Alcator C-Mod Run 980106 Information

Back to runs

Miniproposals
Miniproposal:
Date Filed:
Title:
First Author:
Session Leader: Unknown (shots 1-15)
Miniproposal:196a
Date Filed:12/16/1997
Title:High SOL Density Shots for the Calibration
First Author:Yijun Lin
Session Leader:Yijun Lin (shots 29-34)
Miniproposal:197
Date Filed:11/5/1997
Title:Helium Probe Comparison with Fast Scanning Probe
First Author:Spencer Pitcher
Session Leader:Spencer Pitcher (shots 13-28)

Operators
Session leader(s): Unknown,Spencer Pitcher,Yijun Lin
Physics operator(s):Robert Granetz
Engineering operator(s): Unknown

Engineering Operator Run Comment
^z

Session Leader Plans

Physics Operators Plans

Session Leader Summaries
Entered: Jul 7 2004 03:49:17:727PM
Author: To Be Determined
MP196A High SOL Density Shots for the Calibration of Reflectometer Profiles
------ --------------------------------------------------------------------

The point of this MP was to drive the density up at the edge so high we
could feel confident that all the reflectometer channels were reflecting
from nearly the same location -- the limiter.
The run sequence plan was completed very successfully:

1) produce a linear ramp in density with as high a density
as possible ( up to 4 X 10^20 m-3) with standard outer
gap -- shots 29-30

2) repeat but with 0.5 cm gap -- shot 31

3) scan outer gap during shot from approx 2 cm to < .5 cm -- shot 32

4) repeat -- shot 33

5) try for a plasma limited on the outer limiter -- shot 34

The most obvious effects were seen in step (5) where all the channels
responded to some degree as the plasma moved out onto the limiter. This
shot was also the only one during which the 110 GHz channel really seemed
to be well outside and within range of the other channels.

A problem with the fluctuation measurements was discovered during this
run. A 150 kHz oscillation is present on the fast channels that has
nothing to do with the plasma. We are trying to fix this problem now.

Entered: Jul 7 2004 03:49:17:740PM
Author: To Be Determined
Session Leader's summary of run 980106

MP 197 Helium Beam Check-Out

SL: CSP (Pitcher)
PO: Granetz

-----------------------------------------------------------
Basic Plan:

Today we will be comparing the helium beam with the fast-scanning
probe (FSP). The helium beam has only sparse spatial coverage,
ie delta_r ~ 7 mm, and thus to obtain high resolution profiles we will
be 'scanning the plasma' by changing the outer gap slowly
over 700 ms approx. The FSP will scan at three times,
first to ensure the SOL is not evolving during the gap scan,
and second, to compare with the helium beam results. The
experiment will be performed in Ohmic discharges first
and then in RF discharges, a few densities will be used.

We also have a piggy-back experiment....Jim Weaver will be
trying to make helium II and Halpha drift measurements
in the divertor. Contrasting these will shed light
on the question of plasma pressure loss in the divertor
due to plasma-neutral friction.

In the first part of the day we will be running several
fiducials to re-establish operation after the holidays.

-----------------------------------------------------------


shot 02 - Lower divertor capillary plenum filled to 7 psi of helium
(Bob is wrong.)

shot 07 - density rises at .35 s, also fside ratiomatic,
Halpha looks like entering dithering Ohmic H-mode.
Loop voltage, ie Ohmic power high 1.6 MW.
Dithering Ohmic H-mode??? maybe?

*****************NOW START MP197*****************************

shot 12 - next shot - first helium puff thru AB limiter

shot 13 - helium puff thru AB limiter
delta_t = 0.1 s starting at 0.15 s @ 7 psia
- FSP moving in silo only
- Weaver on HeII, Chromex on helium red lines

* Weaver computer glitch ... no data
* Chromex satruated on 600 line grating

shot 14 - helium puff increased to
dt = 0.2 @ 0.15 s
- chromex changed to 1200 line grating

* plasma disrupts at 0.35 s

shot 15 - ninja plenum changed to 3 psia helium
t = .15 s, dt = 0.2
- chromex on 600 line grating

* fizzle

shot 16 - ninja t = 0.15 s, dt = .1 s
- chromex NOT saturated, although Halpha interference?

shot 17 - FSP to go on this shot
- tweak chromex to reduce Halpha interference
* good shot, nl4 = 1.0e20, FSP and chromex data both good

shot 18 - drop density, keep all else the same
* density not quite flat
* fsp and chromex good

shot 19 - repeat from last shot
* density behaviour no better....abandon

shot 20 - go back to nl4 = 1.0e20
- shift ninja gas to inner wall (B16-001) for background check
* good fsp and chromex data

shot 21 - go to nl4 = 1.5e20
- shift ninja gas to AB limiter again
- fsp stays out
* upper marfe seen on TV
* valves wide open but only got nl4 ~ 1.3e20
* chromex data good

shot 22 - repeat shot 21
- fsp will go
* good chromex and fsp data

shot 23 - repeat shot 22 + 1.5 MW ICRH from t = 0.4 to 1.2 s
- fsp off

* weird shot...starts off ELM-free H-mode with typical
signatures. Following a single giant ELM we go into
a steady-state ELM-free LOW D-alpha H-mode? or is
it enhanced D-alpha H-mode?

