| Session leader(s): | Steve Wolfe |
| Physics operator(s): | John Goetz |
| Engineering operator(s): | Gary Dekow,Bill Cochran,Frank Silva |
| Engineering Operator Run Comment |
| Reverse start up/Wolfe/Silva |
| Session Leader Plans |
| Physics Operators Plans |
| Session Leader Summaries |
| Entered: Jul 7 2004 03:49:22:257PM |
| Author: To Be Determined |
| Run Summary for 991102 MP#206 startup (reverse-field) SL: wolfe PO: Goetz EO: Silva,Cochran Plan: ----- The purpose of this run is to establish operation with reversed (positive) field and current, and carry out discharge development for the reverse-field physics runs later this week. The TF and PF busswork was reversed on Saturday. The HYBRID tree has been modified to reflect the new configuration (smw-991030). Operation will begin with a series of PLC power system tests to verify the configuration changes. HYBRID demands will be monitored during the PLC shots to verify the Hybrid/PCS changes. Plasma operation will start with a reversed current shot based on 991029022, a good 800kA shot from last Friday. Some tweaking will presumably be required to get a good startup, and to adjust offsets for RCUR and perhaps other wires. Check gaps for RF. If the startup goes smoothly, we may try increasing the current to 1.0 and 1.2MA, to allow the RF group to check the coupling in preparation for the run on Wednesday. For the upper x-point run on Thursday, Ian wants to do triangularity scans like those on 990922. We'll start from a static point (RXL=0) based on 990922006, then try to implement the sweep. Results: -------- We succeeded only in the first of the goals, getting a plasma with reversed field. The startup is far from robust, and we only got three plasmas that reached flattop (#24-26). Those shots however lasted to 2sec (longest ever), when they were taken out by power supply inversion. RF turned on on shot 26, and got close to a MW using D&E together. Both transmitters ran without faulting at the 0.4-0.5MW level. Unfortunately, we lost the plasma after #26, so they were unable to increase the power. CLEARIN required substantial re-programming to avoid running with zero gaps. Shot 24 was limited on both sides for most of the long shot, and had an elongation of only 1.3. Shot 25 was better, with elongation of 1.5, and #26 was close to nominal with eout=1.6. The program waveform changed from shot to shot from -.01 to +.01 to +0.015. I suspect we actually should be using an offset in CLEARIN of magnitude 5000 : 10000, but this was not done. A dummy CLEARIN with offset of 10000 was loaded on wire 9 with no gain for shot 26, for comparison. ZCUR also had to be adjusted, from -0.015 to -.005 in order to get an EFIT ZMAXIS of -.008m. The RCUR offsets in segment 2 were changed from +8000,+5000 to -8000,-5000. These were not checked in detail. The Segment 1 RCUR offset was changed from 5500 to negative values in the range -5500 : -10000. It seems to matter a lot, but no robust solution was found. All the successful startups had bad hesitations. The upper null segments set up for development of Ian's run on Thursday are still sitting in segments 3 and 4, as described in the logbook, but were not tried. The RCUR offsets and CLEARIN and ZCUR program waveforms should presumably be modified as was done to segment 2. All the power systems ran well today, except for an unexplained OH2L failure on shot 22. Irby was monitoring the TF resistances, and gave us permission to put the delay on the TF fizzle detector back to 0.1 sec after shot 23. This helped a good deal on cooling after fizzles. The tc's used to indicate central column cooling seemed to have changed, such that we were not getting below about -176 no matter how long we cooled. One possibility is that the reference point junction is off-temp. |
