26-019 |
| May 26 1995 02:45:13:220PM | 950526026 | Gary McCracken | shot 026
Methane injection, B16-001 at inner wall; CII filter, higher density Trigger at 0.55s for 0.3 s ; 6.1 psi Good shot ; nl04 1.29 ; CII plume; |
| May 26 1995 02:53:13:260PM | 950526027 | Gary McCracken | shot 027
Methane injection, B16-001 at inner wall; CIII filter Trigger at 0.55s for 0.3 s ; 6.1 psi Disruption at 0.82 s,but got the CIII plume, faint but clearly towards the upper divertor ; nl04 1.62 |
| May 26 1995 02:55:23:880PM | 950526027 | Gary McCracken | shot 027
Methane injection, B16-001 at inner wall; CIII filter Trigger at 0.55s for 0.3 s ; 6.1 psi Disruption at 0.82 s,but got the CIII plume, faint but clearly towards the upper divertor ; nl04=1.62 |
| May 26 1995 03:03:33:970PM | 950526028 | Gary McCracken | shot 028
Methane injection, B22-022 at inner upper divertor with X-point at top to compare with shot 25 ,(puff from B26-019); CII filter Trigger at 0.55s for 0.3 s ; 6.1 psi Fizzle |
| May 26 1995 03:20:46:040PM | 950526029 | Gary McCracken | shot 029
Methane injection, B22-022 at inner upper divertor with X-point at top to compare with shot 25 ,(puff from B26-019); CII filter Trigger at 0.55s for 0.3 s ; 6.1 psi Very clear flow along the field observed in the B-top camera Good FSP data obtained. This should allow us to compare the flow derived from the plume with that derived from the scanning probe. |
| May 26 1995 03:29:55:710PM | 950526030 | Gary McCracken | shot 030
Changed to limited plasma Methane injection, B16-001 at inner wall ; CII filter Trigger at 0.55s for 0.3 s ; 6.1 psi Disrupted at 0.85 s, but can see the CII at the midplane. Very localized symmetric spot, presumably because the ionization is so fast when it enters in the core plasma |
| May 26 1995 03:43:34:840PM | 950526031 | Gary McCracken | shot 031
Limited plasma Methane injection, B16-001 at inner wall ; CIII filter Trigger at 0.55s for 0.3 s ; 6.1 psi Good shot ; CIII bright light spreads uniformly in both directions |
| May 26 1995 03:52:32:370PM | 950526032 | Gary McCracken | shot 032
Limited plasma Methane injection, B16-001 at inner wall ; CIII filter Trigger at 0.55s for 0.3 s ; 6.1 psi Repeat shot except higher density Disrupted at 0.45 s due to plasma being too resistive. Why? |
| May 26 1995 04:09:15:280PM | 950526033 | Gary McCracken | shot 033
Limited plasma Methane injection, through B-side lower; 0.6 to 0.95 s; Plenum pressure 3.4 psi. CIII filter Good shot . Clear signal on CII at C-port |
| May 26 1995 04:10:13:950PM | 950526033 | Gary McCracken | shot 033
Limited plasma Methane injection, through B-side lower; 0.6 to 0.95 s; Plenum pressure 3.4 psi. CIII filter Good shot . Clear signal on CII at C-port |
| May 26 1995 04:14:46:450PM | 950526034 | Gary McCracken | shot 034
Lower X-point to compare with previous limited plasma Methane injection, through B-side lower; 0.6 to 0.95 s; Plenum pressure 3.4 psi. CIII filter Died early; Power supply problem on lower OH2 |
| May 26 1995 04:27:22:960PM | 950526035 | Gary McCracken | shot 035
Lower X-point to compare with previous limited plasma Methane injection, through B-side lower; 0.6 to 0.95 s; Plenum pressure 3.4 psi. CIII filter Good shot clear signal on Cii at C-port. the delta CIi is about half of the value in the limiter shot |
| May 26 1995 04:45:04:930PM | 950526036 | Gary McCracken | shot 036
Lower X-point Methane injection, through B26-019; Trigger at 0.55 s for 0.3 s; Plenum pressure 6.0 psi. CII filters Good shot ; nl04 =1.1e20 Very faint signal on the B-top camera but not certain it comes from the gas puffing at the PFZ |