shot 24 - repeat of shot 23
- fsp will go
- ninja gas shifted to inner wall for background check
* good fsp and chromex data
* again a new beastie....no large elm this time

shot 25 - repeat of shot 24
- fsp will go
- ninja moved to AB limiter
* good fsp and chromex data
* new beastie again
* Halpha rises towards end of shot, no large elm

shot 26 - attempt at lowering density
* very messy shot, series of elm-free periods

shot 27 - go back to shot 23, nl4 = 1.5e20, Ohmic,
ninja gas to inner wall for background check
* chromex and fsp ok

shot 28 - go back to shot 19, nl4 ~ 0.7e20, Ohmic,
ninja gas to inner wall for background check
* chromex and fsp ok
* LAST SHOT OF THIS MINI-PROPOSAL


-----------------------------------------------------------

WHAT WAS ACTUALLY ACCOMPLISHED:

Start-up was slow after the holiday break. (MP197 didn't
start until shot 13.) The machine had symptoms of being
dirty, and this might have been exacerbated by the presence
of argon. By shot 13, the machine was running better
and we starting mp 197. By this point the argon
didn't seem to affect things. Shots 14,15 were bad, but all
the rest, ie 16 thru 28 were good discharges.

We compared the helium beam and FSP at low, medium and high density
in Ohmic discharges. For each of these densities, a 'background
discharge' was taken, where the AB limiter gas puff was
turned off and the same quantity of helium was
gas puffed at the inner wall. This provided a background
helium recycling level, which is possibly important for some
of the helium beam spatial channels. Both the chromex
data from the helium beam and the FSP data look very
good, with nice robust, noise-free signals. The detailed
comparison between the two remains to be done.

Unfortunately, the piggy-back experiment, ie He II drift
measurements in the divertor, could not be properly
carried out, as the helium level required by Jim Weaver was far
too high for the chromex spectrometer, ie it saturated
the chromex. Instead, Jim focussed on Halpha measurements
and in a few cases B II.

In shots 23 - 26 we injected 1.7 MW ICRH and found what
appeared to be a new type of H-mode (albeit with poor
confinement, ie H_iter89 ~ 1.3 to 1.5). These plasmas (eg shot 24)
achieved steady-state conditions, but were significantly
different than standard EDA H-modes. For example, comparing
shot 24 with 971216031 (-- a reasonable EDA shot),
the neutral density in the divertor and
at the midplane were similar, but the H-alpha was radically
different, ~ 0.7 in shot 24 and ~ 5 in EDA, ie shot 24 was more like
an elm-free H-mode, BUT STEADY-STATE! Further, the radiated
power and Mo impurity level were much higher in shot 24,
ie 60% and 3e15, respectively, compared with 30% and 0.5e15
in EDA. It appears that these new discharges accumulate
Mo impurities only up to a point, where an elevated equilibrium
level is reached. Further study is probably warranted.

Spencer Pitcher





















































Physics Operator Summaries
Entered: Jul 7 2004 04:36:58:380PM
Author: To Be Determined
Physics operator summary for run 980106.

MP 197 -- Helium Probe Edge Measurements & Comparison with Scanning Probe
MP 196A -- Reflectometer Calibration

SL: Pitcher for MP 197, Lin for MP 196A
PO: Granetz

Summary: Despite a troublesome morning, we ended up with a very successful run,
accomplishing two separate mini-proposals.

Physop things learned today:

(1) Reducing the slow ZCUR P and D gains from 1.5 to 1.0 after
t=0.1 s, or even to 0.7 after t=0.25 s, greatly increases
vertical stability in the time range from 0.3 to 0.5 s. A
further (but smaller) improvement is achieved by reducing the Ip
gain by 20%, although the current feedback control is mildly
worsened as well. Thanks goes to Ian for working on this problem.

(2) High densities (80% of density limit, and maybe even higher) can
be achieved with gas puffing by filling both the J-bottom and
K-side plena with 0.2 MPa (30 psi) of D2 and programming both
valves (PG 3 and 5) fully open (in addition to nl_04
programming).

Shots 2-15 were spent cleaning up the machine after the two week
holiday break, and improving early vertical stability (see first
comment above).

Shots 13-28 were devoted to MP 197. These were very successful shots,
with big ramps in the outer gap, at several densities, with and
without RF. Weird H-modes were seen on shots 23-26. They had
some characteristics of ELM-free and some characteristics of EDA,
but with lousy H-factors.

Shots 29-34 were devoted to MP 196A. These were also very successful
shots, with density ramps up to nebar=4.3e20 m-3 (80% of
Greenwald density limit)!! Shot 34 just barely became limited on
the outboard limiters for 50 ms (on purpose).

** Warning **
Most of the shots today had large ramps in outer gap and/or
density, and therefore are not suitable as starting points for
future runs (at least not segment 2).

Scorecard: 29 plasmas (7 disrupted early)
4 fizzles
0 duds
1 test
----------
34 total

------------------------------------------------------------------------------

Run Plan and Engineering Setup for Tuesday, 980106

MP 197 -- Helium Probe Edge Measurements & Comparison with Scanning Probe
MP 196A -- Reflectometer Calibration

Standard ECDC in deuterium for a couple of hours before run.

Engineering setup as on 971218006, with TF under hybrid control
This will be 0.8 MA, 5.3 T, nl_04=1e20 m-2, flattop to 1.2 s.

Gas setup - fill K-side with 30 PSI of D2, hybrid enable
fill B-top with 6 PSI of D2, hybrid enable
fill B-side-lower with 1.5 PSI of Ar, hybrid enable

Ninja system setup:
Fill lower divertor capillary plenum to 5 PSI of helium, and enable the
valve.

The following gate valves should be enabled for the start of the run,
assuming no vacuum problems:

ECE (no others until we verify the status of each system)

------

In addition to the two MP's scheduled for today, it has also been requested
that we begin with a number of fiducial shots in order to evaluate wall
conditions. The setup shot specified above is both a standard 1997
fiducial, as well as the starting point for MP 197.