| Physics Operator Summaries |
| Entered: Jul 7 2004 04:37:09:897PM |
| Author: To Be Determined |
| Physics operator summary for 9911-2 EO: Silva & Cochran & Dekow PO: Wolfe & Goetz SL: Wolfe Run 991102 was in support of MP 206 -- Startup (with reversed field) The purpose of this run is to establish operation with reversed (positive) field and current, and carry out discharge development for the reverse-field physics runs later this week. The TF and PF busswork was reversed on Saturday, October 30. The HYBRID tree has been modified to reflect the new configuration (SMW-991030). Operation will begin with a series of PLC power system tests to verify the configuration changes. HYBRID demands will be monitored during the PLC shots to verify the Hybrid/PCS changes. Plasma operation will start with a reversed current shot based on 991029022, a good 800kA shot from last Friday. Some tweaking will presumably be required to get a good startup, and to adjust offsets for RCUR and perhaps other wires. Check gaps for RF. If the startup goes smoothly, we may try increasing the current to 1.0 and 1.2MA, to allow the RF group to check the coupling in preparation for the run on Wednesday. For the upper x-point run on Thursday, November 4, triangularity scans like those on 990922 need to be done. We'll start from a static point (RXL=0) based on 990922006, then try to implement the sweep. The first thirteen shots today were used as tests to determine if the changeover in the magnet bus was done correctly. The run began with a plasma attempt on shot 14. This was, luckily, a fizzle with positive radial field and 20mT of vertical field. The OH1 was producing too much current and the demand was lowered. Also, the EF1s were lowered 200A in an attempt to get the curvature of the field straightened out. The offset in the RCUR controller in segment 1 was twiddled about to try and find the best number to use. Eventually, an offset of between -5500 and -8000 produced plasmas. It was also necessary to increase the EF4 current to 1425A in order to get the vertical field correct. After all of this, three plasmas were produced. Each had a serious hesitation in the current rise. However, each plasma lasted longer than 2s. The first plasma produced indicated that there was also an offset problem with CLEARIN. This manifested itself in a plasma that was limited on both the inner and outer limiters. The drawn waveform was changed by 2.5cm to obtain gaps similar to those of "normal" field. The vertical position, ZCUR, also had an offset, and a change of +1cm brought ZCUR back to its nominal -0.6cm position. These three plasmas were followed by four fizzles to end the run. These four appeared to be lost vertically (down) as indicated by the MFLUX z calculation. Why this vertical excursion appeared remains unexplained. The last shot was a reload of a plasma shot and it too fizzled with this vertical excursion. Engineering, computing, and diagnostics today: The OH2L supply pulled after the switch closing @ -0.7sec resulting in a dud. All else appeared to be o.k. Scorecard: 30 shots = 3 plasmas + 13 fizzles + 1 duds + 13 tests Shot summary for Run 991020 Steve Wolfe does a bunch of pre-run setup work PCS Setup: ----------- LOAD_FROM 991029022 (acknowledge config mismatch) IMPORT_SEG #3 from 990922006 segment 2 IMPORT_SEG #4 from 990922006 segment 2 (same as #3) Segment #1: All_p's - flip all IC_*, IP All V's - flip everything but pulse gas BR0 offset is already 0.0! RCUR offset is 5500; leave it for now Change title to "Reverse field startup" Segment #2: All P's: flip EF2byEF4, IC_EF4, Ip All V's: flip everything