| May 26 1995 04:51:21:970PM | 950526037 | Gary McCracken | shot 037
Lower X-point No Methane injection, Backgroundshot for spectroscopy Good shot; nl04=1.08e20. Nice low flat CII background signal |
| May 26 1995 05:12:23:790PM | 950526038 | Gary McCracken | shot 038
Limiter shot; Another background shot for spectroscopy No Methane injection, Good shot; nl04=1.15e20. |
| Physics Operator Comments | |||
| May 25 1995 06:27:03:800PM | Robert Granetz | MP #104, Investigation of Plasma Flows Using Gas Injection Plumes
Primary purpose: Inject CH4 and He at the inner wall and divertor with NINJA system and study plumes in plasmas with B x grad B drift down. Compare to data from 950414, when the B x grad B drift was reversed. We will run lower x-pt's, upper x-pt's, and limited plasmas, at several densities and several puffing locations. Secondary purpose: Compare limiter/divertor screening of methane, which is a non-recycling impurity | |
| May 25 1995 06:27:33:940PM | Robert Granetz | Setup: Lower x-pt: 950512027, with density changed to nl_04=0.8e20 m-2,
Upper x-pt: 950322029, with density changed to nl_04=0.8e20 m-2, Limited : 950519021, with density changed to nl_04=1.0e20 m-2. To speed up PCS operations, I setup a PCS the evening before this run with segments 2,3, & 4 loaded with the three configurations, respectively. Engineering preparations: Setup up from shot 950512027 (including commutation resistors, TF start time, TF current, etc). Normal ECDC in D2. Set pre-fill at 1.8e-5 torr of D2. Fill A-side with D2 at 19-20 psi. Fill B-side-lower with methane at 3 psi, BUT LEAVE DISABLED! Fill the NINJA plenum with methane at 5 psi. All other gas valves should be disabled as well. | |
| May 26 1995 08:56:56:830AM | 950526001 | Robert Granetz | Shot 01 -- Setup for a lower x-pt baseline shot. No methane puffing.
Plasma, good shot. Density oscillated a bit. Next shot: repeat with NINJA system puffing methane from the inboard midplane location and from the inner divertor nose location. |
| May 26 1995 09:05:09:530AM | 950526002 | Robert Granetz | Shot 02 -- Plasma, good shot. Methane plumes were seen. On request from
Brian, we will tweak upper and lower x-pt positions on the next shot. Next shot: tweak upper and lower x-pt locations, and puff methane from midplane only. |
| May 26 1995 09:20:49:140AM | 950526003 | Robert Granetz | Shot 03 -- Plasma, good shot. Good plume (CIII).
Next shot: tweak upper x-pt another 5 mm up. NINJA puff from both locations enabled. |
| May 26 1995 09:34:35:090AM | 950526004 | Robert Granetz | Shot 04 -- Plasma, disrupted at 0.59 s. VDE. Probably not due to excessive
radiation. Elongation was very high. Next shot: lower ZCUR by 5 mm, and turn off methane in order to get a background shot. |
| May 26 1995 09:50:45:090AM | 950526005 | Robert Granetz | Shot 05 -- Plasma, good shot. Density got to the requested value of 1.3e20 m-2.
Zxray did move down. Next shot: NINJA methane puff from midplane. |
| May 26 1995 10:03:08:730AM | 950526006 | Robert Granetz | Shot 06 -- Plasma, good shot. Good plume data (CIII). Jim Terry reports
that carbon background level is gradually rising from shot to shot. Next shot: NINJA methane puff from midplane, different filter. |
| May 26 1995 10:03:28:250AM | 950526006 | Robert Granetz | Engineering has enabled the I^2*t circuit starting with the next shot.
|
| May 26 1995 10:14:06:780AM | 950526007 | Robert Granetz | Shot 07 -- Plasma, good shot. Good plume data (CIII).