Although MP 197 will begin with ohmic plasmas, later in the morning we will
want to have very long 1 MW RF pulses, starting at t=0.4 s and ending at
t=1.2 s (RF pulse duration of 0.8 s). The FSP should be setup to scan at
t=0.60, 0.85, and 1.10 s on every shot of this MP. Helium will be puffed
from the A-B (outboard) limiter midplane during the shots, and the outer gap
will be scanned slowly from 15 to 5 mm during each shot. Several densities
will be run, with and without RF.

MP 196A will be run after MP 197. It requires similar plasma magnetic
configurations, but at high density (nebar=4e20 m-3), with no RF.

-------------------------------------------------------------------------------

Detailed shot log:

Note: BAROLO is dead (hardware problem). Steve W has re-assigned EFIT
to run on RHONE, but processes that depend on it will fail. I've
sent mail to Tom and Josh, requesting them to re-assign the logical
queue name MHDQ.

Note: 2 hours of ECDC in deuterium was done prior to today's run.

Hybrid coldstart done; no errors found
($set default cmod$models:[hybrid.tcl]
$@hybinit_dispatch
$@show_bitbus_log
Use editor search to verify no occurrences of "error" nor
"not respond")
Reload fiducial plasma shot 971218006 (0.8 MA, 5.3 T, 1.0e20 m-2)
RF on for the initial fiducial shots.
RF feedback on RCUR is turned off (RF_LOAD_E = 0.0)

Shot 01 -- Standard Tuesday OH test. 10 kA. Coax resistances are good and low.

Next shot: try for fiducial plasma.

Shot 02 -- Fizzle. Good null. CII level was at 0.01 W/cm2/st. Got the
pre-fill gas puff.

Next shot: re-try, no changes.

Shot 03 -- Fizzle. Good null. CII is lower, but still not down to normal
levels.

Next shot: gas-head tweak (increase pre-pulse puff from 15 to 18 ms)

Shot 04 -- Plasma. Disrupted at 0.43 s. Ip=0.8 MA. VDE. Loop voltage was
2.3 volts! I think it's just dirty.

Next shot: repeat, trying to cleanup.

Shot 05 -- Fizzle.

Next shot: reduce pre-pulse puff from 18 to 14 ms.

Shot 06 -- Plasma. Another early VDE disruption. Spencer has noticed that
the F-side ratiomatic pressure is 3 times higher than during
December, even though the divertor pressure is the same.

Next shot: no changes

Shot 07 -- Plasma. Another early VDE disruption. Density takes off at 0.34 s,
even though the feedback turns off the gas. Spencer is speculating
that the plasma is toggling between ohmic H and L modes.

Next shot: lower the density from nl04=1.0 to 0.8e20 m-2.

Shot 08 -- Plasma. Disrupted even earlier (0.34 s.). There was a large moly
brick at 0.26 s.

Next shot: repeat, no changes.

Shot 09 -- Plasma. Disrupted at 0.46 s. N/Ncrit is reaching 1.2, but the
elongation and gaps are not excessive.

Next shot: turn off argon

Shot 10 -- Plasma. Good shot. Full length. Density was high initially, but
got back down to nl04=0.8e20 m-2 after 0.4 s. H-alpha shows
continuous dithering for the entire shot. Only difference from the
previous shot was turning off the argon.

Next shot: reduce early gas valve voltage, and increase nl04 back
to 1.0e20 m-2 (from 0.8). Still no argon.

Shot 11 -- Plasma, but disrupted early. I think it was due to a moly
injection. Early density does not overshoot.

Next shot: repeat, no changes. Still no argon.

Shot 12 -- Plasma. Good shot. nl04=1.0e20 m-2. Only one RF transmitter
worked.

Next shot: program RCUR to ramp the outer gap linearly from 15 to
5 mm over the time range of 0.5 to 1.2 s respectively. Also, enable
NINJA helium puffing.

Shot 13 -- Plasma. Good shot. Outer gap ramped from 15 to 6 mm over the
requested time range. No RF. Helium went in. Chromex saturated.
FSP needs positioning adjustments.

Next shot: increase outer gap ramp by 1 more mm (to get 5 mm instead
of 6 mm at end of ramp). Still no argon.

Shot 14 -- Plasma, but disrupted early again.

Next shot: reduce NINJA plenum pressure from 5 to 3 psi of helium
(purge and re-fill). On a suggestion from Ian, reduce slow ZCUR P
and D gains from 1.5 to 1.0 (for nearly the entire shot). The claim
is that this should help the unstable period around 0.3 s. Also,
re-enable argon.

Shot 15 -- Fizzle. Null has degraded, but should still be good enough.

Next shot: try again, no changes.

Shot 16 -- Plasma. Good shot. Outer gap scan is exactly as requested.
nl04=1.0e20 m-2. JR saw argon. RF is "practicing" after 1.15 s.
Chromex helium data looks pretty good. Vertical instability at
t=0.3 s is gone.

Next shot: repeat. Turn on fast scanning probe.

Shot 17 -- Plasma. Good shot. Good spectrometer data and good scanning probe
data.

Next shot: reduce density to nl04=0.7e20 m-2 (from 1.0). Also, at
Ian's request, further reduce slow ZCUR P and D gains from 1.0 to
0.7 after t=0.25 s.

Shot 18 -- Plasma. Good shot. Density only dropped to nl04=0.8e20 m-2 (we
wanted 0.7). Also, at Ian's request, reduce RCUR P and D gains
from 6 to 5 after t=0.25 s.

Next shot: repeat. Increase argon pulse from 12 to 17 ms.

Shot 19 -- Plasma. Good shot. Density was still >= 0.8e20 m-2.