but pg valves RCUR offsets are 8000,5000. Leave them for now. Change title to "800kA reverse-field lower null" Segment #3: All P's: flip EF2byEF4, IC_EF4,Ip; flip ZCUR for upper null; set RXL,RXU waveforms to zero interchange ZXL and ZXU All V's: flip all voltages except pg valves RCUR offsets are 8000,5000. Leave them for now, change as indicated by seg 2 results. CLEARIN: set keyword ZXA=.394 and CALL ALG | Apply | OK factor Controller: zxa=.394 | call | OK | OK timeconst. STRKSPI: keywords RSOUT=.685,ZSOUT=.60,ZXA=.394 | Call Alg waveform set to zero Pgain to 5 starting at .3sec Controller: zxa=.394,RSOUT=.685,ZSOUT=.60, ALL=['PSI0','RCUR','ZCUR','CLEARIN','STRKPSI','RXL','ZXL','RXU'] Call | set common rate to 9.0 Factor to .0001 (was .0002) RXL: Load gains from RXU (seg 4, wire 7) ZXL: Load gains from ZXU (seg 4, wire 8) RXU: Load gains from RXL (seg 4, wire 5) ZXU: Load gains from ZXL (seg 4, wire 6), set to 0.5 after 0.3sec Change title to reverse-field upper null Segment 4: Import segment from PCS$SCRATCH, shot -1, seg#3 For RXL, RXU load wave from 990922006 RXU, RXL (\PCS::TOP.SEG_02:P_05,\PCS::TOP.SEG_02:P_07). RXL exceeds wavegen limit; reduce factor from 4e-5 to 3.6e-5. Change title to "Reverse field triangularity scan" Build All - No errors or warnings Save As 991102999 (just in case) Load/Clean - put it in the model, ready to go. coldstart hybrid - everything loaded; all matrices respond The plasma INIT scope in the CMOD$MODELS:[HYBRID.SCOPES] area has an offset in the plasma current used in the calculation of R and z. This offset has the wrong sign for this bus configuration. Shot change result ____ _____ ______ 1 EF2U 2kA PLC test shot EF2U current is positive (as intended) EF2U voltage is also ok; Hybrid is calling for +voltage, then negative 2 EF2L 2Ka PLC test shot looks ok 3 EFC 1Ka test shot. negative on EFC_L 4 TF test (50KA, PLC control) o.k.; 50KA positive 5 EF4 test shot; "forward" o.k.; positive current, 1250A in each 6 EF4 test shot; "negative" o.k.; negative current 7 EF3 test shot air flow problem in the air switch 8 EF3 test shot o.k.; -1750A 9 EF1U test shot o.k.; 2kA positive 10 EF1L test shot o.k.; 2kA positive 11 OH1 test shot o.k.; 10kA positive 12 OH2U test shot o.k.; 10kA positive 13 OH2L test shot o.k.; 10kA positive 14 try for plasma; EF4 limit fizzle; b.d. @ .027s!!!; Bz = -.02T @ only set to 8.5kA t=0; Br(0)=.001; OH1 has about 1kA more current than last week 15 IC_OH1=19kA (was 20kA) fizzle; b.d. @ .006s; Bz = -.012T; Br=0 field curvature is wrong 16 IC_OH1=18.5kA; IC_EF1s=5900A fizzle; b.d. @ .005s; Bz=-.005T; Br=0 (was 6000A) field curvature is wrong 17 IC_OH1=19kA; IC_EF4=1380A fizzle; b.d. @ .0045; Bz=-.006; Br=0; (was 1280A) field curvature is wrong 18 IC_EF1s to 5800A; flip the fizzle; b.d. @ .005; Bz=-.008; Br=+.001 sign on the RCUR offset because vert. field is straight; RCUR error the RCUR error is ~14000 is now ~4000 19 IC_EF4=1410A; RCUR offset to fizzle; b.d. @ .006; Bz=-.006; Br=+.001 -10000 in segment 1 RCUR error is now -1500 20 IC_EF4=1440A; RCUR offset to 0 fizzle; b.d. @ .007; Bz=-.005; Br=+.001 RCUR error ~8000 21 IC_EF4=1425A; BR_0 offset = dud; OH2L pulls out @ -0.7 -.001; PG4 = 22ms (was 18ms) 22 try again fizzle; b.d. @ .0055; Bz turns around after breakdown; second plasma forms @ .045s 23 RCUR offset = -2750 fizzle; b.d. @ .006; Bz falls as Ip rises; three plasmas this time 24 RCUR offset = -5500 plasma until 2sec; b.d. @ .006; bad hesitation; both gaps near zero; large offset on CLEARIN 25 RCUR offset = -7000; draw plasma until 2sec; b.d. @ .005; bad CLEARIN to +.01 (was -.01) hesitation; kappa=1.5; some gaps 26 RCUR offset=-8000; CLEARIN plasma until 2sec; b.d.@.0055; very bad out to .015; ZCUR up 1cm hesitation; gaps and position o.k. 27 PG4=20ms; CLEARIN=.01 @ .37S fizzle; b.d. @ .0055; 40kA; current rise too fast for Bz 28 RCUR offset = -9000 fizzle; b.d. @ .006; 40kA 29 PG4=22ms; BR_0=0 fizzle; weaker 30 reload 991102026; CLEARIN=.01 fizzle; seems to be moving down @ 0.37s |