Next shot: increase density to nl_04=1.6e20 m-2. Turn off methane. |
| May 26 1995 10:30:10:470AM | 950526008 | Robert Granetz | Shot 08 -- Plasma, good shot. Density got to nl_04=1.6e20 m-2 by 0.75 s.
Next shot: enable NINJA methane puff. |
| May 26 1995 10:46:05:040AM | 950526009 | Robert Granetz | Shot 09 -- Plasma, good shot. Got plume data (CIII).
Next shot: NINJA methane puff with different filter. |
| May 26 1995 10:51:51:760AM | 950526010 | Robert Granetz | Shot 10 -- Plasma, good shot. Good plume data (CII).
Next shot: raise density to nl_04=1.8e20 m-2. Also puff methane. |
| May 26 1995 11:08:27:120AM | 950526011 | Robert Granetz | Shot 11 -- Plasma, disrupted at 0.86 s. Got good plume data anyway.
Next shot: change to upper x-pt at nl_04=0.8e20 m-2. No methane puffing. |
| May 26 1995 11:27:19:930AM | 950526012 | Robert Granetz | Shot 12 -- Plasma, disrupted at 0.41 s. VDE.
Next shot: slow down the current rise with a little tweak. |
| May 26 1995 11:44:41:820AM | 950526013 | Robert Granetz | Shot 13 -- Plasma, disrupted at 0.96 s. VDE. Some hard x-rays. Brian wants
the plasma lowered by 1 cm so the scanning probe can reach, but we're near the vertical stability limit. Next shot: Turn on EF2_BY_EF4 control, lower ZCUR and ZXU by 1 cm, increase RCUR by 2 mm, and raise pre-fill from 1.8 to 2.0e-5 torr. Also turn on methane puff. |
| May 26 1995 11:58:42:710AM | 950526014 | Robert Granetz | Shot 14 -- Plasma, disrupted at 0.42 s. VDE. No hard x-rays.
Next shot: raise ZCUR by 5 mm. Methane puff enabled. |
| May 26 1995 12:10:40:300PM | 950526015 | Robert Granetz | Shot 15 -- Plasma, disrupted at 0.41 s. TF and PF crowbarred. Vinnie thinks
the OH2L had a fault which caused the disruption. Next shot: repeat with no changes. |
| May 26 1995 12:47:34:390PM | 950526016 | Robert Granetz | Shot 16 -- Fizzle.
Next shot: repeat with no changes. |
| May 26 1995 12:48:29:040PM | 950526017 | Robert Granetz | Shot 17 -- Plasma, good shot. Upper x-pt. No methane.
Next shot: enable NINJA methane puff. |
| May 26 1995 01:01:16:750PM | 950526018 | Robert Granetz | Shot 18 -- Plasma, disrupted at 0.99 s, VDE. Good plume data. The plume
reversed direction. Next shot: another NINJA methane puff, with different filter. |
| May 26 1995 01:09:54:210PM | 950526019 | Robert Granetz | Shot 19 -- Plasma, good shot. Good plume data.
Next shot: raise density to nl_04=1.2e20 m-2, with methane puff. |
| May 26 1995 01:29:59:980PM | 950526020 | Robert Granetz | Shot 20 -- Plasma, disrupted at 0.40, which is before I made any change in
the density programming. VDE. Next shot: repeat with no changes. |
| May 26 1995 01:45:27:580PM | 950526021 | Robert Granetz | Shot 21 -- Plasma, good shot. Good plume data. Density made it to 1.2e20 m-2.
Next shot: repeat with different filter. |
| May 26 1995 01:55:08:560PM | 950526022 | Robert Granetz | Shot 22 -- Plasma, good shot. Good plume data, good probe data.