Next shot: increase density back to nl04=1.0e20 m-2. (Also, change
title of segment 2 to "Spencer's outer gap ramp"

Shot 20 -- Plasma. Good shot. Got the desired density.

Next shot: raise nl04 to 1.5e20 m-2. Still ohmic. Also, at Ian's
request, reduce CLEARIN gain by about 20% after t=0.36 s.

Shot 21 -- Plasma. Good shot. Density got to nl04=1.3e20 m-2 (wanted 1.5).
Scanning probe was turned off. There was a big marfe until 0.5 s.

Next shot: same density, ohmic. Undo the CLEARIN gain tweak. The
scanning probe will be enabled.

Shot 22 -- Plasma. Good shot. nl04=1.3e20 m-2 again. Another big marfe on
the upper inboard wall early in the shot.

Next shot: repeat, with 1.5 MW of ICRF from t=0.4 to 1.2 s. Also,
at Ian's request, reduce Ip gain from 6 to 5 after t=0.41 s.
No scanning probe for the next shot.

Shot 23 -- Plasma. Good shot. Prf=1.6-1.8 MW for 0.8 s. Plasma is in
ELM-free H-mode for the entire RF pulse, with a single ELM at
0.66 s. Density was much higher, nl04=1.8-1.9e20 m-2. Outer gap
is now ramping from 13 to 3 mm. 2pi bolo had almost 1 MW of
radiation, but it reaches a steady state for 0.5 s, even though it's
an ELM-free H-mode. An 0.5 s steady-state ELM-free H-mode! I think
that's a new phenomenon. However, the stored energy is only 100 kJ
(H factor of 1.2).

Next shot: repeat with no changes.

Shot 24 -- Plasma. Good shot. Prf=1.6-1.75 MW for 0.8 s. nl04 reaches
2.1e20 m-2. Prad=0.9 MW on 2pi bolo. It looks like an ELM-free
H-mode with lousy energy and particle confinement.

Next shot: repeat with no changes, except at Ian's request, reduce
CLEARIN P gain by 20% after t=1.2 s.

Shot 25 -- Plasma. Good shot. Prad was lower. H-alpha was rising slightly
as time went by. It's really tough to classify this type of H-mode.

Next shot: try for low density (nl04=0.6e20 m-2). Turn down the
pre-programmed valve voltage also.

Shot 26 -- Plasma. Good shot. Very different looking H-modes. Target density
was indeed lower. There are two ELM-free H-modes, followed by
two EDA-like H-modes, except that the H-factor is still low.

Next shot: repeat one of the earlier high-density shots, but without
RF. Re-load 980106023 and turn off RF.

Shot 27 -- Plasma. Good shot. nl04=1.3-1.4e20 m-2. Ohmic. No H-modes.

Next shot: repeat one of the earlier low-density ohmic shots.
Re-load 980106019 (and rename the segment as done for shot 20).

Shot 28 -- Plasma. Good shot. nl04=0.8e20 m-2. Good background shot.

End of MP 197 (Spencer's run)

Start MP 196A (Lin & Irby).

Next shot: fill pulse gas 5 plenum (J-bottom) with 30 psi of D2.
Reset RCUR demand to give a steady outer gap of about 10 mm.
Program in pulse gas 5 voltage to be fully open from t=0.5 to 1.0 s.
Program density ramp from nl04=0.6 to 2.4e20 m-2 during the time
span of t=0.3 to 1.0 s.

Shot 29 -- Plasma. Good shot. Density ramped up to nl04=2.0e20 m-2 at
t=1.0 s.

Next shot: increase pre-programmed voltage on pulse gas valve 3 to
100 volts from t=0.5 to 1.0 s. Also, turn off argon.

Shot 30 -- Plasma. Good shot. Density ramped up to nl04=2.6e20 at t=1.0 s !!
All the the shots for this MP will undoubtedly have density-limit
disruptions during the rampdown. Outer gap is 12-15 mm. TCI gate
valve closes after t=1.2 s, which is why the density looks like it
falls quickly.

Next shot: reduce outer gap to 5 mm. (Increase RCUR by 4 mm.)

Shot 31 -- Plasma. Good shot. Good density ramp. Outer gap is 5-8 mm.

Next shot: ramp density to nl04=1.8e20 m-2 at t=0.75 s, then hold
density at that value until t=1.0 s. Program the outer gap to
scan from 20 to 5 mm during the time span from t=0.75 to 1.0 s.
Reduce pre-programmed valve voltages to 50 volts after t=0.75 s.

Shot 32 -- Plasma. Good shot. Density flattopped at nl04=1.9e20 m-2 at
t=0.75 s. Outer gap ramped from 16 to 7 mm during the time span
of t=0.75 to 1.0 s.

Next shot: repeat with no changes.

Shot 33 -- Plasma. Good shot. Good repeat of the previous shot.

Next shot: extend outer gap ramp so that the plasma actually becomes
limited on the outboard limiters for 50 ms. Also, extend the
density ramp to get a flattop value of nl04=2.1e20 m-2.

Shot 34 -- Plasma. Good shot. Density got to nl04=2.3e20 m-2. Plasma just
barely became outboard-limited for 50 ms at t=1.0 s. Exactly as
requested! All reflectometer channels responded on this shot.

End of MP 196A

End of run

Session Leader Comments
Jan 6 1998 08:59:37:280AMSpencer PitcherMP 197 Helium Beam Check-Out
---------------------------------------

Today we will be comparing the helium beam with the fast-scanning
probe (FSP). The helium beam has only sparse spatial coverage,
ie 7 mm, and thus to obtain high resolution profiles we will
be 'scanning the plasma' by changing the outer gap slowly
over 700 ms approx. The FSP will scan at three times,
first to ensure the SOL is not evolving during the gap scan,
and second, to compare with the helium beam results. The
experiment will be performed in Ohmic discharges first
and then in RF discharges, a few densities will be used.