| Session Leader Comments | |||
| Nov 2 1999 10:05:37:530AM | 991102007 | Steve Wolfe | Shot 7: EF3 test. Didn't go. |
| Nov 2 1999 10:16:23:483AM | 991102008 | Steve Wolfe | Shot 8: EF3 test. (try again, it had a Rectifier air-flow fault last time?!) OK, -1700A |
| Nov 2 1999 10:19:47:903AM | 991102009 | Steve Wolfe | Shot 9: EF1U test - positive current, 2Ka |
| Nov 2 1999 10:26:19:030AM | 991102010 | Steve Wolfe | Shot 10: EF1L test/ should be same polarity. OK. |
| Nov 2 1999 10:35:11:797AM | 991102011 | Steve Wolfe | Shot 11: OH1 test. Electrons tree back on, to check on ECE and TCI. positive 10KA - ok Next shot, oh2u. |
| Nov 2 1999 10:41:27:090AM | 991102012 | Steve Wolfe | Shot 12: oh2u test - ok, 10kA positive, switch closes normally |
| Nov 2 1999 10:49:58:983AM | 991102013 | Steve Wolfe | Shot 13: oh2l test - looks good. Next shot: Try for plasma. |
| Nov 2 1999 11:21:04:763AM | 991102014 | Steve Wolfe | Shot 14: Try for plasma. Note: Bill points out that the EF4 was reset to 8.5ka on 990922, not at 10kA, as indicated in Ian's logbook entry. Therefore, we will use 8.5ka and hope for the best (it won't matter until we get to the x-point sweep, anyway). Fizzle detector is set to take out the TF early, for this first full current shot. Fizzle. Bz0 is very negative (-20mT). Breakdown is at 30msec. OH1 is at 20.6kA, EF4 is at 1250A; setpoints are 20.0 and 1280, and OH1 is normally around 19.7kA. Next shot: Decrease IC_OH1 program. |
| Nov 2 1999 11:40:52:810AM | 991102015 | Steve Wolfe | Shot 15: OH1 current program down by 1kA. TF fd delay to 1500msec anoter fizzle. Bz is still -10mT. bd at 7msec. Br0 ~0. Has strong negative curvature on MFLUX plots. Next shot: try another 500A off of IC_OH1, 100A off of the EF1's. |
| Nov 2 1999 11:59:52:153AM | 991102016 | Steve Wolfe | Shot 16: try another 500A off of IC_OH1, 100A off of the EF1's. On further reflection, I think we should have used the EF4 instead of OH1. fizzle- bd at 5msec, index is still bad (worse). Next shot- put back OH1, put 100A on EF4. Note: In Ian's INIT scope, change the sign of the Ip offset in RCUR and ZCUR panels. |
| Nov 2 1999 12:20:33:013PM | 991102017 | Steve Wolfe | Shot 17: put OH1 back to 19kA, put +100A on EF4, leave EF1's at 5900 fizzle. bd at 4.7msec curvature still too negative, but better. Next shot: Reverse sign of RCUR offsets (seg 1 and 2), and take another 100A off of EF1's (to 5800A) |
| Nov 2 1999 12:46:02:403PM | 991102018 | Steve Wolfe | Shot 18: Reverse sign of RCUR offsets (seg 1 and 2), and take another 100A off of EF1's (to 5800A) fizzle. Curvature is better, Bz takes off negative after 10ms, probably too much so. Next shot: EF4 more positive by 30A, seg 1 RCUR offset from -5500 => -10000 |
| Nov 2 1999 01:12:42:560PM | 991102019 | Steve Wolfe | Shot 19: EF4 more positive by 30A, seg 1 RCUR offset from -5500 => -10000 And another fizzle! RCUR error is now negative, but it doesn't help; maybe worse than previous. |
| Nov 2 1999 02:00:12:217PM | 991102020 | Steve Wolfe | Shot 20: Seg1 RCUR offset to 0. from -10000; +30A more on EF4 There is a delay while the cooling is looked at. After 45 minutes, another fizzle. Bz evolution certainly changed. Breakdown looks even dimmer than before. |
| Nov 2 1999 02:11:38:437PM | 991102021 | Steve Wolfe | Shot 21: Back off 15A on EF4 and raise gas. BR0 from 0 to -.001 Running out of ideas (and daylight). Dud. OH2L crowbars at -.7sec |