Next shot: repeat with no methane, to get a background shot. |
| May 26 1995 02:07:41:240PM | 950526023 | Robert Granetz | Shot 23 -- Plasma, good shot. Good baseline shot.
Next shot: NINJA methane puff from bottom divertor. |
| May 26 1995 02:17:31:580PM | 950526024 | Robert Granetz | Shot 24 -- Runaway screamer.
Next shot: raise pre-fill from 2.0 to 2.2e-5 torr. |
| May 26 1995 02:29:00:250PM | 950526025 | Robert Granetz | Shot 25 -- Plasma, good shot.
Next shot: raise density to nl_04=1.6e20 m-2. |
| May 26 1995 02:42:49:610PM | 950526026 | Robert Granetz | Shot 26 -- Plasma, good shot. Density made it to 1.6e20 m-2. Good plume data,
Good FSP data. Next shot: repeat with NINJA methane puff and different filter. |
| May 26 1995 02:56:52:010PM | 950526027 | Robert Granetz | Shot 27 -- Plasma, disrupted at 0.83 s. VDE. Probably due to molybdenum
injection. Got plume data. Next shot: reduce density to nl_04=1.2e20 m-2 and puff from the bridge of the divertor. Still upper x-pt configuration. |
| May 26 1995 03:04:17:250PM | 950526028 | Robert Granetz | Shot 28 -- Fizzle.
Next shot: repeat with no changes. |
| May 26 1995 03:16:05:360PM | 950526029 | Robert Granetz | Shot 29 -- Plasma, good shot. Good plume data and FSP data. Nl_04=1.2e20 m-2.
This is the last upper x-pt configuration in this series. Next shot: change to limiter plasma configuration, with nl_04=1.0e20 m-2. |
| May 26 1995 03:33:14:850PM | 950526030 | Robert Granetz | Shot 30 -- Plasma, limited, disrupted at 0.85 s, maybe due to more carbon
getting in. Next shot: repeat with different filter. |
| May 26 1995 03:42:17:140PM | 950526031 | Robert Granetz | Shot 31 -- Plasma, good shot. Good data. Carbon brightness is a factor of
4 to 5 higher than diverted plasmas. Next shot: raise density to nl_04=1.5e20 m-2. Continue methane puffing from inboard midplane. |
| May 26 1995 03:54:45:940PM | 950526032 | Robert Granetz | Shot 32 -- Plasma, disrupted at about 0.45 s. Looks like it got very dirty
and resistive. Next shot: reduce density back to nl_04=1.0e20 m-2, turn off NINJA, but puff methane through B-side-lower. (B-side-lower plenum pressure is 3.4 psi) |
| May 26 1995 04:04:18:490PM | 950526033 | Robert Granetz | Shot 33 -- Plasma, good shot. Limited, with methane puff from B-side-lower
valve. This is the shot I need for screening comparison on non- recycling impurities. nl_04=1.1e20 m-2. Next shot: switch back to lower x-pt, with methane puff from B-side-lower valve. |
| May 26 1995 04:09:03:050PM | 950526034 | Robert Granetz | Shot 34 -- Plasma, disrupted at 0.23 s, prior to becoming diverted. Disruption
was caused by an OH2L fault. Next shot: repeat with no changes. |
| May 26 1995 04:31:05:100PM | 950526035 | Robert Granetz | Shot 35 -- Plasma, good shot. Diverted, lower x-pt, with methane puff from
B-side-lower valve. This is the other shot I need for screening comparison of non-recycling impurity. nl_04=1.1e20 m-2. Next shot: repeat with NINJA puffing of methane from the private flux zone. B-side-valve is disabled. |