We also have a piggy-back experiment....Jim Weaver will be
trying to make helium II and Halpha drift measurements
in the divertor. Contrasting these will shed light
on the question of plasma pressure loss in the divertor
due to plasma-neutral friction.

Jan 6 1998 05:29:33:100PMYijun LinMP196a--reflectometer calibration

The MP is to produce high density shots to get the calibration of the reflectometer.
The main idea is to compare the reflectometer responses to different density ramp and different
outer gap. If the phase measurements from all channels are limited at some constants during the
varying of the density and the outer gap, it is reasonable to introduce some offsets to calibration
the reflectometer.

There are four channels of reflectometer working, i.e. 50GHz, 60GHz, 75GHz, and 110GHz channel.


Jan 6 1998 09:48:18:160AM980106002Spencer Pitchershot 02 -

Lower divertor capillary plenum filled to 7 psi of helium
(Bob is wrong.)
Jan 6 1998 10:52:55:390AM980106007Spencer Pitchershot 07 - density rises at .35 s, also fside ratiomatic,
Halpha looks like entering dithering Ohmic H-mode.
Loop voltage, ie Ohmic power high 1.6 MW.
Dithering Ohmic H-mode??? maybe?
Jan 6 1998 11:55:22:840AM980106012Spencer Pitchernext shot - first helium puff thru AB limiter
Jan 6 1998 12:15:18:460PM980106013Spencer Pitchershot 13 - helium puff thru AB limiter
delta_t = 0.1 s starting at 0.15 s @ 7 psia
- FSP moving in silo only
- Weaver on HeII, Chromex on helium red lines

* Weaver computer glitch ... no data
* Chromex satruated on 600 line grating
Jan 6 1998 12:24:00:960PM980106014Spencer Pitchershot 14 - helium puff increased to
dt = 0.2 @ 0.15 s
- chromex changed to 1200 line grating

* plasma disrupts at 0.35 s
Jan 6 1998 12:38:04:290PM980106015Spencer Pitchershot 15 - ninja plenum changed to 3 psia helium
t = .15 s, dt = 0.2
- chromex on 600 line grating

* fizzle
Jan 6 1998 12:51:24:780PM980106016Spencer Pitchershot 16 - ninja t = 0.15 s, dt = .1 s
- chromex NOT saturated, although Halpha interference?
Jan 6 1998 01:02:33:920PM980106017Spencer Pitchershot 17 - FSP to go on this shot
- tweak chromex to reduce Halpha interference
* good shot, nl4 = 1.0e20, FSP and chromex data both good
Jan 6 1998 01:15:33:560PM980106018Spencer Pitchershot 18 - drop density, keep all else the same
* density not quite flat
* fsp and chromex good

Jan 6 1998 01:33:53:420PM980106019Spencer Pitchershot 19 - repeat from last shot
* density behaviour no better....abandon
Jan 6 1998 01:49:17:760PM980106020Spencer Pitchershot 20 - go back to nl4 = 1.0e20
- shift ninja gas to inner wall (B16-001) for background check
* good fsp and chromex data
Jan 6 1998 01:59:07:900PM980106021Spencer Pitchershot 21 - go to nl4 = 1.5e20
- shift ninja gas to AB limiter again
- fsp stays out
* upper marfe seen on TV
* valves wide open but only got nl4 ~ 1.3e20
* chromex data good
Jan 6 1998 02:14:07:140PM980106022Spencer Pitchershot 22 - repeat shot 21
- fsp will go
* good chromex and fsp data
Jan 6 1998 02:35:01:210PM980106023Spencer Pitchershot 23 - repeat shot 22 + 1.5 MW ICRH from t = 0.4 to 1.2 s
- fsp off

* weird shot...starts off ELM-free H-mode with typical
signatures. Following a single giant ELM we go into
a steady-state ELM-free LOW D-alpha H-mode? or is
it enhanced D-alpha H-mode?
Jan 6 1998 02:48:43:440PM980106024Spencer Pitchershot 24 - repeat of shot 23
- fsp will go
- ninja gas shifted to inner wall for background check
* good fsp and chromex data
* again a new beastie....no large elm this time

Jan 6 1998 03:00:34:340PM980106025Spencer Pitchershot 25 - repeat of shot 24
- fsp will go
- ninja moved to AB limiter
* good fsp and chromex data
* new beastie again
* Halpha rises towards end of shot, no large elm

Jan 6 1998 03:16:50:230PM980106026Spencer Pitchershot 26 - attempt at lowering density
* very messy shot, series of elm-free periods

Jan 6 1998 03:30:12:040PM980106027Spencer Pitchershot 27 - go back to shot 23, nl4 = 1.5e20, Ohmic,
ninja gas to inner wall for background check
* chromex and fsp ok

Jan 6 1998 03:39:37:310PM980106028Spencer Pitchershot 28 - go back to shot 19, nl4 ~ 0.7e20, Ohmic,
ninja gas to inner wall for background check
* chromex and fsp ok
* LAST SHOT OF THIS MINI-PROPOSAL

Jan 6 1998 05:17:04:800PM980106029Yijun Linshot 29--Density is not as high as expected. Some limitings in the phases are observed.
Try higher density next shot.
Jan 6 1998 05:19:42:830PM980106030Yijun Linshot 30--Density is as high as expected. A good shot to provide refrence data for the
following shots.
Jan 6 1998 05:22:34:480PM980106031Yijun Linshot 31--The outer gap is reduced; however, the preliminary results show there is little
variation of the phases in the low frequency channels ( 50 and 60 GHz Channels )
compared with shot 30.
Jan 6 1998 05:24:38:020PM980106032Yijun Linshot 32--During the ramping of the outer gap, preliminary results show little variation
in the phase measure measurement of 50Ghz, 60GHz, and 75 GHz Channels.
Jan 6 1998 05:25:17:360PM980106033Yijun Linshot 33--Similiar result as last shot.
Jan 6 1998 05:27:59:110PM980106034Yijun Linshot 34--Similiar preliminary results as last shot even though the ramping of the outer gap
is much larger.