| Nov 2 1999 02:41:33:170PM | 991102022 | Steve Wolfe | Shot 22: Repeat of 21 Back off 15A on EF4 and raise gas. BR0 from 0 to -.001 fizzle, but it gets to 50KA before it dies. Some runaways, and can see it on ECE and XUV |
| Nov 2 1999 03:06:57:217PM | 991102023 | Steve Wolfe | Shot 23: RCUR offset to -2700 in seg1. tf fizzle delay back to 100msec (from 1500), to see if it will help the cooling Gets to 80kA, eventually, but it takes several bumps on the way. The Bz evolution is very non-monotonic. Next shot: RCUR offset back to -5500. Keep going that way until it stops getting better? |
| Nov 2 1999 03:30:11:623PM | 991102024 | Steve Wolfe | Shot 24: RCUR offset back to -5500. Keep going that way until it stops getting better? Plasma runs to 2sec! Front of the room TV didn't show up. Still has a bad hesitation Both gaps go to zero for most of the shot. Elongation=1.3 CLEARIN seems to have a big offset |
| Nov 2 1999 03:50:06:873PM | 991102025 | Steve Wolfe | Shot 25: Set CLEARIN to +.01 (was -.01). Attack hesitation by another RCUR offset tweak, to -7000 in seg1 plasma to 2sec again Elongation is 1.5, and the gaps are about 0.5cm Zaxis is about 1.5cm low. Some hards early. Next shot: Raise ZCUR by 1cm, Increase CLEARIN by .005, RCUR offset to -8000 Put a dummy CLEARIN wire on wire 9, with 0 gains and 10000. offset RF will try to fire. |
| Nov 2 1999 04:10:34:390PM | 991102026 | Steve Wolfe | Shot 26: Raise ZCUR by 1cm, Increase CLEARIN by .005, RCUR offset to -8000 Put a dummy CLEARIN wire on wire 9, with 0 gains and 10000. offset RF will try to fire. Plasma to 2sec. Hesitation is different, slower rise, then a dip D&E at 0.4MW each. Left gap at 1.5 and dropping, right gap drops to 0 at 0.5, but flattops at 0.75cm No hards. Next shot: Move CLEARIN positive more quickly, lower the fill, turn on the Argon |
| Nov 2 1999 04:23:08:090PM | 991102027 | Steve Wolfe | Shot 27: Move CLEARIN positive more quickly, lower the fill, turn on the Argon fizzle. current rise is faster, Bz doesn't change. Next shot: try RCUR offset to -9000 |
| Nov 2 1999 04:35:44:640PM | 991102028 | Steve Wolfe | Shot 28:try RCUR offset to -9000 Move CLEARIN positive more quickly, lower the fill, turn on the Argon Another fizzle. Next shot: Br0 back to 0 from -.001; raise the fill again. |
| Nov 2 1999 04:46:25:903PM | 991102029 | Steve Wolfe | Shot 29:Br0 back to 0 from -.001; raise the fill again try RCUR offset to -9000 Move CLEARIN positive more quickly, lower the fill, turn on the Argon fizzle. Go back to 26, with the CLEARIN adjusted. |
| Physics Operator Comments | |||
| Nov 1 1999 12:16:58:450PM | Steve Wolfe | Setting up PCS for tomorrow's reversed-field run: LOAD_FROM 991029022 (acknowledge config mismatch) IMPORT_SEG #3 from 990922006 segment 2 IMPORT_SEG #4 from 990922006 segment 2 (same as #3) Segment #1: All_p's - flip all IC_*, IP All V's - flip everything but pulse gas BR0 offset is already 0.0! RCUR offset is 5500; leave it for now Change title to "Reverse field startup" Segment #2: All P's: flip EF2byEF4, IC_EF4, Ip All V's: flip everything but pg valves RCUR offsets are 8000,5000. Leave them for now. Change title to "800kA reverse-field lower null" Segment #3: All P's: flip EF2byEF4, IC_EF4,Ip; flip ZCUR for upper null; set RXL,RXU waveforms to zero interchange ZXL and ZXU All V's: flip all voltages except pg valves RCUR offsets are 8000,5000. Leave them for now, change as indicated by seg 2 results. CLEARIN: set keyword ZXA=.394 and CALL ALG | Apply | OK factor Controller: zxa=.394 | call | OK | OK timeconst. STRKSPI: keywords RSOUT=.685,ZSOUT=.60,ZXA=.394 | Call Alg waveform set to zero Pgain to 5 starting at .3sec Controller: zxa=.394,RSOUT=.685,ZSOUT=.60, ALL=['PSI0','RCUR','ZCUR','CLEARIN','STRKPSI','RXL','ZXL','RXU'] Call | set common rate to 9.0 Factor to .0001 (was .0002) RXL: Load gains from RXU (seg 4, wire 7) ZXL: Load gains from ZXU (seg 4, wire 8) RXU: Load gains from RXL (seg 4, wire 5) ZXU: Load gains from ZXL (seg 4, wire 6), set to 0.5 after 0.3sec Change title to reverse-field upper null Segment 4: Import segment from PCS$SCRATCH, shot -1, seg#3 Didn't pick up the title change, but seems to have the waveforms? For RXL, RXU load wave from 990922006 RXU, RXL (\PCS::TOP.SEG_02:P_05, \PCS::TOP.SEG_02:P_07). RXL exceeds wavegen limit; reduce factor from 4e-5 to 3.6e-5. Change title to "Reverse field triangularity scan" BUild All Save As 991102999 Load/Clean | |
| Nov 2 1999 07:24:04:717AM | Steve Wolfe | Coldstart at 07:23 Logfile ok, all matrices respond. | |
| Nov 2 1999 09:54:02:437AM | John Goetz | Run Plan for Tuesday, 991102 ---------------------------- Engineering setup: ------------------ Engineering Setup for Tuesday, 991102: Reverse field startup and development Run will BEGIN at 09:00 and END at 17:00 Setup power supplies from shot 991029022, except INCREASE EF4 CURRENT LIMIT TO 10kA. Note: FIZZLE DETECTOR OFF TIME at 1.7 s 2 hours ECDC in D2 prior to the run, stopping 1 hour before run. Gas setup - fill B-top with 6 psi of D2 Hybrid enabled fill B-side lower with 1 psi of Ar Hybrid DISABLED fill B-side upper with 10 psi of H2 Hybrid DISABLED fill C-side with 30 psi of D2 Hybrid enabled leave J-bottom as is (pumped out) Assuming no vacuum problems, enable ECE and VUV | |
| Nov 2 1999 09:54:46:840AM | John Goetz | Run Plan for Tuesday, 991102 ---------------------------- Engineering setup: ------------------ Physics Run Plan ________________ MP#206 Startup (with reversed field) The purpose of this run is to establish operation with reversed (positive) field and current, and carry out discharge development for the reverse-field physics runs later this week. The TF and PF busswork was reversed on Saturday. The HYBRID tree has been modified to reflect the new configuration (smw-991030). Operation will begin with a series of PLC power system tests to verify the configuration changes. HYBRID demands will be monitored during the PLC shots to verify the Hybrid/PCS changes. Plasma operation will start with a reversed current shot based on 991029022, a good 800kA shot from last Friday. Some tweaking will presumably be required to get a good startup, and to adjust offsets for RCUR and perhaps other wires. Check gaps for RF. If the startup goes smoothly, we may try increasing the current to 1.0 and 1.2MA, to allow the RF group to check the coupling in preparation for the run on Wednesday. For the upper x-point run on Thursday, Ian wants to do triangularity scans like those on 990922. We'll start from a static point (RXL=0) based on 990922006, then try to implement the sweep. | |
| Nov 2 1999 09:07:37:327AM | 991102001 | Steve Wolfe | Shot 1: EF2U 2Ka PLC test shot. EF2U current is positive (as intended). EF2U voltage is also ok Hybrid is calling for +voltage, then negative |
| Nov 2 1999 09:13:15:077AM | 991102002 | Steve Wolfe | Shot 2: EF2L 2Ka PLC test shot. looks ok |
| Nov 2 1999 09:17:59:327AM | 991102003 | Steve Wolfe | Shot 3: EFC 1Ka test shot. Well, it's the right sign (negative on EFC_L). |
| Nov 2 1999 09:24:01:030AM | 991102004 | Steve Wolfe | Shot 4: TF test (50KA, PLC control) 50KA positive, ok. |
| Nov 2 1999 09:52:34:387AM | 991102005 | Steve Wolfe | Shot 5: EF4 PLC test, "forward" (should be positive) OK, current is positive, 1250A in each. I'm not seeing the supply voltage trace. STORE ACTION, and several inputs of \ENGINEERING::PS_CAMAC.EF4:TR16_1 were turned off in the pulse file. Too quick on the trigger. Next shot, reverse bridge EF4. |