| May 26 1995 05:27:30:320PM | 950526035 | Robert Granetz | Comparison of CII signal shows a factor of two difference between the limiter
plasma (033) and the diverted plasma (035). The CII signal is a measure of carbon flux through the edge. Measurements of CVI brightness taken by the McPherson show a factor of 2-3 difference. |
| May 26 1995 04:35:39:050PM | 950526036 | Robert Granetz | Shot 36 -- Plasma, good shot. Lower x-pt, with methane puff from private
flux zone. Next shot: repeat, but with no methane puffing, to get a baseline. |
| May 26 1995 04:45:21:180PM | 950526037 | Robert Granetz | Shot 37 -- Plasma, good shot. No methane puffing. Good baseline shot for
diverted, lower x-pt. nl_04=1.1e20 m-2. Next shot: switch to limiter configuration again, but without any methane puffing, to get a baseline shot at nl_04=1.0e20 m-2. |
| May 26 1995 05:15:22:300PM | 950526038 | Robert Granetz | Shot 38 -- Plasma, good shot. No methane puffing. Good baseline shot for
limiter configuration. nl_04=1.1e20, with some wobbles. Brian got some fast scanning probe data. |
| May 26 1995 05:15:38:210PM | 950526038 | Robert Granetz | Using these two baseline shots (037 and 038) with the two methane shots (033
and 035), Jim Terry reports a factor of two difference in CVI brightness. |
| Engineering Operator Comments | ||||
| Shot | Time | Type | Status | Comment |
| 1 | 08:35:16:400AM | Plasma | Ok | |
| 2 | 08:51:58:830AM | Plasma | Ok | |
| 3 | 09:07:25:090AM | Plasma | Ok | |
| 4 | 09:23:33:930AM | Plasma | Ok | |
| 5 | 09:37:10:750AM | Plasma | Ok | |
| 6 | 09:51:30:660AM | Plasma | Ok | |
| 7 | 10:06:35:220AM | Plasma | Ok | |
| 8 | 10:20:33:300AM | Plasma | Ok | |
| 9 | 10:34:35:380AM | Plasma | Ok | |
| 10 | 10:48:32:930AM | Plasma | Ok | |
| 11 | 11:01:28:480AM | Plasma | Ok | |
| 12 | 11:13:17:840AM | Plasma | Ok | |
| 13 | 11:29:22:270AM | Plasma | Ok | |
| 14 | 11:45:00:600AM | Plasma | Ok | |
| 15 | 12:00:48:140PM | Plasma | Bad | oh2l comm fault |
| 16 | 12:31:41:170PM | Plasma | Ok | |
| 17 | 12:41:37:860PM | Plasma | Ok | |
| 18 | 12:53:39:100PM | Plasma | Ok | Comm fault OH2U OH1 |
| 19 | 01:05:59:020PM | Plasma | Ok | |
| 20 | 01:17:58:500PM | Plasma | Ok | |
| 21 | 01:30:35:000PM | Plasma | Ok | |
| 22 | 01:49:07:340PM | Plasma | Ok | |
| 23 | 02:03:18:430PM | Plasma | Ok | |
| 24 | 02:14:40:420PM | Plasma | Ok | |
| 25 | 02:24:27:020PM | Plasma | Ok | |
| 26 | 02:35:21:150PM | Plasma | Ok | |
| 27 | 02:46:53:380PM | Plasma | Ok | |
| 28 | 03:00:11:640PM | Plasma | Ok | |
| 29 | 03:09:47:930PM | Plasma | Ok | |
| 30 | 03:20:38:320PM | Plasma | Ok | |
| 31 | 03:32:18:640PM | Plasma | Ok | |
| 32 | 03:42:31:610PM | Plasma | Ok | |
| 33 | 03:54:52:900PM | Plasma | Ok | |
| 34 | 04:04:34:530PM | Plasma | Bad | Comm fault OH2L caused disruption. |
| 35 | 04:16:29:620PM | Plasma | Ok | |
| 36 | 04:27:51:450PM | Plasma | Ok | |
| 37 | 04:40:34:020PM | Plasma | Ok | |
| 38 | 04:51:52:330PM | Plasma | Ok | |