Physics Operator Comments
Jan 5 1998 01:11:18:010PMRobert GranetzRun Plan for Tuesday, 980106 (con't):

In addition to the two MP's scheduled for today, it has also been requested
that we begin with a number of fiducial shots in order to evaluate wall
conditions. The setup shot specified above is both a standard 1997
fiducial, as well as the starting point for MP 197.

Although MP 197 will begin with ohmic plasmas, later in the morning we will
want to have very long 1 MW RF pulses, starting at t=0.4 s and ending at
t=1.2 s (RF pulse duration of 0.8 s). The FSP should be setup to scan at
t=0.60, 0.85, and 1.10 s on every shot of this MP. Helium will be puffed
from the A-B (outboard) limiter midplane during the shots, and the outer gap
will be scanned slowly from 15 to 5 mm during each shot. Several densities
will be run, with and without RF.

MP 196A will be run after MP 197. It requires similar plasma magnetic
configurations, but at high density (nebar=4e20 m-3), with no RF.
Jan 6 1998 09:03:54:910AMRobert Granetz Note: BAROLO is dead (hardware problem). Steve W has re-assigned EFIT
to run on RHONE, but processes that depend on it will fail. I've
sent mail to Tom and Josh, requesting them to re-assign the logical
queue name MHDQ.

Note: 2 hours of ECDC in deuterium was done prior to today's run.

Hybrid coldstart done; no errors found
($set default cmod$models:[hybrid.tcl]
$@hybinit_dispatch
$@show_bitbus_log
Use editor search to verify no occurrences of "error" nor
"not respond")
Reload fiducial plasma shot 971218006 (0.8 MA, 5.3 T, 1.0e20 m-2)
RF on for the initial fiducial shots.
RF feedback on RCUR is turned off (RF_LOAD_E = 0.0)
Jan 6 1998 11:20:17:010AMRobert GranetzRun Plan and Engineering Setup for Tuesday, 980106

MP 197 -- Helium Probe Edge Measurements & Comparison with Scanning Probe
MP 196A -- Reflectometer Calibration

Standard ECDC in deuterium for a couple of hours before run.

Engineering setup as on 971218006, with TF under hybrid control
This will be 0.8 MA, 5.3 T, nl_04=1e20 m-2, flattop to 1.2 s.

Gas setup - fill K-side with 30 PSI of D2, hybrid enable
fill B-top with 6 PSI of D2, hybrid enable
fill B-side-lower with 1.5 PSI of Ar, hybrid enable

Ninja system setup:
Fill lower divertor capillary plenum to 5 PSI of helium, and enable the
valve.

The following gate valves should be enabled for the start of the run,
assuming no vacuum problems:

ECE (no others until we verify the status of each system)
Jan 6 1998 09:00:09:890AM980106001Robert Granetz
Shot 01 -- Standard Tuesday OH test. 10 kA. Coax resistances are good and low.

Next shot: try for fiducial plasma.
Jan 6 1998 09:20:59:100AM980106002Robert GranetzShot 02 -- Fizzle. Good null. CII level was at 0.01 W/cm2/st. Got the
pre-fill gas puff.

Next shot: re-try, no changes.
Jan 6 1998 09:50:45:410AM980106003Robert Granetz
Shot 03 -- Fizzle. Good null. CII is lower, but still not down to normal
levels.

Next shot: gas-head tweak (increase pre-pulse puff from 15 to 18 ms)
Jan 6 1998 10:01:24:090AM980106004Robert GranetzShot 04 -- Plasma. Disrupted at 0.43 s. Ip=0.8 MA. VDE. Loop voltage was
2.3 volts! I think it's just dirty.

Next shot: repeat, trying to cleanup.
Jan 6 1998 10:09:11:220AM980106005Robert GranetzShot 05 -- Fizzle.

Next shot: reduce pre-pulse puff from 18 to 14 ms.
Jan 6 1998 10:29:52:760AM980106006Robert Granetz
Shot 06 -- Plasma. Another early VDE disruption. Spencer has noticed that
the F-side ratiomatic pressure is 3 times higher than during
December, even though the divertor pressure is the same.

Next shot: no changes
Jan 6 1998 10:56:01:690AM980106007Robert Granetz
Shot 07 -- Plasma. Another early VDE disruption. Density takes off at 0.34 s,
even though the feedback turns off the gas. Spencer is speculating
that the plasma is toggling between ohmic H and L modes.

Next shot: lower the density from nl04=1.0 to 0.8e20 m-2.
Jan 6 1998 11:00:12:490AM980106008Robert GranetzShot 08 -- Plasma. Disrupted even earlier (0.34 s.). There was a large moly
brick at 0.26 s.

Next shot: repeat, no changes.
Jan 6 1998 11:32:15:940AM980106009Robert GranetzShot 09 -- Plasma. Disrupted at 0.46 s. N/Ncrit is reaching 1.2, but the
elongation and gaps are not excessive.