| Nov 2 1999 09:56:45:950AM | 991102006 | Steve Wolfe | Shot 6: EF4 PLC test, "reverse" (should be negative) OK, negative current, and the voltage is now there. Goetz is here and taking over as PHYSOP. I'll close my PCS now |
| Nov 2 1999 10:01:44:327AM | 991102006 | John Goetz | Shot 006 EF4 test shot 800 A; looks o.k. |
| Nov 2 1999 10:04:24:890AM | 991102007 | John Goetz | Shot 007 EF3 test shot air flow problem; no pulse |
| Nov 2 1999 10:16:26:060AM | 991102008 | John Goetz | Shot 008 EF3 test shot looks o.k.; -1750A |
| Nov 2 1999 10:21:41:327AM | 991102009 | John Goetz | Shot 009 EF1U 2kA test shot looks o.k., 2kA positive |
| Nov 2 1999 10:26:06:450AM | 991102010 | John Goetz | Shot 010 EF1L 2kA test shot looks o.k., 2kA positive |
| Nov 2 1999 10:34:11:857AM | 991102011 | John Goetz | Shot 011 OH1 test shot 10kA positive |
| Nov 2 1999 10:41:24:653AM | 991102012 | John Goetz | Shot 012 OH2U test shot 10kA positive |
| Nov 2 1999 10:48:55:280AM | 991102013 | John Goetz | Shot 013 OH2L test shot 10kA positive |
| Nov 2 1999 11:20:09:967AM | 991102014 | John Goetz | Shot 014 try for plasma EF4 PLC limit set to 8.5kA only Results fizzle breakdown @ .027s!!! Bz = -.02T @ t=0; Br(0)=.001 OH1 has about 1kA more current than an equivalent shot from last week |
| Nov 2 1999 11:39:05:937AM | 991102015 | John Goetz | Shot 015 draw IC_OH1 to 19kA (was 20kA) Results fizzle breakdown @ .006s Bz = -.012T @ breakdown but it is going more negative after breakdown; Br(0)~0mT field curvature is in the wrong direction |
| Nov 2 1999 11:54:49:733AM | 991102016 | John Goetz | Shot 016 draw IC_OH1 to 18.5kA draw IC_EF1s to 5900A (was 6000A) Results fizzle breakdown @ .005s Bz = -.005T @ breakdown but it is going more negative after breakdown; Br(0)~0mT at breakdown field curvature is in the wrong direction |
| Nov 2 1999 12:22:03:357PM | 991102017 | John Goetz | Shot 017 draw IC_OH1 to 19kA draw IC_EF4 to 1380A (was 1280A) it was noted that in the INIT scope, there is an offset removed from the plasma current in the calculation of R and z. This offset is off by a sign now that the currents have been reversed. Results fizzle breakdown @ .0045s Bz = -.006T @ breakdown but it is going more negative after breakdown; Br(0)~0mT at breakdown field curvature is in the wrong direction |
| Nov 2 1999 12:41:38:153PM | 991102018 | John Goetz | Shot 018 draw IC_EF1s to 5800A the RCUR error is ~14000 before breakdown so change the sign on the RCUR offset (from 5500 to -5500 in segment 1 and from 8000 to -8000 in segment 2) Results fizzle breakdown @ .005s Bz = -.008T @ breakdown and goes the right direction after breakdown; Br(0)~0mT at breakdown field curvature is straight now RCUR error is down to ~4000 |
| Nov 2 1999 01:21:16:840PM | 991102019 | John Goetz | Shot 019 draw IC_EF4 to 1410A keep going on the RCUR offset - change to -10000 in segment 1 Results fizzle breakdown @ .006s Bz = -.006T and Br(0)~0mT at breakdown RCUR error is down to -1500 it is difficult to tell where this plasma is, but the thought is that it is getting snuffed on the inside |
| Nov 2 1999 01:58:04:000PM | 991102020 | John Goetz | Shot 020 draw IC_EF4 to 1440A change the RCUR offset to 0 in segment 1 Results fizzle; very weak; hardly any visible light signals breakdown @ .007s Bz = -.005T but not rising very quickly; Br(0)~+1mT at breakdown RCUR error is 8000 |
| Nov 2 1999 02:19:01:513PM | 991102021 | John Goetz | Shot 021 draw IC_EF4 to 1425A change pre-puff to 22ms from 18ms Br_0 offset = -.001 Results dud - OH2L power supply pulls out at -0.7s - probably something to do with the switch closing |