Next shot: turn off argon
Jan 6 1998 11:32:52:850AM980106010Robert Granetz
Shot 10 -- Plasma. Good shot. Full length. Density was high initially, but
got back down to nl04=0.8e20 m-2 after 0.4 s. H-alpha shows
continuous dithering for the entire shot. Only difference from the
previous shot was turning off the argon.

Next shot: reduce early gas valve voltage, and increase nl04 back
to 1.0e20 m-2 (from 0.8). Still no argon.
Jan 6 1998 11:43:04:830AM980106011Robert GranetzShot 11 -- Plasma, but disrupted early. I think it was due to a moly
injection. Early density does not overshoot.

Next shot: repeat, no changes. Still no argon.
Jan 6 1998 11:58:13:150AM980106012Robert GranetzShot 12 -- Plasma. Good shot. nl04=1.0e20 m-2. Only one RF transmitter
worked.

Next shot: program RCUR to ramp the outer gap linearly from 15 to
5 mm over the time range of 0.5 to 1.2 s respectively. Also, enable
NINJA helium puffing.
Jan 6 1998 12:13:51:260PM980106013Robert GranetzShot 13 -- Plasma. Good shot. Outer gap ramped from 15 to 6 mm over the
requested time range. No RF. Helium went in. Chromex saturated.
FSP needs positioning adjustments.

Next shot: increase outer gap ramp by 1 more mm (to get 5 mm instead
of 6 mm at end of ramp). Still no argon.
Jan 6 1998 12:23:15:500PM980106014Robert GranetzShot 14 -- Plasma, but disrupted early again.

Next shot: reduce NINJA plenum pressure from 5 to 3 psi of helium
(purge and re-fill). On a suggestion from Ian, reduce slow ZCUR P
and D gains from 1.5 to 1.0 (for nearly the entire shot). The claim
is that this should help the unstable period around 0.3 s. Also,
re-enable argon.
Jan 6 1998 12:34:10:480PM980106015Robert GranetzShot 15 -- Fizzle. Null has degraded, but should still be good enough.

Next shot: try again, no changes.
Jan 6 1998 12:50:23:100PM980106016Robert GranetzShot 16 -- Plasma. Good shot. Outer gap scan is exactly as requested.
nl04=1.0e20 m-2. JR saw argon. RF is "practicing" after 1.15 s.
Chromex helium data looks pretty good. Vertical instability at
t=0.3 s is gone.

Next shot: repeat. Turn on fast scanning probe.
Jan 6 1998 01:01:58:670PM980106017Robert GranetzShot 17 -- Plasma. Good shot. Good spectrometer data and good scanning probe
data.

Next shot: reduce density to nl04=0.7e20 m-2 (from 1.0). Also, at
Ian's request, further reduce slow ZCUR P and D gains from 1.0 to
0.7 after t=0.25 s.
Jan 6 1998 01:15:23:980PM980106018Robert GranetzShot 18 -- Plasma. Good shot. Density only dropped to nl04=0.8e20 m-2 (we
wanted 0.7). Also, at Ian's request, reduce RCUR P and D gains
from 6 to 5 after t=0.25 s.

Next shot: repeat. Increase argon pulse from 12 to 17 ms.
Jan 6 1998 01:33:16:990PM980106019Robert GranetzShot 19 -- Plasma. Good shot. Density was still >= 0.8e20 m-2.

Next shot: increase density back to nl04=1.0e20 m-2. (Also, change
title of segment 2 to "Spencer's outer gap ramp"
Jan 6 1998 01:43:35:860PM980106020Robert GranetzShot 20 -- Plasma. Good shot. Got the desired density.

Next shot: raise nl04 to 1.5e20 m-2. Still ohmic. Also, at Ian's
request, reduce CLEARIN gain by about 20% after t=0.36 s.
Jan 6 1998 01:58:55:760PM980106021Robert GranetzShot 21 -- Plasma. Good shot. Density got to nl04=1.3e20 m-2 (wanted 1.5).
Scanning probe was turned off. There was a big marfe until 0.5 s.

Next shot: same density, ohmic. Undo the CLEARIN gain tweak. The
scanning probe will be enabled.
Jan 6 1998 02:13:30:240PM980106022Robert GranetzShot 22 -- Plasma. Good shot. nl04=1.3e20 m-2 again. Another big marfe on
the upper inboard wall early in the shot.

Next shot: repeat, with 1.5 MW of ICRF from t=0.4 to 1.2 s. Also,
at Ian's request, reduce Ip gain from 6 to 5 after t=0.41 s.
No scanning probe for the next shot.
Jan 6 1998 02:29:30:220PM980106023Robert GranetzShot 23 -- Plasma. Good shot. Prf=1.6-1.8 MW for 0.8 s. Plasma is in
ELM-free H-mode for the entire RF pulse, with a single ELM at
0.66 s. Density was much higher, nl04=1.8-1.9e20 m-2. Outer gap
is now ramping from 13 to 3 mm. 2pi bolo had almost 1 MW of
radiation, but it reaches a steady state for 0.5 s, even though it's
an ELM-free H-mode. An 0.5 s steady-state ELM-free H-mode! I think
that's a new phenomenon. However, the stored energy is only 100 kJ
(H factor of 1.2).

Next shot: repeat with no changes.
Jan 6 1998 02:43:13:850PM980106024Robert GranetzShot 24 -- Plasma. Good shot. Prf=1.6-1.75 MW for 0.8 s. nl04 reaches
2.1e20 m-2. Prad=0.9 MW on 2pi bolo. It looks like an ELM-free
H-mode with lousy energy and particle confinement.

Next shot: repeat with no changes, except at Ian's request, reduce
CLEARIN P gain by 20% after t=1.2 s.
Jan 6 1998 02:55:19:660PM980106025Robert GranetzShot 25 -- Plasma. Good shot. Prad was lower. H-alpha was rising slightly
as time went by. It's really tough to classify this type of H-mode.

Next shot: try for low density (nl04=0.6e20 m-2). Turn down the
pre-programmed valve voltage also.
Jan 6 1998 03:13:47:870PM980106026Robert GranetzShot 26 -- Plasma. Good shot. Very different looking H-modes. Target density
was indeed lower. There are two ELM-free H-modes, followed by
two EDA-like H-modes, except that the H-factor is still low.

Next shot: repeat one of the earlier high-density shots, but without
RF. Re-load 980106023 and turn off RF.
Jan 6 1998 03:33:30:280PM980106027Robert GranetzShot 27 -- Plasma. Good shot. nl04=1.3-1.4e20 m-2. Ohmic. No H-modes.

Next shot: repeat one of the earlier low-density ohmic shots.
Re-load 980106019 (and rename the segment as done for shot 20).
Jan 6 1998 03:52:57:470PM980106028Robert GranetzShot 28 -- Plasma. Good shot. nl04=0.8e20 m-2. Good background shot.

End of MP 197 (Spencer's run)
Jan 6 1998 03:53:44:430PM980106029Robert GranetzStart MP 196A (Lin & Irby).

Next shot: fill pulse gas 5 plenum (J-bottom) with 30 psi of D2.
Reset RCUR demand to give a steady outer gap of about 10 mm.
Program in pulse gas 5 voltage to be fully open from t=0.5 to 1.0 s.
Program density ramp from nl04=0.6 to 2.4e20 m-2 during the time
span of t=0.3 to 1.0 s.
Jan 6 1998 04:03:23:220PM980106029Robert GranetzShot 29 -- Plasma. Good shot. Density ramped up to nl04=2.0e20 m-2 at
t=1.0 s.

Next shot: increase pre-programmed voltage on pulse gas valve 3 to
100 volts from t=0.5 to 1.0 s. Also, turn off argon.
Jan 6 1998 04:24:48:320PM980106030Robert GranetzShot 30 -- Plasma. Good shot. Density ramped up to nl04=2.6e20 at t=1.0 s !!
All the the shots for this MP will undoubtedly have density-limit
disruptions during the rampdown. Outer gap is 12-15 mm. TCI gate
valve closes after t=1.2 s, which is why the density looks like it
falls quickly.

Next shot: reduce outer gap to 5 mm. (Increase RCUR by 4 mm.)
Jan 6 1998 04:44:24:360PM980106031Robert GranetzShot 31 -- Plasma. Good shot. Good density ramp. Outer gap is 5-8 mm.

Next shot: ramp density to nl04=1.8e20 m-2 at t=0.75 s, then hold
density at that value until t=1.0 s. Program the outer gap to
scan from 20 to 5 mm during the time span from t=0.75 to 1.0 s.
Reduce pre-programmed valve voltages to 50 volts after t=0.75 s.
Jan 6 1998 04:48:16:440PM980106032Robert GranetzShot 32 -- Plasma. Good shot. Density flattopped at nl04=1.9e20 m-2 at
t=0.75 s. Outer gap ramped from 16 to 7 mm during the time span
of t=0.75 to 1.0 s.

Next shot: repeat with no changes.
Jan 6 1998 05:12:56:860PM980106033Robert GranetzShot 33 -- Plasma. Good shot. Good repeat of the previous shot.

Next shot: extend outer gap ramp so that the plasma actually becomes
limited on the outboard limiters for 50 ms. Also, extend the
density ramp to get a flattop value of nl04=2.1e20 m-2.
Jan 6 1998 05:13:14:850PM980106034Robert GranetzShot 34 -- Plasma. Good shot. Density got to nl04=2.3e20 m-2. Plasma just
barely became outboard-limited for 50 ms at t=1.0 s. Exactly as
requested! All reflectometer channels responded on this shot.

End of MP 196A

End of run

Engineering Operator Comments
ShotTimeTypeStatusComment
108:48:56:390AMTestOk
209:14:46:870AMPlasmaOk
309:39:23:670AMPlasmaBadfizzle
409:52:31:940AMPlasmaOk
510:03:33:310AMPlasmaOk
610:15:48:250AMPlasmaOk
710:31:29:980AMPlasmaOk
810:52:03:460AMPlasmaOk
911:03:16:440AMPlasmaOk
1011:17:17:760AMPlasmaOk
1111:34:38:310AMPlasmaOk
1211:44:45:940AMPlasmaOk
1312:00:39:610PMPlasmaOk
1412:14:53:730PMPlasmaOk
1512:27:52:520PMPlasmaOk
1612:37:23:600PMPlasmaOk
1712:51:27:960PMPlasmaOk
1801:04:19:120PMPlasmaOk
1901:20:19:180PMPlasmaOk
2001:36:44:200PMPlasmaOk
2101:50:05:890PMPlasmaOk
2202:04:31:170PMPlasmaOk
2302:15:41:490PMPlasmaOk
2402:32:17:450PMPlasmaOk
2502:45:16:490PMPlasmaOk
2603:03:55:300PMPlasmaOk
2703:18:09:820PMPlasmaOk
2803:33:15:800PMPlasmaOk
2903:55:50:690PMPlasmaOk
3004:10:33:370PMPlasmaOk
3104:25:18:420PMPlasmaOk
3204:40:12:440PMPlasmaOk
3304:52:04:330PMPlasmaOk
3405:02:56:060PMPlasmaOk