| Nov 2 1999 02:42:26:107PM | 991102022 | John Goetz | Shot 022 draw IC_EF4 to 1425A change pre-puff to 22ms from 18ms Br_0 offset = -.001 Results fizzle; breakdown @ .0055; makes 40kA Bz turns around after the breakdown second plasma forms over the whole chamber at about 45ms |
| Nov 2 1999 03:07:31:980PM | 991102023 | John Goetz | Shot 023 RCUR offset = -2750 Results fizzle; breakdown @ .006; only 30kA Bz turns around as the current rises which shouldn't be three plasmas this time - @ .006, .02, and .034 |
| Nov 2 1999 03:32:03:733PM | 991102024 | John Goetz | Shot 024 RCUR offset = -5500 Results plasma - ran until 2sec; breakdown @ .006; bad hesitation outer gap = 0; inner gap = 0.8cm->0; kappa = 1.3 there appears to be a large offset on CLEARIN |
| Nov 2 1999 03:51:03:967PM | 991102025 | John Goetz | Shot 025 RCUR offset = -7000 draw CLEARIN to +.01 (was -.01) to try and take care of the offset Results plasma - ran until 2sec; breakdown @ .006; bad hesitation outer gap = 0.6; inner gap = 1cm->0.5; kappa = 1.5 ZCUR = -1.3cm |
| Nov 2 1999 04:03:42:047PM | 991102026 | John Goetz | Shot 026 RCUR offset = -8000 draw CLEARIN to +.015 ZCUR up 1cm to -.007 load spare CLEARIN onto wire 9 - turn off gains and put 10000 offset in for diagnostic purposes Results plasma - ran until 2sec; breakdown @ .0055; really bad hesitation! outer gap = 0.75; inner gap = 1.2cm->0.9; kappa = 1.6 ZCUR = -0.6cm |
| Nov 2 1999 04:23:46:700PM | 991102027 | John Goetz | Shot 027 PG4 = 20ms (was 23ms) to try and help out on the current rise CLEARIN = .01 @ .37s to try and get rid of OH1 crossover bump Results fizzle; breakdown @ .0055 current rises to 40kA too quickly for the Bz |
| Nov 2 1999 04:35:35:450PM | 991102028 | John Goetz | Shot 028 RCUR offset = -9000 to try and speed up Bz Results fizzle; breakdown @ .006 current rises to 40kA |
| Nov 2 1999 04:47:03:640PM | 991102029 | John Goetz | Shot 029 PG4 = 22ms (was 20ms) BR_0 offset = 0 (was -.001) Results fizzle; very weak |
| Nov 2 1999 05:04:26:090PM | 991102030 | John Goetz | Shot 030 reload 991102026 CLEARIN = .01 @ .37sec Results fizzle; this one appears to be moving down |
| Engineering Operator Comments | ||||
| Shot | Time | Type | Status | Comment |
| 1 | 08:59:42:513AM | Test | Ok | ef2u only |
| 2 | 09:08:13:090AM | Test | Ok | ef2lower test only |
| 3 | 09:15:21:297AM | Test | Ok | efc chopper test only |
| 4 | 09:22:22:327AM | Test | Ok | TF test only |
| 5 | 09:38:15:810AM | Test | Ok | ef4 forward test |
| 6 | 09:52:57:153AM | Test | Ok | ef4reverse test |
| 7 | 10:01:15:263AM | Test | Ok | ef3 air flow fault , no pulse |
| 8 | 10:11:30:920AM | Test | Ok | ef3 test |
| 9 | 10:17:55:903AM | Test | Ok | ef1upper test |
| 10 | 10:24:43:263AM | Test | Ok | EF1 Lower test ok |
| 11 | 10:31:21:810AM | Test | Ok | oh1 test |
| 12 | 10:39:21:280AM | Test | Ok | oh2upper test |
| 13 | 10:46:52:140AM | Test | Ok | oh2lower test |
| 14 | 11:09:52:217AM | Plasma | Ok | no faults/late fizzel |
| 15 | 11:28:31:030AM | Plasma | Ok | no faults fizzel |
| 16 | 11:47:32:653AM | Plasma | Ok | no faults/fizzel |
| 17 | 12:10:50:060PM | Plasma | Ok | no faults/fizzell |
| 18 | 12:32:14:153PM | Plasma | Ok | (null) |
| 19 | 01:00:56:890PM | Plasma | Ok | (null) |
| 20 | 01:46:40:357PM | Plasma | Ok | fizzle |
| 21 | 02:07:33:873PM | Plasma | Bad | oh2lower comm flt. |
| 22 | 02:31:29:733PM | Plasma | Ok | fizzle |
| 23 | 02:54:36:217PM | Plasma | Ok | fizzle |
| 24 | 03:09:14:560PM | Plasma | Ok | Plasma |
| 25 | 03:32:01:450PM | Plasma | Ok | plasma |
| 26 | 03:51:51:763PM | Plasma | Ok | plasma |
| 27 | 04:11:40:467PM | Plasma | Ok | fizzle |
| 28 | 04:26:11:750PM | Plasma | Ok | fizzle |
| 29 | 04:41:40:950PM | Plasma | Ok | fizzle |
| 30 | 04:55:50:547PM | Plasma | Ok | fizzle |